Annexure I (Reference File No.: MED-16014(12)/1/2024-eoffice)
CATEGORIES OF MEDICAL DEVICES FOR RISK-BASED CLASSIFICATION
S. No. Categories of Medical Devices 1 Interventional Radiology 2 Radiotherapy 3 Oncology 4 Class A (non-sterile and non-measuring) medical devices
Note: All concerned associations/stakeholders are requested to forward their comments by filling the Google form at https://docs.google.com/forms/d/1e- FFEU6wVU9nxgE-B40lJfeFdaMHxO_Y1jxUiQSq2sw/edit?pli=1 within 30 days from the date of publication of this draft.
Name of Category:
Interventional Radiology
Total No. of Devices:
186
Category: Interventional Radiology
S.
No.
Device name
Intended use
Risk
Classification
1 Angiographic x-
ray system
An angiographic x-ray system is a device
intended for radiologic visualization of the
heart, blood vessels, or lymphatic system
during or after injection of a contrast medium.
Class C
2 Angiography
contrast medium
injection system,
battery-powered
A battery-powered assembly of devices
designed to inject contrast media through a
small catheter and into the vascular system
for angiographic procedures (e.g.,
examination of the coronary and renal
arteries, and great vessels and vasculature of
the heart, brain, abdominal organs, and
extremities).
Class C
3 Angiography
contrast medium
injection system,
line-powered,
mobile/stationary
An assembly of devices designed to inject
contrast media through a small catheter and
into the vascular system for angiographic
procedures (e.g., examination of the coronary
and renal arteries, and great vessels and
vasculature of the heart, brain, abdominal
organs, and extremities). It consists of mains
electricity (AC-powered) electromechanical
injectors (electric motor connected to a
jackscrew that moves a syringe piston) with
syringes capable of delivering media with the
pressure, flow range, and volume required for
angiographic studies, and may be stationary
or mobile (e.g., on a wheeled pedestal).
Some types can synchronize media delivery
with the electrocardiographic cycle and/or the
x-ray generator.
Class C
4 Automatic
radiographic film
processor
An automatic radiographic film processor is a
device intended to be used to develop, fix,
wash, and dry automatically and continuously
film exposed for medical purposes.
Class C
5 Automatic-
aperture-control
diagnostic x-ray
system
collimator
A standard, diagnostic x-ray beam-limiting
and shaping device that includes an
automatically controlled motorized shutter
mechanism. It is designed to adjust the
shutters so that the x-ray beam shape
matches the size of the film cassette in the
holder and is attached to the beam exit port
of a diagnostic x-ray system tube housing
assembly. It is used to limit the effects of
scattered radiation on image quality and to
provide patient protection by eliminating
exposure to non-target body areas.
Class B
6 Axial length
measuring
ultrasound
system
An ophthalmic device that uses ultrasound
technology to measure axial length of the eye
(distance along the axis of the eye between
the anterior corneal surface and the anterior
retinal surface).
Class B
S. No. Device name Intended use Risk Classification 7 Basic diagnostic x-ray system table, non- powered A non-powered table that is a component of a basic diagnostic x-ray system designed to position and support a patient during a variety of routine/planar or speciality diagnostic procedures requiring the use of a diagnostic x-ray system; it is not a planar tomography nor computed tomography (CT) table. It is not intended for use during interventional radiology/surgery. Class B 8 Basic diagnostic x-ray system table, powered A powered/programmable table that is a component of a basic diagnostic x-ray system designed to position and support a patient during a variety of routine/planar or speciality diagnostic procedures requiring the use of a diagnostic x-ray system; it is not a planar tomography nor computed tomography (CT) table. It is not intended for use during interventional radiology/surgery. Class B 9 Basic/General- purpose diagnostic x-ray system, mobile/stationary /portable An assembly of devices that comprise an analogue general-purpose mobile diagnostic x-ray system used in a variety of routine x-ray imaging applications. It is typically an x-ray film based system using analogue or analogue-to-digital techniques for image capture and display. It is commonly used for bedside imaging and for interventional and intraoperative imaging. Class C 10 Bladder ultrasound imaging transducer An ultrasound imaging transducer assembly specifically designed to be positioned within the bladder either manually or under endoscopic guidance that steers, focuses, and detects the ultrasound beam and resulting echoes either mechanically or electronically. Class B 11 Bladder volume ultrasound imaging system An assembly of portable, battery-powered devices designed for the extracorporeal ultrasound measurement of the male or female urinary bladder volume. It typically consists of an electronic measuring and imaging display unit with a connected ultrasound transducer which together reproduce digitized data of the bladder, typically bladder volume and post-void residual (PVR), and images. It will generate ultrasound pulses, direct them to the target area, detect the ultrasound echoes, and process the resulting information to produce and display static or dynamic three- dimensional (3-D) data/images. It is used in the diagnosis of urinary retention and postoperative urinary retention (POUR). Class B
S.
No.
Device name
Intended use
Risk
Classification
12 Blood flowmeter
catheter, Doppler
A flexible tube intended to be inserted into
the lumen of a blood vessel to determine
blood-flow velocity by measuring the
ultrasonic frequency shift between
transmitted and reflected signals (Doppler
principle).
Class C
13 Body-orifice
ultrasound
imaging
transducer cover
A sheath intended to be used as a physical
barrier for protection against the effects of
environmental exposure (e.g., body fluids,
gliding creams) and/or to maintain the
required hygienic level of an ultrasound
imaging system transducer (probe) used in
natural body orifices (e.g., vagina, rectum,
oesophagus) during an ultrasound
examination; it has no additional functionality.
Class B
14 Body-orifice
ultrasound
imaging
transducer
cover, image-
enhancement
A sheath containing an extraluminal
pocket/balloon portion intended to cover an
ultrasound imaging system transducer
(probe) used in natural body orifices (e.g.,
vagina, rectum), to both protect the
transducer and improve image quality. The
balloon portion is intended to, when filled with
an aqueous solution (e.g., saline), displace
air and provide an adjustable interface
between the transducer and the anatomy of
interest (e.g., prostate) to improve imaging.
Class B
15 Bone
densitometer
A bone densitometer is a device intended for
medical purposes to measure bone density
and mineral content by x-ray or gamma ray
transmission measurements through the
bone and adjacent tissues.
Class C
16 Bone sonometer
A bone sonometer is a device that transmits
ultrasound energy into the human body to
measure acoustic properties of bone that
indicate overall bone health and fracture risk.
Class B
17 Cardiovascular
MRI system,
Permanent
magnet
A magnetic resonance (MR) diagnostic
imaging device specifically designed to
visualize the cardiovascular system. The
device uses a permanent magnet, and can
be stationary, mobile, or portable. The device
is equipped with a patient support bed
specifically designed to control the
positioning of the patient's body, to optimize
visualization of the cardiovascular system at
rest and during exercise stress tests, or
during interventions.
Class C
S.
No.
Device name
Intended use
Risk
Classification
18 Cardiovascular
MRI system,
resistive magnet
A magnetic resonance (MR) diagnostic
imaging device specifically designed to
visualize the cardiovascular system. The
device uses a resistive magnet and is
equipped with a patient support bed
specifically designed to control the
positioning of the patient's body, to optimize
visualization of the cardiovascular system at
rest and during exercise stress tests, or
during interventions.
Class C
19 Cardiovascular
MRI system,
Superconducting
magnet
A magnetic resonance (MR) diagnostic
imaging device specifically designed to
visualize the cardiovascular system. The
device uses a superconducting magnet, and
can be stationary, mobile, or portable. The
device is typically equipped with a patient
support bed specifically designed to control
the positioning of the patient's body, to
optimize visualization of the cardiovascular
system at rest and during exercise stress
tests, or during interventions.
Class C
20 Cardiovascular
ultrasound
imaging system
An assembly of mains electricity (AC-
powered) devices designed for
extracorporeal and/or intracorporeal
(endosonography or endoscopic) imaging
procedures involving the heart and blood
vessels. Included are software packages that
support a variety of static or real-time cardiac
specific imaging applications used to
diagnose anatomical defects of the heart,
determine blood flow characteristics and
functional/anatomical problems associated
with myocardial infarction. It is used to
generate ultrasound pulses, direct them to a
target area, detect the echoes, and process
the resulting information to produce and
display static or dynamic two or three-
dimensional (3-D) images. May be used in
intraoperative surgical planning.
Class C
21 Cephalometric
digital X-ray
sensor
A sensor for a digital X-ray image processing
system combined with a diagnostic X-ray
system designed for visualization and size
measurement of the human head (skull)
using radiation. The sensor consists of the
charge coupled devices (CCDs) and contains
a sensor driving circuit and signal processing
circuit
Class C
22 Cine or spot
fluorographic x-
ray camera
A cine or spot fluorographic x-ray camera is a
device intended to photograph diagnostic
images produced by x-rays with an
imageintensifier
Class C
S. No. Device name Intended use Risk Classification 23 Computed radiography digital imaging scanner A device/device assembly designed to be used with an analogue x-ray system to capture radiographic images and then generate digital x-ray images from them as part of a two-step process (computed radiography) for image viewing, storage, or hard-copy printing; it is not dedicated to dental use. It consists of an image reader/scanner and may also include an exposure unit with imaging receptor (e.g., a cassette containing a plate), or additional supportive hardware (e.g., a printer). Class B 24 Computer system for ultrasound imaging system A mainframe computer, a personal computer (PC) or a PC-based platform, and related hardware, software and operating system software dedicated used for controlling and monitoring performance of a diagnostic ultrasound system and related image processing, display and analytical functions. Class B 25 Contrast medium conservation system (Contrast management system) An assembly of electromechanical devices that controls the amount of contrast agent supplied to the syringe/injector portion of the contrast medium injector system. It is designed to optimise the use of injectible contrast agents in order to prevent unnecessary waste of contrast media during and/or between patient procedures. Class B 26 Contrast medium injection system administration set A collection of noninvasive devices designed to conduct fluids between a contrast medium injection system and a patient catheter for the delivery of a contrast medium and/or saline solution to a patient undergoing a diagnostic imaging procedure. It consists of tubing with switches/valves and connectors to syringes/fluid/contrast medium injection system, and possibly a pressure transducer. This is a disposable device. Class B 27 Contrast medium injection system control unit A mains electricity (AC-powered) electronic device that may be a component of a contrast medium injector system and designed to initiate and terminate the operation of a contrast medium injection procedure from a remote location (e.g., the shielded operators control room). Class C
S.
No.
Device name
Intended use
Risk
Classification
28 Contrast medium
injection system
hand controller
kit
A collection of non-powered devices
designed to be used and interact with a
contrast medium injection system for the
administration of a contrast medium or saline
solution to a patient undergoing a diagnostic
imaging procedure. It consists of a high
pressure line, a high pressure stopcock, and
a hand-held manual controller used to stop
and start the contrast medium injection
system instantaneously.
Class C
29 Contrast medium
injection system
syringe
A component of a contrast medium injection
system for the administration of a contrast
medium or saline solution to a patient
undergoing a diagnostic imaging procedure.
It is primarily an exchangeable plunger-in-
barrel chamber for the medium/solution; the
movement of the plunger (piston) is
controlled by the contrast medium/solution
injector. This is a reusable device.
Class B
30 Coronary optical
coherence
tomography
system
An assembly of optical and computer-based
devices that use a broad-bandwidth light
beam for optical coherence tomography
(OCT) to provide high-resolution images
(e.g., 5 to 40 microns) of the coronary
intravasculature. The system includes an
optical-imaging catheter which is connected
to the console for transmission of the images
from the intravascular site. The system may
be used during interventional cardiology,
radiology, or diagnostic angiography.
Class C
31 Coronary optical
coherence
tomography
system catheter
A sterile flexible tube with an extremely fine
working end that houses an optical fibre and
imaging lens intended to be transcutaneously
introduced into the coronary vasculature to
perform coronary optical coherence
tomography (OCT). The proximal end of the
device connects to a coronary OCT console
for the transmission of high-resolution images
(e.g., 5 to 40 micron) of the coronary
intravasculature.
Class D
32 CT contrast
medium injection
system,
stationary/mobile
An assembly of devices designed to inject
contrast media through a small catheter and
into the vascular system for computed
tomography (CT) procedures (e.g., spine,
head, gastrointestinal, or vascular diagnostic
procedures using CT scanning). The system
may synchronize media delivery with the x-
ray generator during CT scanning.
Class C
S. No. Device name Intended use Risk Classification 33 Dental computed radiography digital imaging scanner A device/device assembly dedicated to dental use and designed to be used with an analogue dental x-ray system to capture radiographic images and then generate digital x-ray images from them as part of a two-step process (computed radiography) for image viewing, storage, or hard-copy printing. It consists of an image reader/scanner and may also include an exposure unit with imaging receptor (e.g., a cassette containing a plate), or additional supportive hardware (e.g., a printer). Class B 34 Dental digital X- ray imaging sensor A digital dental X-ray sensor used in the oral cavity, in combination with a general- purpose dental X-ray imaging device. The sensor consists of a CCD or other components, and comprises a sensor drive circuit and a signal processing circuit. Class B 35 Dental x-ray film, screen A screen x-ray film specifically sized and designed for use with dental x-ray systems. It is designed to be sensitive primarily to the wavelengths of light emitted from an image intensifying screen. May be used endorally/extraorally. Class B 36 Diagnostic ultrasonic transducer A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Class B 37 Diagnostic x-ray beam-limiting device A diagnostic x-ray beam-limiting device is a device such as a collimator, a cone, or an aperture intended to restrict the dimensions of a diagnostic x-ray field by limiting the size of the primary x-ray beam. Class C 38 Diagnostic x-ray digital imaging conversion system An assembly of medical devices designed to convert an existing analogue x-ray system to digital (i.e., retrofit) through the acquisition and processing of digital images using existing analogue x-ray system components. As a system it consists of both: 1) a filmless cassette-size digital detector (e.g., indirect flat panel detector); and 2) additional processing/viewing hardware [e.g., a central processing unit (CPU) with integrated software and a monitor]. It does not contain the controls for the direct operation of an x- ray imaging system. Class B
S.
No.
Device name
Intended use
Risk
Classification
39 Diagnostic x-ray
digital imaging
system
workstation
A freestanding component of an x-ray-based
diagnostic digital imaging assembly intended
to associate X-ray/CT/fluoroscopic images
with patient and exam information, apply
images processing to facilitate diagnosis,
display the images, and output the resulting
image and exam data for further display,
distribution, or archiving. It may be intended
to be integrated as part of a radiology picture
archiving and communication system (PACS)
and does not have controls for direct
operation of the imaging system.
Class B
40 Diagnostic x-ray
high voltage
generator
A diagnostic x-ray high voltage generator is a
device that is intended to supply and control
the voltage and electrical energy provided to
a diagnostic x-ray tube for medical purposes.
Class B
41 Diagnostic x-ray
tube housing
assembly
A diagnostic x-ray tube housing assembly is
an x-ray generating tube encased in a
radiation-shielded housing that is intended for
diagnostic purposes.
Class B
42 Digital
angiographic x-
ray system,
mobile/stationary
/portable
An assembly of devices intended to visualize
and optimize the macroscopic or quantitative
evaluation of the anatomy and function of the
blood and lymphatic systems in the heart,
brain, or other organs. The device uses
digital technology for real-time image capture,
display, and manipulation, and typically may
have a radiographic function in addition to
fluoroscopic function.
Class C
43 Digital basic
diagnostic x-ray
system,
mobile/portable/s
tationary
An assembly of devices that comprise a
digital general-purpose mobile diagnostic x-
ray system used in a variety of routine planar
x-ray imaging applications. It uses digital
techniques for image capture, display and
manipulation. It is commonly used for
bedside imaging and for interventional and
intraoperative imaging.
Class C
44 Digital
cephalometric x-
ray system
An x-ray system designed to generate and
control x-ray beams and to record the
absorption patterns of x-rays passing through
a patient's head (skull). It is used for: 1)
radiographic visualization and measurement
of the dimensions of the human head, e.g.,
abnormal structures of the skull; 2)
orthodontic assessment of the relationship of
the teeth to the jaws and the jaws to the rest
of the facial skeleton; 3) assessment airway
and other soft tissue structures.
Class C
S.
No.
Device name
Intended use
Risk
Classification
45 Digital general-
purpose
fluoroscopic x-
ray system,
mobile/portable/s
tationary
A general-purpose diagnostic fluoroscopic x-
ray system that uses a C-arm and digital
techniques for image capture, display and
manipulation and is designed to be used in a
variety of general-purpose applications
requiring real-time fluoroscopic imaging
capabilities. It is also intended to optimize the
capability of users to visually and
quantitatively evaluate the anatomy and
physiological function of various targeted
body areas in real-time. It is frequently used
in conjunction with an ingested or injected x-
ray contrast medium. Images can be both
real-time and delayed formats.
Class C
46 Digital imaging
cassette/storage
phosphor
screen, basic
diagnostic x-ray
(CR cassette)
A device that includes both a digital imaging
cassette and a phosphor plate (also known
as screen) intended to receive radiation by a
diagnostic x-ray system, to record a patient
radiation pattern during a non-dental
procedure. The device is then transferred,
with the cassette shielding the plate from
environmental light, to a computed
radiography (CR) scanner for insertion to
produce the digital image of the patient.
Class B
47 Digital intraoral
dental x-ray
system,
mobile/portable/s
tationary/hand-
held
A diagnostic dental x-ray system designed to
generate and control x-ray beams. It records
the absorption pattern of x-ray beams used
for general-purpose, routine, dental
radiography examinations involving the
diagnosis and treatment (e.g., surgical or
interventional) of diseases of the teeth, jaw
and oral cavity structures. The data is either
from analogue-to-digital conversion
techniques imaging or by digital imaging.
Class C
48 Digital
mammographic
x-ray system,
mobile/stationary
/portable
An assembly of devices specifically designed
to be installed permanently and intended to
compress and image the breast. It is primarily
used to optimize the capability of users to
visually evaluate x-ray film images
representing the anatomy and function of
blood and lymphatic vessels within the
human breast. It may be used for breast
cancer screening and in conjunction with the
placement of biopsy markers and stereotactic
biopsy and lesion localization equipment
requiring x-ray guidance.
Class C
S. No. Device name Intended use Risk Classification 49 Digital panoramic/tomo graphic dental x- ray system (OPG) A diagnostic digital dental x-ray system designed for permanent fixture in one location with an extraoral x-ray sensor and source intended to generate and control x-ray beams used in advanced dental imaging applications involving the teeth, jaw, oral cavity, sinus, and/or other maxillofacial structures. Class C 50 Digital uro- gynaecological fluoroscopic x- ray system A diagnostic x-ray system with real-time fluoroscopic capabilities specifically designed for use in urological and/or gynaecological surgical and interventional procedures requiring real-time visualization of the pelvic area. Class C 51 Doppler blood- flow measurement ultrasound system A portable or stationary ultrasonic device (invasive or noninvasive) that does not produce 2-dimensional or 3-dimensional images, and is intended to be used for determining various blood-flow related parameters of the heart, artery, or vein. The device is used to output audible data, or display the Doppler shift as a function of time. Class B 52 Electrostatic x- ray imaging system An electrostatic x-ray imaging system is a device intended for medical purposes that uses an electrostatic field across a semiconductive plate, a gas-filled chamber, or other similar device to convert a pattern of x- radiation into an electrostatic image and, subsequently, into a visible image. Class C 53 Emission computed tomography system An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron- emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data Class C 54 Extracorporeal ultrasound imaging transducer, hand-held (ultrasound probe) A hand-held noninvasive component of an ultrasound imaging assembly designed to be moved over the intact surface of a patient's body, typically with a coupling gel, during a variety of extracorporeal ultrasound imaging procedures (non-dedicated). it converts electric voltages into ultrasound which it transmits, detects resulting echoes, and transmits a proportional signal/data to an ultrasound system controller for image processing and display. This is a reusable device. Class B 55 Extremity MRI system, permanent magnet A diagnostic magnetic resonance imaging (MRI) system employing permanent magnet technology specifically designed to image only the head and neck or limbs. Class B
S. No. Device name Intended use Risk Classification 56 Extremity MRI system, superconducting magnet A diagnostic magnetic resonance imaging (MRI) system employing superconducting magnet technology specifically designed to image only the head and neck or limbs. Class B 57 Flat panel radiography and fluoroscopy system A digital image acquisition system in which X- rays passed through a human body are captured by an X-ray flat panel detector, and digital output signals are supplied to an image processing device to receive digital images. If necessary, image processing technology including gamma correction and edge enhancement are applied to digital images. Image data are either recorded on the media, or output to a server or other external devices for recording. Class C 58 Flexible endoscopic ultrasound imaging transducer A flexible, ultrasound imaging transducer assembly designed to be to be inserted through the working channel of an appropriate flexible endoscope and positioned within the gastrointestinal, respiratory or urinary tract, (i.e., application is not specific to any particular anatomy) to transmit a signal/data to another device for display. This is a reusable device. Class B 59 Flexible ultrasound bronchoscope An endoscope with a flexible inserted portion intended for the visual examination and treatment of the trachea, bronchi, and lungs. It is inserted through the mouth or nose during bronchoscopy. Class B 60 Flexible ultrasound colonoscope An endoscope with a flexible inserted portion intended for the visual examination and treatment of the entire colon [lower gastrointestinal (GI) tract]. It is inserted through the anus during colonoscopy. Class B 61 Flexible ultrasound duodenoscope An endoscope with a flexible inserted portion, combined with an ultrasound probe, intended for the visual examination and treatment of the duodenum (the first part of the small intestine). It is inserted into the body through the mouth during duodenoscopy. Class B 62 Flexible ultrasound gastroduodenosc ope An endoscope with a flexible inserted portion, combined with an ultrasound probe, intended for the visual examination and treatment of the upper gastrointestinal (GI) tract [oesophagus, stomach, and duodenum (the first part of the small intestine), including the pancreas and the bile duct]. It is inserted into the body through the mouth during gastroduodenoscopy. Class B
S. No. Device name Intended use Risk Classification 63 Flexible ultrasound laparoscope An endoscope with a flexible inserted portion, combined with an ultrasound probe, intended for the visual examination, treatment, and ultrasonic imaging of the abdominal/retroperitoneal cavity and its organs. It is inserted through an incision made in the abdominal wall (routinely just below the umbilicus) during laparoscopy. Class B 64 Fluorescent scanner A fluorescent scanner is a device intended to measure the induced fluorescent radiation in the body by exposing the body to certain x- rays or low-energy gamma rays. Class C 65 Foetal cardiac monitor A mains electricity (AC-powered) device designed to detect, measure, and display foetal heart activity during the perinatal period. Class C 66 Foetal Doppler system A portable, hand-held, battery-powered device assembly consisting of a measuring and display unit and an attached probe or interchangeable probes designed to noninvasively detect foetal heart beats using ultrasound/Doppler technology. The heart beats are typically conveyed audibly via the measuring/display unit and attached probe which is applied to the surface of the pregnant woman's abdomen. The device aids in determining foetal viability. Class C 67 Full-body MRI system, permanent magnet A general-purpose magnetic resonance imaging (MRI) system designed to scan any targeted area of the body. It includes a permanent magnet assembly. Class C 68 Full-body MRI system, resistive magnet A diagnostic general-purpose magnetic resonance imaging (MRI) system designed to scan any targeted area of the body (full-body imaging). It includes a resistive magnet assembly. Class C 69 Full-body MRI system, superconducting magnet A diagnostic general-purpose magnetic resonance imaging (MRI) system designed to scan any targeted area of the body (full-body imaging). This system includes a superconducting magnet assembly. Class C 70 General- purpose ultrasound imaging system A stationary or mobile (e.g., on wheels) assembly of devices designed to collect, display, and analyse ultrasound images during a variety of extracorporeal and/or intracorporeal (endosonography or endoscopic) ultrasound imaging procedures (e.g., cardiac, OB/GYN, endoscopy, breast, prostate, vascular, and intra-surgical imaging). Class B
S.
No.
Device name
Intended use
Risk
Classification
71 General-purpose
fluoroscopic x-
ray system,
mobile/portable/s
tationary
A general-purpose, analog diagnostic
fluoroscopic x-ray system designed to be
used in a variety of general-purpose
applications requiring real-time fluoroscopic
imaging capabilities. It is frequently used in
conjunction with an ingested or injected x-ray
contrast medium. Images can be both real-
time and delayed formats.
Class C
72 Grid for MRI
(needle guide
positioner)
A grid used for position the puncture site etc.
when biopsying under the magnetic
resonance imaging (MRI) guide.
Class A
73 Hand-held
intraoral dental
x-ray system
A diagnostic dental x-ray system designed to
generate and control x-ray beams. It records
the absorption pattern of x-ray beams used
for general-purpose, routine, dental
radiography examinations involving the
diagnosis and treatment (e.g., surgical or
interventional) of diseases of the teeth, jaw
and oral cavity structures. The sensor is
placed in the mouth, the purpose being to
visualize a limited region in detail. The data is
either from analogue imaging and digitized
afterwards or by digital imaging.
Class C
74 Hand-held
ultrasound
imaging system,
body-surface
A battery-powered, hand-held device/device
assembly designed to be placed over the
external body surface for collection, display,
and analysis of ultrasound information during
a variety of ultrasound imaging procedures
(i.e., non-dedicated); it may in addition be
intended to be used with a sterile cover for
intraoperative imaging. It consists of a flat- or
concave-lensed ultrasound transducer and
includes an integrated image processing
system; it may include an integrated display
or be intended to wirelessly transmit images
to an off-the-shelf device (e.g., smartphone,
tablet) with dedicated software installed. It is
not designed to be inserted into body orifices.
Class B
75 Hand-held
ultrasound
imaging system,
rectal/vaginal
A battery-powered, hand-held device/device
assembly designed to be inserted in
vaginal/rectal cavity for collection, display,
and analysis of ultrasound information during
a variety of ultrasound imaging procedures
(i.e., non-dedicated); it may in addition be
intended to be used with a sterile cover for
intraoperative imaging.
Class B
S.
No.
Device name
Intended use
Risk
Classification
76 Hand-held-
detector nuclear
medicine system
An assembly of diagnostic devices
comprising a hand-held, non-imaging system
designed primarily to detect, record, quantify
and analyse various radioactive emissions
(gamma, alpha, beta) of injected or ingested
radiopharmaceuticals, radiation emitting
devices, or other materials. It is frequently
used in the monitoring and detection of deep
vein thrombosis (DVT), in surgical
procedures requiring radionuclide localization
and in some radiolabeled monoclonal
antibody applications.
Class B
77 Head/extremity
imaging MRI
system
A magnetic resonance (MR) diagnostic
imaging device specifically designed to
visualize the head, neck, or limbs using a
resistive magnet.
Class C
78 Hepatic
ultrasound
elastography
system
An assembly of mains electricity (AC-
powered) devices, which may include
rechargeable batteries, designed to measure
liver stiffness/ultrasonic attenuation of tissues
based on transient elastography for the
evaluation/diagnosis of hepatic disease (e.g.,
liver fibrosis/steatosis); it may additionally be
intended to measure spleen stiffness as an
adjunct evaluation. It is typically used during
diagnosis of chronic liver conditions such as
viral hepatits, alcoholic hepatitis, and biliary
disease, or for post-transplant evaluation.
Class B
79 Image-
intensified
fluoroscopic x-
ray system
An image-intensified fluoroscopic x-ray
system is a device intended to visualize
anatomical structures by converting a pattern
of x-radiation into a visible image through
electronic amplification.
Class C
80 Indirect flat panel
x-ray detector
An electrically-powered, cassette-like device
intended to be used as part of an x-ray
system to detect x-ray images following
exposure, and create a digital signal; it is not
dedicated to imaging of a specific anatomy. It
includes two types of transducing technology
(i.e., indirect): a scintillator [e.g. caesium
iodide (CsI)] screen converts x-ray energy
into light, followed by the conversion of light
into a digital signal by photodiode array. The
image data can be sent to an appropriate
processing unit through a wired or wireless
connection (e.g., Wi-Fi). It may be used in
place of a fixed detector in x-ray imaging
systems, e.g., as part of a digital imaging
conversion system.
Class B
S.
No.
Device name
Intended use
Risk
Classification
81 Integrated
ultrasound/electr
ocardiograph/
stethoscope unit
An electrically-powered modular device
intended for general medical examination
using ultrasound, electrocardiograph (ECG)
and auscultation modalities, with a user
interface that enables synchronized data
display and recording. It consists of a multi-
purpose transducer designed for
transmitting/receiving ultrasound and sensing
sound signals, to which an ECG lead
assembly can be attached, and a connected
image display monitor (e.g., tablet) with
dedicated software; in addition to the
integrated electronic stethoscope, it may also
allow connection of a stethoscope headset
for the user to directly hear sounds.
Class C
82 Intracardiac
ultrasound
imaging catheter,
steerable, single-
use
A steerable, flexible tube with an ultrasonic
phased-array imaging transducer at its distal
tip designed for intracardiac
echocardiography to enable intracardiac and
possibly great vessel (e.g., pulmonary artery)
visualization for the assessment of cardiac
anatomy and physiology, and visualization of
other devices in the heart; it is not intended
for peripheral vascular or coronary artery
insertion/imaging. It is introduced
percutaneously, and includes controls to
allow it to be manoeuvred (steerable). It is
connected to a compatible ultrasound system
for transmission and display of the images.
This is a single-use device.
Class D
83 Intracorporal
ultrasound
transducer
contouring
attachment
Intended to be used with a hand-held
ultrasound imaging system transducer to
provide means to replace the position of a
needle guide.
Class B
84 Intracorporeal
central nervous
system
ultrasound
transducer/probe
An ultrasonic probe used temporarily in
central nervous system surgery. It refers to a
hand-held ultrasonic transducer assembly
designed to be placed at the surgical site to
take local images during surgery. It includes
the configuration of various transducer
assemblies that consist of single or multiple
elements that convert voltage into an
ultrasonic beam. This category includes
ultrasonic transducers used for mode A,
mode B, mode M, Doppler, Color Doppler
(CD), and dual (combination image, Doppler
or color flow) scanning.
Class D
S.
No.
Device name
Intended use
Risk
Classification
85 Intraoperative
ultrasound
imaging
transducer cover
A sterile sheath intended to be used as a
physical barrier for protection against the
effects of environmental exposure (e.g., body
fluids, gels) and/or to maintain the required
hygienic level of an ultrasound imaging
system transducer (probe) used
intraoperatively, e.g., on internal organs,
during an ultrasound examination. This is a
single-use device.
Class B
86 Intraoral dental
x-ray system,
mobile/portable/s
tationary/hand-
held
A diagnostic dental x-ray system designed to
generate and control x-ray beams. It records
the absorption pattern of x-ray beams used
for general-purpose, routine, dental
radiography examinations involving the
diagnosis and treatment (e.g., surgical or
interventional) of diseases of the teeth, jaw
and oral cavity structures.
Class C
87 Intravascular
optical
coherencetomog
raphy catheter
A catheter and a guide wire used to observe
a cross section of blood vessels using an
optical fiber by light. For instance, optical
coherence tomography visualizes internal
microscopic structure of the tissues without
physical invasion of the outer protective
layers. This utilizes the nature of light that
penetrates and is reflected at various
degrees according to the type of the tissue.
Class D
88 Intravascular
ultrasound
imaging catheter,
non-steerable
A flexible, non-steerable tube with an
ultrasound imaging transducer enclosed in a
fluid-resistant or waterproof acoustically- and
electrically-insulated housing, designed for
insertion into the vascular system (i.e.,
peripheral vascular system, coronary artery)
by an operator; it is not intended for
intracardial insertion/imaging. Also known as
a vascular ultrasound transducer, it includes
either a single or an array of transducer
element(s) (piezoelectric, active, or crystal).
Class C
89 Invasive vascular
ultrasound
system
A mains electricity (AC-powered) device
assembly designed to invasively locate and
assess the extent of vascular flow restriction
(e.g., clots, stenosis, mechanical damage)
using ultrasound/Doppler/transit time
technology during procedures involving
vascular surgery. It may also provide
additional measurements (e.g., blood
pressure, vascular resistance). It consists of
a mobile (on wheels) control unit with
graphical user-interface, typically placed
outside the sterile field; and an attached,
reusable, sterilizable probe intended to be
used within the body (invasive) but not within
the lumen of the blood vessel.
Class C
S.
No.
Device name
Intended use
Risk
Classification
90 Invasive vascular
ultrasound
system probe
A hand-operated component of an invasive
vascular ultrasound system intended to be
used perioperatively/intraoperatively to
measure the flow of blood in an isolated
blood vessel, by being applied to the outside
of the vessel, using ultrasonic/Doppler/transit
time technology. It may also be used for
cross-sectional visualization/imaging. This is
a reusable device.
Class C
91 Limited view field
X-ray computed
tomography
system
A diagnostic X-ray computed tomography
(CT) system equipped with a gantry designed
exclusively for taking images of the area from
the head to the neck or the extremities. The
system is designed to have at least 1 fixed
circular arrangement of multiple X-ray tubes
and detectors, or designed to have an
assembly where single or multiple X-ray
tube(s) and detector(s) rotate around an axis
at high speed within the area that the gantry
can take images. Using this system, 2D or 3D
images are produced. In addition, spiral CT
and special radiography can be performed at
multiple angles set relative to the position of
the body.
Class C
92 Mammographic
biopsy x-ray
table, powered
A powered/programmable x-ray
mammography table with electronic positional
controls specifically designed to position and
support a patient during breast imaging
examinations usually in conjunction with
breast biopsy. It is made of radiolucent
materials with low x-ray attenuation
coefficients and typically has special
openings or attachments to facilitate breast
imaging.
Class B
93 Mammographic
MRI system,
Permanent
magnet
A magnetic resonance (MR) diagnostic
imaging device specifically designed for
breast imaging. The device uses a
permanent magnet, and can be stationary,
mobile, or portable. The MR device for
mammography is typically equipped with a
dedicated patient support bed specifically for
the purpose of controlling the positioning of
the patient's body to optimize visualization of
the breast.
Class C
94 Mammographic
MRI system,
Resistive magnet
A magnetic resonance (MR) diagnostic
imaging device specifically designed for
breast imaging. The device uses a resistive
magnet, and can be stationary, mobile, or
portable. The MR device for mammography
is typically equipped with a dedicated patient
support bed for the purpose of controlling the
Class C
S. No. Device name Intended use Risk Classification positioning of the patient's body to optimize visualization of the breast. 95 Mammographic MRI system, Superconducting magnet A magnetic resonance (MR) diagnostic imaging device specifically designed for breast imaging. The device uses a superconducting magnet, and can be stationary, mobile, or portable. The MRI device for mammography is typically equipped with a dedicated patient support bed for the purpose of controlling the positioning of the patient's body to optimize visualization of the breast. Class C 96 Mammographic stereotactic biopsy system An assembly of devices intended to be used in conjunction with mammography to insert biopsy devices, typically designated needles, into a breast in order to extract a biopsy of a suspect tumour for diagnostic purposes. It consists of a biopsy unit with needle guides, needles, an examination table, and an integral computer for communicating lesion coordinates. It may be a standalone unit or a unit that mounts onto an existing mammographic unit. Class B 97 Mammographic x-ray system (portable/mobile/ stationary) A mammographic x-ray system is a device intended to be used to produce radiographs of the breast for diagnostic purposes. It may be used intraoperatively. Class C 98 Mammographic x-ray system stereotactic unit A device that can only fulfil its purpose when used together with and enhance the function of a diagnostic mammographic x-ray system. It is used to take two pictures of a breast from two different angles. On the basis of this stereoscopic picture set, it is possible to determine the exact position of a lesion in the breast and a cytological sample of this will be taken. This device or technique may be built- in to some diagnostic x-ray systems for mammography. Class B
S. No. Device name Intended use Risk Classification 99 Mammographic x-ray system, mobile/stationary /portable An assembly of devices specifically designed to be installed permanently and intended to compress and image the breast. It is primarily used to optimize the capability of users to visually evaluate x-ray film images representing the anatomy and function of blood and lymphatic vessels within the human breast. It may be used for breast cancer screening and in conjunction with the placement of biopsy markers and stereotactic biopsy and lesion localization equipment requiring x-ray guidance. Class C 100 Manual contrast medium injection system An assembly of devices designed for the manual-only injection of contrast media into the blood or lymphatic vessels of a patient for a diagnostic imaging procedure; it enables the operator to manually control the volume and rate of contrast media administered under pressure. The system typically consists of a graduated injection syringe, a manifold with multiple stopcocks, and a pressure transducer system used by the operator to monitor catheter-tip pressure. Class B 101 Medical image analyzer/Medical image interpretive application software Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, are intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis. Class B 102 Medical x-ray film, screen A screen x-ray film specifically designed for medical diagnostic imaging applications. It is sensitive primarily to wavelengths of light emitted from an image intensifying screen or other visible light source. This film is not limited to use with x-ray imaging systems but can also be used in a variety of diagnostic imaging modalities using image intensifier technology or matrix formatters to output images, e.g., nuclear medicine or ultrasound. Class B
S. No. Device name Intended use Risk Classification 103 MR/CT roadmap (interpretive software) A stand alone software that allows the synchronization of live fluoroscopy images with previously acquired MRA/CTA datasets. This may be used as an aid in applications such as image-guided surgeries, calibrating the contrast medium doses, and needle guidance. Class C 104 MRI contrast medium injection system, mobile/stationary An assembly of devices designed to inject contrast media through a small catheter and into the vascular system for magnetic resonance imaging (MRI) procedures (e.g., spine, head, gastrointestinal, or vascular diagnostic procedures using MRI). It may interface with the MRI system or a monitoring device to control timing of contrast injections; it is made entirely of non-ferromagnetically active materials for MRI compatiblity. Class C 105 MRI image interpretive software An interpretive software program intended to be used to analyse magnetic resonance imaging (MRI) data (e.g., prostate MRI) to detect and localize suspected abnormalities and sometimes provide results as clinically relevant tags; it may additionally be intended for three-dimensional (3-D) model creation slice-based images (segmentation). It typically utilizes artificial intelligence (AI) and deep learning techniques, and may be compatible with radiology information systems, data formats, and medical imaging software programs [e.g., picture archiving and communication system (PACS), digital imaging and communications in medicine (DICOM) format]. Class C 106 MRI planner (radiotherapy planning application software) It is a stand alone software product intended to enable a completely MRI-only workflow in radiotherapy planning. It achieves this by analyzing a dedicated MR image and generating a synthetic CT image that can replace the conventional CT in the treatment planning process. In addition, the software may also performs automated contouring of risk organs in certain anatomies. Class C 107 MRI radio frequency coil This radio frequency coil is intended to act as a transmitter, receiver, or both a transmitter and receiver of RF pulses necessary for diagnostic magnetic resonance imaging (MRI) procedures. It is intended to be used to enhance image resolution by improving signal to noise characteristics. Class B
S. No. Device name Intended use Risk Classification 108 MRI system application program software It is a standalone software application that is intended to be used for viewing, post- processing and quantitative evaluation of magnetic resonance (MR) images for applications such as quantitative cardiovascular computed tomography (CT) evaluation, determination of the triglyceride fat fraction in magnetic resonance images of the liver, multi slice and gradient echo image analysis, etc. It does not drive MRI systems. Class C 109 MRI system synchronizer A physiological monitoring unit used as a component of a diagnostic magnetic resonance imaging (MRI) system, that produces a signal which enables image formation or data collection to be synchronized with a specific measurable physiological parameter, e.g., the beginning of a patient's respiratory or cardiac cycle. It is primarily used for the purpose of artifact reduction or enhancement of signal-to-noise ratios in various real-time or dynamic MRI applications. Class B 110 MRI system table, powered A powered/programmable table designed with electronic and/or software controls to position and support a patient during magnetic resonance imaging (MRI) examinations. It is made with ferromagnetically inactive materials and may have a detachable table top, equipment supports, physiological monitors, mattresses, alarms and a patient positioning system (PPS). Class B 111 MRI system workstation A standalone image-processing workstation that is designed to be networked with one or more magnetic resonance imaging (MRI) systems. The workstation may have any type of hardware, or any type of configuration. It is not intended to control or directly manipulate the diagnostic imaging device. The device allows data transfer either on-line or off-line, and is typically situated away from the MRI operator console. It is intended to process MRI images and provide valuable information that allow decision making, evaluation or diagnosis of the patient's pathologic conditions. Class B
S.
No.
Device name
Intended use
Risk
Classification
112 MRI-patient
physiologic
monitoring
system
A mobile assembly of devices designed for
continuous assessment of several vital
physiologic parameters (e.g., ECG, blood
pressure, heart rate, temperature, cardiac
output, apnoea, and respiratory/anaesthetic
gas concentrations) of infant to adult
intensive care patients during a magnetic
resonance imaging (MRI) procedure.
Class C
113 Multi-modality
diagnostic x-ray
system (full body
CT/fluoroscopic
x-ray system)
An assembly of diagnostic x-ray devices
intended to acquire, display and process
images from any part of the body (full-body)
incorporating the techniques of more than
one x-ray imaging modality, typically
including real-time fluoroscopy, two-
dimensional (2D) radiography, and computed
tomography (CT); it is not a single-modality
system.
Class C
114 Neurosurgical
ultrasound
navigation
system
application
software
An individual software program or group of
programs, routines or algorithms that add
specific image processing and/or analysis
capabilities to a neurosurgical ultrasound
(US) navigation system.
Class C
115 Neurosurgical
ultrasound
navigation
system optical
tracking unit
A component of a neurosurgical ultrasound
(US) navigation system used to track the
position of all the localizers/spatial markers
during intraoperative imaging and to supply
the US system with valid positional data. It
typically uses equidistant-placed cameras
using infrared (IR) light to measure the
position/placement of the reflective spheres
attached to the localizers or spatial marker
holders.
Class B
116 Nonfetal
ultrasonic
monitor
A nonfetal ultrasonic monitor is a device that
projects a continuous high frequency sound
wave into body tissue other than a fetus to
determine frequency changes (doppler shift)
in the reflected wave and is intended for use
in the investigation of nonfetal blood flow and
other nonfetal body tissues in motion.
Class B
117 Non-image-
intensified
fluoroscopic x-
ray system
A non-image-intensified fluoroscopic x-ray
system is a device intended to be used to
visualize anatomical structures by using a
fluorescent screen to convert a pattern of X-
ray radiation into a visible image.
Class C
S. No. Device name Intended use Risk Classification 118 Noninvasive vascular ultrasound probe A hand-operated component of a vascular ultrasound system intended to be used to measure the flow of blood noninvasively by measuring flow in underlying vasculature from the body surface pre-/post-operatively, using ultrasonic/Doppler/transit time technology. It may also be used for cross- sectional visualization/imaging. Class B 119 Noninvasive vascular ultrasound system An electrically-powered device assembly consisting of a control unit and an attached probe or interchangeable probes, designed to noninvasively locate and assess the extent of vascular (venous and arterial) flow restriction (e.g., clots, stenosis, mechanical damage) by measuring blood flow rate using ultrasound/Doppler/transit time technology. It may also be used to detect the foetal heartbeat. Class C 120 Nuclear electrocardiogra m synchronizer/Nu clear medicine system synchronizer A nuclear electrocardiograph synchronizer is a device intended for use in nuclear radiology to relate the time of image formation to the cardiac cycle during the production of dynamic cardiac images. Class B 121 Nuclear medicine data processing system A data processing system specifically dedicated for nuclear medicine. The system is used for various types of filter processing, image display, clinical analysis, management of image storage, etc. The system may have a control system for directly operating a diagnostic imaging unit. This system can deliver or receive data online or offline, and has a configuration that can provide functions such as processing, manipulating, and displaying a patient's images and information collected with the nuclear medicine scan. Class C 122 Nuclear medicine system workstation A standalone imaging workstation specifically designed to create network connection with one or more nuclear imaging devices, such as gamma camera, PET device, and SPECT device. The workstation may have any type of hardware, or any type of configuration. The device may be regarded as a component of a picture archiving and communication system (PACS). It differs from an operator console in that the workstation does not have controls for the direct manipulation of the imaging device. Class B
S.
No.
Device name
Intended use
Risk
Classification
123 Nuclear sealed
source for
absorption
compensation
A sealed radioisotope equipped dedicated for
the correction of absorption of radiation from
a diagnostic radioisotope by organs and
tissues to improve quantitative capability of
diagnostic imaging of the nuclear imaging
system (PET or SPECT devices), thereby
improving accuracy of diagnosis.
Class B
124 Nuclear
tomography
system/nuclear
medicine system
A nuclear tomography/medicine system is a
device intended to detect nuclear radiation in
the body and produce images of a specific
cross-sectional plane of the body by blurring
or eliminating detail from other planes.
Class C
125 Nuclear whole
body scanner
A nuclear whole body scanner is a device
intended to measure and image the
distribution of radionuclides in the body by
means of a wide- aperture detector whose
position moves in one direction with respect
to the patient.
Class C
126 Oesophageal
ultrasound
imaging
transducer
(oesophagal
endosonography
probe/transducer
)
An ultrasound imaging transducer assembly
enclosed in a fluid-resistant, acoustically- and
electrically-insulated housing, designed to be
inserted and positioned in the oesophagus to
obtain ultrasound images of the heart and/or
to navigate devices requiring ultrasound
guidance and placement (e.g., endoscopy
equipment or needle biopsy equipment). It is
composed of either a single transducer
element or an array of transducer elements,
i.e., piezoelectric element(s), active
element(s), or crystal(s), and associated
damping, backing, and matching layer
materials. It is reusable.
Class B
127 Ophthalmic
ultrasound
imaging
transducer
An ultrasound imaging transducer that is a
hand-held device moved over the intact
surface of a patient's eye during imaging
applications involving the examination of the
lens. It may include single or multiple element
transducer assembly configurations that
convert electric voltages into an ultrasound
beam. This group of devices typically
includes ultrasound imaging transducers
used with A-mode scanning capabilities. This
is a reusable device.
Class B
S.
No.
Device name
Intended use
Risk
Classification
128 Optical
coherence
tomography
(OCT) imaging
system
An OCT (optical coherence tomography)
diagnostic imaging system which
diagnostically represents the images of tissue
characteristics and shapes of various sites
(e.g., heart, blood vessel, abdomen, lung)
using near-infrared rays, without the need to
obtain tissues surgically. This system may or
may not contain a software package that
supports the taking of both still and real-time
images, and is used to diagnose anatomic
abnormalities in tissues and to examine
functional and anatomic features.
Class C
129 Optoacoustic/ultr
asound imaging
system
A mobile assembly of electrically-powered
devices intended for imaging and analysis of
soft-tissue and soft-tissue vasculature using
combined optoacoustic (photoacoustic) and
ultrasound (US) imaging. Pulsed near-
infrared laser light of various wavelengths is
emitted and the associated acoustic feedback
(photoacoustic effect) provides
complementary information regarding tissue
composition/functionality based on the
presence of endogenous chromophores,
such as haemoglobin; this photoacoustic data
can be overlaid on the associated US image.
Class B
130 Orthopaedic x-
ray ruler
A device used in an orthopaedic clinical
setting to establish an accurate measurement
of length or size related to a bone structure
(i.e., length, thickness, or other aspects)
when this is being viewed by the surgeon
during fluoroscopy. It includes a series of
notches and holes that clearly show up on
the x-ray image demonstrating the distances
of the viewed bone.
Class A
131 Panoramic digital
X-ray sensor
A sensor used in combination with a
diagnostic digital dental X-ray imaging device
that is specifically designed for panoramic
imaging of teeth, jaw, and oral structures.
The sensor consists of a CCD and other
components, and comprises a sensor drive
circuit and a signal processing circuit.
Class B
132 Panoramic/tomo
graphic dental x-
ray system
(orthopantomogr
aphy system)
application
software
An application or operating data program
designed for use in, or together with a
panoramic/tomographic dental x-ray system
configuration.
Class C
S. No. Device name Intended use Risk Classification 133 Paranasal ultrasound probe A reusable or single-use ultrasound transducer assembly enclosed in a fluid- resistant, acoustically and electrically insulated housing that is intended to be inserted into and positioned in the paranasal sinuses by the operator. The device is also called an endoscopic ultrasound probe or ultrasound endoscopic transducer to be used in the paranasal sinuses. the device is typically incorporated in or used in combination with a device that requires ultrasonography guidance or ultrasonic wave generation, such as an endoscopic device or a needle biopsy device. The device is composed of either a single transducer element or an array of transducer elements (also called piezoelectric elements, active elements, or crystals), damping, backing, and matching materials. Class B 134 PET infusion system An assembly of devices intended to deliver accurate doses of 18F Fluorodeoxyglucose (FDG) radiopharmaceuticals and commonly used flushing solutions to patients during molecular imaging (nuclear medicine) diagnostic procedures. It may also be used to provide effective radiation shielding to medical personnel from Fluorine-18 (18F) radiation exposure during nuclear medicine diagnostic procedures. (Not meant for pediatric populations). Class C 135 PET system An assembly of devices comprising a diagnostic, positron emission tomography (PET) imaging system used to detect, record, quantify and analyse 511 kilo-electronvolt (keV) photon emission patterns resulting from annihilation reactions produced during the decay of positron emitting radiopharmaceuticals. It produces three- dimensional (3-D) tomographic digital cross- sectional physiological images representing distribution patterns of positron emitting radiopharmaceuticals ingested by, or injected into the patient. Class C 136 PET system application software An individual software program or group of programs, routines or algorithms that add specific image processing and/or analysis capabilities to a positron emission tomography (PET) imaging system configuration. A basic set of applications programs and routines is included with such computer-controlled imaging systems and they can be upgraded to correct Class B
S. No. Device name Intended use Risk Classification programming errors or to add new system capabilities. 137 PET/CT system A diagnostic radiological imaging system that is a combination of a positron emission tomography (PET) camera system for nuclear medicine (NM) images, and a computed tomography (CT) camera system for x-ray images. The nuclear medicine images and the x-ray images may be registered and displayed in a fused format (overlaid in the same orientation) for the anatomical localization of the nuclear medicine data (i.e., distribution of radiopharmaceuticals). The PET and CT portions of the system may be used independently or in combination. Class C 138 PET/MRI system A diagnostic radiological imaging system that is a combination of a positron emission tomography (PET) camera system for nuclear medicine (NM) images, and a magnetic resonance imaging (MRI) system for magnetic resonance images. The nuclear medicine images and the magnetic resonance images may be registered and displayed in a fused format (overlaid in the same orientation) for the anatomical localization of the nuclear medicine data (i.e., distribution of radiopharmaceuticals). The PET and MRI portions of the system may be used independently or in combination. Class C 139 Photofluorograph ic x-ray system A photofluorographic x-ray system is a device that includes afluoroscopic x-ray unit and a camera intended to be used to produce, then photograph, a fluoroscopic image of the body. Class C 140 Pneumoencepha lographic chair, non powered A pneumoencephalographic chair is a chair intended to support and position a patient during pneumoencephalography (x-ray imaging of the brain). Class B 141 Positron CT combined SPECT system A system capable of performing both SPECT examination and positron CT (PET) examination Class C
S. No. Device name Intended use Risk Classification 142 Radiographic film cassette/X- ray cassette A radiographic film cassette is a device intended for use during diagnostic x-ray procedures to hold a radiographic film in close contact with an x-ray intensifying screen and to provide a light-proof enclosure for direct exposure of radiographic film Class B 143 Radiographic film/cassette changer A radiographic film/cassette changer is a device intended to be used during a radiologic procedure to move a radiographic film or cassette between x-ray exposures and to position it during the exposure. Class B 144 Radiographic film/cassette changer programmer A radiographic film/cassette changer programmer is a device intended to be used to control the operations of a film or cassette changer during serial medical radiography. Class B 145 Radiologic quality assurance instrument A radiologic quality assurance instrument is a device intended for medical purposes to measure a physical characteristic associated with another radiologic device. Class A 146 Radiological computer aided triage and notification software Radiological computer aided triage and notification software is an image processing software intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. Class C 147 Radiological computer- assisted diagnostic software for lesions suspicious of cancer A radiological computer-assisted diagnostic software for lesions suspicious of cancer is an image processing prescription device intended to aid in the characterization of lesions as suspicious for cancer identified on acquired medical images such as magnetic resonance, mammography, radiography, or computed tomography. The device characterizes lesions based on features or information extracted from the images and provides information about the lesion(s) to the user. Class C 148 Radionuclide dose calibrator A radionuclide dose calibrator is a radiation detection device intended to assay radionuclides before their administration to patients. Class B 149 Radionuclide rebreathing system A radionuclide rebreathing system is a device intended to be used to contain a gaseous or volatile radionuclide or a radionuclide-labeled aerosol and permit it to be respired by the patient during nuclear medicine ventilatory tests (testing process of exchange between the lungs and the atmosphere). Class C
S. No. Device name Intended use Risk Classification 150 Rectal/vaginal ultrasound imaging transducer An invasive component of an ultrasound imaging assembly designed to be positioned within the vagina and/or rectum, either manually or under endoscopic guidance, for ultrasound imaging from within the vagina/rectum. It may be intended for wired or wireless data transfer and may be intended for use with a dedicated ultrasound system or with an off-the-shelf device (e.g., smartphone) with dedicated software installed. This is a reusable device. Class B 151 Software for electrocardiogra ph with ultrasonic diagnostic system A software, which is designed to process data obtained from an electrocardiograph with ultrasound diagnostic system. The resultant data are provided for diagnosis, etc. This term may involve the recording media where the software are stored. Class C 152 SPECT system application software An individual software program or group of programs, routines or algorithms that add specific image processing and/or analysis capabilities to a positron emission tomography (PET) imaging system configuration. A basic set of applications programs and routines is included with such computer-controlled imaging systems and they can be upgraded to correct programming errors or to add new system capabilities. Class B 153 SPECT system, annular detector array An assembly of devices comprising a diagnostic, stationary, single photon emission computed tomography (SPECT) system designed exclusively for use in tomographic imaging applications. It detects, records, quantifies and analyses radionuclide emissions produced during the decay of radiopharmaceuticals or other radiation emitting materials injected into, or ingested by, a patient. Its imaging capabilities include static and dynamic imaging applications and three-dimensional (3-D) digital image reconstruction/display at any angle including axial, coronal and sagittal planes. Class C 154 SPECT system, rotating detector head An assembly of devices comprising a diagnostic single photon emission computed tomography (SPECT) system that is a three- dimensional (3-D) tomographic imaging gamma camera-based system used to detect, record, quantify and analyse radionuclide emissions (primarily gamma rays) produced during the decay of radiopharmaceuticals or other radiation Class C
S. No. Device name Intended use Risk Classification emitting materials injected into, or ingested by, a patient. 155 SPECT/CT system An assembly of diagnostic devices that comprise a radiological imaging system that is a combination of a single photon emission computed tomography (SPECT) camera system for nuclear medicine (NM) images, and a computed tomography (CT) camera system for x-ray images. The nuclear medicine images and the x-ray images may be registered and displayed in a fused format (overlaid in the same orientation) for the anatomical localization of the nuclear medicine data (i.e., distribution of radiopharmaceuticals). The SPECT and CT portions of the system may be used independently or in combination and the images may be transferred to other systems for radiation therapy planning or additional processing. Class C 156 Spot-film device A spot-film device is an electromechanical component of a fluoroscopic x-ray system that is intended to be used for medical purposes to position a radiographic film cassette to obtain radiographs during fluoroscopy. Class B 157 Surgical ultrasound imaging transducer A hand-held ultrasound imaging transducer assembly designed to be positioned within a surgical site for localized intraoperative imaging applications. This group of devices includes ultrasound imaging transducer assemblies used with A-mode, B-mode, M- mode, Doppler, colour Doppler (CD), and duplex (combination imaging, Doppler and/or colour flow) scanning capabilities. This is a reusable device. Class C 158 Thyroid-uptake nuclear medicine system An assembly of diagnostic devices that comprise a non-imaging nuclear medicine system designed primarily for radioactive iodine related thyroid uptake measurement and analysis applications. It typically includes a multichannel analyser, computer, scintillation detector, detector support equipment, an operator's console, video Class B
S. No. Device name Intended use Risk Classification display, well counter detector, collimator, shielding, and application software. 159 Tissue specimen x-ray tomography system An assembly of devices designed to produce an x-ray image of a tissue sample (e.g., stereotactic biopsy cores taken during mammographic procedures) and/or an amputated body part, by generating and controlling x-ray beams and recording the absorption patterns of the x-rays. Class A 160 Transilluminator for breast evaluation/Breas t transilluminator (Diaphanoscope) A transilluminator, also known as a diaphanoscope or lightscanner, is an electrically powered device that uses low intensity emissions of visible light and near- infrared radiation (approximately 700-1050 nanometers (nm)), transmitted through the breast, to visualize translucent tissue for the diagnosis of cancer, other conditions, diseases, or abnormalities. Class C 161 Transoesophage al vascular ultrasound monitor A mains electricity (AC-powered) device intended to be used for monitoring cardiac output and/or fluid status through connection to a dedicated probe, which is placed in the oesophagus to measure blood flow in the descending aorta. It is based on measuring the ultrasonic frequency shift between transmitted and reflected signals (Doppler principle) and is displayed on the monitor as a velocity/time waveform. It is typically used during a surgical procedure or in intensive care. Class B 162 Transoesophage al vascular ultrasound probe A device intended to be inserted into the oesophagus, orally or nasally, to continuously measure blood flow in the descending aorta by ultrasonic Doppler technology for assessment of cardiac output and/or fluid status. It has multiple sensors at the distal end and a connector at the proximal end. It is typically used during general anaesthesia or in intensive care. This is a single-use device. Class B 163 Ultrasonic amplifier A unit used to amplify signals transmitted from the probe of an ultrasound imaging system. This is used with an ultrasound system in the event the quality of original data transfer is lowered by the distance Class B
S.
No.
Device name
Intended use
Risk
Classification
between the console and the imaging probe
in the ultrasound system.
164 Ultrasonic pulsed
doppler imaging
system
An ultrasonic pulsed doppler imaging system
is a device that combines the features of
continuous wave doppler-effect technology
with pulsed-echo effect technology and is
intended to determine stationary body tissue
characteristics, such as depth or location of
tissue interfaces or dynamic tissue
characteristics such as velocity of blood or
tissue motion.
Class B
165 Ultrasonic pulsed
echo imaging
system
An ultrasonic pulsed echo imaging system is
a device intended to project a pulsed sound
beam into body tissue to determine the depth
or location of the tissue interfaces and to
measure the duration of an acoustic pulse
from the transmitter to the tissue interface
and back to the receiver.
Class B
166 Ultrasound
aspiration biopsy
procedure kit
A collection of sterile devices that includes a
biopsy needle and other supplies used to
perform an ultrasound-guided aspiration
sampling of tissue. The needle tip must
provide a clearly visible image during
ultrasound imaging. This is a single-use
device.
Class B
167 Ultrasound
contrast medium,
transuterine
administration
A substance/formulation (e.g., solution, gel,
foam) intended to be administered into the
female reproductive tract, via the
vagina/uterus, to enhance diagnostic
ultrasound images, typically for the evaluation
of the uterine cavity and/or fallopian tube
patency. (Invasive devices for application not
included)
Class A
168 Ultrasound
endoscope
balloon, Hevea-
latex
A sterile balloon-like cap made of Hevea
natural rubber latex (NRL) designed to be
placed on the distal end of an ultrasound
endoscope and inflated to make contact with
the wall of the viscus to enable a higher-
quality ultrasound image during echo-
endoscopy, typically of the gastrointestinal or
respiratory tract. It is inflated with water
through the endoscope.
Class B
S.
No.
Device name
Intended use
Risk
Classification
169 Ultrasound
endoscope
balloon,
synthetic
polymer
A sterile balloon-like cap made of a synthetic-
polymer (e.g., silicone) designed to be placed
on the distal end of an ultrasound endoscope
and inflated to make contact with the wall of
the viscus to enable a higher-quality
ultrasound image during echo-endoscopy,
typically of the gastrointestinal or respiratory
tract. It is inflated with water through the
endoscope.
Class B
170 Ultrasound
imaging system
application
software
An individual software program or group of
programs, routines or algorithms that add
specific image processing and/or analysis
capabilities to a diagnostic ultrasound system
configuration. A basic set of applications
programs and routines are included with such
computer-controlled imaging systems and
they can be upgraded to correct
programming errors or to add new system
capabilities.
Class B
171 Ultrasound
imaging system
multi-transducer
connection unit
An electronic device intended to be used with
an ultrasound (US) imaging system to enable
the simultaneous connection of multiple
hand-held US transducers so the user can
conveniently switch between transducers as
required. It is a unit with multiple transducer
ports and is typically attached to the US
imaging system stand.
Class B
172 Ultrasound
imaging system
operation
software
Mainframe or personal computer (PC) based
operating system software (or firmware)
specific to the central processing unit (CPU)
incorporated into a diagnostic ultrasound
system configuration. This includes the
software or firmware based programs and
routines supplied by the computer
manufacturer that drive a specific computer in
the performance of its tasks and assists the
operators, applications programmers and
programs with various supporting functions.
Class B
173 Ultrasound
imaging system
operator console
A device that is a component of a diagnostic
ultrasound imaging system that functions as
the primary control panel. It includes
hardware and software that allows for image
display, processing, analysis, archiving and
retrieval. It typically includes the capabilities
to be interfaced into a picture archiving and
communication system (PACS), a local area
network (LAN), a radiology information
system (RIS) or a hospital information system
(HIS). It contains the primary controls for the
direct operation of the ultrasound system and
is typically incorporated in the system design
Class B
S. No. Device name Intended use Risk Classification as is the case with mobile or portable ultrasound systems. 174 Ultrasound imaging system tracking positioner, reusable (ultrasound transducer bracket) A metal or plastic bracket intended to be applied to a hand-held ultrasound (US) transducer for the repeatable positioning (detach from/reattach to same place) of an electromagnetic (EM) sensor and/or needle guide on the transducer. It may be used in conjunction with an EM device tracking system and is intended for applications involving needle guidance and tracking such as ablation, core tissue biopsy, fluid aspiration, therapeutic delivery and vascular access. (transducer and sensor are not included) Class B 175 Ultrasound imaging transducer positioning unit A unit used to position an ultrasound system imaging transducer that is inserted into the body through an endoscope. This device is an accessory to a diagnostic ultrasound imaging system and it transmits electrical signals that, after processing, will display the position of a the transducer on a monitor. This unit will assist the endoscopist in the placement of the working transducer (the probe) and to allow monitoring of the movement after placement. Class B 176 Ultrasound imaging transducer washer- disinfector (chemical) A dedicated electrically-powered unit intended to be used to wash and disinfect ultrasound imaging transducers with the use of appropriate cleaning methods and chemical disinfection. It is typically a stand- alone unit with a washing/disinfecting chamber designed to accept an ultrasound imaging transducer (probe). Not intended for end disinfection. Class B
S. No. Device name Intended use Risk Classification 177 Ultrasound system synchronizer A physiological monitoring device used as one component in a diagnostic ultrasound system. This device produces signals which can synchronize imaging acquisition and data collection with specific measurable, physiological parameters, such as the starting point of a patient's respiratory or cardiac cycle. It is used mainly to reduce artifacts and to increase the signal-to-noise ratio in real-time ultrasonic imaging and in video imaging. Class B 178 Ultrasound- guided laser thermal therapy needle guide positioning system An assembly of noninvasive devices intended for use during soft-tissue lesion treatment by ultrasound-guided laser thermal therapy, to provide software-based simulation of tissue thermal damage for assisting the manual positioning of percutaneous needles through which laser applicators (not included) are introduced. It consists of an electrically- powered interface/display and needle guides for attachment to a body-surface ultrasound (US) imaging probe/transducer; the system overlays real-time US images [from a separate US system] with the predicted laser thermal effect Class C 179 Vascular ultrasound system control unit A mains electricity (AC-powered) external device, which may include internal rechargeable batteries, intended to be used to indicate (visually and/or sonically) blood flow in a peripheral blood vessel, through connection to a dedicated probe/catheter (not included), using ultrasound/Doppler/transit time technology. It may be intended to locate and assess the extent of vascular flow restriction (e.g., clots, stenosis, mechanical damage) or to monitor blood flow in association with microvascular surgery. Class B 180 X-ray film cassette, automatic film changing A housing for x-ray film specifically designed to be used in, and work as part of, an automated x-ray film changing system. The housing is intended to be loaded with x-ray film under appropriate darkroom conditions, and is then inserted into the loading magazine of the film changing system. After exposure, it is held in a receiving magazine until the operator removes it for final processing. Class B 181 X-ray film cassette, manual A device used in medical imaging applications to shield x-ray film from exposure to room light during transport and insertion into a diagnostic imaging system, film formatter, or film processor. Class B
S. No. Device name Intended use Risk Classification 182 X-ray film ruler A device used in radiology to superimpose accurate measurements of length (i.e., inches or millimetres) onto an x-ray film so that these measurements remain as a permanent feature of that x-ray against the anatomical structure that has been captured. It is made of x-ray translucent material (e.g., plastic) whilst the graduations are not translucent and therefore show on the exposed x-ray film. Class A 183 X-ray image intensifier A device that converts an x-ray image into a light image, then to an electronic image and finally back to a light-based image of diminished size and increased brightness. It is primarily used in fluoroscopic, angiographic, cine-fluorographic x-ray, and digital x-ray applications as it outputs an image small enough to be coupled to cine, television or spot-film cameras. Class B 184 X-ray system computer, diagnostic, general-purpose A dedicated mainframe computer, personal computer (PC) or PC based platform and associated hardware, firmware, and operating system software used specifically to control and monitor the operation of a general-purpose medical x-ray system and associated image processing, display and analysis functions. Class C 185 X-ray system synchronizer A physiological monitoring unit used as a component of a diagnostic x-ray system, e.g., fluoroscopy, angiography or general-purpose, that produces a signal which enables image formation or data collection to be synchronized with a specific measurable physiological parameter, e.g., the beginning of a patient's respiratory or cardiac cycle. It is primarily used for the purpose of artifact reduction or enhancement of signal-to-noise ratios in various real-time or dynamic diagnostic x-ray applications. It is sometimes referred to as a gated imaging accessory or trigger. Class C 186 X-ray/CT combined cardiovascular diagnostic X-ray system A system combining a diagnostic X-ray computed tomography (CT) and a diagnostic cardiovascular fluoroscopy system. Class C
Name of Category:
Radiotherapy
Total No. of Devices:
114
Category: Radiotherapy
S.No.
Device name
Intended use
Risk Class
1 Absorbable perirectal
spacer
An absorbable perirectal spacer is
composed of biodegradable material
that temporarily positions the anterior
rectal wall away from the prostate
during radiotherapy for prostate cancer
with the intent to reduce the radiation
dose delivered to the anterior rectum.
Class C
2 Absorbable tissue spacer
for radiotherapy
An absorbable material used to
reduce radiation exposure of normal
tissue during radiotherapy by
implanting surgically or percutaneously
between tissue, internal organs, etc.,
to make a space between the
malignant tumor and normal tissue.
Class D
3 Applicator for bile duct
manual brachytherapy
A manual brachytherapy applicator
specifically designed for bile duct
radiation therapy. An applicator designed
to have a configuration that facilitates
manual placement (puncture or
placement and removal using an
endoscope or a diagnostic imaging
system) of single or multiple therapeutic
radiation sources in treatment sites in
the bile duct.
Class C
4 Applicator for bile duct
remote after loading
brachytherapy
A remote controlled brachytherapy
applicator specifically designed for bile
duct radiation therapy. It is designed to
be temporarily implanted in the bile duct.
It serves as a guide for computer-
controlled temporary placement and
removal of a single or multiple
therapeutic radiation sources at
treatment sites.
Class C
5 Applicator for bladder
designed to facilitate manual placement
(puncture or placement and removal
using an endoscope or a diagnostic
imaging system) of single or multiple
therapeutic radiation sources in
treatment sites in the bladder.
Class C
S.No.
Device name
Intended use
Risk Class
6 Applicator for bladder
brachytherapy
A remote controlled
brachytherapy applicator specifically
designed for bladder radiation therapy. It
is designed to be temporarily implanted
in the bladder. It serves as a guide for
computer-controlled temporary
placement and removal of a single
or multiple therapeutic radiation
sources at treatment sites.
Class C
7 Applicator for
brachytherapy non-
central circulatory
general-purpose manual
A general-purpose brachytherapy
applicator used to facilitate
radiotherapy. A single or module
applicator designed to facilitate manual
placement (puncture, local placement,
placement under endoscopy, and
placement and removal using an image
diagnostic system) of single or multiple
therapeutic radiation sources in
treatment sites in the non-central
circulatory system.
Class C
8 Applicator for
brachytherapy non-
central circulatory
general-purpose remote
after loading
A general-purpose remote controlled
brachytherapy applicator used to
facilitate radiotherapy. It is designed to
be temporarily implanted in the body. It
serves as a guide for computer-
controlled temporary placement and
removal of a single or multiple
therapeutic radiation sources at
treatment sites in the non-central
circulatory system.
Class C
9 Applicator for bronchial
applicator
specifically designed for temporarily use
in bronchial radiation therapy. A single
or module applicator designed to
facilitate manual placement (placement
using an endoscope or positioning,
placement and removal using a
diagnostic imaging system) of single
or multiple therapeutic radiation
sources in treatment sites. It may be
designed to be standard in
configuration or to handle specific
radiation sources.
Class C
S.No.
Device name
Intended use
Risk Class
10 Applicator for bronchial
brachytherapy
designed for bronchial radiation therapy.
It is designed to be temporarily
implanted in the body. It serves as a
guide for computer-controlled
temporary placement and removal of
a single or multiple therapeutic
radiation sources in the bronchus.
This device group includes various
applicators such as hollow needles,
tubes, and catheters, as well as
associated devices and connectors.
Class C
11 Applicator for
cervical/endometrial
specifically designed for uterine cervix
or intrauterine radiation therapy. A
single or module applicator designed to
facilitate manual placement (puncture,
placement with an endoscope or a
diagnostic imaging system) of
single or multiple
therapeutic radiation sources in
treatment sites.
Class C
12 Applicator for
Cervical/endometrial
brachytherapy
A remote-controlled brachytherapy
applicator specifically designed for
uterine, cervical, or intrauterine radiation
therapy. It is designed to be temporarily
implanted in the body. It serves as a
guide for computer-controlled temporary
placement and removal of a single or
multiple therapeutic radiation sources in
the uterus, cervix and endometrium.
Class C
13 Applicator for esophagus
specifically designed for esophagus
radiation therapy. A single or module
applicator designed to facilitate
manual placement (puncture or
placement and removal using an
endoscope or a diagnostic imaging
system) of single or multiple therapeutic
radiation sources in treatment sites in
the esophagus. It may be designed to be
standard in configuration or to handle
specific radiation sources.
Class C
S.No.
Device name
Intended use
Risk Class
14 Applicator for esophagus
brachytherapy
designed for esophagus radiation
therapy. It is designed to be
temporarily implanted in the
esophagus. It serves as a guide for
computer-controlled temporary
placement and removal of a single or
multiple therapeutic radiation sources.
This device group includes various
applicators such as hollow needles,
tubes, and catheters, as well as
associated devices and connectors.
Class C
15 Applicator for eye
A template with a groove on the one
side. The groove shows the position of
the brachytherapy source that is
manually, temporarily delivered to the
eye surface. The other side is shielded.
Class C
16 Applicator for eye remote
after loading
brachytherapy
designed for eye radiation therapy. It
serves as a guide for computer-
controlled temporary placement and
removal of a single or multiple
therapeutic radiation sources. This
device group includes various
applicators such as hollow needles,
tubes, and catheters, as well as
associated devices and connectors.
Class C
17 Applicator for
nasopharynx manual
brachytherapy
specifically designed for
nasopharyngeal radiation therapy. A
single or module applicator
designed to facilitate manual
placement (puncture, endoscopic
placement or placement and removal
using a diagnostic imaging system) of
single or multiple therapeutic
radiation sources in treatment sites in
the nasopharynx.
Class C
S.No.
Device name
Intended use
Risk Class
18 Applicator for
nasopharynx remote
after loading
brachytherapy
designed for nasopharyngeal
radiation therapy. It is designed to be
temporarily implanted in the body. It
serves as a guide for computer-
controlled temporary placement and
removal of a single or multiple
therapeutic radiation sources in the
nasopharynx.
Class C
19 Applicator for neck
specifically designed for neck radiation
therapy. A single or module applicator
designed to facilitate manual placement
(puncture, local placement or placement
and removal using a diagnostic imaging
system) of single or multiple
therapeutic radiation sources in
treatment sites in the neck tissues.
Class C
20 Applicator for neck
brachytherapy
designed for neck radiation therapy. It
is designed to be temporarily implanted
in the neck tissues. It serves as a guide
for computer-controlled temporary
placement and removal of a single
or multiple therapeutic radiation
sources.
Class C
21 Applicator for pancreas
specifically designed for pancreatic
radiation therapy. A single or module
applicator designed to facilitate
manual placement (puncture,
endoscopic placement, or placement
and removal using a diagnostic imaging
system) of single or multiple therapeutic
radiation sources in treatment sites in
the pancreas. It may be designed to be
standard in configuration or to handle
specific radiation sources.
Class C
S.No.
Device name
Intended use
Risk Class
22 Applicator for pancreas
brachytherapy
designed for pancreatic radiation
therapy. It is designed to be temporarily
implanted in the pancreas. It serves as a
guide for computer-controlled temporary
placement and removal of a single
or multiple
therapeutic radiation sources at
treatment sites.
Class C
23 Applicator for prostate
specifically designed for prostate
radiation therapy. A single or
module applicator designed to
facilitate manual placement (puncture
or placement or removal with a trigger
loading device, an endoscope or a
diagnostic imaging system) of single or
multiple therapeutic radiation sources
in treatment sites in the prostate gland. It
may be designed to be standard in
configuration or to handle specific
radiation sources.
Class C
24 Applicator for prostate
brachytherapy
designed for prostate radiation therapy.
It is designed to be temporarily
implanted in the prostate gland. It
serves as a guide for computer-
controlled temporary placement and
removal of a single or multiple
therapeutic radiation sources.
Class C
25 Applicator for rectal/anal
specifically designed for rectal and/or
anal radiation therapy. A single or
module applicator designed to
facilitate manual placement (puncture
or placement and removal using an
endoscope or a diagnostic imaging
system) of single or multiple therapeutic
radiation sources in treatment sites in
the rectum and/or anus.
Class C
S.No.
Device name
Intended use
Risk Class
26 Applicator for rectal/anal
brachytherapy
designed for rectal or anal radiation
therapy. It is designed to be temporarily
implanted in the rectum or anus. It
serves as a guide for computer-
controlled temporary placement and
removal of a single or multiple
therapeutic radiation sources.
Class C
27 Applicator for tongue
specifically designed for lingual
radiation therapy. A single or
module applicator designed to
facilitate manual placement (puncture,
local placement or placement and
removal using a diagnostic imaging
system) of single or multiple therapeutic
radiation sources in treatment sites in
the tongue and the surrounding tissues.
It may be designed to be standard in
configuration or to handle specific
radiation sources.
Class C
28 Applicator for tongue
brachytherapy
designed for tongue or oral cavity
radiation therapy. It is designed to be
temporarily implanted in the tongue or
the surrounding tissues. It serves as a
guide for computer-controlled
temporary placement and removal of
a single or multiple therapeutic
radiation sources.
Class C
29 Applicator for vaginal
specifically designed for vaginal or
transvaginal radiation therapy. A single
or module applicator designed to
facilitate manual placement
(puncture, local placement,
endoscopic placement or placement and
removal using a diagnostic
imaging system) of single or
multiple therapeutic radiation sources in
treatment sites.
Class C
S.No.
Device name
Intended use
Risk Class
30 Applicator for vaginal
brachytherapy
designed for vaginal or transvaginal
radiation therapy. It is designed to be
temporarily implanted in the body. It
serves as a guide for computer-
controlled temporary placement and
removal of a single or multiple
therapeutic radiation sources in the
vagina.
Class C
31 Blood vessel manual
brachytherapy applicator
designed exclusively for blood vessel
radiotherapy. Most commonly, it is used
to prevent formation of plaque, stenosis
and restenosis in blood vessels after
surgery. A single or module device
designed to facilitate manual placement
(placement and removal under
endoscopy or using an image diagnostic
system) of single or multiple
therapeutic radiation sources in the
treatment site.
Class D
32 Blood vessel remote
after loading
brachytherapy applicator
A remote controlled brachytherapy
applicator designed exclusively for
radiotherapy in blood vessels. This
treatment is used to prevent plaque
formation. and stenosis in blood vessels
after surgery. It is designed for
temporary implantation in a blood vessel,
and serves as a guide for computer-
controlled temporary placement and
removal of single or multiple therapeutic
radiation sources at treatment sites.
Class D
33 Brachytherapy needle
A sterile, sharp bevel-edged, hollow
tubular metal instrument that is used to
inject radionuclide into a body cavity or
tissue as a source of nuclear
radiation for cancer therapy
(brachytherapy).
Class B
34 Brachytherapy
radionuclide phantom,
anthropomorphic
A device that consists of preserved
human or animal tissue, or a two or
three-dimensional (3-D) tissue-
equivalent model designed to simulate
the functional, physical, or a
combination of these characteristics of
normal or diseased human organs.
Class B
S.No.
Device name
Intended use
Risk Class
35 Brachytherapy system
chair
A mains electricity (AC-powered)
device (a chair or stool) that is a
component of a brachytherapy system
and which is specifically designed
to support and position a patient
during brachytherapy radiation
treatments given by either a manual
applicator or a
remote after loading brachytherapy
system applicator.
Class B
36 Brachytherapy system
source safe
A component of a remote after
loading brachytherapy system
consisting of a shielded vault, and
associated source retraction and
extrusion mechanisms, alarms, and
related mechanical, electronic and
software controls, used to shield the
brachytherapy sources in order to
protect system operators,
brachytherapy patients and others from
the continuous emissions of the
radioactive brachytherapy source(s)
when they are not in use.
Class C
37 Brachytherapy system
source transfer tube
The transfer tube, when attached to the
applicator and the after loading system,
provides a continuous closed passage
that allows for moving either a
radioactive source(s) and/or positioning
markers from the shielded source
storage compartment of the remote after
loading brachytherapy system into
appropriate positions within a
brachytherapy applicator that has been
positioned at a location either on the
surface of, or within, the patient.
Class C
38 Brain manual
brachytherapy applicator
An applicator specifically
designed for brain radiotherapy. A
single or module applicator designed to
facilitate manual placement (puncture,
placement under endoscopy, or
placement and removal using an image
diagnostic system) of single or
multiple therapeutic radiation sources in
the brain.
Class D
S.No.
Device name
Intended use
Risk Class
39 Brain remote after
loading brachytherapy
applicator
A remote controlled brachytherapy
applicator designed exclusively for
brain radiotherapy for temporary
implantation in the body. It serves as
a computer- controlled guide for
temporary placement and removal
of a single or multiple therapeutic
radiation sources in the brain.
Class D
40 Breast brachytherapy
system applicator,
manual
specifically designed for use in radiation
therapy treatments of the breast.
Class C
41 Breast brachytherapy
system applicator,
remote-afterloading
A sterile, remote-afterloading
brachytherapy applicator specifically
designed for use in radiation therapy
treatments of the breast. It is typically
designed for temporary implantation
within the breast and serves as a guide
for computer-controlled placement and
removal of single or multiple radioactive
sources.
Class C
42 Breast ductography
cannula
A thin, sterile, semi-rigid or rigid
metal tube that is inserted into the
nipple of the female breast to inject a
contrast medium into the lactiferous
ducts to enhance their visualization
during a radiographic procedure.
Class C
43 Central circulatory
general-purpose manual
brachytherapy applicator
A single or module applicator
designed to facilitate manual
placement (puncture, local
placement, placement under endoscopy
or placement and removal using an
image diagnostic system) of single or
multiple therapeutic radiation sources
in treatment sites in the central
circulatory system.
Class D
44 Central circulatory
general-purpose remote
after loading
brachytherapy applicator
Intended to be temporarily implanted
in the body. It serves as a guide for
computer-controlled temporary
placement and removal of a single
or multiple therapeutic radiation
sources at treatment sites in the
central circulatory system.
Class D
S.No.
Device name
Intended use
Risk Class
45 Central circulatory
therapeutic radionuclide
system
A device that places a radiation
source manually or automatically at
the treatment site in the central
circulatory system for providing a
required radiation dose during
radiotherapy. This device does not equip
a
remotely controlled radiation source
transporter.
Class D
46 Central circulatory
permanent implant
source
A device for the central circulatory
system to be placed permanently in the
body for radiotherapy which is
necessary for treatment and
symptomatic therapy, and uses natural
radioisotopes or radioisotopes produced
by an accelerator or a nuclear
reactor. The radiation source, which
is permanently placed manually, is
designed to achieve compatibility with
tissues. The radiation source can be
selected from the following forms –
e.g., microsphere, globe, stent, seed,
and wire- in order to generate low-
energy photons, beta particles, or alpha
particles.
Class D
47 Central circulatory
brachytherapy
source
A device for the central circulatory
system used as radiation source to
deliver a high or low dose rate with an
after-loading brachytherapy device
designed for radiotherapy which is
necessary for treatment and
symptomatic therapy, and uses natural
radioisotopes or radioisotopes produced
by an accelerator or a nuclear reactor.
Class D
48 Central circulatory
brachytherapy
system
A device that places a radiation source
temporarily at the treatment site in the
central circulatory system for providing
a required radiation dose during
radiotherapy. This device equips a
remotely controlled radiation source
transporter.
Class D
S.No.
Device name
Intended use
Risk Class
49 Central circulatory
temporary placement
source
A device for the central circulatory
system that uses natural radioisotopes
or radioisotopes produced by an
accelerator or a nuclear reactor, and
is placed in the body temporarily, and
removed after the pre-determined
treatment period. The radiation
source, which is temporarily inserted
manually, is supplied in various forms
– e.g., encapsulated, sealed, plated,
foiled, or embedded.
Class D
50 Compact
thermoluminescent
dosimetry electrometer
Thermoluminescence dosimeter (TLD)
is used to measure the radiation
dose emitted to the phantom, eyes
and other organs with high radio
sensitivity.
Class A
51 Conformal
Brachytherapy Source
The intended use of the device is for the
treatment of cancer by temporary
intraoperative or surface irradiation.
The device contains radioactive material
with activity up to 200 mci and is
indicated for treatment of temporary
intraoperative, interstitial, intracavitary
or surface application to treat selected
localized tumors.
Class C
52 Coronary artery
brachytherapy system
applicator, manual
A sterile flexible tube intended to
deliver/remove radiation therapy sources
into a coronary artery, typically into the
lumen of an implanted stent, as part of a
manual afterloading brachytherapy
system
Class D
53 Coronary artery
applicator, remote
afterloading
A sterile flexible tube intended to
deliver/remove radiation therapy sources
into a coronary artery, typically into the
lumen of an implanted stent, as part of a
remote-controlled afterloading
Class D
S.No.
Device name
Intended use
Risk Class
54 Fluoroscopy/angiography
table, powered
A powered table designed to support
and position a patient during
fluoroscopy/angiography typically in
conjunction with a cardiovascular
intervention or other minimally-invasive
surgical intervention (e.g., biopsy). It has
a flat table top with electronic controls for
table top positioning and/or motion
relative to the x-ray beam. It is not
dedicated to breast biopsy, and is not
intended for other specialized surgeries
(i.e., it is not a specialized operating
table with x-ray transparent features).
Class B
55 Gamma knife for
radiotherapy/surgery
A set of devices that are designed to
uses very precise beams of gamma rays
to treat an area of disease (lesion) or
growth (tumor), especially in the brain,
upper spine and in certain cases,
vascular abnormalities.
Class D
56 General-purpose
applicator, manual
A general-purpose brachytherapy
applicator designed to facilitate delivery
of radiation therapy treatments that are
not specific to a particular anatomical
region. It is an individual or modular
device designed to facilitate manual
placement, e.g., puncture, topical
placement, endoscopically guided
placement or diagnostic imaging system
guided placement, and removal of single
or multiple radioactive sources at a
treatment site.
Class C
57 General-purpose
applicator, remote-
afterloading
A general-purpose remote-afterloading
brachytherapy applicator intended to be
used to facilitate the delivery of radiation
therapy treatments in a range of
anatomical regions (e.g., liver, kidneys,
lungs, gastrointestinal tract). It is
designed for temporary introduction into
the body and functions as a guide for
computer-controlled placement and
removal of single or multiple radioactive
sources in a treatment area.
Class C
S.No.
Device name
Intended use
Risk Class
58 High-frequency
hyperthermia system
A system used to generate high-
temperatures and to control the
provision of heat to the body in the
treatment of malignant and benign
tumors, or other diseases.
Class C
59 Interventional radiology
percutaneous-access kit
A collection of various
instruments/devices (e.g., introducer
sheath, dilator, guides, cannula, needle,
angiographic catheter) intended to be
used for percutaneous body-access to
facilitate a radiological image-guided
interventional radiology or surgical
procedure [e.g., bile duct biopsy,
percutaneous transhepatic
cholangiogram, non-vascular drainage
procedures, ultrasound-guided spinal
surgery]. Some types are designed to
function as a channel to facilitate the
introduction of radiological image
guidance devices (e.g., ultrasound
probe/transducer) [not included]. It does
not contain pharmaceuticals.
Class C
60 Intra-vaginal organ
positioning device for
diagnostic imaging and
radiotherapy
A device that is specifically designed to
be inserted in the vagina to properly
position and fix the surrounding organs
such as uterine cervix, rectum, and
urinary bladder for image diagnosis or
radiotherapy. This device is used to
facilitate reproducible positioning for
continuous image examination or
continuous radiotherapy.
Class B
61 Laser irradiation therapy
kit
A kit includes a puncture needle, a
guide wire, and a guiding sheath for
guiding probes (used for laser
irradiation therapy, for example).
Not all of the components are
included; in some products, two or more
of components are integrated into one.
Class B
62 Living tissue
radiotherapy system
A low energy X-ray therapy system
designed to treat adjacent tumor
lesions with high dose X-rays by placing
soft X-ray beams from 5 to 50 kV inside
the tumor tissue. It is used in both
intraoperative radiation and stereotactic
localized radiation therapy.
Class C
S.No.
Device name
Intended use
Risk Class
63 Manual brachytherapy
source, permanent
implant
A device that is a naturally-occurring,
accelerator- or reactor-produced
radioactive isotope and intended to be
permanently implanted into the body in
order to deliver curative or palliative
radiotherapy treatments. Manually
implanted sources intended for
permanent implantation are designed to
be histocompatible. It may take the form
of, e.g., microspheres, spheres, stents,
seeds or wires selected for producing
low energy photons, beta or alpha
particles.
Class D
64 Manual brachytherapy
source, temporary
placement
A device that is a naturally-occurring,
accelerator- or reactor-produced
radioactive isotope in a form that is
intended for temporary placement in the
body and removal after a specified
treatment time. It is used in
brachytherapy applications that require
direct manual or endoscopically-guided
placement and removal. Manually
inserted sources intended for temporary
placement come in a variety of forms,
e.g., encapsulated, sealed, plated, foil or
embedded sources.
Class C
65 Manual brachytherapy
system
An assembly of therapeutic, independent
devices, manually brought together and
intended to deliver a palliative or
therapeutic radiation dose to an
anatomical region from radionuclide
sources. Medical personnel manually
place multiple radioactive sources in
appliances and applicators for either
permanent or temporary delivery at the
topical, interstitial, intraluminal, or
intracavitary brachytherapy treatment
site.
Class C
66 Manual-afterloading
source delivery
procedure kit
A collection of sterile devices intended to
dispense/collect hydraulic fluid used to
deliver the radiation source train of a
manual-afterloading brachytherapy
system source transfer device to/from
the delivery catheter/applicator. It
consists of disposable devices such as
syringes, fluid collection bags,
adaptors/connectors.
Class B
S.No.
Device name
Intended use
Risk Class
67 Manual-afterloading
storage container
A non-sterile radiation-shielding plastic
box intended to store a manual-
afterloading brachytherapy system. It is
typically temporarily used as a safety
container to prevent radiation leakage in
the event of procedural difficulties.
Class C
68 Medical charged- particle
radiation therapy system
A medical charged-particle radiation
therapy system is a device that
produces by acceleration high
energy charged particles (e.g., electrons
and protons) intended for use in
radiation therapy.
Class C
69 Medical neutron radiation
therapy system
A medical neutron radiation therapy
system is a device intended to generate
high-energy neutrons for radiation
therapy
Class C
70 MOSFET radiation
therapy dosimetry
system
An assembly of devices using
metal oxide semiconductor field-
effect transistor (MOSFET)
technology intended to be used for on-
the-spot patient or anthropomorphic
radiation dose verification and
monitoring during radiation therapy
and radiology procedures.
Applications typically include radiation
oncology therapy and dosimetry,
treatment plan verification for in
vivo dosimetry, brachytherapy,
intraoperative radiation therapy, image-
guided radiation therapy, and research.
Class B
71 Non-central circulatory
system
A device that places a radiation
source manually or automatically at
the treatment site in the non-central
circulatory system for providing a
required radiation dose during
radiotherapy.
Class C
72 Non-central circulatory
permanent implant
source
A non-central cardiovascular
device which is histocompactible
and containing an isotope naturally
occurring or produced by an
accelerator or a nuclear reactor,
intended to be permanently implanted
in the body for radiation therapy
requiring treatment or symptomatic
treatment.
Class C
S.No.
Device name
Intended use
Risk Class
73 Non-central circulatory
brachytherapy
source
A device for the non-central circulatory
system used as radiation source to
deliver a high or low dose rate with an
after-loading brachytherapy device
designed for radiotherapy which is
necessary for treatment and
symptomatic therapy, and uses natural
radioisotopes or radioisotopes produced
by an accelerator or a nuclear
reactor.
Class C
74 Non-central circulatory
brachytherapy
system
A device that places a radiation source
temporarily at the treatment site in the
non-central circulatory system for
providing a required radiation dose
during radiotherapy. This device
equips a remotely controlled
radiation
source transporter.
Class C
75 Non-central circulatory
temporary placement
source
A non-central cardiovascular device
containing an isotope naturally
occurring or produced by an
accelerator or a nuclear reactor,
intended to be temporarily implanted
in the body and to be removed after
a prescribed duration of treatment.
Used in brachytherapy, the device
is placed and removed
manually or under endoscopic
observation.
Class C
76 Patient positioning
device for breast
diagnostic imaging and
radiotherapy
A device that is specifically designed to
properly position and fix a female
patient's breasts and chest for image
diagnosis, image-guided surgery,
interventional therapy, or radiotherapy.
Class B
77 Patient positioning
device for extremity
diagnostic imaging and
radiotherapy
A device that is specifically designed to
properly position and fix a patient's arms
and legs for image diagnosis, image-
guided surgery, interventional therapy, or
radiotherapy.
Class B
78 Patient positioning
device for pelvis
diagnostic imaging and
radiotherapy
The device that consists of frames,
plates, or other parts, and is specifically
designed to properly position and fix the
patient's abdomen and pelvic region
for image diagnosis, image-guided
surgery, interventional therapy,
or radiotherapy.
Class B
S.No.
Device name
Intended use
Risk Class
79 Patient positioning
device for whole body
diagnostic imaging and
radiotherapy
A device that consists of fixed or
adjustable parts (e.g., frames and
plates), and is specifically designed to
properly position and fix the patient's
whole body for image diagnosis,
image-guided surgery, interventional
therapy, or radiotherapy.
Class B
80 Patient Positioning
System, Ultrasound
An assembly of devices used to locate,
with ultrasound, internal soft-tissue
anatomy that moves relative to
external or bony landmarks, to enable
subsequent adjustment of the patient
for precise external beam radiation
treatment of the target tissue. It
typically includes an ultrasound
imaging system, computerized
workstation(s), optical tracking devices,
and dedicated software.
Class C
81 Post Breast Biopsy
Hemostatic Breast
Compression Device
Intended to achieve and maintain
hemostasis of a breast biopsy wound
site.
Class C
82 Powered neutron therapy
table
A programmable bed for radiotherapy
designed to adjust the patient's posture
and immobilize the patient for
treatment that uses neutron rays that are
generated from the nuclear reactor, etc.
Class B
83 Powered patient table for
accelerator
A bed operates by programmable
for electric radiotherapy designed to
adjust the patient's posture and
immobilize the patient for radiotherapy
that uses medical linear accelerator or
non-linear accelerator.
Class B
84 Powered radiation
therapy patient support
assembly
A powered radiation therapy patient
support assembly is an electrically
powered adjustable couch intended to
support a patient during radiation
therapy
Class C
S.No.
Device name
Intended use
Risk Class
85 Powered radionuclide
brachytherapy table
A programmable bed for radiotherapy
designed to adjust the patient's posture
and immobilize the patient for
treatment that uses an after loading
short-distance irradiation treatment
apparatus that is operated
manually or electrically.
Class B
86 Powered remote
irradiation therapy table
A programmable electrically
operated bed for radiotherapy
designed to adjust the patient's
posture and immobilize the patient for
treatment that uses a remote cobalt
60 radiotherapy apparatus and other
remote radionuclide radiotherapy
apparatuses.
Class B
87 Powered X-rays radiation
therapy table
A programmable electrically
operated bed for radiotherapy
designed to adjust the patient's
posture and immobilize the patient for
treatment that uses an X-ray therapy
apparatus.
Class B
88 Radiation therapy beam-
shaping block
A radiation therapy beam-shaping
block is a device made of a highly
attenuating material (such as lead)
intended for medical purposes to modify
the shape of a beam from a radiation
therapy source
Class C
89 Radiation therapy digital
imager
An automated device that is typically
mounted on the gantry of a linear
accelerator and intended to produce
digital images of x-rayed anatomical
landmarks to guide radiation treatment
(e.g., tracking/targeting tumours). The
device may be a digital imaging panel
(e.g., of silicon) on robotic arms; it may
also have an x-ray source to generate
higher quality images.
Class C
90 Radiation therapy
simulation system
A radiation therapy simulation system is
a fluoroscopic or radiographic x-ray
system intended for use in
localizing the volume to be exposed
during radiation therapy and confirming
the position and size of the
therapeutic irradiation field produced.
Class C
S.No.
Device name
Intended use
Risk Class
91 Radiological image
marker, implantable
A device intended to be implanted within
the body, either temporarily or
permanently, to create identifying marks
that can be seen on radiographic film or
digital images. It is typically in the form of
a wire, needle, bead, clip, stent-like tube,
washer or fluid, and used to locate and
delineate a tumour, lesion, or other site
of interest. It is made from materials
compatible with the imaging system with
which it is intended to be used [e.g.,
magnetic resonance imaging (MRI), x-
ray, or nuclear medicine]; it may be
intended for use during radiotherapy
procedures.
Class C
92 Radionuclide
brachytherapy source
A radionuclide brachytherapy source
is a device that consists of a
radionuclide which may be enclosed in a
sealed container made of gold, titanium,
stainless steel, or platinum and
intended for medical purposes to be
placed onto a body surface or into a
body cavity or tissue as a source of
nuclear radiation for therapy
Class C
93 Radionuclide dynamic
function testing
equipment
A device used to measure and
record temporal variations of
radioisotope concentrations in the
body. Specialized devices, such as
devices for thyroid uptake measurement,
renograms, and radioisotope blood
volume measurement, are included.
Class B
94 Radionuclide radiation
therapy system
A radionuclide radiation therapy
system is a device intended to permit
an operator to administer gamma
radiation therapy, with the radiation
source located at a distance from the
patient's body.
Class C
95 Radionuclide source for
remote irradiation
therapy
Radiation sources generated in a reactor
and used as in a remote after loading
system designed to deliver a
therapeutic radiation beam to a target
anatomical area. The radiation sources
incorporated as a component of the
remote after loading system are
generally sealed.
Class C
S.No.
Device name
Intended use
Risk Class
96 Radionuclide system
contour detector for
remote irradiation
therapy
Intended to precisely determine the
outline of the area of the body to be
irradiated. Usually, the information
obtained from this device is entered into
a radiotherapy planning system and
utilized for the radiotherapy plan.
Class C
97 Real-time position
management respiratory
gating system, optical
An assembly of electronic devices
designed to track the respiratory pattern
of a patient by means of optical
technology to correlate tumour
position with the respiratory cycle
during radiation treatment planning,
radiotherapy, computed tomography
(CT) imaging, or other radiation
procedures.
Class C
98 Rectal balloon for
prostate immobilization
A rectal balloon for prostate
immobilization is a single use, inflatable,
non-powered positioning device placed
in the rectum to immobilize the
prostate in patients undergoing
radiation therapy.
Class C
99 Remote controlled
radionuclide applicator
system
A remote controlled radionuclide
applicator system is an
electromechanical or pneumatic
device intended to enable an operator
to apply, by remote control, a
radionuclide source into the body or to
the surface of the body for radiation
therapy.
Class C
100 Remote-afterloading
brachytherapy source
A device that is a naturally-occurring,
accelerator or reactor produced-
radioactive isotope used as a source of
radiation in high or low dose rate remote
afterloading brachytherapy systems
designed to deliver curative or palliative
radiotherapy treatments. Sources used
in remote afterloading radiotherapy
systems may come in a variety of
strengths and physical forms, e.g., single
encapsulated source (sealed source),
ribbons, plated, foil or embedded
sources or contained radioactive liquids
or gels.
Class C
S.No.
Device name
Intended use
Risk Class
101 Remote-afterloading
An assembly of devices using a
remotely-controlled radioactive source
transport system designed to deliver a
therapeutic or palliative radiation dose
from a single source or source train to an
anatomical region by placing the
radioactive source(s) temporarily on or in
the treatment site.
Class C
102 Stereotactic radiosurgical
system
An assembly of devices designed to
deliver a therapeutic radiation dose to an
anatomical region from external beams
produced from multiple radionuclide
sources arranged in a fixed focal point
collimated array; typically used to treat
brain, neck, breast and spinal tumours. It
typically consists of: 1) a stereotactic
positioning device attached to the patient
to define the three-dimensional (3-D) co-
ordinates of the treatment site; 2) fixed
diameter collimators; 3) high-density
beam blockers; 4) a shielded housing to
hold the sources, reducing room
radiation; 5) a precision aligned array to
cause the source beams to intersect at a
single point; and 6) a moveable patient
table.
Class C
103 Stereotactic radiotherapy
accelerator system
A stereotactic radiation therapy
system for treatment based on a linear
accelerator (or microtron). The device
may be used to inactivate lymphocytes.
Class C
S.No.
Device name
Intended use
Risk Class
104 Teletherapy radionuclide
system table, powered
A device that is a component of a
teletherapy radionuclide system
(commonly known as a cobalt therapy
machine) and that is a
powered/programmable table specifically
designed to position and support a
patient during treatments administered
using a therapeutic radionuclide
teletherapy system (e.g., a Cobalt-60
teletherapy system). It has electronic
and/or software controls for table top
height and positioning. It can be a
stationary or mobile unit, or incorporated
as an integral component of a
radionuclide teletherapy system or
gantry configuration.
Class B
105 Teletherapy radionuclide
system,
conventional/Cobalt
therapy machine
A stationary assembly of computer-
based devices designed to deliver a
therapeutic radiation dose to an
anatomical region from a single external
radiation beam produced by a
radionuclide source, typically to treat
malignant tumours; it does not provide
image-guided radiation therapy (IGRT)
functionality during treatment delivery.
Class C
106 Therapeutic x-ray high
voltage generator
It is a device that is intended to supply
and control the voltage and electrical
energy to provide a therapeutic x-ray
system (e.g., brachytherapy) with the
power needed to produce an x-ray beam
of the desired voltage (kV) and current
(mA).
Class B
107 Ultraviolet Extracorporeal
Photopheresis system
An assembly of devices used for
extracorporeal photoimmunetherapy to
treat immune disorders, especially
cutaneous T-cell lymphoma (CTCL). It
irradiates the leukocyte-rich fraction of
peripheral blood by UVA radiation
extracorporeally and returns back the
treated and untreated blood to the
patient. This system may or may not use
a UV-active drug (either ingested by
patient or injected into the leukocyte-rich
fraction of the extracted peripheral
blood).
Class C
S.No.
Device name
Intended use
Risk Class
108 Ultraviolet Photopheresis
system blood set
A collection of devices used as part of a
photopheresis system for extracorporeal
photoimmunetherapy to treat immune
disorders, especially cutaneous T-cell
lymphoma (CTCL). It typically consists of
tubing (patient and fluid lines), fluid/blood
component bags, a dedicated centrifuge
bowl, a photoactivation chamber,
connectors, and clamps. It conducts
blood from the patient to the system's
main unit, aids in blood processing, and
returns leukocyte-enriched and
untreated components of the blood after
irradiation. This is a sterile, single-use
device.
Class C
109 Ultraviolet Photopheresis
system lamp assembly
An assembly of ultraviolet A (UVA)-
emitting tubular strip lights that is an
exchangeable component of a
photopheresis system, and used to
irradiate blood components during
extracorporeal photoimmunetherapy to
treat immune disorders, especially
cutaneous T-cell lymphoma (CTCL).
Class C
110 X-ray CT system for
radiotherapy planning
A X-ray CT system that has a
special configuration, containing
hardware, software, etc. used in
radiotherapy planning. It is used to
determine the size and positioning of
the therapeutic radiation field based
on a series of treatment parameters to
be generated.
Class C
111 X-ray radiation therapy
system
An x-ray radiation therapy system is a
device intended to produce and control
x-rays used for radiation therapy.
Class C
112 X-ray/CT combined
linear accelerator system
A combined system of a linear
accelerator system and an X-ray CT
system for radiotherapy planning.
Class C
113 X-ray/CT combined
particle radiotherapy
equipment
A combined system of particle
radiotherapy equipment and an X-ray CT
system for radiotherapy planning.
Class C
S.No.
Device name
Intended use
Risk Class
114 X-ray/MR combined
linear accelerator system
A system intended to provide treatment
planning, image- guided stereotactic
radiosurgery and precision radiotherapy
for lesions, tumors and conditions
anywhere in the body where radiation
treatment in indicated. The system
operates on the principle of linear
acceleration of electrons, providing a
predictable radiation field in a beam of
well-defined dimensions.
Class C
Name of Category:
Oncology
Total No. of Devices:
75
Category: Oncology S. No. Device name Intended use Risk Class 1 Accelerator system chair A seat, typically with legs, that is a component of a therapeutic accelerator system, and used to support and position a seated patient during radiation therapy treatments involving the use of either a medical linear accelerator or non-linear accelerator. Class C 2 Accelerator system quality assurance device An instrument specifically designed to be used to check the calibration and performance of linear and non-linear medical accelerator systems used for radiation therapy applications, for quality assurance (QA) purposes. Class C 3 Alternating electric field antimitotic cancer treatment system An assembly of portable devices designed to apply low-intensity, intermediate-frequency (100-300 kHz) alternating electric fields to treat certain forms of recurrent or newly-diagnosed cancer; typically glioblastoma multiforme (GBM) [malignant brain tumour]. Class D 4 Alternating electric field antimitotic cancer treatment system transducer array Alternating electric fields therapy is a novel anticancer treatment that disrupts tumor cell mitosis. Class C 5 Alternating electric field antimitotic cancer treatment system generator Alternating electric fields therapy is a novel anticancer treatment that disrupts tumor cell mitosis. Class C 6 Anorectal brachytherapy system applicator, remote-afterloading A remote afterloading brachytherapy applicator specifically designed for use in radiation therapy treatments of the rectum and/or anus. Class C 7 Anorectal brachytherapy system applicator, manual A manual brachytherapy applicator specifically designed to be used in radiation therapy treatments of the rectum and/or anus. Class C 8 Balloon kyphoplasty kit A collection of sterile surgical instruments and devices used for the reduction of a vertebral compression fractures (VCFs) caused by trauma, cancer, or osteoporosis during a minimally invasive procedure commonly known as balloon kyphoplasty. Class C 9 Bladder instillation buffer solution A sterile buffer solution intended to be used exclusively for bladder instillation to help create an optimal environment necessary for the effective treatment of superficial bladder cancer with a chemotherapy agent. Class B
S. No. Device name Intended use Risk Class 10 Brachytherapy source spacer A sterile, bioabsorbable device designed to separate radioactive sources of the seed type that are permanently implanted in close proximity to a selected localized tumour, to increase the distribution of radioactivity to the tumour. Class C 11 Brachytherapy system remote- afterloading operator console A mains electricity (AC-powered) component of a remote- afterloading brachytherapy system intended to function as the primary control panel for the remote afterloader. It typically includes hardware and software that allows for information display and/or transfer, data processing, analysis, and information archiving functions; it may also be intended to interface with other devices (e.g., radiation therapy treatment planning computer) as part of a picture archiving and communication system (PACS). Class C 12 Breast 3-D infrared imaging/vascular analysis system An assembly of mains electricity (AC-powered) devices intended for three-dimensional (3-D) breast imaging and breast vascular analysis, typically used with mammography screening to perform a breast cancer risk examination. Class C 13 Breast brachytherapy system applicator, remote-afterloading A sterile, remote-afterloading brachytherapy applicator specifically designed for use in radiation therapy treatments of the breast. It is typically designed for temporary implantation within the breast and serves as a guide for computer-controlled placement and removal of single or multiple radioactive sources. Included are various types of applicators such as hollow needles, tubes, or catheters, and their associated components. This is a single-use device. Class C 14 Breast transilluminator A mains electricity (AC-powered) transilluminating device with a built-in light source using low intensity emissions of visible light and near-infrared radiation (700 to 1050 nm) that is transmitted through the female breast to visualize translucent tissue for the diagnosis of cancer, or other conditions, diseases or abnormalities. This device may also be known as a diaphanoscope. Class C 15 Breast ultrasound imaging system An assembly of mains electricity (AC-powered) devices designed for intracorporeal (endosonography or endoscopic) ultrasound imaging procedures involving the breast. It typically includes special imaging tables used to optimize the ability to give reproducible images of the breast. Class C 16 Breast ultrasound imaging system An assembly of mains electricity (AC-powered) devices designed for extracorporeal ultrasound imaging procedures involving the breast. It typically includes special imaging tables used to optimize the ability to give reproducible images of the breast Class B
S. No. Device name Intended use Risk Class 17 Cancer diagnostic probe A device intended for detection of breast cance lesions of various subtypes intraoperatively by checking for tumour side margins. Class C 18 Capsular tension ring A circular band intended to be used to enhance the mechanical stability of a subluxated crystalline lens capsule in the presence of weak or absent supporting zonules. Class C 19 Cervical cone knife A surgical, manually-operated, instrument that is inserted into the vagina and designed for excising a sample of abnormal tissue, e.g., indicated by the presence of precancerous changes, from the cervix. Class C 20 Cervical cytology scraper, single-use A hand-held, manual, blunt surgical instrument designed to scrape and retrieve cytological material from the surface of the cervix (neck of the uterus) or vaginal area for pathological examination and diagnosis, often for the detection of cervical cancer. This is a single-use device. Class B 21 Colonic cytology sampling set A collection of non-sterile devices intended to collect exfoliated colonic cells (colonocytes) from the surface of human rectal mucosa for colorectal cancer investigation and/or patient screening. Class B 22 Computer vision/Machine learning-aided software application for cancer detection A standalone software application intended to be used to scan radiological images (X-ray, CT scan, etc.) in order to screen/diagnose for cancerous tissues using machine learning-based training models and computer vision. Class C 23 Coronary artery brachytherapy system applicator, manual- afterloading A sterile flexible tube intended to deliver/remove radiation therapy sources into a coronary artery, typically into the lumen of an implanted stent, as part of a manual-afterloading brachytherapy system. It is introduced into the patient and subsequently connected to the brachytherapy system source transfer device; it includes radiopaque markers to monitor the position of the radiation source. Disposable devices associated with the procedure may be included (e.g., syringe, connectors). This is a single-use device. Class D 24 Cryosurgical set A sterile collection of disposable devices used in conjunction with a cryosurgical unit as well as monitoring and other devices to perform a surgical technique that involves freezing a targeted area of tissue to damage and destroy cancer cells in the unwanted portions. Class C 25 Electro cancer therapy system An assembly of devices designed for the treatment of tumours and the destruction of their cancerous cells using low-voltage direct current of small intensity delivered via electrodes placed across the affected body area. Class C
S. No. Device name Intended use Risk Class 26 Electronic clinical breast examination system A portable assembly of devices designed to electronically measure, map, document and store information about breast lesions/masses with regard to shape, size, location, consistency/relative hardness during a clinical breast examination (CBE). Class B 27 Electroporation therapy system A mobile assembly of devices designed to apply electrical impulses to the tissue to enable electroporation, a phenomenon that induces alteration in the structure of cell membranes to increase their permeability and allow molecules that usually cannot enter the cell membrane, such as drugs [electrochemotherapy (ECT)] and genetic materials [electrogenetherapy (EGT)], to reach the cytoplasm. Class C 28 Electroporation therapy system endoscopic applicator A sterile, patient-contact component of an electroporation therapy system intended to fit onto the distal tip of an endoscope and connect to an electroporation therapy system generator to deliver electrical impulses to tissues during endoscopy as part of electroporation, a phenomenon that induces alteration in the structure of cell membranes to increase their permeability and allow molecules that usually cannot enter the cell membrane, such as drugs [electrochemotherapy (ECT)], to reach the cytoplasm. Class C 29 Embolization particle, bioabsorbable A bioabsorbable, implantable bead/microsphere intended to be used to temporarily occlude an artery supplying hyperplastic/neoplastic tissue in a variety of anatomies (e.g., liver, lung, breast, bladder, uterus, head or neck); it does not include a pharmaceutical agent. It may be used independently to create ischemia in the tissue, or as an adjuvant in combination with cytostatic agents and other antitumour drugs to help optimize intra-tumour drug accumulation for interventional radiology and visceral surgery procedures. It is typically available as an injectable solution containing numerous microspheres [e.g., degradable starch microspheres (DSM)]. Class D 30 Embolization particle, non- bioabsorbable A non-bioabsorbable, implantable bead/microsphere intended to be introduced into the peripheral vasculature during an interventional radiology procedure to treat hypervascularized tumours and arteriovenous malformation in a variety of anatomies (e.g., head, neck, spine, liver, genitourinary tract, uterus, gastrointestinal, limbs and lungs). It is typically available as injectable solution containing numerous microspheres [e.g., compressible polyvinyl alcohol (PVA) microspheres] intended to permanently obstruct blood flow to the tumour/malformation. Class D
S. No. Device name Intended use Risk Class 31 Endocervical aspirator A collection of devices designed to remove superficial tissue from the mucous membrane lining the cervical canal (endometrium) through manually- powered suction. Class C 32 Externally- propelled flexible video colonoscope A non-sterile endoscope with a highly flexible sleeve and distal tip intended for the visual examination of the entire adult colon [lower gastrointestinal (GI) tract].It is used for the screening of colorectal cancer and the detection of other diseases of the lower GI tract. This is a single-use device. Class B 33 Extravascular- circulation hyperthermia system An assembly of devices designed to produce and control heated fluids circulated within a vessel applied to the body (e.g., vest, mattress, jacket, band, pad, body wrap, catheter, probe) for systemic or localized heating to treat malignant tumours, benign growths, or other disease-related conditions. Class B 34 Extravascular- circulation hyperthermia system applicator, extracorporeal A vessel applied to the outside of the body (e.g., in the form of a jacket, vest, body wrap, cushion, blanket, or mattress) that incorporates tubing through which heated fluids are circulated for systemic or localized heating to treat malignant tumours, benign growths, or other disease-related conditions. The applicator typically includes a thermometry component that monitors the temperature of the applicator during operation. The applicator includes tubing, cables, and connectors that interface with the hyperthermia system's control unit during treatments. It is typically used in an oncology department. This is a reusable device. Class A 35 Extravascular- circulation hyperthermia system applicator, intracorporeal A component of a hyperthermia system that typically consists of catheter-enclosed tubing which is introduced into the body either manually or endoscopically. Heated fluid is circulated through the applicator's tubing for localized heating to treat malignant tumours, benign growths, or other disease-related conditions. The applicator (also called an interstitial applicator or probe) typically includes a thermometry component that monitors the temperature of the applicator during operation; it also includes tubing, cables, and connectors that interface with the hyperthermia system's control unit during treatments. It is typically used in an oncology department. This is a single-use device. Class C
S. No. Device name Intended use Risk Class 36 Facial prosthesis An externally-applied device intended to be used as an artificial substitute for parts or sections of the face [e.g., nose, eye(s), eye brows, upper lip] to help restore facial appearance. The device may be customized to meet the patient's needs and may be held in position with magnets or screw-like implants embedded into the patient's facial bone. It is used in cases of severe facial disfigurement typically caused by trauma (e.g., gunshot wounds) or major facial cancer surgery. Class B 37 Fixed-aperture therapeutic x-ray system collimator A non-automated, x-ray beam-limiting device that is a component of a therapeutic x-ray system and whose opening size/length/shutter assembly is fixed. It is used in radiation therapy applications to limit the effects of scattered radiation and to protect the patient by limiting or eliminating exposure to non- target body areas during treatment. This device is specifically designed for use with an x-ray simulation or therapeutic x-ray system. Class C 38 Flexible fibre optic bronchoscope An endoscope with a flexible inserted portion intended for the visual examination and treatment of the trachea, bronchi, and lungs. It is inserted through the mouth or nose during bronchoscopy. Anatomical images are transmitted to the user by the device through a fibre optic bundle. This device is commonly used to diagnose lung infections, pneumonia, or lung cancer, and allows physicians to view the insides of the lungs and take biopsies and samples of secretions. This is a reusable device. Class B 39 Flexible fibre optic mediastinoscope An endoscope with a flexible inserted portion intended for the visual examination and treatment of the mediastinum (the intrapleural space located behind the sternum). It is inserted into the body through an artificial orifice created by an incision made during mediastinoscopy. Anatomical images are transmitted to the user by the device through a fibre optic bundle. This device is commonly used to examine structures such as lymph nodes during a staging evaluation of lung cancer, or to establish the diagnosis of a tumour that is localized to the mediastinum. This is a reusable device. Class C
S. No. Device name Intended use Risk Class 40 Flexible ultrasound bronchoscope An endoscope with a flexible inserted portion intended for the visual examination and treatment of the trachea, bronchi, and lungs. It is inserted through the mouth or nose during bronchoscopy. Anatomical images are transmitted to the user by the device typically through a fibre optic bundle or a video system, and an ultrasound probe. The probe may be built-in or inserted through a dedicated lumen so that its distal tip is positioned adjacent to that of the endoscope. It is commonly used to diagnose lung infections, pneumonia, or lung cancer, and allows physicians to view the insides of the lungs and take biopsies and samples of secretions. This is a reusable device. Class B 41 Flexible video bronchoscope, reusable An endoscope with a flexible inserted portion for endoscopic procedures of the airways and tracheobronchial tree (i.e., bronchoscopy). It is inserted through the mouth or nose during bronchoscopy. Anatomical images are transmitted to the user by a video system with a charge-coupled device (CCD) chip at the distal end and the images showing on a monitor. It is commonly used to diagnose lung infections, pneumonia, or lung cancer, and allows physicians to view the insides of the lungs and take biopsies and samples of secretions. This is a reusable device. Class B 42 General-purpose infusion pump, mechanical, reusable A non-electric, mechanically-powered (e.g., a spring mechanism) device designed for the continuous or intermittent infusion of medication, typically for antibiotic therapy, chemotherapy, or pain management by intravenous (IV), subcutaneous, intramuscular, or epidural routes. It is primarily designed to be worn by the patient during ambulation in the home. It may be used for patient- controlled analgesia (PCA), and may include mechanical indicators for flow and fluid level status. This is a reusable device. Class C 43 General-purpose infusion pump, mechanical, single- use A portable, non-electric, mechanically-powered device designed to be operated by healthcare professionals for dispensing a single dose of fluid medication (e.g., for antibiotic therapy, chemotherapy, analgesia). It consists of an empty reservoir intended to be filled with medication, a flow-rate regulator and a non-sterile (sterilizable) administration line intended to be connected to an infusion catheter (not included) for intravenous (IV), subcutaneous, intramuscular, or epidural administration. It may include flow and fluid level mechanical indicators and may be worn by the patient in and outside of healthcare settings. This is a single-use device. Class C
S. No. Device name
Intended use
Risk
Class
44 Hyperthermia
system
temperature probe
A device (a probe) intended to be used exclusively to
monitor tissue or body temperature during
hyperthermia treatments. Depending on the kind of
hyperthermia system it is used with, e.g., ultrasound,
radio-frequency (RF) or microwave, this probe is
shielded in such a way that its operation is not
affected by the energy from the hyperthermia
applicator(s) in use during the treatment. This is a
reusable device.
Class C
45 Implantable
Vascular
port/catheter
A fully-implantable device assembly intended to
provide access to arteries/veins (excludes coronary
and intracerebral circulation) for infusion (e.g.,
chemotherapeutic agents, blood transfusions) and/or
drainage (e.g., blood). It consists of: 1) a
subcutaneous chamber, with a self-sealing
puncturable septum for percutaneous insertion of a
hypodermic needle; and 2) an attached catheter
which passes into the arteries/veins.
Class C
46 Intracavitary-
circulation
hyperthermia
system
An assembly of electrically-powered devices
designed to continuously lavage body cavities (e.g.,
pleural cavity, peritoneal cavity, bladder lumen) with
warmed fluids/chemotherapeutic agents to raise the
local temperature and/or enhance the effect of
drugs, typically in the treatment of malignancy. It
typically includes a water heater/cooler, fluid pump,
controls, inputs for monitoring temperature, a circuit
with heat exchanger, fluid reservoir, and tubing
connectors. Introduction/drainage catheters are
connected to the system and placed in the body
cavity.
Class C
47 Intraoperative
gamma radiation
detection system
probe, reusable
An electrically-powered, invasive component of an
intraoperative gamma radiation detection system
intended to be introduced into the body during a
surgical procedure to detect and quantify gamma
radiation emitted from previously-administered
radiopharmaceuticals (e.g., localized in sentinel
lymph nodes), in conjunction with a dedicated control
unit (not included). It is typically manipulated using
surgical forceps and incorporates a crystal scintillator
and photomultiplier. This is a reusable device.
Class C
S. No. Device name
Intended use
Risk
Class
48 Intraoperative
tumour margin
fluorescence
imaging system
A set of devices designed for the excitation of an
injected fluorophore (e.g., pegulicianine), and
subsequent imaging of the illuminated area, to allow
for intraoperative detection of residual cancerous
tissue in the resection cavity following removal of the
primary specimen during surgical procedure (e.g.,
lumpectomy). It includes a probe, a handpiece with
an optical component including a digital camera
which is inserted into the cavity for image capture,
and a cable which includes a connection to the
computerised system and a fibre optic light guide.
This is a reusable device.
Class C
49 Intravascular/intrac
avitary-circulation
hyperthermia
system
An assembly of devices designed to circulate
warmed fluids or autologous blood through the
vasculature or cavity of a targeted anatomical area
(e.g., peritoneal/pleural cavities, thorax, abdomen,
section of a limb) to treat cancer. It is typically
computer-controlled and consists of pumps, heat
exchanger, valves, detectors/sensors, clamps,
monitors, alarms, disposable tubing/filtration set and
oxygenator. It heats the blood/fluids to temperatures
up to 43 °C to enhance the effect of drugs [e.g., for
intraperitoneal hyperthermic chemotherapy (IPHC)]
and help destroy cancer cells.
Class C
50 Intravascular/intrac
avitary-circulation
hyperthermia
system set/Kit for
hyperthermic
perfusion
A collection of sterile devices used with an
intravascular/intracavitary-circulation hyperthermia
system for the circulation of blood/fluid between the
patient and the system's extracorporeal circuit during
cancer therapy. It typically consists of
drains/catheters, tubing, connectors, manual clamps,
and hard or soft reservoirs; some types may include
in-line filters, an oxygenator, and temperature
sensors. This is a single-use device.
Class C
51 Microwave ablation
system
An assembly of devices consisting of a generator
and a probe (and other accessories) intended to
generate and transmit microwave energy for
localized non-vascular soft-tissue ablation, typically
to treat tumours, hydatid cysts and/or menorrhagia.
The generator connects via a delivery cable to a
probe for delivery of microwaves to the target
tissues. It is intended to be used in percutaneous,
laparoscopic, natural orifice or open surgery
procedures to ablate tissue typically in the liver, lung,
pancreas, kidney, and uterus (e.g., endometrial
ablation).
Class C
S. No. Device name Intended use Risk Class 52 Microwave ablation system generator A mains electricity (AC-powered) device intended to generate microwave energy for localized non- vascular soft-tissue ablation, typically to treat tumours, hydatid cysts and/or menorrhagia (excludes cerebral and coronary tissues). The generator connects via a delivery cable to a probe (not included) for delivery of microwaves to the target tissues. It is intended to be used in percutaneous, laparoscopic, natural orifice or open surgery procedures to ablate tissue typically in the liver, lung, pancreas, kidney, and uterus (e.g., endometrial ablation). Class C 53 Microwave ablation system probe/microwave ablation antenna A hand-held surgical instrument designed to connect to a microwave ablation system generator to deliver microwaves to a targeted operative site for localized soft-tissue ablation, typically to treat tumours, hydatid cysts and/or menorrhagia. The device is intended to be used in percutaneous, endoscopic [e.g., gastroscopic, laparoscopic], natural orifice or open surgery procedures to ablate tissues (e.g., endometrial ablation). Class C 54 Microwave hyperthermia system A mains electricity (AC-powered) device assembly designed for controlled heating (i.e., temperatures above 43° Celsius) of the body using microwaves, for the treatment of malignant or benign tumours or other disease conditions [e.g., benign prostatic hyperplasia (BPH), prostatitis]. It includes a microwave generator, controls, software (e.g., treatment planning software) and a microwave applicator. It is intended for both whole-body and localized heating of tissues/organs through either externally-mounted components or from catheter- or probe-type applicators inserted topically, endoscopically, or surgically. Class C 55 Microwave hyperthermia system applicator, extracorporeal A component of a microwave hyperthermia system designed to direct and deliver microwave energy used to produce a systemic or local heating effect. It is typically mounted on a floor, wall, or ceiling suspension that allows the operator to position the microwave source from a point outside the patient's body. The applicator assembly includes cables and connectors that interface with the energy-generating unit of the hyperthermia system during treatments. Class C
S. No. Device name
Intended use
Risk
Class
56 Microwave
hyperthermia
system applicator,
intracorporeal
A component of a microwave hyperthermia system
designed to be placed inside the body to deliver
microwave energy to produce a systemic or local
heating effect. It is also referred to as an interstitial
applicator or probe and is typically a catheter-
enclosed conductor of microwave energy that is
introduced into the body either manually or
endoscopically. It typically includes a thermometry
component that monitors the temperature of the
device during operation.
Class C
57 Microwave/electros
urgical system
generator
An electrically-powered component of a microwave
ablation/electrosurgical system intended to generate
both: 1) radio-frequency (RF) electrical current; and
2) microwave energy for subsequent cutting,
coagulation, and ablation of soft tissues during an
endoscopic or open surgical procedure; it is not
dedicated to focal ablation of specific tissues. It is
not intended for inert gas electrosurgery.
Class C
58 Multi-modality
hyperthermia
system
A mains electricity (AC-powered) device assembly
designed for controlled heating (i.e., temperatures
above 43° Celsius) of the body using multiple energy
sources, for the treatment of malignant or benign
tumours, or other disease conditions (excludes
cerebral and coronary tissues). It includes a patient-
end applicator, software, and a multi-modality
generator(s)/control unit. It is intended to achieve
whole-body and localized heating of tissues using
two or more energy sources [e.g., ultrasound, radio-
frequency (RF), microwave, circulating heated fluid]
through either externally-mounted components, or
from catheter- or probe-type applicators inserted
endoscopically or surgically.
Class C
59 Photopheresis
system blood set
A collection of devices used as part of a
photopheresis system for extracorporeal
photoimmunetherapy to treat immune disorders,
especially cutaneous T-cell lymphoma (CTCL). It
typically consists of tubing (patient and fluid lines),
fluid/blood component bags, a dedicated centrifuge
bowl, a photoactivation chamber, connectors, and
clamps. It conducts blood from the patient to the
system's main unit, aids in blood processing, and
returns leukocyte-enriched and untreated
components of the blood after irradiation. This is a
sterile, single-use device.
Class C
60 Photopheresis
system lamp
assembly
An assembly of ultraviolet A (UVA)-emitting tubular
strip lights that is an exchangeable component of a
photopheresis system, and used to irradiate blood
components during extracorporeal
photoimmunetherapy to treat immune disorders,
especially cutaneous T-cell lymphoma (CTCL).
Class C
S. No. Device name Intended use Risk Class 61 Polymer-metal oesophageal stent A non-bioabsorbable, expandable, tubular device intended to be implanted into the oesophagus to maintain luminal patency in strictures and prevent tumour in-growth. It may also be intended to seal oesophageal fistulas, reduce acute bleeding from oesophageal varices, and/or to treat other lesions causing oesophageal leakage (e.g., anastomotic). It is typically expanded in situ (e.g., self-expands) and disposable devices intended to assist implantation may be included. Class C 62 Radiation therapy digital imager An automated device that is typically mounted on the gantry of a linear accelerator and intended to produce digital images of x-rayed anatomical landmarks to guide radiation treatment (e.g., tracking/targeting tumours). The device may be a digital imaging panel (e.g., of silicon) on robotic arms; it may also have an x-ray source to generate higher quality images. Class C 63 Real-time position management respiratory gating system, optical/Respiratory gating system, radiation procedure An assembly of electronic devices designed to track the respiratory pattern of a patient by means of optical technology to correlate tumour position with the respiratory cycle during radiation treatment planning, radiotherapy, computed tomography (CT) imaging, or other radiation procedures. It provides real-time position management (RPM) and involves optical-based tracking (e.g., video, infrared, laser) of the respiratory cycle to enable irradiation control during the procedure. It may include a tracking camera connected to a dedicated personal computer workstation and a reflective chest wall/abdominal marker. Class C 64 Robotic Guidance system for image Guided procedures It is an accessory to an imaging system (CT, CT- PET) intended for the spatial positioning and orientation of an instrument guide. A surgeon then manually advances one or more instruments for percutaneous image guided interventional procedures through the instrument guide. The device is not intended to make any contact with the patient. Class B 65 Scalp cooling system, hair loss An assembly of electrically-powered devices intended to reduce scalp hair loss associated with intravenous chemotherapy treatment by cooling the scalp to cause vasoconstriction and reduction of chemotherapy drug uptake by the follicular cells. It consists of a control unit which pumps a cooling agent to a head-worn cap containing coolant channels and temperature monitoring sensors. An insulating cap (e.g., neoprene) may be worn over the cooling cap to secure it and prevent condensation. Class B
S. No. Device name Intended use Risk Class 66 Stereotactic radiosurgery system for central nervous system (CNS) A set of devices that are designed to uses very precise beams of gamma rays to treat an area of disease (lesion) or growth (tumor), especially in the brain, upper spine and in certain cases, vascular abnormalities. Class D 67 Teletherapy radionuclide system table, powered A device that is a component of a teletherapy radionuclide system (commonly known as a cobalt therapy machine) and that is a powered/programmable table specifically designed to position and support a patient during treatments administered using a therapeutic radionuclide teletherapy system (e.g., a Cobalt-60 teletherapy system). It has electronic and/or software controls for table top height and positioning. It can be a stationary or mobile unit, or incorporated as an integral component of a radionuclide teletherapy system or gantry configuration. Class B 68 Teletherapy radionuclide system, conventional/Cobal t therapy machine A stationary assembly of computer-based devices designed to deliver a therapeutic radiation dose to an anatomical region from a single external radiation beam produced by a radionuclide source, typically to treat malignant tumours; it does not provide image- guided radiation therapy (IGRT) functionality during treatment delivery. Class C 69 Telethermographic system A telethermographic system intended for adjunctive diagnostic screening for detection of breast cancer or other uses is an electrically powered device with a detector that is intended to measure, without touching the patient's skin, the self-emanating infrared radiation that reveals the temperature variations of the surface of the body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. Class A 70 Therapeutic oncological/gynaec ological ultrasound system (High intensity focused ultrasound (HIFU) ablation system) An assembly of electrically-powered devices designed to treat solid tumours of hard (bone) and/or soft tissue [e.g., liver, kidney, breast, prostate] and/or gynaecological disorders (e.g., uterine fibroid, adenomyosis) through noninvasive or non-surgically invasive localized application of high intensity focused ultrasound (HIFU) or high intensity therapeutic ultrasound (HITU) intended to gradually denature/ablate tissue lesions. Optimal patient positioning is achieved using an ultrasound scanner or MRI system. Class C
S. No. Device name Intended use Risk Class 71 Tumour-therapy radio-frequency hyperthermia system A mains electricity (AC-powered) device assembly designed for controlled heating (i.e., temperatures around 43° Celsius) of the body using radio- frequency (RF) energy, for the treatment of malignant or benign tumours, or other disease conditions; it is not intended for direct tissue ablation. It is intended for both whole-body and localized heating of tissues/organs through either externally- mounted components or from catheter- or probe- type applicators inserted topically, endoscopically, or surgically. Class C 72 Tumour-therapy radio-frequency hyperthermia system applicator, intracorporeal A component of a radio-frequency (RF) hyperthermia system designed to be placed in the body to deliver radio-frequency (RF) energy to produce a systemic or local heating effect for the treatment of malignant or benign tumours, or other disease conditions. It is also referred to as an interstitial applicator or probe and is typically a catheter-enclosed conductor of RF energy that is introduced into the body either manually or endoscopically. This is a single use device. Class C 73 Ultrasonic hyperthermia system/Ultrasonic ablation system An assembly of devices designed to produce and control the delivery of high heat (i.e., temperatures greater than 43 deg Celsius) to the body using ultrasonic energy for the intracorporeal or extracorporeal treatment of malignant or benign tumours, or other disease conditions. Energy delivered to the patient is from either an externally- mounted transducer or a transducer-containing catheter or probe used topically or inserted endoscopically or surgically. It is not intended for physiotherapy applications. Class C 74 Ultrasound hyperthermia system transducer, extracorporeal An ultrasound (US) transducer assembly that is a component of an ultrasound hyperthermia system used to induce systemic or local body heating of sufficient magnitude to create a targeted therapeutic effect. It is mounted within the gantry or housing of the hyperthermia system or mounted on a floor, wall, or ceiling suspension that allows the operator to position the transducer assembly external to the patient during treatment. Class C 75 Ultraviolet Extracorporeal Photopheresis system An assembly of devices used for extracorporeal photoimmunetherapy to treat immune disorders, especially cutaneous T-cell lymphoma (CTCL). It irradiates the leukocyte-rich fraction of peripheral blood by UVA radiation extracorporeally and returns back the treated and untreated blood to the patient. This system may or may not use a UV-active drug (either ingested by patient or injected into the leukocyte-rich fraction of the extracted peripheral blood) Class C
Name of Category:
Class
A
(non-sterile
and
non-
measuring) medical devices
Total No. of Devices:
803
List of Class A (Non-sterile and non-measuring) medical devices (other than
In-vitro diagnostic (IVD) medical devices)
S.No. Device Name
Intended Use
1 Abdominal support/belt/binders,
reusable
Intended to hold abdominal muscles together to
relieve pain and increase circulation at surgical site
to promote healing and decrease swelling. May also
make physical activity more comfortable.
2 Abdominal support/belt/binders,
single use
Intended to hold abdominal muscles together to
relieve pain and increase circulation at surgical site
to promote healing and decrease swelling.
3 Abrasive polishing agent
A substance, or combination of abrasive substances
(e.g. silica pumice), intended to be used for cleaning
the accessible surfaces of the teeth by polishing. It
does not contain medication and used by dental
professionals.
4 Absorbent cotton wool
Intended to used for applying medication to, or
absorbing small amounts of body fluids
5 Accupressure calf band
Intended to wear around the calf to apply pressure
to an acupressure point to relieve low back pain,
including sciatica and piriformis syndrome.
6 Accupressure wristband
Intended for the application of pressure to the Nei-
kuan (P6) acupressure point on the wrist, the area
identified to help relieve the sensation of nausea.
7 Acoustic chamber for audiometric
testing
Intended for use in conducting diagnostic hearing
evaluations, while eliminating sound reflections and
providing isolation from outside sounds.
8 Acoustic Stethoscope
A mechanical listening device used for listening to
sounds from the heart and lungs and their qualitative
assessment.
9 Activated-medium surgical warming
blanket
A non-sterile device intended to help prevent a
patient from cooling prior to, during, and after a
surgical intervention by serving as a heat-generating
cover.
10 Acupressure applicator, reusable
Applied to the skin without insertion into the body to
apply pressure and/or friction to acupressure points.
11 Acupressure applicator, single-use
To be used for acupressure treatment at home by
applying non-invasive pressure to acupuncture
points. Applied to the skin without insertion into the
body.
12 Acupuncture unit, home use
It is intended to provide mild and temporary
stimulation of accupressure points resulting in
improvement of energy flow and overall vitality.
S.No. Device Name
Intended Use
13 Adaptor, light source
Intended to permit the connection of the devices that
together deliver light to a work field from a light
source.
14 Adaptor, luer
Adapters which connect or stopper enteral feeding
tubes.
15 Adhesive felt pad
Intended to offload pressure from ulcer areas,
recurrent plantar ulcers, calluses and corns
16 Adhesive plaster (Zinc oxide
adhesive plaster)/Tape
It is intended to secure dressings and for strapping
on intact skin.
17 Adhesive strip
A flexible strip (of fabric, plastic, paper, or other
material) coated on one side with a pressure-
sensitive adhesive, typically used to cover or
approximate the edges of superficial wounds or fix
dressings to skin. The device may include an
adhesive pad and have qualities such as waterproof.
18 Adhesive tape remover
Help reduce pain and ease removal of medical
adhesives used in stoma care products, dressings
and other medical appliances from the skin surface
19 Adult Diaper/Under Pad
Intended to be used by an incontinent patient's for
absorbing excreta (e.g, urine, etc.). It may be
intended for home-use or/and hospital settings.
20 Aerosol adhesive for compression
garments
A sticky substance that is applied on the skin to
anchor compression garments in place.
21 Aerosol delivery tubing
A flexible tube used in conjunction with an oxygen
mask, endotracheal (ET) tube, humidifier, or
nebulizer, intended for the delivery of aerosolized
humidification, typically oxygen enriched.
22 Air cushion/Pressure relief cushions
Intended to relieve pressure on an affected area
when a person has hemorrhoids or when he has
undergone surgery in the back or buttock
area/prevention of pressure sores/ulcers in patients
23 Airway device cleaning
utensil
A hand-held device intended to be used to clean an
in situ airway device (not disinfection). It can also be
used to clean an airway device ex situ (i.e., once it
has been removed from the patient).
24 Airway protection face mask
Intended to provide respiratory protection
S.No. Device Name
Intended Use
25 Airway tube forceps, reusable
Used for grasping a tube for its insertion and/or
extraction into/from the airways, or for grasping
obstructive objects for their removal from the
airways.
26 Alarm, bed occupancy
Bed Floor Mat/Crash Mat Alarm/Alert Patient
Movement
27 Allergy testing adhesive strip,
epicutaneous
A band of adhesive material designed to be used for
epicutaneous allergy testing.
28 Alternating-pressure bed mattress
overlay, reusable
A pad or thin mattress, usually full-body length,
which is placed on top of a bed mattress and
designed with parallel chambers that are alternately
filled with forced-air to actively change an occupant's
bed contact points, typically to relieve pressure
points for comfort and to prevent pressure sores. It
is commonly used for elderly immobilized (especially
in cases of decubitus ulcers), patients with
disabilities, or patients with low body fat. This is a
reusable device.
29 Alternating-pressure bed mattress
overlay, single-use
A pad or thin mattress, usually full-body length,
which is placed on top of a bed mattress and
designed with parallel chambers that are alternately
filled with forced-air to actively change an occupant's
bed contact points, typically to relieve pressure
points for comfort and to prevent pressure sores. It
is commonly used for elderly immobilized (especially
in cases of decubitus ulcers), patients with
disabilities, or patients with low body fat. This is a
single use device.
30 Alternating-pressure bed mattress
system
An alternating pressure system designed to be used
on top of a bed to provide therapeutic or protective
effect.
31 Amsler grid
Intended to rapidly detect central and paracentral
irregularities in the visual field.
32 Anaesthesia catheter Luer connector
Intended to create a mechanical union between an
anaesthesia catheter and an external device
33 Anaesthesia depth simulator
intended to simulate the electroencephalography
(EEG) signals of an unconscious patient
34 Anaesthesia instrument
table
It is intended as a support for Anaesthesia
instruments used during general anaesthesia
surgical procedures.
S.No. Device Name
Intended Use
35 Anaesthesia mask stabilizer
A device intended to secure an anaesthesia mask
on the face of a patient typically by providing
anchorage for the fixation of the mask's headstrap.
36 Anaesthesia mask stabilizer, single-
use
Accessory to anesthesia face mask to assist in
locating the mask headgear (strap) to the mask.
37 Anaesthesia system leakage tester
A device intended to test an anaesthesia system for
leakage.
38 Anaesthesia tube holder
The intended purpose for this device is for holding
and supporting respiration tubing and breathing
circuits connected to a patient, to prevent
disconnection.
39 Anaesthesia warmer
A device intended to warm the anaesthetic solutions
prior to it being administered to a patient for
anaesthesia.
40 Anaesthetic gas scavenging terminal
unit
A device intended to function as an outlet assembly
to which the operator can connect/disconnect an
anaesthetic gas scavenging system (AGSS).
41 Analytical non- scalp cutaneous lead
Intended to conduct electrical signals between a skin
electrode(s) or needle electrode(s) [electrode not
included] and a device designed for
electrophysiological recording/monitoring [e.g.,
electromyography (EMG), evoked potentials (EP),
bioelectrical impedance].
42 Analytical non-scalp cutaneous
electrode
Electrical conductor designed to be attached to the
skin surface of a patient outside of the hair line (i.e.,
non-scalp) to conduct electrical signals to a parent
device for electrophysiological recording/monitoring.
43 Analytical non-scalp cutaneous
lead
Intended to conduct electrical signals between a skin
electrode(s) or needle electrode(s) [electrode not
included] and a device designed for
electrophysiological recording/monitoring [e.g.,
electromyography (EMG),
evoked potentials (EP), bioelectrical impedance].
44 Anatomical training model
A 3-D printed patient specific model of an organ
(e.g., heart, kidney, lung) or bone(s) intended to be
used to assist peri-operative anatomical visualisation
and surgical treatment planning. It is created using
3-D Printing processes based on patient medical
images.
45 Anesthesia face mask, reusable
A device designed to be placed over a patient's nose
and/or mouth to administer anaesthetic gases to the
upper airway.
S.No. Device Name
Intended Use
46 Anesthesia face mask, single use
A device designed to be placed over a patient's nose
and/or mouth to administer anaesthetic gases to the
upper airway.
47 Ankle binder, reusable
A strip or roll of fabric or plastic material designed to
be wrapped around the ankle to help stabilize and
support the joint.
48 Ankle cap
Intended to provide mild compression & support to
the ankle joint and to manage ankle pain and
inflammation.
49 Ankle sleeves/brace
Intended as the external supports to limit certain
motions, such as plantar flexion/inversion and
manage postoperative/posttraumatic pain
50 Ankle traction belt
Intended to immobilize the ankle and adjoining joints
after injury / surgery to support fast healing and
prevents further damage and also helps to decrease
pain, spasms and swelling.
51 Ankle/foot orthosis
A prefabricated (non-customized) externally applied
and wearable appliance or apparatus intended to
encompass the ankle joint, or the ankle and foot, to
support, align, prevent, or correct orthopaedic
deformities/injuries or to improve function of the
ankle and/or foot through actions such as alignment,
hinge-support, or dorsiflexion assistance.
52 Ankle/foot orthosis
A prefabricated externally applied and wearable
appliance or apparatus intended to offload and
redistribute foot pressures that affect pedal
circulation to improve blood flow and help heal
diabetic foot ulcers or postsurgical wounds.
53 Anomaloscope
Intended to test a patient for abnormal red/green
colour vision by differentiating the red/green colour
vision defects.
54 Anterior chamber physical simulator
Designed to mimic the fluid dynamics in the anterior
chamber of the eye during priming/calibration of a
phacoemulsification system irrigation/aspiration
pump.
55 Anterior wedge footwear/Off loading
shoes
Intended for offloading the pressure from forefoot to
heel after correction of conditions such as Hallux
Valgus, Hammer Toes, Tailors Bunions and Ulcers
of the forefoot
56 Antimicrobial postsurgical brassiere
Intended to support and/or contour the breast(s) or
hold a dressing in place after surgical intervention
(e.g. thoracic surgery, mastectomy, lumpectomy).
May be used to prevent the formation of oedema
and to support venous return and lymphatic
drainage.
S.No. Device Name
Intended Use
57 Antimicrobial postsurgical female
underpants
It is intended for use during medical treatment (e.g.,
chemotherapy) or be used to protect the skin
following treatment with a medication (e.g. ointment,
cream).
58 Anti-shock body support/Garment
Intended to prevent excessive blood loss and
prevent post partum haemorrhage complications.
59 Anti-tremor gyroscopic orthotic glove
A battery-powered device designed to be worn on
the hand/forearm of a patient with hand tremors
[e.g., due to Parkinson’s disease, Essential tremor
(ET)] to reduce tremors and facilitate activities of
daily living using the hand. It is in the form of a glove
with forearm sleeve and built-in gyroscopes which
generate tremor-countering forces to promote hand
stabilization.
60 Arm board pad
It is intended to provide support to the patient's arm
during prolonged surgeries.
61 Arm board, pediatric
This medical device is intended to immobilize and
hold steady infants’ limbs when securing intravenous
lines.
62 Arm prosthesis socket
A device that fits over the terminal end of a residual
upper limb and serves as an interface between the
limb and an artificial replacement arm segment
(prosthesis).
63 Arm prosthesis, pediatric
Artificial hand or arm prosthesis for physically
disabled pediatric patients.
64 Arm sling
Intended to immobilize the arm, by means of a fabric
band suspended from around the neck.
65 Arm/leg procedure positioner padded
strap
To protect the patient's foot/leg in stirrups.
66 Artificial airway stylet
A device intended for insertion within the lumen of
an artificial airway tube to stiffen and/or maintain the
shape of the tube to facilitate intubation.
67 Assistive ergonomic chair mobility
base
Intended for use as a support to an assistive
ergonomic chair used by a healthcare provider/carer
to provide mobility for a disabled (often paediatric)
patient.
S.No. Device Name
Intended Use
68 Assistive ergonomic chair,
electrically-adjusted
An electrically-powered device intended to support
and position a patient (disabled and/or discomforted)
in a sitting posture while performing activities of daily
living, including work-related activities and to provide
improved personal body posture and/or to help
prevent/reduce physical problems associated with
poor mobility or sedentary lifestyle. It is not a
wheelchair
69 Assistive foot-propelled bicycle
A therapy bike designed for children and teenagers
with neuromuscular diseases permitting independent
mobility, assisting physiotherapist treatments,
providing training for support and balance reactions
as well as movement coordination.
70 Assistive gripping/twisting tool,
reusable
The devices are handheld grips with attached
wristbands intended for stimulation of the sensory
neuronal network (hand/fingers) as part of therapies
for patients suffering from neurodeneration such as
ASD, ADHD and Parkinson's, or other nervous
issues.
71 Assistive shoes/boots
It is a shoe attachment that is primarily intended to
assist people with Parkinson’s to overcome a
symptom called Freezing of Gait (FOG) through
visual cueing by providing visual cues in the form of
laser lines. It can act as an intervention for falls
prevention in the Parkinson’s population.
72 Atomizer
A device that is intended to provide liquid medication
in aerosolized form into the air that a patient will
breathe.
73 Attention assistive training device
A device designed to be used by a person with a
disorder to develop their attention ability to select,
focus, control and inhibit.
74 Automated perimeter
Intended to be used to measure the visual field of
the eye.
75 Back Support/Brace
Intended to prevent unnecessary movements and
pain that may further damage the back and to help
in alignment of spine and strengthening of back
muscles.
76 Bagolini lens
Intended to determine harmonious/anomalous
retinal correspondence.
77 Balance Board
Intended to train patient with difficulties in balance
(proprioreception) ( e.g. a paraplegic or a stroke
patient)
S.No. Device Name
Intended Use
78 Balance/mobility management
system
This device uses sensors to determine the centre of
pressure and postural sway of an individual during
standing for balance management.
79 Bandage scissors, reusable
Heavy scissors for cutting pads, plaster and plastic
rail associations.
80 Bar dental precision attachment
It connects removable partial dentures to fixed
bridgework under a male/female locking mechanism.
81 Basket stretcher
Intended to transfer patients in emergency situations
such as in the mountains, and marine offshore
application.
82 Bed board
A flat, stiff device designed to increase the firmness
of a hospital bed's mattress to provide an
appropriate rest surface for the user typically with
spinal problems.
83 Bed exit monitor, fall prevention
Intended to be placed under mattress and used to
indicate by an alarm or other signal when a patient
attempts to leave the bed.
84 Bed traction frame
Intended to provide overhead support framing for
hospital beds to assist in the treatment and
movement of bed patients
85 Bedrail pad
Intended to protect the patient from coming into
contact with the bedrails and inadvertently hurting or
injuring themselves.
86 Bicycle exerciser
Device is used by patients in a physical therapy
setting under the direction of a medical professional
for passive, assisted & active training of upper and
low limbs during rehabilitation, to improve motor
function.
87 Binocular loupes
It is intended for viewing details of objects with some
magnification during ophthalmic/dental and other
examinations and surgeries
88 Biofeedback system for rehabilitation
A system, comprising of wearable motion capture
devices that allows patients with upper and lower
extremity movement dysfunction to undertake
intensive practice of single and multi-joints using
movement biofeedback. The system detects
movement and provides this as feedback to the
patient. This does not display a quantifiable property
in measuring units.
89 Birthing bath
Intended to be filled with heated water for use before
and/or during child birth and to provide a
comfortable environment for the expecting mother
prior to birth.
90 Birthing bed/table, non-powered
Intended to be used as a support surface during
labor and delivery.
S.No. Device Name
Intended Use
91 Birthing bed/table, powered
Intended to be used as a support surface during
labor and delivery.
92 Bite block, dental
A non-sterile device inserted into a patient's mouth
for the prevention of oral tissue damage (e.g., the
teeth, lips, tongue and buccal mucosa) and to allow
them to bite down on for greater comfort during a
dental procedure. This is a reusable device.
93 Bite block, endoscopy
A single use accessory used to protect both the
patient and the endoscope, when an endoscope is
inserted into a patients mouth for a procedure. It
may also be used during other treatments such as
intubation of a tracheal tube.
94 Bite block, radiotherapy
Bite Blocks are devices which are used to
reproducibly position the mouth, tongue or lips
during radiotherapy treatment of the head and neck.
95 Bite registration rim wax
A dental material (modelling wax) with or without
reinforcing foils (metal, polymer) for registration of
jaw relation (making bite rims).
96 Bite relief pad
Intended to help relieve pain or discomfort after
orthodontic treatment by placement between the
upper and lower sets of teeth and bitten down on.
97 Blood/body fluid solidifier
A solidifier substance is intended to be used to
absorb and solidify blood and/or other body fluids
collected in surgical suction canisters. The
substance is often added to the bottle/canister
before or after collection of the fluid.
98 Bobath table
Intended for postural exercises, rolling, turning and
balance training of patients with loss of sensation or
poor balance and other neurological conditions such
as stroke, head injury and orthopedic indications
99 Body arch traction table
Intended to support the body of a patient and
provide traction for the back muscles and spine by
flexing the patient into a reverse supine body arch
100 Body waste receptacle
Intended for medical purposes that is not attached to
the body and that is used to collect the body wastes
of a bed patient.
101 Brachytherapy radionuclide
phantom, test object
Intended to mimic the functional/physical
characteristics of normal or diseased human organs
during performance evaluations of brachytherapy
system
S.No. Device Name
Intended Use
102 Breast binder
A non-sterile device intended to be applied to the
breast or breasts for soft tissue support or securing
a dressing in place after surgical procedures.
103 Breast pad
Used to absorb excess milk and protect the breast
during nursing.
104 Breathing circuit bag/Reservoir bag
A device intended to store breathing gas during the
respiratory cycle.
105 Breathing circuit condenser
Intended to be integrated within the expiratory limb
of a breathing circuit to remove excess moisture
through cooling and condensing, whilst also
reheating the dried gases to an appropriate
machine-compatible temperature.
106 Breathing circuit dryer
A device that is used for the purpose of drying
breathing circuit equipment that have been washed
in order to prevent bacteria growth and deterioration.
107 Breathing mouthpiece
A device intended to be inserted into a patient's
mouth to facilitate access to the respiratory system.
108 Cable-controlled hand prosthesis
Intended as an artificial hand for physically disabled
persons, whose movement can be controlled by a
cable
109 Camera cover
Provide an infection-control protective barrier for
dental intra-oral camera
110 Camera for dental use
Intended to be used to record that location of posts
and or abutments in a lab model or the mouth
allowing for a dental treatment plan to be produced
during dental implant planning.
111 Camera for medical applications
A camera to be used with microscopes, colposcopes
and on endoscopic procedures.
112 Canalith repositioning procedure
chair, manual
A device intended to be occupied by a patient and
manually operated by a healthcare professional to
facilitate therapeutic physical manoeuvres of the
patient’s head and trunk (accelerations and
decelerations in a multiaxial plan) to treat balance
disorders [e.g., benign paroxysmal positional vertigo
(BPPV), canalithiasis] caused by displaced canaliths
(otoconia) in the inner ear of the patient.
113 Capnography oxygen mask
A device intended to be placed over the nose and
mouth to deliver oxygen (O2) to a patient's airway
and to sample exhaled respiratory gases for
monitoring the patient’s ventilatory status.
S.No. Device Name
Intended Use
114 Capnography sampling
adaptor
The device is intended for sampling CO2 and use
with monitors enabled with capnography technology
.
115 Cardiac compression board
A flat, rigid device that is placed under a patient to
act as a support during cardiopulmonary
resuscitation.
116 Cardiopulmonary resuscitation mask
A flexible, form-shaped device that is placed over a
patient's mouth to administer "mouth-to-mask"
exhaled air from the user to the patient during
cardiopulmonary resuscitation (CPR).
117 Cardiopulmonary resuscitation
mouthpiece/cover
A non-sterile device that consists of a mouthpiece
with an integrated face/body covering sheet
intended to be placed over a patient during
cardiopulmonary resuscitation (CPR) to prevent
cross-contamination between patient and CPR
administrator. This is a single-use device.
118 Cast boot/shoes
A boot-like cover for a foot enclosed in a leg cast or
bulky bandages intended to support postsurgical
healing, injury/trauma care or any situation that
requires a pressure off-load of the foot.
119 Cast removal protective tape
To be placed between a cast stockinette and the
lining of a cast, along the line where subsequent
cast cutting is intended to occur during cast removal,
to prevent cuts and burns to the skin caused by the
cast cutter/saw.
120 Cast stockinette
Intended to be used with casting material to provide
under-cast padding and cushioning.
121 Cast walking heel
A rubber device designed to be attached to or
incorporated into a cast to support the patient while
walking.
122 Cerebral Cavernous Malformations
(CAM) vision stimulator
Intended for stimulation of the eye for treatment of
lazy/ambyopic eye
123 Cerebral Palsy (CP) walker
Intended as a walking aid for patients affected by
cerebral palsy, for postoperative rehabilitation, polio,
and other neurodevelopmental conditions
124 Cervical cytology scraper, reusable
Intended to scrape and retrieve cytological material
from the surface of the cervix (neck of the uterus) or
vaginal area for pathological examination and
diagnosis, often for the detection of cervical cancer.
S.No. Device Name
Intended Use
125 Cervical pillow
Intended to provide support to people who have
trouble keeping their spine straight and management
of neck/cervical pain
126 Cervical spine collar
Intended to support or immobilize the cervical spine
to treat deformities, fractures, sprains, or strains
(often to treat whiplash resulting from an
automobile accident).
127 Cervical traction head holder/head
halter
Intended to be used with cervical traction kit to halt
the head for physical support.
128 Cervical traction kit (with/without
weights)
Intended for traction of cervical region in conditions
such as neck or back pain, post surgery
rehabilitation, reducing neck pressure, etc.
129 Cervical traction spreader bar/
Spreader bar
Intended to serve as a component of a traction kit to
relieve pressure by stretching the musculature and
connective tissue components of the cervical spine.
130 Cervicothoracic spine orthosis
Intended to support or immobilize deformities,
fractures, sprains, or strains of the cervicothoracic
spine.
131 Cervicothoracolumbosacral spine
orthosis
Intended to encompass the
cervicothoracolumbosacral spine region of the neck
and trunk.
132 Chair cushion, pediatric use
A supportive seat shell designed for use with a
disabled child to allow them to sit independently on a
chair/swing/shopping trolley
133 Chair, flotation therapy
Intended to aid in the prevention of pressure injuries
as a result of reduced mobility and extended sitting
postures.
134 Chemotherapy spill clean-up kit
Intended to safely clean and dispose of
Chemotherapy drug spills.
135 Chest stand
Intended to provide a stable and adjustable platform
to position patients for accurate and high-quality
chest/full body X-ray examinations.
136 Chest/Limb Restrainer
Intended to stop or limit a patient's movement
137 Cholera bed
Intended as a physical suppport for severely ill and
weak patients including those suffering from cholera
138 Circulating-fluid thermal therapy
system
Designed to provide extended cold therapy to assist
in a variety of applications as directed by a medical
professional. It is designed not to be applied directly
to the skin and is to be used as part of recovery from
surgery either in the hospital setting or at home.
S.No. Device Name
Intended Use
139 Clamp for rubber dam
A dental device with a buccal and lingual wing, or
flange, used to anchor a rubber dam down to the
cervical region of an exposed tooth.
140 Clavicle support/brace/belt
Intended to support and limit the movement of
clavicle bone in case of injury, fracture, post surgery,
etc.
141 Cognitive test equipment
Evaluation and training in stereognosis, form
constancy, position in space, shape recognition and
sequencing as well as teaching compensatory
techniques for loss of visual field.
142 Cold compression therapy cervical
spine collar
Intended to facilitate, through cooling and
compression, the treatment of a variety of conditions
resulting from injury/surgery to the neck region (e.g.,
inflammation, stiffness, whiplash).
143 Cold/cool therapy gel
Intended for localized topical skin application to
provide a cooling effect for underlying muscles/joints
to reduce pain and swelling.
144 Collar and cuff arm sling
Intended to immobilize forearm, elbow, humerus or
shoulder injuries.
145 Colour discrimination non-
prescription spectacles
A non-prescription ophthalmic device intended for
use by people with color blindness to improve color
discrimination and blocks UVA and UVB radiation
from sunlight.
146 Colour discrimination tester
Intended to test a person's ability to differentiate
between colours.
147 Colour-discrimination eye chart
Intended for testing colour vision.
148 Compression bandage: two
component system
Intended for management/prevention of venous
insufficiency, chronic oedema, venous ulcers, etc.
149 Compression/pressure sock/stocking
applicator
Intended to assist in wearing/removing of medical
compression stockings.
150 Compression/pressure tubular
garment
To tightly fit over and apply compression/pressure
(i.e., graduated or even-force) to a part of a limb,
typically for the treatment/prevention of a disorder(s)
of circulation [e.g., venous insufficiency, deep vein
thrombosis (DVT)], to control scarring, and/or to
manage lymphoedema.
S.No. Device Name
Intended Use
151 Conductive abrasive gel for skin
preparation
Designed to reduce the resistance (impedance) and
defatting the skin in preparation for
electrophysiological examinations:
electroencephalography (EEG), long-term EEG
monitoring, polysomnography (PSG), evoked
potentials (EP), stress test ECG, ECG monitoring by
Holter in clinics, hospitals, diagnostic centers
152 Conductive electrode gel/paste
(adhesive/non-adhesive)
Intended to reduce resistance between skin and
electrodes during examinations such as ECG, EEG,
EMG
153 Congenital hip dislocation abduction
splint
Intended to stabilize the hips of a young child with
dislocated hips in an abducted position (away from
the midline).
154 Contact lens inserter/remover
Intended to insert and remove contact lenses from
the eye.
155 Continuous glucose monitor
retrospective data analysis software
Intended to analyze and correlate retrospective data
from a continuous glucose monitoring device.
156 Controlled ankle motion (CAM)
walker
Intended for preventing/limiting foot/ankle movement
in cases of serious sprain, surgery or removal of
cast
157 Core-body mechanical weight
exerciser
Intended to enable a patient with a lumbar spine
injury to perform controlled extension, contraction,
and/or twisting movements of the lumbar/thoracic
spine back region and the abdomen, for testing and
rehabilitation.
158 Corn/callus dressing, reusable
The device is intended to offer symptomatic relief of
physical foot disorders and similar conditions, by
providing a mechanical barrier which relieves the
effects of pressure and friction and or support.
159 Corn/callus dressing, single use
A non-sterile adhesive dressing intended for
protection and/or relief of corns or calluses
(hardened areas of skin on the feet or toes).
160 Cotton and rubber elastic bandage
It is and elasticated device intended to be worn on
limbs for managements of sprains, strains and
painful joints on intact skin.
161 Cotton crepe bandage
It is intended to be worn on limbs for managements
of sprains, strains and painful joints on intact skin.
162 Cottonoid pad
A device that is placed next to the parotid gland in
order to absorb the saliva that is being secreted.
This aids in dental procedures as it helps maintain a
working field dry.
S.No. Device Name
Intended Use
163 Cover for thermomete probe
It acts as a physical barrier between the patient and
an in-the-ear thermometer to prevent cross-
contamination between patients.
164 Cover, dental
To cover and protect a dental handpiece during
dental procedures and prevent cross-transmission of
infection between patients.
165 CPAP/BPAP chin strap, reusable
A device in the form of a strap or band intended to
be fastened around a patient’s head (vertically from
chin to crown) during positive airway pressure (PAP)
therapy to control mouth air leaks by gently holding
the mouth closed.
166 CPAP/BPAP chin strap,single use
A device in the form of a strap or band intended to
be fastened around a patient’s head (vertically from
chin to crown) during positive airway pressure (PAP)
therapy to control mouth air leaks by gently holding
the mouth closed.
167 Cranial orthosis, pediatric
A patient-matched helmet-like device, intended to be
worn on the head of an infant with an abnormal head
shape (e.g. due to plagiocephaly, brachycephaly) to
improve cranial symmetry/shape during growth over
a period of months.
168 Crash cart/Emergency recovery
trolley
Intended for use in hospitals for transportation and
dispensing of emergency medication/ equipment at
the site of a medical/surgical emergency for life
support protocols.
169 Crutches
Intended to help transfer of load from the legs to the
upper body of people who cannot use their legs to
fully support the weight of their body due to
temporary or permanent disabilities.
170 CT phantom, test object
Intended to establish, validate, and provide for
routine and periodic verification of patient imaging
CT system performance consistent with
internationally recognized standard.
171 Custom-made anatomy model
A model of the bone of a specific patient intended for
use to facilitate accurate preoperative planning and
introperative placement of a device during
orthopaedic surgical procedures.
172 Cytotoxic waste receptacle
Intended to safely deposit, collect and store
cytotoxic materials (e.g.,
chemotherapy/antineoplastic drugs).
173 Dappen dish, reusable
Intended to hold and mix dental materials.
174 Dappen dish, single-use
Used to hold and knead dental materials.
S.No. Device Name
Intended Use
175 Denis Brown splint
Intended to immobilize the foot, particularly on
young children with tibial torsion (excessive rotation
of the lower leg) or club foot.
176 Dental abrasive disk guard
A dental device intended to be fitted around an
abrasive disk when rotating at high-speed to protect
tissues not intended to come in contact with the
abrasive disk.
177 Dental abrasive disk, reusable
Intended for use in contouring, finishing, and
polishing dental restorations.
178 Dental abrasive disk, single-use
Intended for use in contouring, finishing, and
polishing dental restorations.
179 Dental abrasive strip, reusable
It is intended to be used for cleanup, polish or finish
dental crowns and bridges after cementation of
crowns, veneers, bridges, inlays or onlays.
180 Dental abrasive strip, single-use
It is intended to be used for cleanup, polish or finish
dental crowns and bridges after cementation of
crowns, veneers, bridges, inlays or onlays.
181 Dental amalgam mixer
This is a dental benchtop amalgamator that is
intended to mix by vibration/shaking, encapsulated
materials to be used for dental restorative
procedures in the mouth.
182 Dental articulation liquid
A range of pressure indicating material used to
locate areas of denture or removable dental device
discomfort in the patients mouth. Used for seating
crowns, bridges, full dentures, immediate dentures
and partial dentures.
183 Dental articulation paper
Intended for marking areas of contact between
opposite teeth, restorations or appliances.
184 Dental automated x-ray film
processor
A mains electricity (AC-powered) automated dental
x-ray film processor designed to be used in the
dental practice for the rapid processing of films.
185 Dental cheek retractor, home-use
This is a hands-free, self-retaining frame that is
intended for home use by a patient to retract the
lips/cheeks to facilitate viewing of the teeth.
186 Dental cotton roll
It is intended as an absorbent, hard-packed cylinder
(a roll) that is used as a saliva absorber from the oral
cavity during dental procedures. It may also be used
as a packing between the lip/cheek and the gum to
give better examination/operative exposure.
187 Dental cuspidor
This is a vessel, e.g. a round bow, into which the
patient may expectorate during and after dental
procedures.
S.No. Device Name
Intended Use
188 Dental floss
It is a tape/thread-like device intended for removing
plaque, food debris, etc. from between the teeth
and aid in prevention of dental caries
189 Dental implant abutment analog
A rod-like device intended to be used as a copy of a
dental abutment, for a dental laboratory working
model, to confirm the orientation/position of the final
dental abutment placement; it may be used during a
scanning procedure
190 Dental implant abutment impression
cap
A device intended to be placed intraorally on the
abutment head of a dental implant during dental
impression procedures, to facilitate the removal of
the impression material after setting and to ensure
the position of the implant in order to transfer it to a
working model for the dental technician.
191 Dental implant/prosthesis handling
forceps
This is a hand-held manual dental instrument
designed for grasping and holding a dental
devices(e.g., crown, bridge, screw) during its
application to a patient's oral cavity
192 Dental impression adhesive
A liquid material specifically made to give adhesion
of an impression material to the impression tray.
193 Dental impression material
Intended to take an oral impression.
194 Dental impression material kit,
reusable
A collection of non-sterile devices designed to obtain
a negative imprint of the teeth. The kit typically
includes dental impression materials and a dental
impression tray(s).
195 Dental impression material kit,
single-use
A collection of non-sterile devices designed to obtain
a negative imprint of the teeth. The kit typically
includes dental impression materials and a dental
impression tray(s).
196 Dental impression material mixer
An electric device used to mix impression materials
immediately before use at the chair side
197 Dental impression tray material
A material intended to be used to create a custom
impression tray intended for filling with dental
impression materials; it is not intended for the
fabrication of a patient-worn dental appliance.
198 Dental impression tray,
reusable
A impression tray is a metal or plastic device
intended to hold impression material, to make an
impression of a patient's teeth to reproduce the
structure of a patient's teeth.
S.No. Device Name
Intended Use
199 Dental impression tray,
single-use
A impression tray is a metal or plastic device
intended to hold impression material, to make an
impression of a patient's teeth to reproduce the
structure of a patient's teeth.
200 Dental laboratory CAD/CAM unit
This is a laboratory based device that is intended to
be used to aid in the computer-aided design (CAD)
and manufacturing/production (CAM) of dental
restorations in the dental laboratory.
201 Dental material application tool,
reusable
Intended to be used in dentistry for the application of
a dental material (e.g., composite resin, etchants,
etc.) to a dental prosthesis or to a patient’s
tooth/teeth or elsewhere in the oral cavity
202 Dental material heater
Intended to heat regular (paste) and flowable
composite material, including bulk fill composites,
compules, porcelain veneers and onlays
203 Dental material mixing surface,
reusable
Intended to be used as a surface to mix dental
materials.
204 Dental material mixing surface,
single-use
Intended to be used as a surface to knead or mix
dental material (impression material, cement, etc).
205 Dental matrix wedge
Used to wedge between teeth to create space when
applying restorative materials
206 Dental pick
Intended for the removal of plaque and/or debris
from the proximal surfaces of teeth and the gingival
surfaces of fixed prostheses.
207 Dental precision attachment
fitting/adjustment tool
This is a hand held dental laboratory instrument that
is intended to be used in a dental laboratory or
dental surgery during fabrication or revision of a final
dental prosthesis (e.g., denture, crown, bridge) to
integrate and/or adjust the fitting between male and
female precision attachments.
208 Dental resin applicator
Intended to be used for the application of a dental
restorative material (e.g., amalgam, resin) to a
patient’s tooth/teeth; the restorative material is not
included.
209 Dental rubber dam
Dental rubber dam is a stretchable sheet used to
isolate a field of work from the rest of the oral cavity
in dentistry.
210 Dental spatula, reusable
Used for kneading or mixing dental material
(impression material, cement, etc).
211 Dental spatula, single- use
Used for kneading or mixing dental material
(impression material, cement, etc).
S.No. Device Name
Intended Use
212 Dental suction system fluid-
separation unit
A separator used in the oral cavity for the separation
of fluids (saliva, blood) from gases to avoid liquids
from entering the suction pump (i.e., dry
suction).
213 Dental surgical microscope
This is a dental device that is intended to be used
to magnify and visualize minute intraoral structures
during dental examination and dental surgical
procedures in the mouth.
214 Dental torque wrench adaptor
It is used to enable connection of the main body of
the wrench to a specific endpiece(s).
215 Dental X-ray beam aligner
Intended to assist the dental radiographer position
the positioning indicator device relative to the tooth
and receptor
216 Dental X-ray film holder
It is used to hold the dental xray films and position it
to the right angle inside the mouth to aim taking an
accurate xray
217 Dental/oral lavage unit (water
flosser)
Intended to inject a stream of water on teeth to
remove dental plaque in between teeth and improve
gum health. It is intended as an OTC (home use)
device.
218 Dental-professional prophylaxis
brush/Dental polishing brush
A rotary dental instrument with natural or synthetic
bristle brushes intended to be used by a dental
hygienist and/or dentist to clean and polish teeth.
219 Dentifrice
Prophylaxis pastes without fluoride intended to be
used by professionals for tooth cleaning and
polishing procedure by abrasion/mechanical
cleaning only.
220 Dentures cup
This cup is intended to hold patient dentures when
they are not in the mouth.
221 Dermatome skin approximation tape
Intended for attachment onto a skin graft knife to
collect a skin graft tissue.
222 Dermatoscope
Intended for non-invasive, visual inspection of intact
skin by a healthcare professional.
223 Derotation bar with Traction
Intended for lower limb traction and aid in early
healing, correction of drop foot
224 Diagnostic condensing lens
Intended for use in binocular indirect
ophthalmoscopy to focus reflected light from the
fundus of the eye.
225 Diagnostic X-ray system positioning
frame
Diagnostic X-Ray System positioning frame
226 Diagnostic x-ray tube mount
Intended to support and to position the diagnostic x-
ray tube housing assembly for a medical
radiographic procedure.
S.No. Device Name
Intended Use
227 Dialysis bag clamp/clip
Intended to be used for opening or closing the
dialysis tube.
228 Dialysis set holder
To hold and secure in position a dedicated set being
used during a dialysis procedure.
229 Diaphragm wearable urinal
A non-sterile urine drainage device designed for
men with incontinence consisting of a front piece
with a scrotal support and a leak-proof, flexible
diaphragm through which the penis passes into a
closed coneshaped tube connected to a leg bag into
which the urine is collected.
230 Digital examination camera
Digital camera for close-up skin inspections or
general examinations.
231 Direct ophthalmoscope eye cup
The device is intended to be attached to the distal
(patient-contact) end of a direct ophthalmoscope
head to facilitate appropriate positioning around the
patient’s eye socket during an examination.
232 Disability vehicle, moped
A moped that provides mobility to a disabled person
233 Disposable Kelly pad
Intended to funnel blood to a collection device in
order to help detect postpartum hemorrhage
234 Dissection board
A device on which an object is placed for support
during dissection procedures
235 Douche kit
A packaged collection of items (douche solution
prefilled in a douche unit) used to provide cleansing
of the vaginal area of the user's body.
236 Drainage tube
A hollow device used as a conduit to remove fluids
or purulent material from a cavity, wound, or infected
area.
237 Drape adhesive
Intended to be placed on the skin to attach a
surgical drape.
238 Draping incontinence liner
A non-sterile padded sheet intended to cover a
device/piece of furniture (e.g., wheelchair, bed, sofa)
occupied by an incontinent person to catch and
retain urine.
239 Dressing retention garments
Intended for use in wet and/or dry wrapping and
dressing retention in the treatment of atopic eczema.
240 Drip wearable urinal
A non-sterile urine drainage device designed for
men to contain a moderate leakage consisting of a
front piece with an integrated tubular sheath that fits
fully over the penis and into which dribbling urine is
collected.
241 Dry powder inhaler
Intended to administer powdered medicine through
the mouth and into the bronchial airways.
S.No. Device Name
Intended Use
242 Dry salt therapy device
Intended to provide an inhalation of prefilled salt
particles through the mouth and into the bronchial
airways to induce natural self-cleansing
mechanisms that flush away the impurities from the
surface of cells and mechanically clean the airways.
243 Drying chamber for breathing circuits
A chamber designed for drying breathing circuit
equipment.
244 Ear bowl
Intended to fit the curvature of the head and sit
closely under the ear lobe to enable ear treatments.
245 Ear canal impression tray
Intended to hold and confine the impression material
in opposition to the surfaces to be recorded, and to
control the impression material while it sets to form
the impression of the ear canal.
246 Ear canal light
Intended to illuminate the ear canal.
247 Ear Irrigation kit
Irrigates the ear canal with solution to clear wax and
debris.
248 Ear microsystem needle
Intended to apply continuous pressure to stimulate
meridian points during auriculotherapy
249 Ear wick
Intended to minimize bleeding during ear surgery.
250 Elastic adhesive Bandage
It is intended to secure dressings and for strapping
on intact skin
251 Elasticated cohesive Bandage
It is intended to secure dressings and for strapping
on intact skin
252 Elbow immobilizer
A non-rigid external device, usually made of a fabric,
used to temporarily render the elbow immovable to
support healing of an injury or a surgical wound.
253 Elbow orthosis
A prefabricated (non-customized) externally applied
and wearable orthopaedic appliance designed to
encompass the elbow joint to support, align, prevent,
or correct deformities/injuries or to improve function
of the elbow.
254 Electric dental chair without
operative unit
A mains electricity (AC-powered) device designed to
support a patient in a seated position to facilitate
dental examination, treatment, and/or minor surgery
procedures. It does not include operative unit.
255 Electric stander
Standing frames for those with standing difficulties
256 Electric toothbrush
Intended to remove plaque, tartar, food debris, etc.
from teeth and gums and aid in prevention of dental
caries, etc.
S.No. Device Name
Intended Use
257 Electrical-only medical device
connection adaptor
This device connects incompatible plugs/sockets on
related medical devices (e.g., video control centers,
peripheral devices and controller) and/or cables and
relates the control signals.
258 Electronic occlusion spectacles
Intended to test and train vision for conditions where
decreased visual acuity may be due to unequal
vision in the eyes.
259 Electrosurgical electrode holder
A device intended to be used to house active hand-
controlled electrosurgical electrodes (pencil
electrodes) when not in use during the surgical
procedure.
260 Emesis bag
A disposable device intended to collect and contain
a patient's vomit and eliminate exposure to vomit
until it is convenient to dispose of the bag and its
contents.
261 Endaural wax extraction curette
This device is intended to simply extract ear wax
using a soft silicone tip that expands in ear cavity to
aid extraction.
262 Endoscope calibration cover,
reusable
A plastic cap/cover designed to be placed onto the
distal end of an endoscope to allow white balance
and/or 3-D adjustment to be calibrated in the
electronic control unit of the endoscope. This is a
reusable device.
263 Endoscope channel cleaning valve
An adapter or valve to be fitted to an endoscope for
cleaning purposes
264 Endoscope cleaning brush
A device used to remove dirt or debris from the
working channel (s) and irregular surface of the
endoscope prior to disinfection and sterilization.
265 Endoscope collector
A device intended to collect the excreted endoscopic
video capsule camera.
266 Endoscope flushing adaptor, single-
use
An adapter intended to be fitted to an endoscope
air/water channel for flushing of the endoscope
during post-procedural ‘precleaning’ step of
endoscope reprocessing after endoscopic
procedure.
267 Endoscope leak tester tubing
Leak tester tubing for use during endoscope leak
test.
268 Endoscope/instrument holder,
manual-adjusted
A manually adjusted device intended to be used to
hold, maintain in a desired position, and enable
positional adjustment of an endoscope (e.g.,
laparoscope) and/or associated instruments/devices
during an endoscopic examination/surgical
procedure.
269 Endoscopic irrigation/aspiration foot-
pedal
A manual, foot-operated device intended to be used
in conjunction with an irrigation tubing set to facilitate
delivery of sterile irrigation fluid.
S.No. Device Name
Intended Use
270 Endoscopic video image interpretive
unit
The device is intended to be used by trained
clinicians as an adjunct to video Endoscopy for the
purpose to highlighting regions with visual
characteristics consistent with different types of
mucosal abnormalities.
271 Endoscopic video image processor
A unit designed to receive the electronic signals sent
from a video endoscope and may/may not
compensate for/enhance the colour and light
qualities provided.
272 Endotracheal tube holder
Intended to provide a reliable stabilization of
endotracheal tube after intubation
273 Endotracheal tube introducer
Intended to help with the insertion of an
endotracheal (ET) tube into the airways of a patient
during intubation
274 Enema bag
Intended to be used to hold an enema solution for
administration into the rectum
275 ENT chair, electric
Intended to support a patient in a seated position,
electrically control position in a way to facilitate ear,
nose, throat examination, treatment, and/or minor
surgery.
276 ENT chair, mechanical
Intended to support a patient in a seated position,
mechanically control position in a way to facilitate
ear, nose, throat examination, treatment, and/or
minor surgery.
277 ENT surgical microscope
Intended to magnify minute structures (e.g., nerves,
vessels) in the performance of ear, nose, and/or
throat (ENT) surgery requiring high magnification
and adjustable focusing.
278 ENT transilluminator
Intended to illuminate sinus tissue during an
ear/nose/throat (ENT) procedure to render the tissue
translucent for examination
279 Enteric stomal shield
A non-sterile cap intended to be placed over an
enteric stoma to protect it from harmful external
influences (e.g., knocks, friction).
280 Enuresis alarm
The device that activates a signal when a small
quantity of urine is detected by a sensing
mechanism to help a child with training the bladder.
281 Epileptic seizure alarm
Intended to alert or notify healthcare professional or
next of kin when an epileptic patient (often pediatric)
gets a seizure attack.
282 Euthyscope
Intended to project a bright light encompassing an
arc of approximately 30 degrees on the fundus of
the eye for the treatment of amblyopia.
S.No. Device Name
Intended Use
283 Exothermic heat therapy pack
Intended to be applied to the body surface,
sometimes with pressure, to provide heat therapy to
reduce muscle spasms and cramps and/or for joint
and muscle stiffness and pain.
284 External asembled lower limb
prosthesis
Intended for use as an preassembled external
artificial limb for the lower extremity.
285 External limb prosthesis socket liner
Intended for use as an interface that goes between a
person's skin and his or her prosthetic to protect the
wearer's skin while enhancing comfort and
maintaining a more consistent fit.
286 External lower-limb prosthesis
socket
A component of an external lower limb prosthesis
designed to fit over a residual limb after amputation
at any hip, knee, or ankle-foot level, and intended to
function as the limb-prosthesis interface and to
provide support, stabilization and suspension.
287 External nasal splint
Intended for external use for immobilization of parts
of the nose.
288 External silicone breast prosthesis
Intended to restore normal contour and improve
function for patients who have experienced partial or
total loss of their breast to traumatic injury, disease
or due to surgical removal of the nose (rhinectomy).
289 Extractor
Used to aid in the removal of foreign objects from
superficial skin tissue.
290 Eye cup
Intended to fit around the eye socket and placed
over the eye to allow washing of the affected eye.
291 Eye irrigation shield
Intended for use with an eye irrigation kit/system to
direct irrigation solution to the surface of the eye and
allow the solution to gently lavage the surface of the
eye.
292 Eye pad, non sterile
Intended for use as a bandage over the eye for
protection or absorption of secretions.
293 Eye patch, non-sterile
A device with a synthetic origin used to close/cover
one or both eyes, for the purpose of protecting the
eye after a procedure or as part of an ophthalmic
therapeutic intervention.
294 Eye patch, non-sterile
Intended to provide temporary occlusion for the eye
in the treatment of Amblyopia, diplopia or as a
occluder for vision or field testing.
295 Eye spud
Intended to remove a foreign body/object embedded
in or adhering to the surface of the eye globe.
S.No. Device Name
Intended Use
296 Eyelid clamp
Intended to automatically grasp and hold the eyelid
during an ophthalmic surgical intervention.
297 Eyelid weight, external
Intended to be applied to the outside of the upper
eyelid to "lidload" the eyelid to restore upper eyelid
muscle function.
298 Face shield
Intended to protect the wearer's entire face (or part
of it) from hazards such as chemical splashes (in
laboratories or in industry), or potentially infectious
materials (in medical and laboratory environments).
299 Feeding nipple
Intended for use as a feeding nipple for infants with
oral or facial abnormalities
300 Felt tangent screen
Intended for assessing the extent of the patient's
peripheral visual field by mapping the visual
response to a test object moved from the periphery
towards the centre of the screen.
301 Female wearable urinal, reusable
A non-sterile, urine drainage device designed for
women that typically consists of a flexible tube
attached to a collector that is specially formed to
securely fit around the female genitals to provide a
route to channel urine, via a tube, into a collection
bag.
302 Fiber cast bandage
Intended to be used as a medical dressing made of
synthetic materials to support and protect bones that
have been broken or joints that need stabilization.
303 Filter for dental suction equipment
Filter that ensures debris does not enter dental
suction equipment
304 Filter for gas central supply system
Designed to remove particulates and volatile organic
compounds (VOCs) from the gas flowing into
incubators used in IVF procedures.
305 Finger and toe bandage
An elastic, tubular device intended to be worn
over/around the limb
(arm/leg) and/or finger of a patient affected by a skin
condition/disorder (e.g., eczema, wound, fragile
skin) to protect the skin typically after treatment with
a medication (e.g., ointment, cream), and/or to
support a dressing over a wound/injury.
306 Finger Cot/brace
Intended to support, align, prevent or correct
deformities/injuries or to improve function of finger
S.No. Device Name
Intended Use
307 Finger/hand exerciser
Intended primarily to strengthen the muscles of the
fingers and hands through repetitive finger/hand
motion against resistance from elasticity or density
(post fracture/stress/weak forearm and grip
strength).
308 Finger/thumb prosthesis
Intended as an artificial substitute for a missing
finger or thumb
309 Flatus tube, single-use
A soft and pliable tube intended to be inserted
through the anus to relieve colic in infants. It is a
single-use device.
310 Flex bar
Intended to increase tendon strength and reduce
elbow pain for those who suffer from tennis and
golfer's elbow.
311 Flotation therapy bed, adult
A mains electricity (AC-powered) bed designed to
minimize pressure points on a patient's body by
providing contact with as much of the body surface
as possible, typically through a mattress that
contains a large volume of constantly moving media,
e.g., water, air, or mud that lifts the patient to
simulate a floating effect. It is used in cases of
decubitus ulcers or where a patient has little
remaining body fat and the displacement of body
weight is vital for treatment and/or comfort.
312 Flotation therapy bed, neonatal
A fixed (non-adjustable) device designed with a
mattress or cushions containing air, water, gel, or
other appropriate material used for the continuous
care of newborns, and sick and/or premature babies.
313 Fluid delivery mount, general-
purpose
The device is designed to hold the fluid bags when
gravity irrigation is needed for scope surgery.
314 Foot traction kit
It is used in relieving pain due to muscle spasm,
maintaining the limb in a position of comfort and to
restore and maintain the alignment of bone due to
fracture or dislocation.
315 Footstool, conductive
A low stool for standing on to compensate for height
differences. It is intended to be used in, e.g.
examination rooms or operating theatres.
316 Forceps, dressing
A surgical instrument with scissor like handles used
to grasp dressings, e.g. cotton swabs, towel, lints
317 Forceps, sterilizer transfer
Forceps with specially designed blades used to
handle sterile instruments or implants.
S.No. Device Name
Intended Use
318 Fornixscope
Intended to provide indirect access and viewing of
the upper conjunctival fornix and inner surface of the
eyelid as an alternative to eyelid eversion.
319 Fresnel lens
Intended to be applied to the back of spectacle
lenses to focus light to a focal point to help manage
various vision conditions.
320 Fresnel prism
Intended to be applied to spectacle lenses to give a
prismatic effect typically to manage strabismus or
other eye muscle dysfunction.
321 Fundus camera
It is intended to capture and document digital
images of the posterior and anterior segments of the
human eye.
322 Fundus camera, pediatric
Intended to capture digital color images of the retina
of infants having less than 55 weeks of Post
Menstrual Age (PMA), while there eyes are dilated
using mydriatic drops under clinical control.
323 Gait rehabilitation electronic walker
A mobility and transfer device that helps a health
care professional raise a patient to a standing
position, providing support for standing or
ambulation, or lifting a patient in the seated posture
and transferring them between wheelchairs, chairs,
toilets, beds or therapy tables.
324 Gait Trainer
It is used to assist and train a person with physical
disabilities to relearn/improve walking.
325 Gait trainers-Unweighing System
Intended for partial weight-bearing therapy in
rehabilitation
326 Gas sampling line filter, reusable
The device is indicated for adult, pediatric,
infant/neonatal patients for removal of moisture from
the gas sample.
327 Gastric pH strip
pH strips to indicate the pH level of human gastric
aspirate
328 Gel base for metallic neuro head
frame
It is used for providing support patient's head in a
suitable and stable position (e.g., supine, lateral, or
prone) during intracranial and spinal surgeries.
329 Gel knee ring
Intended to be worn around the knee for pain relief
and immobilization
S.No. Device Name
Intended Use
330 Gel-filled bed mattress
A large, durable, fabric case filled with various
densities of foam that support a gel structure,
designed to be placed on a bed mattress support
platform to provide a comfortable surface on which
an occupant can lie, rest, and sleep, and intended to
distribute body weight over a large surface area
typically to help relieve pressure points for comfort
and improve blood circulation to prevent pressure
sores.
331 Gel-filled bed mattress overlay
A gel-filled pad or thin mattress, usually full-body
length, which is placed on top of a bed mattress to
distribute an occupant's body weight over a large
surface area, typically to relieve pressure points for
comfort and to prevent pressure sores.
332 Gel-filled chair cushion
A gel-filled pad designed to be placed on the seat of
a chair to distribute an occupant's body weight over
the surface of the buttocks, typically to relieve
pressure points for comfort and to prevent pressure
sores.
333 General-gas low pressure tubing,
reusable
Intended to enable nitrogen or instrument air to be
delivered to the console from the wall in the hospital.
334 General-purpose diagnostic x-ray
phantom, anthropomorphic
A device that consists of preserved human or animal
tissue, or a two or three-dimensional (3-D) tissue-
equivalent model designed to mimic the functional,
physical, or a combination of these characteristics of
a normal or diseased human organ or organ
systems. It is a quality assurance (QA) device
intended to permit both subjective and quantitative
evaluation of the overall function of both
conventional and digital general-purpose x-ray
systems and associated image acquisition, display
and analysis computer programs.
335 General-purpose diagnostic X-ray
phantom, test object
Intended to provide for routine and periodic
verification of the energy, hardness and intensity of
an X-ray beam
336 General-purpose manually-operated
operation table
A mobile, manually-operated hydraulic table
designed to be adjusted to support a patient during
many types of surgical interventions.
337 General-purpose
diagnosis/treatment table
Intended for use in general diagnosis and
procedures in examination room to aid patient
examination.
338 Gingival barrier (non-medicated)
Intended for protection of gingival area during tooth
bleaching process
S.No. Device Name
Intended Use
339 Gonadal radiation protector
Intended to attenuate unnecessary radiation
exposure in diagnostic, medical or dental
procedures, and shield the gonad of the patient and
the operator.
340 Gonio lens
Intended for viewing the anterior chamber angle and
retina
341 Gravity enema set
Intended to deliver an injection of fluid (e.g., saline
solution) into the rectum to facilitate evacuation of
the large intestine.
342 Grip-enhancing exoskeleton
Intended to improve grip strength and hand function
for people (from children to elderly people) with
impaired hand function. The condition may be (but is
not limited to) due to orthopaedic or neurological
problems, from birth or acquired. The device mimics
the user’s grip movements and supports the users’
natural grip.
343 Gynaecological bib
Intended to be worn by the patient during
gynecological examinations and procedures.
344 Gynaecological operating table top
Intended to support a patient during gynaecological
surgical procedures
345 Gynaecological operating table,
electrohydraulic or
electromechanical or hydraulic
Intended to support a patient during gynaecological
surgical procedures
346 Gynaecological surgical microscope
Intended to improve visualization of anatomical
structures via
transmitted light during gynaecological surgery
(e.g., on the fallopian tubes).
347 Gynaecological
examination/treatment table
Intended to support a woman's body in the
appropriate positions during gynaecological
examinations.
348 Haidinger brush imager
Intended to produce an image which facilitates
visual function evaluation, particularly the macular
integrity.
349 Hallux valgus metatarsophalangeal
orthosis
A prefabricated (non-customized) non-rigid, home-
use device intended to be worn on the foot to help
prevent and/or reduce hallux valgus deformity,
thereby reducing walking pain and bunion formation.
350 Hand exerciser table
This modality exercise is meant for rehabilitation of
the hands, fingers, and wrist disorders injuries &
fractures.
351 Hand/finger splint
Intended to immobilize an injured hand to protect
from injuries to the digits, metacarpals, and wrist
during the healing process.
S.No. Device Name
Intended Use
352 Hand-carried stretcher
Intended to carry patient as an assistance for injury
or disability.
353 Handgrip cover for laryngoscope
Intended to be placed over the reusable
laryngoscope device during intubation to create a
barrier in order to reduce the cleaning time and
turnaround time of the laryngoscope blade for reuse.
354 Handgrip cover for surgical light
Intended to provide physical barrier on a handgrip to
prevent cross-contamination between the operating
room surgical lights (non-sterile) and the operating
room personnel.
355 Handgrip cover for vein illumination
device
The cover is intended to isolate vein illumination
devices and simplify cleaning after use.
356 Hand-held telescope
Intended to enlarge images for a visually impaired
patient/person.
357 Hand-held urinal, female
Intended as a container/set to be directly urinated
into by a female (typically bed-bound) patient for
excretory purposes.
358 Hand-held urinal, male, reusable
Intended as a container to be directly urinated into
by a male (typically bed-bound) patient for excretory
purposes.
359 Head marking cap
A head covering designed as a cap intended to be
worn on the head of a patient to allow a healthcare
professional to mark specific reference points on the
head during a procedure [e.g., transcranial magnetic
stimulation (TMS)] as an alternative to marking the
scalp directly or to avoid cutting the patient’s hair.
360 Head reflector
A head-worn device consisting of a circular concave
mirror (reflector) attached to a headband designed
to reflect light to a site of examination/minor
intervention (e.g., the nose, throat, or body surface)
on a patient.
361 Head Rest/Ring
Intended to position and protect a patient's head,
face and neck during prolonged surgeries
362 Heal cushion/Spur pad/cup
Intended to worn on the heel to relieve heel pain,
plantar fasciitis, foot pain, pain during walking,
calcaneal spur, sprain and strain of the foot, ankle,
feet and knee pain.
363 Heat-sealing device
Intended for medical purposes that uses heat to seal
plastic bags containing blood or blood components.
S.No. Device Name
Intended Use
364 Heel stirrup
Intended to position and support the patient’s foot,
lower leg and upper leg in a variety of surgical
procedures including, but not limited to gynecology,
urology, laparoscopy, general and robotic surgery.
365 Hip pad positioning device
It is used to support the patient hip during prolonged
surgeries.
366 Hip protector (Belt)
Intended to be worn by older adults or patients who
have balance/vision problems, bone conditions, or at
risk of falling to prevent hip injuries and fractures.
367 Hip shield
A mechanical guard worn over the hip to reduce risk
of injury in the case of a fall
368 Horizontal non-powered traction
system
Intended to be attached to a table for the application
of constant horizontal traction forces to the cervical
or lumbar vertebrae of patient by means of attached
harnesses.
369 Hospital Backrest
Intended to support patients with spinal problems
and pain management
370 Hruby fundus lens
Intended for use in the examination of the vitreous
body and the fundus of the eye under slit lamp
illumination and magnification.
371 Hydraulic/pneumatic external knee
prosthesis
This device is intended for transfemoral amputees,
is a prosthetic device designed to restore some
function of an anatomical knee joint.
372 Hydrocollator pack
Intended to be applied to the body surface to provide
heat therapy to reduce muscle spasms and cramps
and/or for joint/muscle stiffness and pain
373 Hydrotherapy equipment/tank
Used for rehab treatment requiring resistance
created by water to aid in injury recovery, improve
blood circulation, and decrease inflammation,
speeding recovery time
374 Hydrotherapy treadmill
Intended for use in partially immersed in water, e.g.,
in a hydrotherapy tank, to provide additional
resistance to the treadmill walking exercise without
increasing the impact and/or stress on the patient's
joints.
375 Hypothermia-prevention cap
Intended to prevent heat loss from the patient’s head
during surgery and under anesthesia.
376 Ice bag
Intended to alleviate pain and/or promote healing in
minor injuries of the body or for application around
the neck or limbs.
S.No. Device Name
Intended Use
377 Image display monitor mount
A dedicated mount onto which a visual display unit
(VDU), e.g., a patient monitor, can be placed and
secured so that it is safely mounted in a position
suitable for viewing. manufactured for hospital use.
378 Incentive spirometer
Intended to be used in respiratory therapy to
encourage and motivate deep-breathing
manoeuvres, typically for the postsurgical treatment
and prevention of atelectasis (lung collapse) and to
help facilitate airway opening and clearing.
379 Incontinence device suspender
A device used by a person with a disability to secure
a body-worn incontinence device [e.g., a nappy
(diaper) or a pad] in a stable and safe position
380 Incontinence penis clamp, reusable
Intended to gently compress the penis, either the
proximal shaft or the glans penis, to occlude the
urethra and prevent involuntary urination/dribbling
for men who are incontinent of the bladder.
381 Indirect binocular ophthalmoscope
An ophthalmic instrument designed to examine the
interior of the eye allowing the examiner to clearly
see a wide angle, stereoscopic impression of the
details of the fundus (retina) and other structures.
382 Infant acoustic earmuffs
A device consisting of two ear-covering cups (one
for each ear) intended to provide noise insulation for
neonates/infants in a special care unit.
383 Infant cap
To be used with an infant nasal CPAP respiratory
support breathing system to act as a stabilizing head
strap.
384 Infant care table
Intended for use in nursing, e.g., washing or
changing of nappies, of newborn babies.
385 Infant incubator warming hood
Intended to provide warmth to an incubator chamber
by placement above the chamber.
386 Infant inguinal hernia truss
Intended to be worn over the groin to prevent
protrusion of abdominal contents in an infant with an
inguinal hernia.
387 Infant limb immobilizer, reusable
Intended to temporarily render parts of an infant's
body immovable, e.g., the arms and/or feet while the
patient undergoes therapeutic or diagnostic
interventions.
388 Infant limb immobilizer, single- use
Intended to temporarily render parts of an infant's
body immovable, e.g., the arms and/or feet while the
patient undergoes therapeutic or diagnostic
interventions.
S.No. Device Name
Intended Use
389 Infant oral pacifier
To be used to calm the premature infant for longer
duration specifically during Nasal CPAP or when
infant is intubated and on mechanical ventilation. For
hospital use only.
390 Infant resuscitation table
It is a patient platform that can be used to facilitate
neonatal transition at birth. It provides a stable
platform for mother side resuscitation and
stabilization practices.
391 Infant sleep positioner
Intended to modify the sleeping position/posture of
infants to prevent deformational plagiocephaly, a
flattening of the back of the skull from a consistent
back-sleeping position.
392 Infant-hammock bed mattress
Intended to be placed in a
cot/crib/bassinet/bed/incubator and to cradle a
young infant during sleep/rest, and can used for
phototherapy, transportation and burns patients.
Intended to help reduce the risk of infant
injuries/disorders such as suffocation, flat head
(plagiocephaly), sudden infant death syndrome
(SIDS), pressure sores, and hyperthermia.
393 Infant-hammock bed mattress
overlay
Intended to be placed on a cot/crib/bed mattress and
to cradle a young infant during sleep/rest to help
reduce the risk of infant injuries/disorders such as
suffocation, flat head (plagiocephaly), sudden infant
death syndrome (SIDS), and hyperthermia.
394 Inflatable hot/cold therapy
pack/manual pump
A non sterile assembly of devices that are intended
to be applied with pressure to a body surface to
provide cold or heat therapy to that surface and/or
underlying tissue, e.g. the muscle. This device
typically consists of a cold/heat therapy pack, air
inflation sleeve and a manual air pump with
connection tubing. It is reusable and intended in
both the professional and home setting.
395 Inflatable limb splint
A sleeve placed around an arm or leg and inflated to
immobilize the limb.
396 Inflatable patient transfer sliding
mattress, single-use
An inflatable device designed to support a
recumbent patient on a bed or table and assist in the
physical movement of a patient for safer, easier
transfer and positioning in Trendelenburg and
Reverse Trendelenburg.
397 Infusion line clamp
A device intended for the clamping of medical tubes.
S.No. Device Name
Intended Use
398 Infusion line securement clip, single-
use
The device is intended to hold tubing to avoid
kinking of the tube during infusion and patient
moving.
399 Inguinal hernia belt (Adult)/surgical
truss, reusable
A reusable device Intended to relieve discomfort and
hernia pain caused by an inguinal or sports hernia
400 Inguinal hernia belt (Adult)/surgical
truss, singe use
A single use device intended to relieve discomfort
and hernia pain caused by an inguinal or sports
hernia
401 Injection site desensitizer, tactile
A non-sterile, noninvasive reusable device intended
to be used to manually apply tactile pressure stimuli
to a skin-surface injection site, prior to/during
injection, to reduce the pain/anxiety/stress
associated with injecting.
402 In-line arterial blood sampling set
A collection of devices designed to obtain an in-line
arterial blood specimen while maintaining a closed
system.
403 In-line backflow valve
A general-purpose device used in medical tubing or
pipelines to prevent the backflow of gases or liquids.
404 Instrument lubricant
Intended to be applied to a surgical instrument prior
to sterilization for purposes of
lubrication/maintenance (e.g., to prevent moving or
mating parts from sticking together during use,
inhibit corrosion, and to improve longevity of the
instrument)
405 Integration Unit for Flat panel x-ray
detector
A portable, electrically-powered device intended to
enable a direct flat panel detector (not included) to
be integrated as part of an analogue x-ray [i.e.,
screen-film, computerized radiography (plate)]
system to convert the analogue x-ray system into a
digital x-ray system (retrofit) using the same
modality components.
406 Intestinal ostomy base plate
The base plate is intended to fix the ostomy bag and
adhere to intact skin around the stoma.
407 Intestinal stoma shield/support belt,
reusable
A non-sterile plate intended to be placed over an
enteric stoma to protect it from harmful external
influences (e.g., knocks, friction), to reduce the risk
of stomal herniation, or to help maintain adhesion of
the base plate to the skin.
408 Intraoperative chest positioner
A preformed reusable pad designed to be attached
to an operating table for stabilizing a patient's chest
to adequately position, support, and reduce pressure
on the bony prominences of the body part during
intraoperative procedures
S.No. Device Name
Intended Use
409 Intravenous access monitoring
system, line-disconnection
An alarm system that connects to a sensors
around/on a venous puncture site dressing or
catheter connection to detect blood or disconnection
of the catheter.
410 Intravenous catheter holder, non-
sterile
It is a skin adhesive device for the securement of
urinary catheters and medical tubing as part of
ongoing catheter care.
411 Intravenous hanger, wall-mounted
The device is intended to be used within the medical
application to hold infusion bags within the approved
load range.
412 Intubation teeth protector
A device designed to fit over the upper and lower
sets of teeth to protect them from damage during
endotracheal (ET) tube intubation procedures.
413 Inversion table
Intended to support the body of a patient and
provide traction for the back muscles and spine by
allowing the patient manually invert their entire body
in a supine position (feet up and head down),
thereby assist in alleviating back pain.
414 Irrigation kit
To irrigate the nasal passages of infants and young
children.
415 Irrigation tip, surgical mist
A device that is designed to connect to an air/water
syringe to blow a mixture of air and/or water into a
patient's mouth during dental procedures.
416 Jaw exerciser
Hard gum is used to exercise jaw muscles for
temporomandibular disorders, increased saliva
production, helping to improve efficient food chewing
and swallowing.
417 Kinesiology tape
used to support and relieve pain in muscles, joints,
and/or ligaments. It reduces swelling, increases
mobility and enhances recovery.
418 Kinetic bed
Intended to enable continuous change of the
patient's lying position.
419 Knee cap/brace/wrap
Intended to limit the mobility of the knee to relieve
pain in conditions of arthritic pain, ligament/muscle
injury, surgical procedures, fractures, sprains, etc.
420 Knee ice wrap
Intended to relieve pain, swelling and inflammation
in the knee due to ACL, arthritis, sports injuries,
sprains, strains, muscle fatigue and surgery.
421 Knife handle
A device that is an interchangeable component of a
scalpel and that functions as a handle designed to
mount a compatible blade.
S.No. Device Name
Intended Use
422 Knife sharpener
A device used to create or re-establish a thin, keen
edge or fine point on an instrument used to cut, such
as a knife
423 Laparoscope laser adaptor
Intended to connect the laparoscope to the laser or
the laser arm for laparoscopic laser treatment.
424 Laparoscopic surgery handle
A manually operated handle of various
configurations to hold instruments used in
laparoscopic surgery.
425 Laparoscopic tray
Intended for holding laparoscopy instruments during
a laparoscopic intervention or sterilization.
426 Laryngeal airway introducer
A device intended to aid insertion of a laryngeal
airway into the pharyngeal cavity of a patient while
reducing or eliminating the need for finger
manipulation within the mouth.
427 Laser protection eyewear
Intended to be used by the staff in an operating or
treatment room where a laser is being used, in order
to protect their eyes from injury during the laser
surgery.
428 LED transilluminator
Intended to provide optimal LED light in the
incubator
429 Leg intraoperative repositioning
support
A mechanical device intended to be used to help
support the weight of a patient's leg when the leg
needs to be repositioned during an orthopaedic
procedure. It is designed to be attached to the
operating table.
430 Lens system, high magnification
A device (lens) used to observe the retina or anterior
chanber of the eye.
431 Lice comb
Removes head lice eggs and nits and can be used
in conjunction with all lice treatments
432 Light beam patient position indicator
Intended to project a beam of light (incoherent
light or laser) to determine the alignment of the
patient with a radiation beam.
433 Limb/torso/head restraint, reusable
The devices are designed to position and support
patients in a variety of surgical procedures including,
but not limited to gynecology, urology, laparoscopy,
general and robotic surgery during Reverse
Trendelenburg positioning.
434 Liquid barrier dressing, dental
Protection of gingival area during tooth bleaching
process
435 Liquid crystal vein locator
Intended to indicate the location of a vein by
revealing variations in the surface temperature of the
skin.
S.No. Device Name
Intended Use
436 Lower limb exerciser
Intended to provide above and below knee
amputees the ability to improve muscle and
strengthen residual limbs
437 Lower limb sliding exerciser, non-
powered
A non-powered device intended to encourage active
movement of a patient’s lower limb(s) by assisting
flexion/extension at the knee for rehabilitation. This
is a reusable device.
438 Lower-limb orthosis
The device is an externally-applied wearable
appliance intended to support and correct
deformities of the leg and/or feet. It may be one of
numerous different designs and is typically made of
synthetic and textile materials. This is a single-
patient reusable device.
439 Lower-limb warming sock/stocking,
single-use
Intended to prevent heat loss during surgery and
under anaesthesia. The patient’s lower extremity is
placed in the legwarmer which is equipped the same
way as socks.
440 Maddox trial lens
Intended as a device that change the size, shape,
and colour of an image to dissociate the eyes in the
evaluation of eye muscle dysfunction.
441 Magnetic field shielding material for
cochlear implants
An accessory intended to be used on Cochlear
implant recipients to prevent implant magnet
dislodgement during MRI scans. The device is
placed externally on the recipient's head by a
healthcare professional prior to undergoing an MRI
procedure.
442 Mammographic x-ray phantom, test
object
A device that consists of a uniform density or a two
or three-dimensional (3-D) pattern of objects,
materials or openings of varying dimensions. It is a
quality assurance (QA) device used in both
subjective and quantitative evaluations of various
calibration and performance characteristics of
dedicated digital or conventional diagnostic x-ray
mammography systems or general-purpose x-ray
systems adaptable for mammography applications.
443 Mammographic x-ray system
compression paddle
A mechanically, electronically or software-controlled
component of a mammographic x-ray system used
to flatten and stabilize the breast during
mammography studies or mammography system-
guided biopsy procedures.
S.No. Device Name
Intended Use
444 Mandible-repositioning sleep-
disordered breathing orthosis
A removable intraoral device designed to alleviate
sleep-disordered breathing conditions (e.g., snoring,
obstructive sleep apnoea) by repositioning and/or
controlling the lower jaw (mandible), typically in a
downward and forward position.
445 Manual adjustable hospital bed
Intended for medical purposes that consists of a bed
with a manual mechanism operated by an attendant
to adjust the height and surface contour of the bed.
446 Manual dental burnisher, reusable
Used to polish the surface and smooth the margins
of restorations during dental procedures.
447 Manual dexterity assessment
software
The software is intended to detect micrographia
(smallness of writing), bradykinesia (slowness of
movement) and tremor with the
help of an off-the-shelf tablet and a digital pen. This
is not intended for self-assessment.
448 Manual dexterity test
Dexterity devices intended to evaluate motor
coordination, depth perception and eye-hand
coordination.
449 Manual radionuclide applicator
system
Intended to apply a radionuclide source into the
body or to the surface of the body for radiation
therapy.
450 Manual stander
A manually-powered, mobile device designed to
support and lift a disabled patient to an upright
standing position to provide the health benefits
associated with standing (e.g., improved breathing,
circulation, and flexibility); it is not intended to
facilitate gait.
451 Manual toothbrush
An instrument designed to be used by a patient to
manually brush clean their teeth and gums (i.e.,
remove plaque, tartar, stains, etc) and/or aid in
prevention of dental disorders such as caries.
452 Manual wheeled stretcher
Intended to transport patients in a horizontal position
as an assistance to injury or disability.
453 Manual-aperture-control diagnostic
x-ray system collimator, non-
motorized
A standard, non-motorized, diagnostic x-ray beam-
limiting device whose aperture size/length/shutter
assembly must be manually adjusted in order to
match the size of the x-ray beam to the size of the x-
ray cassette in use. An x-ray collimator is used to
limit the effects of scattered radiation on image
quality and to provide patient protection by
eliminating exposure to non-target body areas.
S.No. Device Name
Intended Use
454 Marker, dental use
This device range helps mark plaque areas on a
tooth and is thus intended to be used to aid
disclose/make marks that allows identify dental
plaque.
455 Mask strap
A head strap used to hold and stabilise a CPAP or
BiPAP face mask onto a patient's face.
456 Massage table/couch
Intended for supporting body during therapeutic
massages (not intended for general wellness
purpose).
457 Maternity belt
It is used to support the belly bump during the
different prenatal periods to avoid forward &
downwards stomach pulling, provide maximum
support and relieve back & pelvic ache.
458 Mechanical dental chair without
operative unit
Intended to support a patient in a seated position to
facilitate dental examination, treatment, and/or minor
surgery procedures. It does not include operative
unit.
459 Mechanical-resistance external
ankle-foot external prosthesis
External prosthetic components for the purpose of
replacement for a foot/ankle. The foot/ankle is
intended to be used for a lower limb external
prosthetic system
460 Mechanical-resistance external knee
external prosthesis
An externally applied component for use in
replacement of the knee joint (or other) in persons
with limb amputation or difference.
461 Mechatronic external ankle-foot
prosthesis
The device is intended as a part of an external
prosthetic system that replaces the foot and ankle
function of a missing lower limb.
462 Medical compression stockings
Intended for management/prophylaxis of leg pain,
deep vein thrombosis, varicose veins and related
complications
463 Medical equipment drape, single-
use, non-sterile
A flexible polymer sleeve designed to form a
protective water-resistant enclosure around a piece
of medical equipment such as intra-oral cameras
and its equipment components. It is intended as a
hygienic barrier to protect the contents from soiling
and contamination when used in the vicinity of, or
entered into, a hygienic area. It is neither a
dedicated handle cover, device cap nor an
ultrasound transducer cover.
464 Medical examination light (AC
powered or Battery)
Intended for medical purposes that is used to
illuminate body surfaces and cavities during a
medical examination.
S.No. Device Name
Intended Use
465 Medical gas terminal unit
A device that is a component of a medical gas
pipeline system or a medical gas/vacuum pipeline
system that has a gas-specific outlet connection for
a single/mixture of gas to which the operator can
connect and disconnect a
medica device.
466 Medical image printer
A non-invasive, reusable printer that produces hard
copies of medical video images.
467 Medical image storage device
Intended to provides electronic storage and retrieval
functions for medical images.
468 Medical image storage device
A medical image storage device is a device that
provides electronic storage and retrieval functions
for medical images. Examples include devices
employing magnetic and optical discs, magnetic
tape, and digital memory.
469 Medical imaging recorder
A non-invasive, reusable electrically powered device
designed to be used in conjunction with video
endoscopes, and/or surgical video imaging devices
to digitally record video images.
470 Medical instrument cable/lead
A reusable cable that functions as a conductor
between power source and the medical instrument
used for the purpose of transmitting an energy that
may, or may not, contain information.
471 Medical print paper
Thermal paper to be used to produce hard copy of
printed data generated by a medical device. For
recording the EEG,ECG, EMG output from the ECT
device.
472 Medical support stocking
Intended to be worn on the upper or lower extremity
to support, correct, prevent deformity, or to align
body structures for functional improvement.
473 Medical x-ray film, non-screen
A non-screen x-ray film specifically designed for use
in medical imaging applications. Non-screen x-ray
film is designed for direct exposure to x-rays and is
relatively insensitive to the visible light emitted from
screens.
474 Medical/cadaver body bags, non
sterile
Intended for transporting a human corpse without
any spread of biohazard/infectious agent.
475 Medical/Orthotic insoles
Intended to be worn inside shoes/footwear to relieve
heel pain, plantar fasciitis, foot pain, pain during
walking, calcaneal spur, sprain and strain of the foot,
ankle, feet and knee pain.
S.No. Device Name
Intended Use
476 Medicine chamber spacer
A device intended to be placed between a nebulizer
or a metered dose inhaler (MDI) and the patient's
mouth, to function as a reservoir into which an
aerosol medication is dispensed in order to minimize
delivery of large aerosolized particles.
477 Medicine cup
A specially designed cup used by the nursing staff
when they are administrating tablets or pills to
patients. It may be used both for tablets and fluids.
478 Meibomian gland camera
It is an ophthalmic imaging device intended for use
by a physician in adult patients to capture, archive,
manipulate and store digital images of the
meibomian glands.
479 Meibomian gland
expressor/evaluator, manual
A non-powered, hand-held manual ophthalmic
instrument intended to be used to apply pressure to
the lower eyelid to facilitate natural release and/or
expression of meibomian gland secretions (meibum)
to treat dry eye syndrome.
480 Metatarsal pad
Intended to relieve chronic metatarsal pain, corns,
hard skin, morton’s neuroma, forefoot soreness, and
painful balls of the feet.
481 Micropore tape
Intended for dressing retention on non-flexing areas
and may also be used for patients who have
exhibited an adverse reaction to adhesives based
upon zinc oxide, rubber, or resin.
482 Microscope cover
This is a device which is usually shaped to fit the
surgical camera/microscope that is intended to be
used as a physical barrier to prevent cross-
contamination between staff or patients during
examination.
483 Mirror-prism spectacles
Intended to enable the patient to see over the top of
their head enabling them to look forward in the
direction their head is pointing.
484 Mobile radiation protection barrier
Intended to protect the operator etc. from
unnecessary exposure to radiation used for medical
diagnosis, treatment and dental procedures.
485 Moist-cold pressure bandage
A pressure bandage intended to provide a cool and
to compress a local area of the body.
486 Moldable splint
A moldable splint device that is intended for use as
an immobilization of non- or minimally displaced
wrist fractures, contusions, and sprains
S.No. Device Name
Intended Use
487 Motorized diagnostic imaging view
box
Intended to retain, retrieve, and project light for
direct observation of medical images taken using a
variety of methods such as X‐ray, magnetic
resonance (MR), CT, and ultrasound, and recorded
in radiographic film.
488 Mouth gag
To be used in dentistry to be placed between the
teeth of a patient in order to maintain an open oral
cavity during dental procedures.
489 MRI enhancement padding
Dielectric pads that are intended for use with
magnetic resonance imaging systems to aid in body
positioning and help reduce shading artifacts
associated with the air/tissue interface during
abdominal, pelvic, and neck imaging.
490 MRI examination catheter holder
The device is intended to be used with ICP catheter
kits to enable the catheters to be left in during MRI
scanning.
491 MRI phantom, test object
Intended for both subjective and quantitative
evaluations of various calibration and performance
characteristics of magnetic resonance imaging
systems and associated image acquisition, display,
and analysis computer programs.
492 MRI system chair
A chair or stool specifically designed to support and
position a patient during examinations involving the
use of a diagnostic magnetic resonance imaging
(MRI) system. For MRI system compatibility these
chairs/stools are made of ferromagnetically inactive
materials.
493 Multi-chamber venous compression
system garment, reprocessed
Deep Vein Thrombosis (DVT) Garments are part of
an external compression system, in which
intermittent or sequential compression is provided
using a pump/controller and limb garment. When
coupled with an appropriate inflation system,
garments are intended for use in preventing deep
vein thrombosis (DVT) as well as the treatment of
edema secondary to venous insufficiency.
494 Multi-modality therapeutic radiation
phantom, test object
This device is intended to establish and provide for
routine and periodic verification with use of target
motion of patient dose and machine output in
therapeutic radiation oncology system performance.
S.No. Device Name
Intended Use
495 Multiple spinal region orthosis
This is a body worn device that is intended to
provide full circumferential containment when
reduction in gross spinal
motion about the lower thoracic and lumbar spine is
required. This may include post operative spinal
stabilisation, fracture and related soft tissue injuries
496 Nappy changing table, portable
Intended to support an infant, child or an adult
during nappy (diaper) changing, or for a patient with
a disability who is incontinent and requires regular
changing of their nappies.
497 Nasal aspirator, electric
Intended for superficial use to enable an adult to
gently suction and clear excessive mucus from the
nasal passages of an infant or child to facilitate
easier breathing.
498 Nasal aspirator, manual
A portable, hand-held, manual suction device
designed to enable an adult to gently suction and
clear excessive mucus from the nasal passages of
an infant or child to facilitate easier breathing.
499 Nasal dilator
Intended to provide temporary relief from transient
causes of breathing difficulties resulting from
structural abnormalities and/or transient causes of
nasal congestion associated with reduced nasal
airflow.
500 Nasal Irrigation kit
To cleanse or flush the nasal passages and sinus
cavity with saline for postoperative, preventive, or
symptomatic nasal care.
501 Nasal scissors
A surgical instrument used to cut tissue during
surgery on the nose and its associated structures. It
comprises two pivoted blades usually provided with
finger and thumb holes. The distal end of the blades
may be different shapes.
502 Nasal septum straightening forceps
Intended to straighten the nasal septum through
grasping and manipulation of the bone/cartilage of
the septum during nasal reconstructive procedures.
503 Nasogastric tube holder
This device is intended to secure nasogastric tubes
to skin near and around the nose.
504 Nasogastric tube holder, intranasal
The device is used with enteral feeding tubes to
prevent inadvertent removal or displacement of the
tubes for adult patients.
505 Nebulizer tubing set
A device intended to be connected to a nebulizer
and transfer medication from the nebulizer to the
patient with respiratory disorders.
S.No. Device Name
Intended Use
506 Needle guard
Intended to encapsulate needles withrawn from vein
or fistula thereby aiding in the prevention of needle
stick injuries.
507 Needle holder
Intended to secure needles during suturing.
508 Needle holder, ophthalmology
It is intended to hold needle for procedure during
eye surgeries.
509 Neonatal eye pad
Used to cover and protect the eye of an infant during
therapeutic procedures, such as phototherapy.
510 Neurosurgical chair
Intended to support and position a patient in a sitting
or
reclined position during neurosurgery.
511 Neurosurgical headrests
Intended to support the patient's head during a
surgical procedure
512 Newborn-infant bed
Intended as support surface to newborn babies.
513 Nipple shield
Intended to aid infants with trouble latching onto the
breast by acting like a longer, firmer nipple
514 Non-powered accelerator system
table
Intended to adjust the patient's posture and
immobilize the patient for radiotherapy.
515 Nonpowered flotation therapy
mattress
Intended for medical purposes which contains air,
fluid, or other materials, to treat or prevent
decubitus ulcers (bed sores).
516 Non-powered neutron therapy table
Intended to adjust the patient's posture and
immobilize the patient for treatment that uses
neutron rays.
517 Non-powered remote irradiation
therapy table
Intended to adjust the patient's posture and
immobilize the patient for treatment that uses a
remote radionuclide radiotherapy apparatus.
518 Non-powered X-rays radiation
therapy table
Intended to adjust the patient's posture and
immobilize the patient for treatment using X-ray
radiation therapy
519 Non-rebreathing oxygen face mask
A flexible, form-shaped device designed with valve
to control rebreathing and contamination of gas,
placed over the nose and mouth to deliver air of high
oxygen (O2) concentration to a patient's airway for
oxygen therapy.
520 Nose clip
It is a device used to help prevent air movement
through
the nose. It is typically used during pulmonary
function tests to ensure that airflow is focused
through the mouthpiece for accurate measurement.
S.No. Device Name
Intended Use
521 Nuclear scanning bed
A nuclear scanning bed is an adjustable bed
intended to support a patient during a nuclear
medicine procedure.
522 Nuclear sealed calibration source
A nuclear sealed calibration source is a device that
consists of an encapsulated reference radionuclide
intended for calibration of medical nuclear radiation
detectors.
523 Nystagmus inducing optokinetic
drum
Intended to elicit nystagmus.
524 Nystagmus inducing tape
Intended to be moved across a patient's field of
vision to elicit optokinetic nystagmus and to test for
blindness.
525 One-piece urostomy bag
The ostomy bag is intended to collect output from a
stoma. The adhesive is intended to fix the bag and
adhere to intact skin around the stoma.
526 One-piece urostomy bag, open-
ended
The product is intended for the management of
stomal output.
527 Operating room shoes cover, non-
sterile
Intended to be worn by operating room personnel
during surgical procedures to protect foot of both the
surgical patient & the operating room personnel from
transfer of micro-organisms, body fluids, and
particulate material.
528 Operating table traction frame
Designed to position, support and/or distract the
patient’s hand, wrist, forearm and arm in a variety of
surgical procedures including, but not limited to
arthroscopic or Orthopedic surgery.
529 Operation theatre pendant
Intended to provide ergonomic solution for the
positioning and management of medical devices,
equipment, and utilities during surgeries and other
medical interventions
530 Operator radiation protection
spectacles
Intended to protect the eyes of the operator and
other personnel from unnecessary exposure to
primary radiation and scattered radiation associated
with diagnosis and treatment.
531 Ophthalmic equipment stand
It is used to keep the ophthalmic equipment like Slit
Lamp, Keratometer, Laser etc.
532 Ophthalmic head reflector
Intended to reflect light onto the eye of a patient to
allow examination of the eye and its associated
structures.
533 Ophthalmic implant handling forceps,
reusable
A reusable instrument used to grasp and manipulate
ophthalmic implants (excluding sutures) during a
surgical procedure.
534 Ophthalmic operating table top
Intended to provide support for and stabilization of
the head of the patient and to help provide optimal
access for the surgeon(s) during the intervention.
S.No. Device Name
Intended Use
535 Ophthalmic surgery fluid collection
bag
A fluid collection bag connected to a surgical
system, used to catch aspirated material during an
ophthalmic surgical procedure.
536 Ophthalmic surgical device handling
forceps
Intended to grasp and manipulate a nonimplantable
invasive ophthalmic surgical device (e.g., ophthalmic
cannula, handless iris retractor)
537 Ophthalmoleukoscope
Intended for testing colour perception by means of
colours produced by polarized light.
538 Opthalmic examintaion station
A device assembly intended to provide complete
ophthalmic examination position and support
conditions for the patient and clinician
539 Oral care swab
A foam headed oral swab impregnated with a
sodium bicarbonate cleaning solution designed to be
used wet to clean the teeth and oral cavity.
540 Oral exerciser
A device that is used to exercise tongue muscles as
part of oral myology physiotherapy
541 Oral medicine dropper
Intended for aspirating a small volume of liquid
medicine so that it can be dispensed in single drops
into a patient's mouth, typically an infant or small
child. It is a non-measuring device (non-graduated).
542 Orbital depressor
Intended to displace tissue to facilitate examination
of the
surrounding area in the orbital cavity during eye
surgery.
543 Orthodontic aligner/retainer seater,
reusable
A small cylindrical device intended to aid in
positioning and seating an orthodontic progressive
aligner or teeth retainer into place. This is a reusable
device intended for single-patient use.
544 Orthodontic band pusher
A hand-held dental instrument designed for
positioning and adapting metal bands on teeth in
orthodontics.
545 Orthodontic positioner support kit
A collection of devices intended to be used to
facilitate use of an orthodontic teeth positioner,
including placement, removal and cleaning.
It typically consists of an orthodontic appliance
remover, an orthodontic positioner seater, and an
orthodontic appliance cleanser. The positioner is not
included. This is a single-patient device.
S.No. Device Name
Intended Use
546 Orthodontic resin positioning tray
A custom-made dental device in the form of a
patient’s dentition which contains cavities intended
to hold and facilitate positioning of light-cured dental
resin onto teeth for subsequent attachment of an
orthodontic appliance (e.g. aligner) after curing.
547 Orthodontic spring
The intended purpose of the orthodontic spring is to
open and close space between teeth.
548 Orthodontic vibrator
Orthodontic predictability enhancement device
designed to maximise the accuracy of fit and
effectiveness of clear aligner treatment.
549 Orthopaedic bed
Intended to provide support for skeletal traction to
stabilize fracture sites.
550 Orthopaedic bone pin holder
A non-sterile stainless steel cylindrical device
intended to temporarily hold one or more
orthopaedic pins (e.g K-Wires) during sterilisation &
surgical procedure against damage.
551 Orthopaedic countersink
used for bone preparation to enable the placement
of gingival part of components during dental
procedures in the mouth.
552 Orthopaedic instrument extractor
A reusable surgical instrument designed to extract
another orthopaedic instrument during an
orthopaedic surgical procedure.
553 Orthopedic cast padding
It is intended to protect long prominences under the
plaster on intact skin .
554 Orthopedic walker shoes/boots/Air
cast boots
Intended for the treatment and rehabilitation of foot
and ankle injuries and to provide support, stability,
and protection to the foot and ankle during the
healing process in cases of fractures, sprains, post-
surgery, or severe foot or ankle injuries.
555 Ostomy adhesive
Adhesive to be placed around stoma to attach 2
piece ostomy collection device.
556 Otoscope cover
A disposable ear specula is a plastic cover which fits
onto an otoscope to form a physical barrier
protecting patients against cross contamination.
557 Otoscope, direct
Intended for examination of the outer ear canal and
tympanic membrane (eardrum) by direct viewing
through the ear opening.
558 Otoscope, endoscopic
Intended to be used in otology mainly for
observation, diagnosis, and treatment of the outer
and/or middle ear.
559 Oxygen administration hood
Intended to provide an enriched oxygen (O2)
environment to increase the patient's O2 uptake.
S.No. Device Name
Intended Use
560 Oxygen administration hood,
paediatric
Intended to be used as an enclosure over an infant's
whole body, or the head only, in order to provide an
enriched environment of oxygen (O2) to increase the
patient's O2 uptake.
561 Oxygen administration tent,
neonatal/paediatric
Intended to cover the bed of a neonatal or small
child to provide an enriched environment of oxygen
(O2) to increase the patient's O2 uptake. It is used
for the treatment of breathing disorders in infant and
paediatric patients permitting them movement
without restriction.
562 Oxygen mask respiration indicator
Designed to be attached to the exhalation vent hole
on the exterior of a standard oxygen face mask and
provide visual indication of patient respiration.
563 Pacemaker test magnet
Intended to test an inhibited or triggered type of
pacemaker pulse generator and cause an inhibited
or triggered generator to revert to asynchronous
operation.
564 Packer, gauze
To be used in dentistry and dental surgery for
introducing and packing gauze or dressings into
cavities.
565 Paediatric dorsiflexion slant board
It is intended to be used in the treatment of various
medical conditions (e.g., congenital, neurological,
post-traumatic) where tendon tightness and muscle
contracture affect the ability to dorsiflex the foot,
possibly leading to an abnormal gait.
566 Panniculus skin-adhesive retractor,
reusable
This retractor is intended to be used on intact skin
for positioning redundant tissue (i.e., post c-section,
post laparotomy) away from the lower abdominal
incision/wound.
567 Parallel bar exerciser
Intended to assist users in maintaining good walking
posture, particularly a person with a disability, a
paraplegic, or a patient who has suffered a stroke
and is learning to walk.
568 Parapodium walking frame
Intended to encompasse and provide support for the
body of a patient who is unable to stand unassisted
to help them move (walk) by changing their centre of
gravity (COG).
569 Partial hand radiation protector
Intended to protect part of hands and fingers from
unnecessary exposure to primary radiation and
scattered radiation associated with diagnosis and
therapeutic measures.
570 Patient audiovisual communication
system
It is intended for use by trained personnel in
functional, interventional and clinical MRI
environments to facilitate audio communications
during a scanning session.
S.No. Device Name
Intended Use
571 Patient enclosure system
A freestanding, framed, tent-like structure designed
to be constructed on-site to function as a temporary
enclosure for safe playing, resting or sleeping,
typically for a patient with autism, epilepsy,
challenging behaviours and/or learning difficulties. It
has a soft fabric/plastic interior with padded regions
to prevent injury, and a limiting and low-stimulation
environment.
572 Patient gown, non-sterile
Intended to be worn by patients in a clinical setting
(e.g., during hospitalization, during examination in a
doctor's office).
573 Patient holder
The device is used to support patient's trunk and or
lower limbs into a more symmetrical position. The
intended patient group are those with disabilities
who require assistance to maintain an optimal
position, therefore preventing secondary
complications such as joint contractures, pressure
areas and pain.
574 Patient isolation tent canopy
A non-sterile covering designed to be used with a
frame to provide environmental isolation of a
contagious patient to protect hospital personnel and
other patients from exposure to non-airborne
infections (e.g. diphtheria, meningococcus) whilst
the patient is being treated.
575 Patient lift
Intended as an assistive device to lift and transfer a
patient with limited mobility from one place to
another.
576 Patient transfer board
Intended to reduce the friction and difficulty during
the task of moving or transferring a patient from a
surface to another e.g. Bed to stretcher and reduce
risks associated with manual handling by the staff
577 Pedal exerciser
Intended for exercise therapy for upper and/or lower
body.
578 Pediatric gait trainers
Intended to assist and train children unable to walk
or stand independently to enhance movement and
independent ambulation
579 Pelvic floor exercise biofeedback
device
It is intended for encouraging regular pelvic floor
exercises; strengthening of the pelvic floor muscles;
rehabilitation and training of weak pelvic floor
muscle; treatment of stress, urge and mixed urinary
incontinence in women; improving sexual pleasure
or function in women with existing pelvic floor
dysfunction; improving pelvic floor function in the
post-partum period; improving mild to moderate
prolapse symptoms
S.No. Device Name
Intended Use
580 Pelvic traction kit (with/without
weights)
Intended to apply uniform and controlled traction to
the lumbar, sacral, and thoracic areas of the
vertebral column.
581 Pen torch
Intended to assist with examinations by illuminating
regions of the eye mount or ear.
582 Percussion hammer, manual
Intended to be used by an examining physician to
gently tap near a patient's joints to test reflexes.
583 Periodontic hoe
A hand-held manual dental instrument with a small
blade at an acute angle (e.g., 90 degrees) to its
stem intended to be used with a pulling action to
remove gross amounts of supragingival calculus,
and some subgingival calculus, in large wide-open
pockets where the gingival tissue is soft and easily
cleaned.
584 Philadelphia collar (cervical)
Intended to be used in cervical fracture to support
neck bones and ligaments by restricting the
movement of the neck muscles
585 Phoropter arm
Used to mount the phoroptor during an ophthalmic
examination; typically to mount the phoroptor to
determine a patient's prescription for glasses.
586 Phototherapy eye protector, reusable Intended to cover and protect the eyes of a patient
or user from potentially harmful rays [e.g., ultraviolet
(UV)] to which parts, or all, of their body is
intentionally exposed during light therapy treatment.
587 Phototherapy eye protector, single-
use
Intended to cover and protect the eyes of a patient
from potentially harmful rays [e.g., ultraviolet (UV)] to
which their body is intentionally exposed during light
therapy treatment.
588 Phototherapy infant blanket
A piece of fabric, made from ultraviolet (UV) light
transmittable material, intended to be used to wrap a
newborn/infant during UV phototherapy for the
treatment of neonatal jaundice (neonatal
hyperbilirubinemia) to enable therapeutic light to
pass through to the patient while simultaneously
providing comfort. It does not include a light source.
589 Physical restraint
Intended to secure patient to chair, wheelchair, bed
etc. to prevent self-harm or injuries to the health
care professional handling the patient
S.No. Device Name
Intended Use
590 Physical therapy table, line powered
Intended to facilitate functional training of
patients/subjects with loss of lower extremity, upper
extremity or core functionality, caused by conditions
of orthopedic, musculoskeletal, neurological, or
other origin by providing muscular load in
combination with visual feedback.
591 Physio Motion Therapy BTL-CP
Motion
Intended to be used during the first phase of
rehabilitation following lower limb joint surgeries or
traumas.
592 Picture archiving and communication
system
A device intended for the recording, viewing,
presentation and storage of digital still frames and
video sequences for use with a microsurgical
operating system.
593 Pill crusher/splitter
A hand-held device used for crushing or splitting
medication in the form of a tablet for easier ingestion
by a patient who cannot swallow the tablet whole.
594 Pinwheel
Intended for testing pain sensation
595 Plaster instruments/plaster saw
A instrument used to cut or shave or put plaster.
596 Plaster of Paris bandage
It is intended for support of fractured body part.
597 Plaster spreader
used to separate and spread hardened plaster
598 Pleoptophor
Intended for use during the treatment of eccentric
eye fixation (casts in the eye) by dazzling the
perimacular retina, thereby relatively enhancing the
visual capabilities of the fovea.
599 Pneumatic external knee joint
Intended to replace knee joint in persons with lower
limb amputation and connecting to above knee
prosthesis
600 Pneumatic stander
Standing aid chair for Disabled Person
601 Pneumatic tourniquet and its
attachment
Used for stopping flow of blood through artery by
compression.
602 Pneumatic walker
To provide pneumatic compression and support to
the ankle joint.
603 Polatest
Intended for evaluating hidden (latent) squinting, i.e.,
when the patient is not aware of the condition, and
also when it cannot be seen.
604 Portable system mount
Intended to be fixed vertically within a clinical setting
to enable the easy and secure mounting of a
portable medical device.
S.No. Device Name
Intended Use
605 Positioning aid system
A mattress system specifically designed for the
disabled person allowing various positions to be
maintained with the aid of positioning aids. designed
to allow air flow to reduce sweating/heat and
therefore the possibility of spasm.
606 Positron camera
A positron camera is a device intended to image the
distribution of positron-emitting radionuclides in the
body.
607 Postpartum maternity belt
It is intended to help reduce swelling, provide
support to the lower back and abdominal muscles,
and promote healing after delivery.
608 Postsurgical head/neck garment
It is a wearable contoured cranial cover intended for
the restoration or augmentation of cosmetic
appearance in patients with bony defects of the
cranium.
609 Postsurgical transparent eye shield,
vented
A device designed to be worn over the eye to protect
the eye while allowing the wearer to see, typically
following ophthalmic surgery. It is designed as a
transparent self-adhesive dome which includes vent
holes to allow air to circulate around the eye. This is
a single-use device.
610 Power wheelchair
Intended to aid in mobility of individuals with
physical disability, lack of senses in lower
extremities and having difficulty in walking.
611 Power-assisted foot/hand-propelled
tricycle
Power assisted foot/hand propelled tricycle designed
specifically for people with disabilities
612 Prescription spectacles
An ophthalmic device consisting of a spectacle
frame with a pair of optical spectacle lenses
(typically made of plastic or glass, also known as
eyeglasses) designed to be worn in front of the
user's eyes to provide refractive corrections to
eyesight conditions such as presbyopia (the eye's
diminished power of accommodation that occurs
with ageing), or refractive ametropia (myopia,
hyperopia and astigmatism). The lenses may be
tinted (sunglasses) to protect the eyes from bright
light and radiation.
613 Professional examination gown, non-
sterile
Intended to be worn by healthcare providers,
sometimes over scrub suits, while examining
patients to protect both the patient and staff from the
transfer of contaminants such as microorganisms or
body fluids.
S.No. Device Name
Intended Use
614 Prophylaxis cup/ Dental polishing
cup
This is a device that is intended to be used as a
polishing cup to apply polishing agents during dental
prophylaxis (cleaning). The device is held in a dental
handpiece.
615 Proprioceptive nest
Intended to give physical support to the preterm
newborn. It is designed to provide preterm babies
with age-appropriate proprioceptive stimulation as
well as comfort, a sense of security, and tactile
stimulation.
616 Prosthetic external arm
Intended to be used as an external prosthesis for
upper limb distal to shoulder joint by patients with
upper limb loss or deficiency.
617 Prosthetic external foot (Winged
Foot)
Intended for persons with a lower-limb amputation to
restore lost mobility
618 Prosthetic foot, pediatric
Intended for children with a lower-limb amputation to
restore lost mobility
619 Protective garment for Incontinence
Intended to protect an incontinent patient's garment
from the patient's excreta.
620 Psychophysiological biofeedback
system
Intended to improve certain medical conditions
through user interaction, including anxiety,
depression, pain, loneliness, mood and/or
motivation; and for symptoms such as agitation,
aggression, walking about found in some mental
health disorders and/or neurocognitive disorders
such as dementia.
621 Pulmonary equipment filter
Intended to prevent biological contaminants from
entering pulmonary function testing equipment to
help protect the patient and equipment during lung
function tests.
622 Pulmonary exerciser
The device is a handheld device that is intended to
improve expiratory/exhalation and
inspiratory/inhalation muscle strength for the
enhancement of breathing, coughing, swallow, voice
and fitness.
623 Pupillograph
Intended for recording the response of the pupil to
reflected light for ophthalmic diagnostic purposes.
624 Radiation face protector
Intended to protect the face and eyes of medical
personnel and other personnel from unnecessary
exposure to primary radiation and scattered
radiation associated with diagnosis and treatment.
625 Radiation protection apron
Intended to protect the patient, the operator, and
other personnel from radiation exposure during a
medical or dental procedure.
S.No. Device Name
Intended Use
626 Radiation protection blanket
Intended to protect specific body parts of the patient,
operator, and other personnel from unnecessary
radiation exposure in medical/dental procedures for
diagnosis and treatment.
627 Radiation protection cap
Intended to protect the head of the operator and
other personnel from unnecessary exposure to
primary radiation and scattered radiation associated
with medical procedures for diagnosis and
treatment.
628 Radiation protection collar
Intended to protect the neck or thyroid of the patient,
the operator, and other personnel from unnecessary
radiation exposure in medical/dental procedures for
diagnosis and treatment.
629 Radiation protection gloves
Intended for complete protection of the hands of the
operator and other personnel from unnecessary
exposure to primary radiation and scattered
radiation associated with diagnosis and therapeutic
measures.
630 Radiation protection goggles
Intended to protect the eyes of the operator and
other personnel from unnecessary exposure to
primary radiation and scattered radiation associated
with medical/dental procedures for diagnosis and
treatment.
631 Radiation protection mitten
Intended to protect the hands of the operator and
other personnel from unnecessary exposure to
primary radiation and scattered radiation associated
with diagnosis and therapeutic measures.
632 Radiographic ECG/respirator
synchronizer/X-ray synchronizer
A radiographic ECG/respirator synchronizer/X-ray
synchronizer is a device intended to be used to
coordinate an x-ray film exposure with the signal
from an electrocardiograph (ECG) or respirator at a
predetermined phase of the cardiac or respiratory
cycle.
633 Radiographic film illuminator/X-ray
LED view panel
A radiographic film illuminator is a device containing
a visible light source (e.g., LED) covered with a
translucent front that is intended to be used to view
medical radiographs.
634 Radiographic film marking
system/radiographic film marker
A radiographic film marking system is a device
intended for medical purposes to add identification
and other information onto radiographic film by
means of exposure to visible light.
S.No. Device Name
Intended Use
635 Radiographic film processor,
automatic
It transports dental x-ray film from one solution to the
next in the film developing process without any
manual labour except the insertion of the
film/cassette.
636 Radiographic film, non-screen
Intended to record images during diagnostic
radiologic procedures.
637 Radiographic grid/X-ray grid/Anti-
scatter grid
A radiographic grid is a device that consists of
alternating radiolucent and radiopaque strips
intended to be placed between the patient and the
image receptor to reduce the amount of scattered
radiation reaching the image receptor
638 Radiographic head holder
Intended to position the patient's head during a
radiographic procedure.
639 Radiographic intensifying screen
It is used to reduce the X-ray dose to the patient and
to allow for shorter exposure times to reduce motion
artefact on exposed film.
640 Radiographic/X-ray cassette holder
(wall-mounted/ceiling mounted/floor
standing)
Intended to be used as a support to hold and
position radiographic cassettes for a radiographic
exposure for medical use.
641 Radiologic patient cradle
Intended to be used for rotational positioning about
the longitudinal axis of a patient during radiologic
procedures.
642 Radiologic table
Intended for medical purposes to support a patient
during radiologic procedures.
643 Radiological breast positioning
device, wearable
The device is indicated to aid in supporting
adolescent and adult patients undergoing radiation
therapy including electron, photon, and proton
treatments. The device is also used during image
acquisition to support treatment planning.
644 Radionuclide test pattern phantom
A radionuclide test pattern phantom is a device that
consists of an arrangement of radiopaque material
sealed in a solid pattern intended to serve as a test
for a performance characteristic of a imaging
device.
645 Rebreathing oxygen face mask
Intended to be placed over the nose and mouth to
deliver a proportional mixture of air/oxygen (O2) to a
patient's airway.
646 Reciprocating walking frame
Intended to be used a walking aid by people with
physical disabilities or poor balance
647 Reclining wheelchair
A walking aid that reclines enabling the person to
change positions from sitting to lying down.
S.No. Device Name
Intended Use
648 Recording paper
This device is specially produced paper intended to
be used for recording the output of devices
measuring physiologic parameters, e.g.
electrocardiogram (ECG), electroencephalogram
(EEG) or ultrasound imaging or for recording fetal
images.
649 Rectal dilator
Intended to dilate the anal sphincter and canal when
the size of the anal opening may interfere with its
function. This is a single-patient use device and is
intended for transient use.
650 Rectoscope/proctoscope handle
Intended to be used as a handle to hold a
rectoscope or proctoscope during an examination
procedure.
651 Rectosocope/proctoscope handle
end-cap
Intended to be fitted to the proximal end of the
handle of a rectoscope or proctoscope during an
examination procedure to provide an airtight seal so
that air can be pumped through the scope to
inflate the bowel.
652 Rescue blanket
Intended to keep a person warm and/or to prevent
the further loss of body heat in emergency/rescue
situations.
653 Retractor holder
Intended for mounting of surgical retractors.
654 Reverse-image mirror, dental
A device designed to be placed at the back of the
teeth while taking image for dental examination
purposes.
655 Rib belt
Intended for providing support and compression to
the ribcage and relief from rib pain and discomfort,
especially after injuries or surgeries, supporting
breathing and coughing during recovery, and
preventing rib injuries in certain sports or physically
demanding professions.
656 Rod cutter/reducer
Intended to be used to reduce or seat an
implantable rod into the saddle of the implant.
657 Rollator
Intended as a walking aid for balance and stability in
individuals with limited mobility.
658 Rongeur, dental
To be used in dental surgery to remove hard or
tough body tissues such as cartilage or bone, by
applying a biting/cutting action. This is a reusable
device.
659 Rotating vestibular chair
A tilting and rotating chair intended to be used in
clinical procedures for testing vestibular-ocular
disorders.
660 Scalpel blade remover, non sterile
Intended to assist the removal of used/blunt blades
from a scalpel handle during a surgical procedure
S.No. Device Name
Intended Use
661 Scintillation (gamma) camera
A scintillation (gamma) camera is a device intended
to image the distribution of radionuclides in the body
by means of a photon radiation detector.
662 Scoop stretcher
Intended to transfer a trauma patient and reduce the
chance of undesirable movement of injured areas.
663 Shaped ankle wedge prosthesis,
adult
Intended to be used as external artificial joint that is
designed to replace the ankle joint in adults
664 Shaped ankle wedge prosthesis,
pediatric
Intended to be used as external artificial joint that is
designed to replace the ankle joint in children
665 Shield, resuscitation
To minimise tissue contact and exchange of fluids
between parties during mouth to mouth
resuscitation.
666 Shoulder immobilizer
Intended to temporarily immobilize or limit abduction
of the shoulder joint to support healing of an injury or
a surgical wound.
667 Shoulder wheel
It is a device that allows patients to perform
resistance exercises to improve range of motion and
relieve pain (intended for hospital use)
668 Silicone prosthetic ear
Intended to adhere to the skin as an artificial
external ear using safe, biocompatible glue for
patients born with microtia and other birth
differences such as Treacher Collins or Goldenhar
syndrome; as well as those who have suffered
traumatic injury or undergone cancer resection
surgery.
669 Silicone prosthetic nose
Intended to restore normal contour and improve
function for patients who have experienced partial or
total loss of their nose to traumatic injury, disease or
due to surgical removal of the nose (rhinectomy).
670 Single use binocular indirect lenses
Intended for visualization of the human retina
(fundus) using a binocular indirect ophthalmoscope.
671 Single-chamber venous compression
system garment, reprocessed
Deep Vein Thrombosis (DVT) Garments are part of
an external compression system, in which
intermittent compression is provided using a
pump/controller and limb garment. When coupled
with an appropriate inflation system, garments are
intended for use in preventing deep vein thrombosis
(DVT) as well as the treatment of edema secondary
to venous insufficiency.
S.No. Device Name
Intended Use
672 Sitz bath
Intended for use in external hydrotherapy to relieve
pain or pruritus and to accelerate the healing of
inflamed or traumatized tissues of the perianal and
perineal areas.
673 Sitz bath chair
Intended to support a person in sitting position for
external hydrotherapy to relieve pain or pruritus and
to accelerate the healing of inflamed or traumatized
tissues of the perianal and perineal areas.
674 Skin abrasion device/file
Intended for debridement of thickened nails and
callus.
675 Skin adhesive protective material
Intended to protect healthy skin from friction during
removal of tapes and films. It provides a thin
protective layer on the skin and helps tape and film
adhesion.
676 Skin marker guide
A non-sterile device intended to be used as a guide
to create markings on the skin surface with a
drawing skin marker, typically to outline a surgical
incision site or for dermatology applications. It is
typically shaped specifically for its intended
application.
677 Skin marking transfer paper
A non-sterile, non-invasive device intended to
transfer a grid of ink dots onto the skin surface to
assist a clinical procedure. This is a single-use
device.
678 Skin pressure protectors
Intended to reduce pressure on the skin over a bony
prominence to reduce the likelihood of the patient's
developing decubitus ulcers (bedsores)
679 Skin-cover sealant/Synthetic polymer
liquid barrier cream
This is a non-cytotoxic cyanoacrylate based liquid
skin barrier film for the skin to form a transparent
flexible film to prevent skin irritation caused by
moisture eg. urine, friction from bedding and by
adhesive products.
680 Slit Lamp
Intended to be used as a specialized magnifying
microscope to examine the structures of the eye
(including the cornea, iris, vitreous, and retina).
681 Snoring sensor
Intended for use in the detection of snoring, muscle
movement, respiratory effort and/or airflow.
682 Soft dental reliner
Intended to be used as a tissue conditioner and a
reliner for short term use for removable dentures. It
may be intended for professional and/or home use.
683 Soft rehabilitation glove
Intended for rehabilitation training of patients with
finger joint dysfunction.
S.No. Device Name
Intended Use
684 Solid-ankle-cushion-heel (SACH)
foot
Intended for persons with a lower-limb amputation to
restore lost mobility
685 Spatula
An instrument, usually made of stainless steel,
which is used to spread a material on a surface, to
fill material into a cavity, manipulate tissue or to
remove a material from a surface. This is a reusable
device.
686 Spine orthosis
An externally-applied wearable appliance intended
to support and correct deformities of the spine. It
may be one of numerous different designs and is
typically made of synthetic and textile materials. This
is a single-patient reusable device.
687 Stair chair
Supports and transports a patient occupant in a
seated position up or down a set of stairs.
688 Standing wheelchair
Intended as a walking aid that allows users to raise
themselves in upright position using upper limbs
689 Steam inhaler
Intended for the inhalation of steam for the relief of
the symptoms (congestion and pressure) associated
with upper respiratory disorders.
690 Sterilization/disinfection containers
Intended to hold wrapped and/or unwrapped
medical/dental devices in a chemical
washer/disinfector and/or sterilizer during the
washing/disinfection and/or sterilization cycle(s).
691 Sterilizer/disinfector rack
A non-sterile rack intended to hold medical devices
in a steriliser or chemical washer/disinfector during
the sterilisation or washing/disinfection cycle. It may
also be used to transport medical devices to and
from the process and for their storage. This is a
reusable device.
692 Sternal binder
Intended to stabilize and support the sternum after
sternotomies
693 Stress exercise trolley
This treadmill is intended to be used for cardiac
stress testing.
694 Suction system filter, non sterile
A filter used with a suction system to separate tissue
from the fluid waste collected from a medical
procedure.
695 Surgical binoculars
Intended to be mounted onto a surgeon's spectacles
to function as small telescope and provide a
magnified image of the visual field during patient
examination or surgical intervention.
696 Surgical camera and accessories
Intended to be used to record operative procedures.
S.No. Device Name
Intended Use
697 Surgical camera holder
Intended to hold and position a camera, or camera
system and/or other devices during surgical
procedures.
698 Surgical cap, non-sterile
Intended to protect face of patient & the operating
room personnel from the transfer of micro-
organisms, body fluids & particulate material.
699 Surgical eye pressure shield
A non-sterile device intended to cover and
mechanically protect the eyes of a patient during
surgery to prevent potential ocular pressure
hazards/injury associated with surgery (e.g.,
resulting from inadvertent pressure being applied to
the eye by healthcare professionals, equipment, or
due to the orientation of the patient).
700 Surgical guillotine
Intended for severing tissue as a means to prevent
infection/drainage of infected site
701 Surgical helmet, non-sterile
Intended to protect healthcare workers from
contaminants & debris on head.
702 Surgical implant handling forceps
A hand-held manual surgical instrument with blades
designed to grasp and manipulate a surgical implant
(excluding sutures) during implantation; it is not
intended for use on tissues, is not a dental or
ophthalmic device, and is not a dedicated tissue clip
applier. This is a reusable device intended to be
sterilized prior to use.
703 Surgical instrument assist arm
system
The device assists in maintaining spatial positioning
of instruments during surgical procedures
704 Surgical instrument holder, reusable
device intended to be used in the operating room
(OR) by staff to safely contain a delicate or sharp
hand-held, manual, surgical instrument(s)
705 Surgical instrument rack
Intended for the safe storage and counting of used
sharps during and prior to the the end of a surgical
procedure. The used sharps are held in place by
foam, magnet or other mechanisms and the device
can be closed securely for safe sharps disposal.
706 Surgical knee cover
Reusable surgical cover used as a protective cover
to protect the patella during knee surgery.
707 Surgical lamp
Intended to provide visible illumination of the
surgical field or the patient.
708 Surgical lamp (ophthalmic)
Intended to serve as a cold light source for
illumination of the intraocular posterior segment in
optical surgery in case of vitrectomies.
S.No. Device Name
Intended Use
709 Surgical lens (camera accessory)
A device designed to be used with a camera, or a
camera system being used during medical
procedures.
710 Surgical light/illumination system/OT
lights
Intended to providing optimal light and colour
rendering to aid surgery.
711 Surgical material compression
device
A reusable manual instrument intended to flatten
material, such as tissue, for use in a surgical
procedure (e.g., tympanoplasty) after it has been
flattened.
712 Surgical microscope and
accessories
Intended for use during surgery to provide a
magnified view of the surgical field.
713 Surgical sleeve cover, non-sterile
It is a forearm covering made of non-woven
synthetic fabric materials intended to be worn over a
scrub suit by healthcare staff during procedures to
maintain hygiene and prevent cross-contamination
of microorganisms, body fluids, and particulate
materials between the patient and staff.
714 Surgical torque wrench adaptor
This is a hand held component of a surgical wrench
that is intended to connect to the main body of the
wrench to a variety of distal end pieces (bits) and
assist with the use of a torque wench in the mouth
during a dental procedure.
715 Surgical trays
Intended to provide a suitable platform for general
containment of many medical/surgical instruments
and related items during a clinical procedure.
716 Surgical video camera
A video camera used to generate video images of
surgical procedures, which are fed to a monitor
screen for local or external viewing.
717 Swivel-walker
Intended to encompasses and provide support for
the body of a patient who is unable to stand
unassisted, to help them move (walk) by rocking
sideways (shifting their weight from side-to-side with
a shoulder movement).
718 Synoptophor
Intended for the evaluation and training of a patient's
binocular function.
719 Tachistoscope
Intended to flash words or images at different
speeds, for the purposes of ophthalmic diagnostic
testing.
720 Tearable adhesive plaster
It is intended to secure dressings and for strapping
on intact skin.
721 Tearable adhesive tape
It is intended to secure dressings and for strapping
on intact skin.
722 Teething device, non- fluid-filled
Intended to be bitten by a patient (infant or adult) to
soothe gums during the teething process.
S.No. Device Name
Intended Use
723 Tester, colour discrimination
Devices for colour vision testing
724 Testicular temperature regulation
underpants
This is a garment that is body worn that is intended
to be placed around the pelvic region in the same
manner as conventional male underwear that is
designed to keep the testicles cool for increased
spermatogenesis, as part of an fertility treatment,
through ventilation features (e.g., breathable mesh)
that prevent excessive insulation of the testes.
725 Therapeutic medical binder, general
purpose
Intended for medical purposes and that can be
secured by ties so that it supports the underlying
part of the body or holds a dressing in place like
abdominal, breast and perineal binder.
726 Therapeutic scrotal support
Intended for medical purposes to support the
scrotum (the sac that contains the testicles)
727 Thermographic camera
It is a thermal based imaging device, intended for
viewing and digitally storing thermal patterns
generated by the human body in a clinic, hospital,
acute care setting, surgery, healthcare practitioner
facility or in any environment where healthcare is
provided by a qualified healthcare professional.
728 Thermoplastic mask for
immobilization
Intended to immobilize the patient to deliver the
correct Radiation dose
729 Thumb spica splint
Intended to minimize movement and provide support
and comfort by stabilizing an injury of the thumb
730 Toe separator
Intended to space the toes of the foot to relieve pain,
pressure/friction between toes, and/or to facilitate
realignment of the toes to a natural position.
731 Tongue depressor
Intended to move the tongue to facilitate
examination of surrounding organs and tissue.
732 Tongue scraper
A device used to scrape the upper surface of the
tongue to promote oral hygiene (e.g., to remove
plaque, food debris, and bacteria, and to prevent
bad breath). It may be single use or reusable and is
an OTC device.
733 Tongue-retaining sleep-disordered
breathing orthosis
A removable intraoral device designed to alleviate
sleep-disordered breathing conditions (e.g., snoring,
obstructive sleep apnoea) by retention of the tongue,
frequently keeping the tongue in a forward position
during sleep.
S.No. Device Name
Intended Use
734 Tracheal surgery dilator
Intended to be used during surgical intervention of
the trachea to dilate tracheal structures/passages,
typically during the creation of a tracheostoma
and/or for expanding the margins of a tracheostoma
to assist in the insertion of a tracheostomy tube.
735 Tracheostoma protector
Intended to absorb secretions and to provide
protection and aesthetic coverage of the
tracheostoma.
736 Tracheostoma shower shield
A non-sterile, waterproof, patient-worn device
intended to cover a tracheal stoma or tracheostomy
tube to prevent inhalation of water during showering.
737 Tracheostomy tube holder
Intended to secure a patient’s tracheostomy tube for
use on pediatric (neonates, infants and children)
patients.
738 Tracheostomy tube lubricant
Intended to be applied by healthcare staff or a
patient to the outer surface of a tracheostomy tube
inner cannula to facilitate its easy sliding into the
outer cannula.
739 Tracheostomy tube neck holder,
single-use
To secure a tracheostomy tube by fastening around
the neck of the patient. Composed of cotton/velour
and a velcro fastening.
740 Training aid for communication and
speaking
Designed for the assessment of disordered voice
and speech, and voice and speech therapy.
741 Training aid for perceptual sense
and co-ordination
This device is an exercise system that promotes
therapeutic exercises through use of one or more
sensorized movement controllers.
This device is indicated for use by patients with
neurological or other movement impairment,
typically from stroke, who are undergoing
rehabilitation to facilitate muscle re-education,
increased range of motion, or increased
coordination.
742 Transcutaneous electrical stimulation
electrode adhesive pad
Reusable gel pads that allow electrodes to adhere
gently to the skin.
743 Trial frame
A device used in ophthalmic work for holding trial
lenses in front of the eyes during the sight-testing
procedure.
744 Truncal orthosis
Intended to support or to immobilize fractures,
strains, or sprains of the neck or trunk of the body.
S.No. Device Name
Intended Use
745 Tuning fork
Intended to test the hearing acuity of a patient, to
diagnose hearing disorders, and to test for vibratory
sense.
746 Ulnar/brachial nerve protector
It is use to provide cushioning, support and
protection to elbow and arms during prolonged
surgeries.
747 Ultrasonic coupling gel
A coupling medium for ultrasound transmissions.
748 Ultrasonic safety goggle
Intended to be worn over the eyes to protect (shield)
the healthcare worker from blood and other body
fluid splashes while performing a clinical or
laboratory procedure.
749 Ultrasonic scanner calibrator
It is a block of material with known properties
intended to be used to calibrate ultrasonic scanning
devices
750 Ultrasound imaging phantom, test
object
Intended to provide a standard of reference for
routine and periodic verification of ultrasound
doppler imaging equipment used for the
identification of disease and pathologies.
751 Ultrasound imaging system table,
non-powered
A non-powered patient table designed to support
and position a patient specifically for ultrasound
imaging procedures.
752 Ultrasound imaging system table,
powered
An electrically-powered patient table designed to
support and position a patient specifically for
ultrasound imaging procedures.
753 Ultrasound imaging system
workstation
A freestanding image processing workstation
specifically designed to be networked with one or
more diagnostic ultrasound imaging systems. It is
typically considered to be a component to a picture
archiving and communication system (PACS) and
does not contain the controls for the direct operation
of the diagnostic imaging system.
754 Ultrasound imaging transducer
positioning unit
A unit used to position an ultrasound system
imaging transducer that is inserted into the body
through an endoscope. This device is an accessory
to a diagnostic ultrasound imaging system and it
transmits electrical signals that, after processing, will
display the position of a the transducer on a monitor.
755 Umbilical cord clamp
Used for clamping the umbilical cord after birth.
756 Umbilical ligator
Intended for umbilical cord ligation during surgical
procedures.
S.No. Device Name
Intended Use
757 Upper limbs gravity-compensating
rehabilitation system
Intended to increase muscle strength, motion in
different joints and improve motor function in
patients who have lost their function or have
restricted function in their upper extremities.
758 Urinary dilator
Intended to dilate the urinary duct
759 Urinary-incontinence
vaginal insert, reusable
A non-sterile, device intended to be inserted into the
vagina in order to relieve mixed or stress urinary
incontinence in an adult female by providing urethral
support when pressure is transferred from the
abdomen to the pelvic floor area (e.g. upon
coughing, laughing, sneezing, exertion).
760 Urine drainage extension tubing
A conduit for urine collection, via a pouch and
tubing, into a gravity drainage collection bottle or
bag.
761 Urological operating table top
A component of a modular operating table designed
as a detachable table top to support a patient during
urological examination and surgical procedures,
including cystoscopic procedures and both
transurethral and open surgery of the urinary tract.
762 Urological operating table,
hydraulic/Transurethral resection
table
A mobile table powered by manually-operated
hydraulic mechanisms designed to support a patient
during urological examination and surgical
procedures, including cystoscopic procedures and
both transurethral and open surgery of the urinary
tract.
763 Uterine packer
Intended for introducing dressings into the uterus or
vagina.
764 Vacuum cushion
Intended for positioning and re-positioning patients’
head & neck, thorax, hip and full body during
radiation therapy and other diseases treatment.
765 Vacuum splint
Intended to immobilize and protect injured limbs so
that the patient can be transported safely and easily
in emergency conditions.
766 Vaginal applicator, reusable
Intended to apply medication to the vagina.
767 Vaginal applicator, single use
A single use device intended to apply medication to
the vagina
768 Vaginal exerciser/Pelvic floor muscle
exerciser
Designed to provide Kegel exercises to strengthen
and tone the pelvic floor muscle, which in turn
improves conditions such as female incontinence,
sexual dysfunction, prolapse, etc.
S.No. Device Name
Intended Use
769 Vaginal illuminator
A device used to illuminate the vagina during a
gynaecological examination or procedure.
770 Valsalva manoeuvre mouthpiece
Intended to be inserted into a patient's mouth to
facilitate performance of the Valsalva manoeuvre.
771 Vice grip
Intended to hold pliers during surgical procedures.
772 Video indirect ophthalmoscope
Intended as an illumination system to direct
appropriately focused light into the eye in order to
obtain an intermediate image of the fundus (retina)
that is viewed by the observer to diagnose the eye
and record the images and videos of the patient.
773 Video intubation laryngoscope
handle/monitor
Intended to enable the positioning of the blade into
the oral cavity to manipulate the tongue, preventing
it from obstructing the oropharynx and enabling a
clear view of the trachea for the insertion of an
endotracheal (ET) tube prior to the delivery of
inhalation anaesthesia and/or ventilation.
774 Video laryngoscope adaptor
Intended to serve as a connection between the
laryngoscope and the display/monitoring device
during intubation
775 Vision testing/training flipper
A device in the form of flipper lenses (two lenses on
either side of a rotatable stem) intended to be used
for the diagnosis and treatment of ophthalmic
accommodation disorders.
776 Vision training software
A patient-interactive software package intended to
use visual images (e.g., Gabor patch) to help
improve vision based on noninvasive perceptual
learning therapy (neuro-vision therapy), typically in
patients with amblyopia, or who use corrective
eyewear, or following ophthalmic surgery.
777 Visual chart
Intended for use in testing visual acuity.
778 Visual light box
Intended for use in testing visual acuity.
779 Visual projector
Intended to project an image on a screen to test
visual acuity .
780 Visual field plotter
Device for visual field examination
781 Vitrectomy lenses
Intended to facilitate the visualization of the retina
during Vitrectomy surgery.
782 Warming chamber
A chamber and accessories designed to warm
various items prior to use, e.g. bags of irrigation or
intravenous solutions, blankets, or clothing for
operating theatre personnel
S.No. Device Name
Intended Use
783 Washers for body waste receptacles
Intended for medical purposes that is used to clean
and sanitize a body waste receptacle, such as a
bedpan.
784 Weight activated knee joint external
prosthesis
Intended as an artificial knee with a braking
mechanism that prevents the knee from bending as
long as the user has weight on the prosthesis
785 Weight-bearing protector for flat
panel x-ray detector/film cassette
The device is a protector for a digital X-ray flat panel
detector
786 Wheelchair
Intended to be used by physically disabled
individuals and old age patients with limited mobility
787 Wheelchair accessory, power
assisted conversion kit
A device that can only fulfil its purpose when used
together with and to enhance the function of a
wheelchair. It will provide the occupant/assistant
with a powered driving assistance.
788 Wire tensioner
Intended to be used during orthopaedic surgery to
apply an appropriate tension to a wire that is being
implanted, usually as part of a system to provide
corrective surgery to the spine. This is a reusable
device.
789 Woven gauze pad, non-sterile
A non-medicated, non-sterile device in the form of a
patch or swab (also referred to as a sponge) made
from woven material (e.g., cotton, cellulose) and
primarily designed to absorb fluids for medical
purposes; it does not include petrolatum. It is
typically used to cover/clean superficial wounds or
abrasions and absorb their exudates, absorb body-
surface exudates, or to apply topical medications.
790 Woven gauze roll, non-sterile
A non-sterile device in the form of a long strip of
stretchable, absorbent, woven material (e.g., cotton,
cellulose), wound into a roll, and typically designed
for a variety of applications (non-dedicated) such as
a primary wound dressing, dressing retention, injury
padding and compression.
791 Woven gauze roll/sheet
A device in the form of a long length of stretchable,
absorbent, woven material (e.g., cotton, cellulose),
wound into a roll or folded, and typically designed for
a variety of applications (non-dedicated) such as a
primary superficial wound dressing, dressing
retention, and injury padding and compression.
S.No. Device Name
Intended Use
792 Wrist immobilizer
Intended to temporarily render the wrist immovable
as therapy for non- displaced fractures, strains,
sprains, and muscle injuries of the wrist.
793 Wrist splint
Intended to temporarily render the wrist immovable
as therapy for non-displaced fractures sprains &
muscle injuries of the wrist.
794 Wrist/hand/finger flexion orthosis,
manual
An externally applied orthopedic appliance intended
to assist in the flexion of the wrist, hand, and fingers
795 Wrist/hand/finger stable-position
orthosis
An externally applied orthopaedic appliance or
apparatus designed to maintain both the transverse
and longitudinal arches of the hand, and to hold the
thumb in a functional position. This is a reusable
device
796 Wrist/hand/metacarpophalangeal
joint extension orthosis
Device is designed to assist in the extension of the
wrist, hand, and metacarpophalangeal joints (MP
joints or knuckles)
797 X- ray marker for strut
Intended to be used in Six axis correction apparatus
with software.
798 X-ray film cover
A small envelope cover used to hold and protect x-
ray films that goes into a patient's mouth
momentarily when taking an x-ray and then
processed.
799 X-ray film magnifying glass
A convex lens used for magnifying the details of a
viewed radiograph (x-ray) and shade against strong
interfering light. It can be hand-held or mounted on
the end of an adjustable arm (e.g., a flexible tube or
an articulating arm) which is attached to a stable
base or clamped to a surface worktop.
800 X-ray film processing chemical,
automatic
A chemical, e.g., a fixer or developer, which is
necessary for automatic processing of x-ray film or
other radiographic films used in, e.g., computed
tomography (CT), magnetic resonance (MR),
ultrasound or nuclear medicine.
801 X-ray film processing chemical,
manual
A chemical, e.g., a fixer or developer, used in the
manual processing of x-ray film or other kinds of
radiographic film used in medical applications, e.g.,
computed tomography (CT), magnetic resonance
imaging (MRI), ultrasound and nuclear medicine
applications.
802 X-Ray hanger
Intended to position and hold the radiograph during
X ray film development, eliminate streaking and
maintain alignment.
S.No. Device Name
Intended Use
803 Y-Connector (accessory to perfusion
sets)
Intended to position and hold the radiograph during
X ray film development, eliminate streaking and
maintain alignment.