44 ICMR-CDSCO Standard Performance Evaluation Protocols 2025-Sep-17 7383 KB

Document text

Page 1 of 459

Contents

  1. Performance evaluation protocol for Chikungunya IgM ELISA kits ...................................... 5
  2. Performance evaluation protocol for Chikungunya IgM RDT kits ........................................ 15
  3. Performance evaluation protocol for Chikungunya real-time PCR kits ................................. 25
  4. Performance evaluation protocol for Dengue NS1 RDT kits ................................................. 35
  5. Field evaluation protocol for Dengue NS1 RDT kits ............................................................. 47
  6. Performance evaluation protocol for Dengue NS1 ELISA kits ............................................. 57
  7. Field evaluation protocol for Dengue NS1 ELISA kits.......................................................... 68
  8. Performance evaluation protocol for Dengue IgM RDT kits ................................................. 78
  9. Performance evaluation protocol for Dengue IgM ELISA kits .............................................. 90
  10. Performance evaluation protocol for Dengue NS1/IgM combo RDT kits ........................... 101
  11. Field evaluation protocol for Dengue NS1 and IgM combo RDT kits ................................ 114
  12. Performance evaluation protocol for Dengue real-time PCR kit ......................................... 126
  13. Field evaluation protocol for Dengue real-time PCR kits .................................................... 138
  14. Performance evaluation protocol for Real-time PCR tests for Zika virus ............................ 148
  15. Performance evaluation protocol for Dengue IgG RDT kits ................................................ 161
  16. Performance evaluation protocol for Dengue IgM and IgG RDT combo kits ..................... 172
  17. Performance evaluation protocol for Dengue IgG ELISA kits ............................................ 185
  18. Information on Operational and Test Performance Characteristics Required from Manufacturers for Dengue/Chikungunya/ Zika IVD ........................................................... 197
  19. General Guidelines ............................................................................................................... 201
  20. Performance Evaluation of Molecular IVD Kit detecting influenza A & B viruses, and subtyping into A (H1N1) pdm 09, A(H3N2), B(Yamagata) & B(Victoria) in single plex or multiplex format ................................................................................................................... 205
  21. Performance Evaluation of Molecular IVD Kit detecting SARS-CoV-2 in single plex or multiplex format ................................................................................................................... 213
  22. Performance Evaluation of Molecular IVD Kit detecting Respiratory Syncytial Virus (RSV /RSV A/RSV B) in single plex or multiplex format ............................................................ 220
  23. Performance Evaluation of Molecular IVD Kit detecting Influenza virus and SARS-CoV-2 in multiplex format ............................................................................................................... 227
  24. Performance Evaluation of Molecular IVD Kit detecting Influenza virus, SARS-CoV-2 and Respiratory Syncytial Virus (RSV) in multiplex format ...................................................... 230
  25. Performance Evaluation Report Format ............................................................................... 233
  26. Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers .............................................................................................................. 239
  27. Performance evaluation protocol for Human Metapneumovirus real-time PCR kit ............ 243
  28. Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers .............................................................................................................. 254

Page 2 of 459

  1. Performance evaluation protocol for multiplex respiratory virus real-time PCR kit ........... 258
  2. Performance evaluation report for multiplex respiratory virus real-time PCR kits ............. 270
  3. Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers .............................................................................................................. 272
  4. Performance evaluation protocol for Malaria Rapid diagnostic test (RDT) kits .................. 276
  5. Performance evaluation protocol for Malaria ELISA kits ................................................... 287
  6. Performance evaluation protocol for Malaria real-time PCR kits ........................................ 298
  7. Field evaluation protocol for combo Malaria Rapid Diagnostic Test (RDT) kits (detecting P vivax and P falciparum) ....................................................................................................... 310
  8. Information on Operational and Test Performance Characteristics Required from Manufacturers for Malaria IVD ........................................................................................... 320
  9. Performance evaluation protocol for Nipah virus real-time PCR kit ................................... 324
  10. Performance evaluation report for Nipah virus real-time PCR kits ..................................... 333
  11. Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers .............................................................................................................. 335
  12. Performance evaluation protocol for Chandipura virus real-time PCR kits ......................... 339
  13. Performance evaluation report for Chandipura virus real-time PCR kits ........................... 348
  14. Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers .............................................................................................................. 350
  15. Field Evaluation Protocol for Typhoid Molecular IVDs ...................................................... 354
  16. Performance Evaluation Protocol for Typhoid Molecular IVDs ......................................... 363
  17. Field Evaluation Protocol for Typhoid Antigen-based IVDs ............................................... 371
  18. Performance Evaluation Protocol for Typhoid Antigen-based IVDs ................................... 382
  19. Field Evaluation Protocol for Typhoid Antibody-based IVDs............................................. 391
  20. Performance Evaluation Protocol for Typhoid Antibody-based IVDs ................................ 402
  21. Performance evaluation report format for Typhoid IVDs .................................................... 410
  22. Information on Operational and Test Performance Characteristics Required from Manufacturers ...................................................................................................................... 412
  23. Analytical Performance Evaluation of IVD for Pulmonary Tuberculosis ................. 416
  24. Field Performance Evaluation of IVD for Pulmonary Tuberculosis .......................... 433
  25. Field Performance Evaluation of IVD for Pulmonary DR-TB .................................... 447

Page 3 of 459

ARBOVIRUS IN-VITRO DIAGNOSTICS

Page 4 of 459

List of Contributors: A. Working Group:

  1. Dr. Alagarasu Kalichamy, Scientist-E, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India
  2. Dr. Megha Brijwal, Additional Professor, Department of Microbiology, All India Institute of Medical Sciences, Delhi
  3. Dr. Gururaj Rao Deshpande, Scientist-C, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India
  4. Dr. M. Santhosh Kumar, Scientist D, ICMR-National Institute of Epidemiology, Department of Health Research, Ministry of Health and Family Welfare, Government of India
  5. Ms. Krittika Bhattacharyya, Statistical Officer (Planning), Directorate of Economics and Statistics, Government of National Capital Territory of Delhi
  6. Dr. Vishal Deo, Scientist-C, ICMR-ICMR- National Institute for Research in Digital Health and Data Science, Department of Health Research, Ministry of Health and Family Welfare, Government of India
  7. Dr. Labanya Mukhopadhyay, Scientist-C, Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

B. Review Committee:

  1. Dr. Vasanthapuram Ravi, Former Dean Research and Head of Neurovirology, National Institute of Mental Health and Neuro-Sciences, Bengaluru, Karnataka
  2. Dr. Amita Jain, Former Professor and Head of the Department of Microbiology, King George’s Medical University, Lucknow, Uttar Pradesh
  3. Dr. Manoj Murhekar, Scientist-G and Director, ICMR-National Institute of Epidemiology, Department of Health Research, Ministry of Health and Family
  4. Dr. Lalit Dar, Professor and Head of the Department of Microbiology, All India Institute of Medical Sciences, Delhi
  5. Dr. Gajanan Sapkal, Scientist-F, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India
  6. Dr. Guruprasad Medigeshi, Professor, Indian Institute of Science Education and Research, Tirupati, Andhra Pradesh
  7. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control Organization, Directorate General of Health Services, Ministry of Health and Family
  8. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control
  9. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central Drugs Standard Control Organization, Government of India
  10. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family

Page 5 of 459

Performance evaluation protocol for Chikungunya IgM ELISA kits I. Background:
CDSCO and ICMR, New Delhi, have aimed at facilitating the availability of Quality-Assured diagnostic kits appropriate for use in India. Hence the following guidelines shall establish uniformity in performance evaluation of in-vitro diagnostic kits (IVD). The performance evaluation is to independently verify the manufacturer’s claim regarding IVD performance. II. Purpose: To evaluate the performance characteristics of Chikungunya IgM ELISA kits in the diagnosis of Chikungunya infection using irreversibly de-identified leftover archived/ spiked clinical samples. III. Requirements:

  1. Supply of kits under evaluation (Along with batch/lot No. Expiry & required details). If the kit to be evaluated works in a closed system format, the manufacturer needs to supply the required equipment.
  2. Evaluation sites/laboratories (With required equipment)
  3. Reference test kits
  4. Characterised Evaluation panel
  5. Laboratory supplies
    IV. Ethical approvals:
    Performance evaluation activities using irreversibly de-identified leftover clinical samples are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024.
    Investigators are required to submit a self-declaration form, as outlined in the ICMR guidelines, to the institutional authorities and ethics committee for information. V. Procedure:
  6. Study design/type: Diagnostic accuracy study using irreversibly de-identified leftover clinical/spiked samples.
  7. Preparation of Evaluation sites/laboratories:
    Identified IVD kit evaluation laboratories should establish their proficiency through
    A. Accreditation for at least one Quality management system (accreditation for Testing Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory. B. Staff training: All the staff involved in IVD kit evaluation should undergo hands on training and competency testing on following
    ⮚ Preparation & characterization of kit evaluation panel

Page 6 of 459

⮚ Handling of Chikungunya IgM ELISA kits received for performance evaluation (Verification/Storage/Unpacking etc). ⮚ Testing, interpreting, recording of results & reporting ⮚ Data handling, data safety & confidentiality 3. Preparation of Chikungunya IgM ELISA IVD kit evaluation panel: Well characterised Chikungunya IVD kit evaluation panel is a critical requirement for performance evaluation of IVD kits. Hence statistically significant number of sera samples should be available from Chikungunya confirmed cases. Further characterised for Chikungunya IgM positivity by using approved reference kits having high sensitivity and specificity. Chikungunya IgM performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples. 4. Reference assay:
All the samples will be tested by CDC/NIV real-time (RT-PCR) assay. Samples which are positive by RT-PCR assay will be further tested by any two of the following Chikungunya IgM ELISA kits: i. ICMR-NIV MAC ELISA kit ii. Inbios CHIKjj Detect™ IgM ELISA iii. Anti-Chikungunya virus ELISA (IgM) Test (Euroimmun, Luebeck, Germany) iv. Any WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved IgM ELISA (as and when available) Samples positive by PCR and at least two kits will be considered positive. If sufficient RT- PCR positive samples are not available, samples positive by at least 2 ELISA kits (of the kits mentioned above) can be considered as true positive samples.
Samples which are negative by RT-PCR and at least two IgM ELISA kits mentioned above will be considered as Chikungunya negative samples. 5. Sample size and sample panel composition: Sample sizes of positive and negative samples and sample panel composition against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming 95% level of significance, and an absolute precision of 5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.

Page 7 of 459

Sample sizes are calculated using the formulae:

𝑛𝑠𝑒 ≥𝑍2 x 𝑆𝑒 (1 −𝑆𝑒) 𝑑2

𝑛𝑠𝑝 ≥𝑍2 x 𝑆𝑝 (1 −𝑆𝑝) 𝑑2

· n (se) is the minimum number of positive samples. · n (sp) is the minimum number of negative samples. · Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96). · Se is the predetermined sensitivity. · Sp is the predetermined specificity. · d is the predetermined marginal error (5%)

Positive samples: Positive samples should be positive by RT-PCR at least two ELISA kits from the three mentioned above. If sufficient RT-PCR positive samples are not available, samples positive by at least 2 ELISA kits (of the kits mentioned above) can be considered as true positive samples. Negative samples: Samples which are negative by RT-PCR and at least two IgM ELISA kits mentioned above will be considered as Chikungunya negative samples. Table 1. Sample sizes and panel composition of positive chikungunya samples for different values of sensitivity claimed by the manufacturer Sensitivity Calculated sample size Minimum no. of Positive Samples required [Sample size rounded off] # Sample Panel Composition 99% 15 20 Strong positive: 6 Moderate positive: 7 Weak positive: 7 95% 73 80 Strong positive: 24 Moderate positive: 28 Weak positive: 28 90% 138 140 Strong positive: 42 Moderate positive: 49 Weak positive: 49 85% 196 200 Strong positive: 60 Moderate positive: 70 Weak positive: 70

Page 8 of 459

80% 246 250 Strong positive: 75 Moderate positive: 87 Weak positive: 88 The samples need to be classified as strong, moderate and weak positives based on ELISA units of the reference assay.

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics.

Table 2. Sample sizes and panel composition of negative chikungunya samples for different values of specificity claimed by the manufacturer. Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off]# 99% 15 20

  1. Cross reactivity panel:  Dengue IgM positive: 3 @  Rubella IgM positive: 1 *
  2. aAcute febrile illness cases: 8
  3. bHealthy subjects from endemic regions: 8
    95% 73 80
  4. Cross reactivity panel:  Dengue IgM positive: 15 @  Rubella IgM positive: 5 *
  5. aAcute febrile illness cases: 30
  6. bHealthy subjects from endemic regions: 30 90% 138 140
  7. Cross reactivity panel:  Dengue IgM positive: 26 @  Rubella IgM positive: 8 *
  8. aAcute febrile illness cases: 53
  9. bHealthy subjects from endemic regions: 53 85% 196 200
  10. Cross reactivity panel:  Dengue IgM positive: 38 @  Rubella IgM positive: 12 *
  11. aAcute febrile illness cases: 75
  12. bHealthy subjects from endemic regions: 75 80% 246 250
  13. Cross reactivity panel:  Dengue IgM positive: 47 @  Rubella IgM positive: 15 *
  14. aAcute febrile illness cases: 94
  15. bHealthy subjects from endemic regions: 94 a Acute febrile illness cases negative for above pathogens AND Chikungunya IgM & PCR b Samples from healthy subjects from endemic regions negative for all Chikungunya markers (IgM, RNA)

Page 9 of 459

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics.

Cross reactivity panel is arranged in descending order of priority. The pathogens marked @ are essentially to be tested. It is recommended to test for all pathogens listed in the cross reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

  1. Evaluation method:

The index test and the reference tests should be run simultaneously on the sample panel, and results should be recorded.

  1. Interpretation of results:
    Reference test and index test results will be interpreted as per kit IFU.
  2. Resolution of discrepant results:

True positive samples: These are samples positive by reference assay and index test. True negative samples: These are samples negative by reference assay and index test. False positive samples: These are samples negative by reference assay and positive by index test. False negative samples: These are samples positive by reference assay and negative by index test.

  1. Repeatability and Reproducibility Assessment:

A. Repeatability Assessment This should be done to assess the repeatability of the detection of target analyte using the kit under evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.
Concordance should be 100% based on positive and negative test result (qualitative).

B. Reproducibility Assessment: Reproducibility testing should include the following: a. Lot-to-lot reproducibility

Page 10 of 459

Three lots of an IVD kit shall be evaluated. Sample size for lot-to-lot reproducibility should be as follows:
 First lot of the kit: should be tested on statistically significant number of positive and negative samples as calculated in the protocol.
 Second lot of the kit: should be tested on 25 samples (15 positive samples comprising 10 low positive AND 5 moderate/strong positive samples, and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples (15 positive samples comprising 10 low positive AND 5 moderate/strong positive samples, and 10 negative samples).  There should be no lot-to-lot variability (qualitative). Refer Figure 1.

b. Inter-Operator variability: Testing should be conducted by two different operators, keeping all other testing parameters undisturbed. Within-run and between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.

Page 11 of 459

c. Day-to-day variability: Testing should be performed on at least two non-consecutive days. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.

d. Machine-to-machine variability: It is desirable (not mandatory) to evaluate the IVD kit using two different manufacturer recommended platforms (if applicable). 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.

  1. Blinding of laboratory staff:

To ensure rigor of the evaluation process, laboratory staff performing the evaluation should be blinded to the status of the clinical samples. The PI of the evaluation exercise should remain unblinded, i.e., privy to the status of the samples. Another senior laboratory staff selected by the PI may remain unblinded and carry out coding of samples and dispensing them into similar-looking vials to be used for testing, and maintaining the database of results. Staff performing the reference test and the test under evaluation, interpretation of the test result, and entering the results against the coded samples in the database, should remain blinded to the status of samples till the completion of evaluation. The data should be analyzed only by the PI of the evaluating lab, and the PI should maintain confidentiality of data. Refer to Fig. 2.

Fig.2: Blinding in evaluation exercise

Page 12 of 459

  1. Acceptance Criteria: Sensitivity: ≥90% Specificity: ≥95% Cross-reactivity with other pathogens listed in the negative sample panel: Minimal To achieve at least the performance characteristics outlined in the acceptance criteria, ≥140 positive samples and ≥80 negative samples should be used for evaluation.
  2. Publication Rights: The PI(s) of the evaluating labs shall retain publication rights of the evaluation as lead author(s). After following due procedure as defined in this document, once any kit is found to be Not of Standard Quality, thereafter, no request for repeat testing of the same kit will be acceptable.
    Any request of re-validation from the same manufacturer for the same test type will only be entertained after a minimum of 3 months and only if a high-level technical summary of modifications or functional improvements to the kit design is submitted, without explicit disclosure of proprietary information. Clinical samples are precious, therefore, repeat evaluation of a kit using the same/ different well-characterized sample panel at a different laboratory may be considered only for kits which claim high performance characteristics (sensitivity and specificity 95% and above), but which fail the performance evaluation by a margin of 5%.

VI. References:

  1. Kikuti M, Tauro LB, Moreira PSS, et al. Evaluation of two commercially available Chikungunya virus IgM enzyme-linked immunoassays (ELISA) in a setting of concomitant transmission of Chikungunya, Dengue and Zika viruses. Int J Infect Dis. 2020 Feb;91:38-43.
  2. Johnson BW, Goodman CH, Holloway K, de Salazar PM, Valadere AM, Drebot MA. Evaluation of Commercially Available Chikungunya Virus Immunoglobulin M Detection Assays. Am J Trop Med Hyg. 2016 Jul 6;95(1):182-192. doi: 10.4269/ajtmh.16-0013. Epub 2016 Mar 14.
  3. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: https://iris.who.int/bitstream/handle/10665/258985/WHO-EMP-RHT-PQT-TGS3-2017.03- eng.pdf;sequence=1
4. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability and diagnostic performance: a go-to review. European radiology experimental, 8(1), 79. https://doi.org/10.1186/s41747-024-00474-w

VII. Performance evaluation report format

Page 13 of 459

PERFORMANCE EVALUATION REPORT FOR CHIKUNGUNYA IgM ELISA KIT

Name of the product (Brand /generic) Name and address of the legal manufacturer Name and address of the actual manufacturing site Name and address of the Importer Name of supplier: Manufacturer/Importer/Port office of CDSCO/State licensing Authority Lot No / Batch No.: Product Reference No/ Catalogue No Type of Assay Kit components Manufacturing Date Expiry Date Pack size (Number of tests per kit) Intended Use Number of Tests Received Regulatory Approval: Import license / Manufacturing license/ Test license

License Number: Issue date:

Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: strong, moderate, weak/simulated samples)
Negative samples (provide detail: clinical/spiked, including cross reactivity panel/simulated samples)

Results:

Reference assay ……….……………… (name)

Positive Negative Total Name of Chikungunya IgM ELISA kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

Conclusions: o Cross-reactivity: o Invalid test rate:

Page 14 of 459

o Performance: Satisfactory / Not satisfactory (Sensitivity and specificity have been assessed in controlled lab setting using kits provided by the manufacturer from the batch mentioned above using ….. sample. Results should not be extrapolated to other sample types.)

Disclaimers

  1. This validation process does not approve / disapprove the kit design
  2. This validation process does not certify user friendliness of the kit / assay Note: This report is exclusively for ………………………Kit (Lot No……) manufactured by …………… (Supplied by ……….) Evaluation Done on …………………… Evaluation Done by …………………………. Signature of Director/ Director-In-charge …………………… Seal …………………………

Page 15 of 459

Performance evaluation protocol for Chikungunya IgM RDT kits I. Background:
Diagnostics kits appropriate for use in India. Hence the following guidelines shall establish II. Purpose: To evaluate the performance characteristics of Chikungunya IgM RDT kits in the diagnosis of Chikungunya infection using irreversibly de-identified leftover archived/ spiked clinical samples. III. Requirements:
V. Procedure: through
A. Accreditation for at least one Quality management system (accreditation for Testing Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory. B. Staff training: All the staff involved in IVD kit evaluation should undergo hands on

Page 16 of 459

⮚ Handling of Chikungunya IgM RDT kits received for performance evaluation 3. Preparation of Chikungunya IgM Rapid IVD kit evaluation panel: Well characterised Chikungunya IVD kit evaluation panel is a critical requirement for should be available from Chikungunya confirmed cases. Further characterised for Chikungunya IgM positivity by using approved reference kits having high sensitivity and specificity. Chikungunya IgM performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples. 4. Reference assay:
All the samples will be tested by CDC/NIV real-time Chikungunya PCR assay. Samples which are positive by RT-PCR assay will be further tested by any two of the following IgM ELISA kits: i. ICMR-NIV MAC ELISA kit ii. Inbios CHIKjj Detect™ IgM ELISA iii. Anti-Chikungunya virus ELISA (IgM) Test (Euroimmun, Luebeck, Germany) iv. Any WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved IgM ELISA (as and when available) Samples positive by PCR and at least two kits will be considered positive. If sufficient RT- PCR positive samples are not available, samples positive by at least 2 ELISA kits (of the kits mentioned above) can be considered as true positive samples.
Samples which are negative by RT-PCR and at least two IgM ELISA kits mentioned above will be considered as Chikungunya negative samples. 5. samples and sample panel composition against different values of sensitivity and 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.

Page 17 of 459

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· IR is the invalid test rate

Positive samples: Positive samples should be positive by RT-PCR at least two ELISA kits from the three mentioned above. If sufficient RT-PCR positive samples are not available, samples positive by at least 2 ELISA kits (of the kits mentioned above) can be considered as true positive samples. Negative samples: Samples which are negative by RT-PCR and at least two IgM ELISA kits mentioned above will be considered as Chikungunya negative samples. Table 1. Sample sizes and panel composition of positive chikungunya samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off] # 99% 16 20 Strong Positive: 6 Moderate Positive: 7 Weak Positive: 7 95% 77 80 Strong Positive: 24 Moderate Positive: 28 Weak Positive: 28 90% 145 150 Strong Positive: 44 Moderate Positive: 53 Weak Positive: 53 85% 206 210 Strong Positive: 62

Page 18 of 459

Moderate Positive: 74 Weak Positive: 74 80% 258 260 Strong Positive: 78 Moderate Positive: 91 Weak Positive: 91 The samples need to be classified as strong, moderate and weak positives based on ELISA units of the reference assay.

#It is recommended to calculate the sample size as per manufacturer's claims of

Table 2. Sample sizes and panel composition of negative chikungunya samples for different values of specificity claimed by the manufacturer. Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off]

99% 16 20

  1. Cross-reactivity panel  Dengue IgM positive: 3 @  Rubella IgM positive: 1 *
  2. aAcute febrile illness cases: 12
  3. bHealthy subjects from endemic regions: 4
    95% 77 80  Dengue IgM positive: 13 @  Rubella IgM positive: 3 *
  4. aAcute febrile illness cases: 48
  5. bHealthy subjects from endemic regions: 16 90% 145 150  Dengue IgM positive: 25 @  Rubella IgM positive: 5 *
  6. aAcute febrile illness cases: 90 30 85% 206 210  Dengue IgM positive: 35 @  Rubella IgM positive: 7 *
  7. aAcute febrile illness cases: 126 42 80% 258 260  Dengue IgM positive: 43 @  Rubella IgM positive: 9 *
  8. aAcute febrile illness cases: 156

Page 19 of 459

52 a Acute febrile illness cases negative for above pathogens AND Chikungunya IgM & PCR b Samples from healthy subjects from endemic regions negative for all Chikungunya markers (IgM, RNA)

#It is recommended to calculate the sample size as per manufacturer's claims of

It is recommended to test for all pathogens listed in the cross reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross reactive sample panel remain the same by compensating with the

Evaluation method:

Interpretation of results:
8. Resolution of discrepant results: test. test.

under evaluation.

Page 20 of 459

Three lots of an IVD kit shall be evaluated. Sample size for lot-to-lot reproducibility negative samples).
samples).

b. Inter-Operator variability: Testing and result interpretation should be conducted by two different operators, keeping all other testing parameters undisturbed. Within-run and between-run imprecision (if applicable) should be measured. 3 positive samples

Page 21 of 459

(strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.

c. Day-to-day variability: Testing should be performed on at least two non-consecutive days. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.

kit using two different manufacturer recommended platforms (if applicable). 3 positive samples (strong, moderate and weak positive samples) and 3 negative

Page 22 of 459

  1. Acceptance criteria: Sensitivity: ≥80% Specificity: ≥90% Invalid test rate: ≤5% To achieve at least the performance characteristics outlined in the acceptance criteria, ≥260 positive samples and ≥150 negative samples should be used for evaluation.
  2. Publication Rights: author(s). acceptable.
    VI. References:
  3. Kikuti M, Tauro LB, Moreira PSS, et al. Evaluation of two commercially available Chikungunya virus IgM enzyme-linked immunoassays (ELISA) in a setting of concomitant transmission of Chikungunya, Dengue and Zika viruses. Int J Infect Dis. 2020 Feb;91:38-43.
  4. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
3. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

VII.

Page 23 of 459

PERFORMANCE EVALUATION REPORT FOR CHIKUNGUNYA IgM RDT KIT Name and address of the actual manufacturing site Name of supplier: Manufacturer/Importer/Port office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type Negative samples (provide details: clinical/spiked,

Results:

Positive Negative Total Name of Chikungunya IgM RDT kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

Conclusions: o Cross-reactivity:

Page 24 of 459

Disclaimers Note: This report is exclusively for ………………………Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ……………………..

Page 25 of 459

Performance evaluation protocol for Chikungunya real-time PCR kits I. Background:
the uniformity in performance evaluation of in-vitro diagnostic kits (IVD). The performance II. Purpose: To evaluate the performance characteristics of Chikungunya PCR kits in the diagnosis of Chikungunya infection using irreversibly de-identified leftover archived/ spiked clinical samples. III. Requirements:
1. Supply of kits under evaluation (Along with batch/lot No. Expiry & required details). 2. Evaluation sites/laboratories (With required equipment) 3. Reference test kits 4. Characterised Evaluation panel 5. Laboratory supplies V. Procedure: through
Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory.

Page 26 of 459

⮚ Handling of Chikungunya PCR kits received for performance evaluation 3. Preparation of Chikungunya RNA evaluation panel: Well characterised Chikungunya sample panel positive for RNA is a critical requirement for performance evaluation of IVD kits utilizing genome detection. Hence statistically significant number of sera/whole blood samples should be available from Chikungunya PCR confirmed cases.
4. RNA extraction: RNA extraction should be performed as per reference test and index test Instructions for Use (IFU). If the manufacturer of the index test recommends a specific RNA extraction kit/system, the same needs to be provided by the manufacturer if the evaluating lab is unable to procure the same/it is not available within the lab’s IVD evaluation scope. 5. Real-Time PCR System: PCR should be performed using IVD-approved machines. If any equipment(s) is specified in the IFU of the index test, it should be used for the evaluation, and it should be provided by the manufacturer if not available within the lab’s IVD evaluation scope.
Real-time closed systems/devices awaiting evaluation should be provided by the manufacturer along with all necessary components, supplies and reagents. 6. Internal control/Extraction control: The test under evaluation should have an internal control or extraction control (RNA added before extraction to a sample).
7. Reference assay:
Any WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Chikungunya PCR assay or CDC/NIV protocol for detection of Chikungunya RNA should be used as the reference assay. All positive samples should be confirmed positive for Chikungunya by reference assay. All negative samples should be negative for all markers of Chikungunya infection (RNA using reference assay AND IgM using any two of the following kits - ICMR-NIV MAC ELISA kit/Inbios CHIKjj Detect™ IgM ELISA/Anti-Chikungunya virus ELISA (IgM) Test (Euroimmun, Luebeck, Germany)/ any WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved IgM ELISA (as and when available) 8. Sample size and sample panel composition: Sample sizes of positive and negative samples and sample panel composition against different values of sensitivity and

Page 27 of 459

95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

Table 1. Sample sizes and panel composition of positive chikungunya samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off]# 99% 16 20 Strong Positive: 6 Weak Positive: 7 95% 77 80 Strong Positive: 24 Weak Positive: 28 90% 145 150 Strong Positive: 44 Moderate Positive: 53

Page 28 of 459

Weak Positive: 53 85% 206 210 Strong Positive: 62 Moderate Positive: 74 Weak Positive: 74 #It is recommended to calculate the sample size as per manufacturer's claims of Strong positive (Ct value between <25)
Moderate positive (Ct value between 25-30)
Weak positive (Ct value between >30 to 34)
Table 2. Sample sizes and panel composition of negative chikungunya samples for different values of specificity claimed by the manufacturer. Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off]

99% 16 20  Dengue PCR positive: 4 @  Rubella PCR positive: 1 * 2. aAcute febrile illness cases: 10 3. bHealthy subjects from endemic regions: 5
95% 77 80  Dengue PCR positive: 15 @  Rubella PCR positive: 5 * 2.aAcute febrile illness cases: 40 3.bHealthy subjects from endemic regions: 20 90% 145 150  Dengue PCR positive: 28 @  Rubella PCR positive: 9 * 2. aAcute febrile illness cases: 75 38 85% 206 210  Dengue PCR positive: 39 @  Rubella PCR positive: 13 * 2. aAcute febrile illness cases: 105 3.bHealthy subjects from endemic regions: 53 a Acute febrile illness cases negative for above pathogens AND Chikungunya IgM & PCR b Samples from healthy subjects from endemic regions negative for all Chikungunya markers (IgM, RNA).

Page 29 of 459

#It is recommended to calculate the sample size as per manufacturer's claims of

It is recommended to test for all pathogens listed in the cross reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

Evaluation method:

  1. Interpretation of results:
  2. Resolution of discrepant results:

test. test.

  1. Repeatability and Reproducibility Assessment:

This should be done to assess the repeatability of the detection of target using the kit under evaluation.

Page 30 of 459

Three lots of an IVD kit shall be evaluated. Sample size for lot-to-lot reproducibility should be as follows:
negative samples).
samples).

imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.
Concordance should be 100% based on positive and negative test result (qualitative).

Page 31 of 459

c. Day-to-day variability: Testing should be performed on at least two non- consecutive days. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

d. Machine-to-machine variability: It is desirable (not mandatory) to evaluate the IVD kit using two different manufacturer recommended platforms (if applicable). 3 positive samples (strong, moderate and weak positive samples) and 3 negative (qualitative).

It is desirable to report Ct standard deviation across repeats (≤1cycle deviation recommended).

  1. Blinding of laboratory staff:

Page 32 of 459

  1. Acceptance criteria: Sensitivity: ≥95% Specificity: ≥98% To achieve at least the performance characteristics outlined in the acceptance criteria, ≥80 positive samples and ≥80 negative samples should be used for evaluation.
  2. Publication Rights: author(s). acceptable.

VI. References:

  1. Santiago, G.A., Vázquez, J., Courtney, S. et al. Performance of the Trioplex real-time RT- PCR assay for detection of Zika, Dengue, and Chikungunya viruses. Nat Commun 9, 1391 (2018). https://doi.org/10.1038/s41467-018-03772-1
  2. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
3. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

VII.

Page 33 of 459

PERFORMANCE EVALUATION REPORT FOR CHIKUNGUNYA REAL-TIME PCR KITS Name and address of the actual manufacturing site Name of supplier: Manufacturer/Importer/Port office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: clinical/spiked, strong, moderate, weak/simulated samples)

Results

Positive Negative Total Name of Chikungunya real-time PCR kits Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity: o Performance: Satisfactory / Not satisfactory

Page 34 of 459

(Sensitivity and specificity have been assessed in controlled lab setting using kits provided by the manufacturer from the batch mentioned above using ….. sample. Results should not be extrapolated to other sample types.)

Disclaimers Note: This report is exclusively for Chikungunya………….. Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal …………………………..

Page 35 of 459

Performance evaluation protocol for Dengue NS1 RDT kits I. Background:
evaluation is to independently verify the manufacturer’s claim IVD performance. II. Purpose: To evaluate the performance characteristics of Dengue NS1 RDT kits in the diagnosis of Dengue infection using irreversibly de-identified leftover archived/ spiked clinical samples. III. Requirements:
V. Procedure: through
A. Accreditation for one Quality management system (accreditation for Testing Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory. B. Staff training: All the staff involved in IVD kit evaluation should undergo hands on ⮚ Handling of Dengue NS1 Rapid IVD kits received for performance evaluation

Page 36 of 459

Preparation of Dengue RDT IVD kit evaluation panel: Well characterised Dengue NS1 RDT IVD kit evaluation panel is a critical requirement for should be available from Dengue confirmed cases. Further characterised for Dengue NS1 positivity by using approved reference kits having high sensitivity and specificity. Dengue NS1 performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples. 4. Reference assay:
WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Dengue NS1 ELISA kit should be used as reference assay. Serotype status to be assessed using CDC/NIV real-time PCR serotyping protocols.
5. samples and sample panel composition against different values of sensitivity and 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

Page 37 of 459

Positive samples: The panel of positive samples should include samples positive by the reference assay and real-time PCR assay (True positives). Samples should be representative of all 4 serotypes and varying degrees of positivity. The samples should be classified as strong, moderate and weak positives based on ELISA units of the reference assay.

Negative samples: These should include samples negative by the reference NS1 ELISA assay and real-time PCR using CDC/NIV serotyping protocol (True negatives). Table 1. Sample sizes and panel composition of positive Dengue samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off] # 99% 16 20 Strong Positive: 6 Weak Positive: 7 95% 77 80 Strong Positive: 24 Weak Positive: 28 90% 145 150 Strong Positive: 44 Moderate Positive: 53 Weak Positive: 53 85% 206 210 Strong Positive: 62 Moderate Positive: 74 Weak Positive: 74 80% 258 260 Strong Positive: 78 Moderate Positive: 91 Weak Positive: 91

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. Table 2. Sample sizes and panel composition of negative Dengue samples for different values of specificity claimed by the manufacturer. Specificity Calcula ted sample size Minimum no. of Negative Samples required

Page 38 of 459

[Sample size rounded off] # 99% 16 20 1.PCR/RT-PCR positive samples from other acute febrile illness cases  Chikungunya positive: 4  Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR): 8

2.Cross-reactivity panel (Samples from other flavivirus disease cases)  Japanese Encephalitis PCR/antigen positive: 1@  Zika Virus PCR/antigen positive: 1*  West Nile Virus PCR/antigen positive: 1*

3.Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, nucleic acid): 5

95% 77 80 1.PCR/RT-PCR positive samples from other acute febrile illness cases  Chikungunya positive: 15  Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR): 30

  1. Cross-reactivity panel (Samples from other flavivirus disease cases) positive: 5 @  Zika Virus PCR/antigen positive: 5 *  West Nile Virus PCR/antigen positive: 5 *

  2. Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, nucleic acid): 20

90% 145 150 1.PCR/RT-PCR positive samples from other acute febrile illness cases  Chikungunya positive: 28  Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR): 57

  1. Cross-reactivity panel (Samples from other flavivirus disease cases)

Page 39 of 459

positive: 9@  Zika Virus PCR/antigen positive: 9 *  West Nile Virus PCR/antigen positive: 9 *

3.Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, nucleic acid): 38

85% 206 210 1.PCR/RT-PCR positive samples from other acute febrile illness cases  Chikungunya positive: 39  Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR): 79

  1. Cross-reactivity panel (Samples from other flavivirus disease cases) positive: 13 @  Zika Virus PCR/antigen positive: 13 *  West Nile Virus PCR/antigen positive 13 *

3.Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, nucleic acid): 53

80% 258 260

  1. PCR/RT-PCR positive samples from other acute febrile illness cases  Chikungunya positive: 49  Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR): 98

  2. Cross-reactivity panel (Samples from other flavivirus disease cases) positive: 16 @  Zika Virus PCR/antigen positive: 16 *  West Nile Virus PCR/antigen positive: 16 *

  3. Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, nucleic acid): 65

Page 40 of 459

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics.

It is recommended to test for all pathogens listed in the cross reactivity panel. However, if reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

In the absence of natural samples, spiked samples may be used, as per details provided in the note below.

Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. Recombinant NS1 antigen of cross reactive flaviviruses (Zika, West Nile and Japanese Encephalitis viruses) expressed in mammalian cells can be obtained commercially and reconstituted in serum samples (100 ng -1 µg/ml) and diluted in the ratio of 1:2 and used accordingly (at least five dilutions for each virus specific NS1).
Before used for evaluation, flavivirus NS1 reconstituted in serum samples needs to be tested by the dengue NS1 reference assay, and dilutions which are negative for dengue should be used for evaluation. The serum samples used for reconstitution should be negative for Dengue NS1, RNA and IgM antibody.

Evaluation method:

Resolution of discrepant results:

test. test.

Page 41 of 459

under evaluation.

Three lots of an IVD kit shall be evaluated. Sample size for lot-to-lot reproducibility negative samples).
10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

Page 42 of 459

two different operators, keeping all other testing parameters undisturbed. Within- run and between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples (qualitative).

(qualitative).

(qualitative).

Page 43 of 459

  1. Acceptance criteria: Sensitivity: ≥80% Specificity: ≥95% Cross reactivity with other flavivirus antigens: Minimal To achieve at least the performance characteristics outlined in the acceptance criteria, ≥260 positive samples and ≥80 negative samples should be used for evaluation. author(s). acceptable.

VI. References:

  1. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Vazquez S, Cartozian E, Pelegrino JL, Artsob H, Guzman MG, Olliaro P, Zwang J, Guillerm M, Kliks S, Halstead S, Peeling RW, Margolis HS. Evaluation of commercially available diagnostic tests for the detection of Dengue virus NS1 antigen and anti-Dengue virus IgM antibody. PLoSNegl Trop Dis. 2014 Oct 16;8(10):e3171. doi: 10.1371/journal.pntd.0003171.
  2. Hermann LL, Thaisomboonsuk B, Poolpanichupatam Y, Jarman RG, Kalayanarooj S, Nisalak A, Yoon IK, Fernandez S. Evaluation of a Dengue NS1 antigen detection assay sensitivity and specificity for the diagnosis of acute Dengue virus infection. PLoSNegl Trop Dis. 2014 Oct 2;8(10):e3193. doi: 10.1371/journal.pntd.0003193.
  3. Yow KS, Aik J, Tan EY, Ng LC, Lai YL. Rapid diagnostic tests for the detection of recent Dengue infections: An evaluation of six kits on clinical specimens. PLoS One. 2021 Apr 1;16(4): e0249602. doi: 10.1371/journal.pone.0249602.
  4. Mat Jusoh TNA, Shueb RH. Performance Evaluation of Commercial Dengue Diagnostic Tests for Early Detection of Dengue in Clinical Samples. J Trop Med. 2017; 2017: 4687182. doi: 10.1155/2017/4687182. Epub 2017 Dec 12. PMID: 29379526; PMCID: PMC5742879.

Page 44 of 459

  1. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
6. Mahajan R, Nair M, Saldanha AM, Harshana A, Pereira AL, Basu N, Goswami RP, Bhattacharya N, Bandyopadhay B, SenGupta M, Day M, Flevaud L, Boelaert M, Burza S. Diagnostic accuracy of commercially available immunochromatographic rapid tests for diagnosis of dengue in India. J Vector Borne Dis. 2021 Apr- Jun;58(2):159-164. doi: 10.4103/0972-9062.321747. PMID: 35074951. 7. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

VII.

Page 45 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE NS1 RDT KIT Name and address of the actual manufacturing site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: clinical/spiked, strong, moderate, weak/simulated samples)

Results:

Positive Negative Total Name of Dengue NS1 RDT kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity: ● Conclusions: o Sensitivity, specificity o

Page 46 of 459

(Sensitivity and specificity have been assessed in controlled lab setting using kits provided by the manufacturer from the batch mentioned above using ….. sample. Results should not be extrapolated to other sample types.)

Disclaimers

Note: This report is exclusively for ………….. Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ………………………..

Page 47 of 459

Field evaluation protocol for Dengue NS1 RDT kits I. Background:
II. Purpose: To evaluate the performance characteristics of Dengue NS1 RDT kits in the diagnosis of Dengue infection in individuals with unknown disease status. III. Requirements:
4. Laboratory supplies
IV. Ethical approval:
The study will be initiated after approval from the institutional human ethics committee.
V. Procedure:

  1. Study design/type: Cross-sectional study through
    Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC ⮚ Handling of Dengue NS1 RDT IVD kits received for performance evaluation

Sample size for performance evaluation: Sample sizes of positive and negative samples of Dengue against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming

Page 48 of 459

95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. It is further assumed that 30% of the individuals attending the health care facilities for acute febrile illness and suspected for Dengue will be positive for Dengue. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.
Sample size has to be calculated based on both the sensitivity and the specificity. The final sample size will be the maximum of the two. For example, at 95% sensitivity and 95% specificity, the sample size required will be 260 (maximum of 260 and 110).

𝑑2 x P x (1 −IR) 𝑑2 x P x (1 −IR)

· n (se) is the minimum number of individuals to be enrolled to obtain the requisite number of positive samples. · n (sp) is the minimum number of individuals to be enrolled to obtain the requisite number of negative samples.

P is the prevalence of the disease · d is the predetermined marginal error (5%).

Table 1. Sample sizes for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of individuals to be enrolled* off] # 99% 53 60

Page 49 of 459

95% 255 260 90% 484 490 85% 686 690 80% 861 870

  • Individuals attending the health care facilities for acute febrile illness and suspected for Dengue meeting the inclusion criteria

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

Table 2. Sample sizes for different values of specificity claimed by the manufacturer. Specificity Calculated sample size Minimum no. of individuals to be enrolled* off] # 99% 23 30 95% 109 110 90% 207 210 85% 294 300 80% 369 370

  • Individuals attending the health care facilities for

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity Recruitment of cases shall be halted once desired number of positive and negative samples are reached. 4. Inclusion criteria: Individuals with Dengue like illness (An individual with acute febrile illness of 2-7 days with two or more manifestations: Head ache, retro-orbital pain, myalgia, arthralgia, rash, hemorrhagic manifestations) 5. Exclusion criteria: Individuals with already known positive history for other pathogens 6. Reference assay:
WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Dengue NS1 ELISA kit should be used as reference assay. Serotype status to be assessed using CDC/NIV real-time PCR serotyping protocols.

Page 50 of 459

Study implementation: The individuals with Dengue like illness will be recruited into the study and five ml of whole blood will be collected in vacutainer tubes and the serum will be separated by centrifugation and used for the study. The serum sample will be subjected to the reference tests and the index test. 8. Evaluation method:

  1. Positive samples: Samples positive by the reference NS1 ELISA assay and real-time PCR assay will be considered as true positive sample.
  2. Negative samples:
    Samples negative by the reference NS1 ELISA assay and real-time PCR using CDC/NIV serotyping protocol will be considered as true negative. test. test. A. Cross reactivity: Clinical samples or commercially available NS1 antigens from other flaviviruses will be used to test cross reactivity of the index test. i. Chikungunya PCR positive: 5 samples @ ii. Japanese Encephalitis PCR/antigen positive: 5 samples @ iii. Zika Virus PCR/antigen: 5 samples * iv. West Nile Virus PCR/antigen: 5 samples *

It is recommended to test for all pathogens listed in the cross reactivity panel. However, if reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross reactive sample panel remain the same by compensating with the

Page 51 of 459

In the absence of natural samples, spiked samples may be used, as per details provided in the note below. Note:
Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. Recombinant NS1 antigen of cross reactive flaviviruses (Zika, West Nile and Japanese Encephalitis viruses) expressed in mammalian cells can be obtained commercially and reconstituted in serum samples (100 ng -1 µg/ml) and diluted in the ratio of 1:2 and used accordingly (at least five dilutions for each virus specific NS1). Before used for evaluation, flavivirus NS1 reconstituted in serum samples needs to be tested by the dengue NS1 reference assay, and dilutions which are negative for dengue should be used for evaluation. The serum samples used for reconstitution should be negative for Dengue NS1, RNA and IgM antibody. 12. Statistical analysis:
Sensitivity and specificity will be calculated. Interim analysis of data shall be conducted on completing evaluation of 25%, 50% and 75% of samples. If, at any point, the performance of the assay is found to be not satisfactory, the assay shall not be evaluated further. Evaluation fee shall be charged accordingly. 13. Repeatability and Reproducibility Assessment:

under evaluation.

be as follows:
 First lot of the kit: should be tested on statistically significant number of positive and negative samples as calculated in the protocol.
 Second lot of the kit: should be tested on 25 samples (15 positive samples comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples (15 positive samples comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

Page 52 of 459

two different operators, keeping all other testing parameters undisturbed. Within- run and between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples (qualitative).

(qualitative).

(qualitative).

Page 53 of 459

  1. Resolution of discrepant results:

test. test.

  1. Blinding of laboratory staff:

be blinded to the results of the reference assay. The PI of the evaluation exercise should remain unblinded, i.e., privy to the results of the reference test. Another senior laboratory staff selected by the PI may remain unblinded for overseeing the activity and maintaining the database of results.. The data should be analyzed only by the PI of the evaluating lab, and the PI should maintain confidentiality of data.

  1. Acceptance Criteria: Sensitivity: ≥80% Specificity: ≥95%

To achieve at least the performance characteristics outlined in the acceptance criteria, ≥870 individuals satisfying the case definition need to be enrolled to obtain the requisite number of positive samples. This sample size is sufficient for required number of negative samples. Recruitment should be terminated once the desired number of positive cases is enrolled and tested. 17. Publication Rights: author(s). acceptable.

Page 54 of 459

VI. References:

  1. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Vazquez S, for the detection of Dengue virus NS1 antigen and anti-Dengue virus IgM antibody. PLoSNegl Trop Dis. 2014 Oct 16;8(10):e3171. doi: 10.1371/journal.pntd.0003171.

  2. Hermann LL, Thaisomboonsuk B, Poolpanichupatam Y, Jarman RG, Kalayanarooj S, Nisalak A, Yoon IK, Fernandez S. Evaluation of a Dengue NS1 antigen detection assay sensitivity and specificity for the diagnosis of acute Dengue virus infection. PLoSNegl Trop Dis. 2014 Oct 2;8(10):e3193. doi: 10.1371/journal.pntd.0003193.

  3. Ganeshkumar P, Murhekar MV, Poornima V, Saravanakumar V, Sukumaran K, Anandaselvasankar A, John D, Mehendale SM. Dengue infection in India: A systematic review and meta-analysis. PLoSNegl Trop Dis. 2018 Jul 16;12(7):e0006618. doi: 10.1371/journal.pntd.0006618.

  4. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
5. Mahajan R, Nair M, Saldanha AM, Harshana A, Pereira AL, Basu N, Goswami RP, Bhattacharya N, Bandyopadhay B, SenGupta M, Day M, Flevaud L, Boelaert M, Burza S. Diagnostic accuracy of commercially available immunochromatographic rapid tests for diagnosis of dengue in India. J Vector Borne Dis. 2021 Apr-Jun;58(2):159-164. doi: 10.4103/0972-9062.321747. PMID: 35074951. 6. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

VII.

Page 55 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE NS1 RDT KIT Name and address of the actual manufacturing site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type Positive samples: Not applicable, may categorize cases as per duration of illness Negative samples (may categorize as per duration of illness, must include cross reactivity panel)

Results:

Positive Negative Total Name of Dengue NS1 RDT kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity: ● Conclusions: o Sensitivity, specificity o

Page 56 of 459

(Sensitivity and specificity have been assessed in using kits provided by the manufacturer from the batch mentioned above using ….. sample in ……. (field/controlled lab). Results should not be extrapolated to other sample types.)

Disclaimers

Note: This report is exclusively for NS1…………………Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In charge …………………… Seal ……………………………..

Page 57 of 459

Performance evaluation protocol for Dengue NS1 ELISA kits I. Background:
II. Purpose: To evaluate the performance characteristics of Dengue NS1 ELISA kits in the diagnosis of Dengue infection using irreversibly de-identified leftover archived/ spiked clinical samples. III. Requirements:
V. Procedure: through
Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC ⮚ Handling of Dengue NS1 ELISA kits received for performance evaluation

Page 58 of 459

Preparation of Dengue NS1 ELISA IVD kit evaluation panel: Well characterised Dengue NS1 ELISA IVD kit evaluation panel is a critical requirement for should be available from Dengue confirmed cases. Further characterised for Dengue NS1 positivity by using approved reference kits having high sensitivity and specificity. Dengue NS1 performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples. 4. Reference assay:
WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Dengue NS1 ELISA kit should be used as reference assay. Serotype status to be assessed using CDC/NIV real-time PCR serotyping protocols.
5. samples and sample panel composition against different values of sensitivity and 95% level of significance, and an absolute precision of 5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For require usage of the sample size outlined for 85% specificity.

𝑑2

𝑑2

Page 59 of 459

Positive samples: The panel of positive samples should include samples positive by the reference assay and real-time PCR assay (True positives). Samples should be representative of all 4 serotypes and varying degrees of positivity. The samples should be classified as strong, moderate and weak positives based on ELISA units of the reference assay.

Negative samples: These should include samples negative by the reference NS1 ELISA assay and real-time PCR using CDC/NIV serotyping protocol (True negatives). Table 1. Sample sizes and panel composition of positive Dengue samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off] # 99% 15 20 Strong Positive: 6 Weak Positive: 7 95% 73 80 Strong Positive: 24 Weak Positive: 28 90% 138 140 Strong Positive: 42 Moderate Positive: 49 Weak Positive: 49 85% 196 200 Strong Positive: 60 Moderate Positive: 70 Weak Positive: 70 80% 246 250 Strong Positive: 75 Moderate Positive: 87 Weak Positive: 88

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size

Page 60 of 459

rounded off] # 99% 15 20  Chikungunya PCR positive: 4 @ positive: 1 @  Zika Virus PCR/antigen positive: 1 *  West Nile Virus PCR/antigen positive: 1 * 2. aAcute febrile cases negative for Dengue: 8
3. bHealthy subjects from endemic regions: 5
95% 73 80  Chikungunya PCR positive: 15 @ positive: 5 @  Zika Virus PCR/antigen positive: 5 *  West Nile Virus PCR/antigen positive: 5 * 2.aAcute febrile cases negative for Dengue: 30
3.bHealthy subjects from endemic regions: 20 90% 138 140  Chikungunya PCR positive: 26 @ positive: 9 @  Zika Virus PCR/antigen positive: 9 *  West Nile Virus PCR/antigen positive: 9 * 2.aAcute febrile cases negative for Dengue: 52
3.bHealthy subjects from endemic regions: 35 85% 196 200  Chikungunya PCR positive: 37 @ positive: 13 @  Zika Virus PCR/antigen positive: 13 *  West Nile Virus PCR/antigen positive:13 * 2.aAcute febrile cases negative for Dengue: 74
3.bHealthy subjects from endemic regions: 50 80% 246 250  Chikungunya PCR positive: 46 @ positive: 16 @  Zika Virus PCR/antigen positive: 16 *  West Nile Virus PCR/antigen positive: 16 * 2.aAcute febrile cases negative for Dengue: 94
3.bHealthy subjects from endemic regions: 62 a Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR) b Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, nucleic acid)

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

Page 61 of 459

It is recommended to test for all pathogens listed in the cross-reactivity panel. However, if reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

*In the absence of natural samples, spiked samples may be used, as per details provided in the note below. Note:
Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. Recombinant NS1 antigen of cross reactive flaviviruses (Zika, West Nile and Japanese Encephalitis viruses) expressed in mammalian cells can be obtained commercially and reconstituted in serum samples (100 ng -1 µg/ml) and diluted in the ratio of 1:2 and used accordingly (at least five dilutions for each virus specific NS1). Before used for evaluation, flavivirus NS1 reconstituted in serum samples needs to be tested by the dengue NS1 reference assay, and dilutions which are negative for dengue should be used for evaluation. The serum samples used for reconstitution should be negative for Dengue NS1, RNA and IgM antibody. 6. Evaluation method:

Resolution of discrepant results: test. test.

under evaluation.

Page 62 of 459

be as follows:
 First lot of the kit: should be tested on statistically significant number of positive and negative samples as calculated in the protocol.
 Second lot of the kit: should be tested on 25 samples (15 positive samples comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples (15 positive samples comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

imprecision (if applicable) should be measured. 3 positive samples (strong, moderate

Page 63 of 459

and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.

c. Day-to-day variability: Testing should be performed on at least two non-consecutive days. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.

positive samples (strong, moderate and weak positive samples) and 3 negative

Page 64 of 459

  1. Acceptance Criteria: Sensitivity: ≥90% Specificity: ≥95% To achieve at least the performance characteristics outlined in the acceptance criteria, ≥140 positive samples and ≥80 negative samples should be used for evaluation. author(s). acceptable.

VI. References:

  1. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Vazquez S, for the detection of Dengue virus NS1 antigen and anti-Dengue virus IgM antibody. PLoS Negl Trop Dis. 2014 Oct 16;8(10):e3171. doi: 10.1371/journal.pntd.0003171.
  2. Hermann LL, Thaisomboonsuk B, Poolpanichupatam Y, Jarman RG, Kalayanarooj S, Nisalak A, Yoon IK, Fernandez S. Evaluation of a Dengue NS1 antigen detection assay sensitivity and specificity for the diagnosis of acute Dengue virus infection. PLoS Negl Trop Dis. 2014 Oct 2;8(10):e3193. doi: 10.1371/journal.pntd.0003193.
  3. Central Drugs Standard Control Organization. In-Vitro Diagnostic (IVD) Medical Devices Frequently Asked Questions.

Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf- documents/IVD/FAQs/CDSCO-IVD-FAQ-03-2022-.pdf
4. U.S. Food and Drug Administration. Dengue Virus Serological Reagents - Class II Special Controls Guideline for Industry and Food and Drug Administration Staff. 2014. Available at: https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation- emitting-products/Dengue-virus-serological-reagents-class-ii-special-controls-guideline- industry-and-food-and-drug

Page 65 of 459

  1. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
6. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

VII.

Page 66 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE NS1 ELISA KIT site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type

Results

Positive Negative Total Name of Dengue NS1
ELISA kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity: ● Conclusions: o Sensitivity, specificity

Page 67 of 459

(Sensitivity and specificity have been assessed in controlled lab setting using kits provided by the manufacturer from the batch mentioned above using ….. sample. Results should not be extrapolated to other sample types.) Disclaimers Note: This report is exclusively for …………. Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ………………………

Page 68 of 459

Field evaluation protocol for Dengue NS1 ELISA kits I. Background:
II. Purpose: To evaluate the performance characteristics of Dengue NS1 ELISA kits in the diagnosis of Dengue infection in individuals with unknown disease status. III. Requirements:
The study will be initiated after approval from the institutional human ethics committee.
V. Procedure:

  1. Study design/type: Cross-sectional study through
    Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC ⮚ Handling of Dengue NS1 ELISA kits received for performance evaluation

Sample size for performance evaluation: Sample sizes of positive and negative samples of Dengue against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming

Page 69 of 459

95% level of significance, and an absolute precision of 5%. It is further assumed that 30% of the individuals attending the health care facilities for acute febrile illness and suspected for Dengue will be positive for Dengue. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.
Sample size has to be calculated based on both the sensitivity and the specificity. The final sample size will be the maximum of the two. For example, at 95% sensitivity and 95% specificity, the sample size required will be 245 (maximum of 245 and 105).

𝑑2 x P

𝑑2 x P

· n (se) is the minimum number of number of individuals to be enrolled to obtain the requisite number of positive samples.
· n (sp) is the minimum number of number of individuals to be enrolled to obtain the requisite number of negative samples.

P is the prevalence of the disease · d is the predetermined marginal error (5%).

Table 1. Sample sizes for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of individuals to be enrolled* off] # 99% 51 55 95% 243 245 90% 461 465

Page 70 of 459

85% 653 655 80% 820 820

  • Individuals attending the health care facilities for

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

Table 2. Sample sizes for different values of specificity claimed by the manufacturer. Specificity Calculated sample size Minimum no. of individuals to be enrolled* off] # 99% 22 25 95% 104 105 90% 198 200 85% 280 280 80% 351 355

  • Individuals attending the health care facilities for

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity Recruitment of cases shall be halted once desired number of positive and negative samples are reached. 4. Inclusion criteria: Individuals with Dengue like illness (A patient with acute febrile illness of 2-7 days with two or more manifestations: Head ache, retro-orbital pain, myalgia, arthralgia, rash, hemorrhagic manifestations) 5. Exclusion criteria: Individuals with already known positive history for other pathogens 6. Reference assay:
WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Dengue NS1 ELISA kit should be used as reference assay. Serotype status to be assessed using CDC / NIV real-time PCR serotyping protocols.

Page 71 of 459

Study implementation: The individuals with Dengue like illness will be recruited into the study and five ml of whole blood will be collected in vacutainer tubes and the serum will be separated by centrifugation and used for the study. The serum sample will be subjected to the following reference tests and the index test. 8. Evaluation method:

  1. Positive samples: Samples positive by the reference NS1 ELISA assay and real-time PCR assay (True positives). will be considered as true positive sample.
  2. Negative samples:
    Samples negative by the reference NS1 ELISA assay and real-time PCR using CDC/NIV serotyping protocol will be considered as true negative. N.B.:
    test. test. A. Cross reactivity: Clinical samples or commercially available NS1 antigens from other flaviviruses will be used to test cross reactivity of the index test.
  3. Chikungunya PCR positive samples: 5 @
  4. Japanese Encephalitis PCR/antigen positive: 5 samples @
  5. Zika Virus PCR/antigen: 5 samples *
  6. West Nile Virus PCR/antigen: 5 samples *

reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the

Page 72 of 459

In the absence of natural samples, spiked samples may be used, as per details provided in the note below. Note:
Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. Recombinant NS1 antigen of cross reactive flaviviruses (Zika, West Nile and Japanese Encephalitis viruses) expressed in mammalian cells can be obtained commercially and reconstituted in serum samples (100 ng -1 µg/ml) and diluted in the ratio of 1:2 and used accordingly (at least five dilutions for each virus specific NS1). Before used for evaluation, NS1 reconstituted in serum samples needs to be tested by the reference assay and dilution which are positive only should be used for evaluation. The serum samples used for reconstitution should be negative for Dengue NS1, RNA and IgM antibody. 12. Statistical analysis:
Sensitivity and specificity will be calculated. Interim analysis of data shall be conducted on completing evaluation of 25%, 50% and 75% of samples. If, at any point, the performance of the assay is found to be not satisfactory, the assay shall not be evaluated further. Evaluation fee shall be charged accordingly.

under evaluation.

be as follows:
comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

Page 73 of 459

10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

moderate and weak positive samples) and 3 negative samples should be tested 5 (qualitative).

(qualitative).

Page 74 of 459

(qualitative).

  1. Resolution of discrepant results:

test. test.

  1. Blinding of laboratory staff:

be blinded to the results of the reference assay. The PI of the evaluation exercise should remain unblinded, i.e., privy to the results of the reference test. Another senior laboratory staff selected by the PI may remain unblinded for overseeing the activity and maintaining the database of results. The data should be analyzed only by the PI of the evaluating lab, and the PI should maintain confidentiality of data.

  1. Acceptance Criteria: Sensitivity: ≥90% Specificity: ≥95% To achieve at least the performance characteristics outlined in the acceptance criteria, ≥465 individuals satisfying the case definition need to be enrolled to obtain the requisite number of positive samples. This sample size is sufficient for required number of negative samples. Recruitment should be terminated once the desired number of positive cases is enrolled and tested.
  2. Publication Rights: author(s). acceptable.

Page 75 of 459

VI. References:

  1. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Vazquez S, for the detection of Dengue virus NS1 antigen and anti-Dengue virus IgM antibody. PLoSNegl Trop Dis. 2014 Oct 16;8(10):e3171. doi: 10.1371/journal.pntd.0003171.
  2. Hermann LL, Thaisomboonsuk B, Poolpanichupatam Y, Jarman RG, Kalayanarooj S, Nisalak A, Yoon IK, Fernandez S. Evaluation of a Dengue NS1 antigen detection assay sensitivity and specificity for the diagnosis of acute Dengue virus infection. PLoSNegl Trop Dis. 2014 Oct 2;8(10):e3193. doi: 10.1371/journal.pntd.0003193.
  3. Ganeshkumar P, Murhekar MV, Poornima V, Saravanakumar V, Sukumaran K, Anandaselvasankar A, John D, Mehendale SM. Dengue infection in India: A systematic review and meta-analysis. PLoSNegl Trop Dis. 2018 Jul 16;12(7):e0006618. doi: 10.1371/journal.pntd.0006618.
  4. Castro-Jorge LA, Machado PR, Fávero CA, Borges MC, Passos LM, de Oliveira RM, Fonseca BA. Clinical evaluation of the NS1 antigen-capture ELISA for early diagnosis of Dengue virus infection in Brazil. J Med Virol. 2010 Aug;82(8):1400-5. doi: 10.1002/jmv.21814.
  5. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
6. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

VII.

Page 76 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE NS1 ELISA KIT site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type Positive samples: Not applicable, may categorize cases as per duration of illness Negative samples (may categorize as per duration of illness, must include cross reactivity panel)

Results

Positive Negative Total Name of Dengue NS1 ELISA kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity: ● Conclusions: o Sensitivity, specificity o

Page 77 of 459

(Sensitivity and specificity have been assessed in using kits provided by the manufacturer from the batch mentioned above using ….. sample in controlled lab setting. Results should not be extrapolated to other sample types.)

Disclaimers Note: This report is exclusively for NS1…………………Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In charge …………………… Seal ……………………………..

Page 78 of 459

Performance evaluation protocol for Dengue IgM RDT kits I. Background:
II. Purpose: To evaluate the performance characteristics of Dengue IgM RDT kits in the diagnosis of Dengue infection using irreversibly de-identified leftover archived/ spiked clinical samples.. III. Requirements:
Performance evaluation activities using irreversibly de-identified leftover clinical are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024.
V. Procedure: through
Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC ⮚ Handling of Dengue IgM Rapid IVD kits received for performance evaluation

Page 79 of 459

Preparation of Dengue IgM Rapid IVD kit evaluation panel: Well characterised Dengue IVD kit evaluation panel is a critical requirement for performance evaluation of IVD kits. Hence statistically significant number of sera samples should be available from Dengue confirmed cases. Further characterised for Dengue IgM positivity by using approved reference kits having high sensitivity and specificity. Dengue IgM performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples. 4. Reference assay:
WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Dengue IgM ELISA kit should be used as reference assay. NS1 antigen status to be assessed using WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved NS1 ELISA kit. Serotype status to be assessed using a combination of CDC/NIV real-time PCR serotyping protocols.
At least 50% of the samples should be positive by real-time PCR or NS1 antigen and IgM ELISA. Primary and Secondary status to be assessed by Panbio Dengue IgG capture ELISA kit.

samples of Dengue against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For require usage of the sample size outlined for 85% specificity.

𝑑2 x (1 −IR)

Page 80 of 459

𝑑2 x (1 −IR)

Positive samples: The panel of positive samples should include samples positive by the reference assay, with 50% samples positive for Dengue NS1/RT-PCR assay (True positives). Samples should be representative of all 4 serotypes and varying degrees of positivity. The samples should be classified as strong, moderate and weak positives based on ELISA units of the reference assay.

Negative samples: These should include samples negative by the reference assay, NS1 ELISA assay and/or real-time PCR using CDC/NIV serotyping protocol (True negatives).

Table 1. Sample sizes and panel composition of positive Dengue samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off] # 99% 16 20 Strong Positive: 6 Weak Positive: 7 95% 77 80 Strong Positive: 24 Weak Positive: 28 90% 145 150 Strong Positive: 44 Moderate Positive: 53 Weak Positive: 53 85% 206 210 Strong Positive: 62 Moderate Positive: 74 Weak Positive: 74 80% 258 260 Strong Positive: 78

Page 81 of 459

Moderate Positive: 91 Weak Positive: 91

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off]# 99%# 16 20 1.Cross-reactivity panel  Chikungunya IgM positive: 4 @  Japanese Encephalitis IgM positive: 1 @
 Zika Virus IgM positive: 1 *  West Nile Virus IgM positive: 1 * 2.aAcute febrile cases: 5
3.Rheumatoid Arthritis/other autoimmune disease cases: 4 4.bHealthy subjects from endemic regions: 4
95% 77 80  Chikungunya IgM positive: 16 @  Japanese Encephalitis IgM positive: 3 @  Zika Virus IgM positive: 3 *  West Nile Virus IgM positive: 3 * 2.aAcute febrile cases: 23
3.Rheumatoid Arthritis/other autoimmune disease cases: 16 4.bHealthy subjects from endemic regions: 16
90% 145 150  Chikungunya IgM positive: 30 @  Japanese Encephalitis IgM positive: 5 @  Zika Virus IgM positive: 5 *  West Nile Virus IgM positive: 5 * 2.aAcute febrile cases: 45
3.Rheumatoid Arthritis/other autoimmune disease cases: 30
4.bHealthy subjects from endemic regions: 30

Page 82 of 459

85% 206 210  Chikungunya IgM positive: 42 @  Japanese Encephalitis IgM positive: 7 @  Zika Virus IgM positive: 7 *  West Nile Virus IgM positive: 7 * 2.aAcute febrile cases: 63
3.Rheumatoid Arthritis/other autoimmune disease cases: 42 4.bHealthy subjects from endemic regions: 42

80% 258 260  Chikungunya IgM positive: 52 @  Japanese Encephalitis IgM positive: 9 @  Zika Virus IgM positive: 9 *  West Nile Virus IgM positive: 9 * 2.aAcute febrile cases: 77
3.Rheumatoid Arthritis/other autoimmune disease cases: 52 4.bHealthy subjects from endemic regions: 52

a Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR) b Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, RNA) #It is recommended to calculate the sample size as per manufacturer's claims of

It is recommended to test for all pathogens listed in the cross-reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

Note: Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. If IgM positive samples for cross reactive flaviviruses are not available, commercially available IgM sera panel for different viruses can be procured and used to test cross reactivity.

Page 83 of 459

Evaluation method:

Resolution of discrepant results:

test. test.

This should be done to assess the repeatability of the detection of target analyte(s) using the kit under evaluation.

be as follows:
comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).
10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

Page 84 of 459

two different operators, keeping all other testing parameters undisturbed. Within- run and between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples (qualitative).

(qualitative).

(qualitative).

Page 85 of 459

  1. Acceptance Criteria: Sensitivity: ≥80% Specificity: ≥90% To achieve at least the performance characteristics outlined in the acceptance criteria, ≥260 positive samples and ≥150 negative samples should be used for evaluation.

Page 86 of 459

author(s). acceptable.
VI. References:

  1. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Pelegrino JL, Vázquez S, Artsob H, Drebot M, Gubler DJ, Halstead SB, Guzmán MG, Margolis HS, Nathanson CM, Rizzo Lic NR, Bessoff KE, Kliks S, Peeling RW. Evaluation of commercially available anti-Dengue virus immunoglobulin M tests. Emerg Infect Dis. 2009 Mar;15(3):436-
  2. doi: 10.3201/eid1503.080923.
  3. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Vazquez S, for the detection of Dengue virus NS1 antigen and anti-Dengue virus IgM antibody. PLoS Negl Trop Dis. 2014 Oct 16;8(10):e3171. doi: 10.1371/journal.pntd.0003171.
  4. WHO, Evaluation of commercially available anti-Dengue virus immunoglobulin M tests. (Diagnostics evaluation series, 3). ISBN 978 92 4 159775 3.
  5. Central Drugs Standard Control Organization. Guidance on Performance Evaluation of In- vitro Diagnostic Medical Devices.

Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/medical device/guidanceperformanceivd.pdf 5. Central Drugs Standard Control Organization. In-Vitro Diagnostic (IVD) Medical Devices Frequently Asked Questions. 2022. Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf- documents/IVD/FAQs/CDSCO-IVD-FAQ-03-2022-.pdf
6. U.S. Food and Drug Administration. Dengue Virus Serological Reagents - Class II Special Controls Guideline for Industry and Food and Drug Administration Staff. 2014. Available at: https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation- emitting-products/Dengue-virus-serological-reagents-class-ii-special-controls-guideline- industry-and-food-and-drug
7. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3. 2017. Available at: eng.pdf;sequence=1

Page 87 of 459

  1. Yow KS, Aik J, Tan EY, Ng LC, Lai YL. Rapid diagnostic tests for the detection of recent Dengue infections: An evaluation of six kits on clinical specimens. PLoS One. 2021 Apr 1;16(4):e0249602. doi: 10.1371/journal.pone.0249602.
  2. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

*The validation protocols need to be revisited after introduction of Dengue vaccines and the acceptance criteria needs revisiting every year so as to enable the availability of best diagnostic kits.

VII. Performance evaluation report format

Page 88 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE IgM RDT KIT site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type

Results:

Positive Negative Total Name of Dengue IgM RDT kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity:

Page 89 of 459

Disclaimers Note: This report is exclusively for ………………………Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ……………………….

Page 90 of 459

Performance evaluation protocol for Dengue IgM ELISA kits I. Background:
II. Purpose: To evaluate the performance characteristics of Dengue IgM ELISA kits in the diagnosis of Dengue infection using irreversibly de-identified leftover archived/ spiked clinical samples. III. Requirements:
V. Procedure: through
Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC ⮚ Handling of Dengue IgM ELISA IVD kits received for performance evaluation

Page 91 of 459

Preparation of Dengue IgM ELISA IVD kit evaluation panel: Well characterised Dengue IVD kit evaluation panel is a critical requirement for performance evaluation of IVD kits. Hence statistically significant number of sera samples should be available from Dengue confirmed cases. Further characterised for Dengue IgM positivity by using approved reference kits having high sensitivity and specificity. Dengue IgM performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples. 4. Reference assay:
WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Dengue IgM ELISA kit should be used as reference assay. NS1 antigen status to be assessed using WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved NS1 ELISA kit. Serotype status to be assessed using a combination of CDC/NIV real-time PCR serotyping protocols.
At least 50% of the samples should be positive by real-time PCR or NS1 antigen and IgM ELISA. Primary and Secondary status to be assessed by Panbio Dengue IgG capture ELISA kit. 5. samples and sample panel composition against different values of sensitivity and 95% level of significance, and an absolute precision of 5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For require usage of the sample size outlined for 85% specificity.

𝑑2

𝑑2

Page 92 of 459

Positive samples: The panel of positive samples should include samples positive by the reference assay, with 50% samples positive for Dengue NS1/ RT-PCR assay (True positives). Samples should be representative of primary/secondary Dengue and all 4 Dengue virus serotypes, with varying degrees of positivity. The samples should be classified as strong, moderate and weak positives based on ELISA units of the reference assay.

Negative samples: These should include samples negative by the reference assay, NS1 ELISA and/or real-time PCR using CDC and/or NIV serotyping protocols. (True negatives). Table 1. Sample sizes and panel composition of positive Dengue samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off]# 99% 15 20 Strong Positive: 6 Weak Positive: 7 95% 73 80 Strong Positive: 24 Weak Positive: 28 90% 138 140 Strong Positive: 42 Moderate Positive: 49 Weak Positive: 49 85% 196 200 Strong Positive: 60 Moderate Positive: 70 Weak Positive: 70 80% 246 250 Strong Positive: 75 Moderate Positive: 87 Weak Positive: 88

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

Page 93 of 459

Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off]# 99%# 15 20  Chikungunya IgM positive: 3 @
 Japanese Encephalitis IgM positive: 1 @  West Nile Virus IgM positive: 1 *  Zika Virus IgM positive: 1 * 2.Rheumatoid Arthritis/other autoimmune disease cases: 4
3.aAcute febrile cases: 6
4.bHealthy subjects from endemic regions: 4
95% 73 80  Chikungunya IgM positive: 10 @  Japanese Encephalitis IgM positive: 5 @  West Nile Virus IgM positive: 5 *  Zika Virus IgM positive: 5 * disease cases: 15
3.aAcute febrile cases: 25
4.bHealthy subjects from endemic regions: 15
90% 138 140  Chikungunya IgM positive: 18 @
 Japanese Encephalitis IgM positive: 9 @  West Nile Virus IgM positive: 9 *  Zika Virus IgM positive: 9 * disease cases: 26
3.aAcute febrile cases: 43
4.bHealthy subjects from endemic regions: 26
85% 196 200  Chikungunya IgM positive: 25 @
 Japanese Encephalitis IgM positive: 12 @  West Nile Virus IgM positive: 12 *  Zika Virus IgM positive: 12 * disease cases: 38
3.aAcute febrile cases: 63
4.bHealthy subjects from endemic regions: 38

Page 94 of 459

80% 246 250  Chikungunya IgM positive: 31@
 Japanese Encephalitis IgM positive: 16 @  West Nile Virus IgM positive: 16 *  Zika Virus IgM positive: 16 * disease cases: 47
3.aAcute febrile cases: 77
4.bHealthy subjects from endemic regions: 47
a Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR) b Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, RNA)

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

Note: Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. If IgM positive samples for cross reactive flaviviruses are not available, commercially available IgM sera panel for different viruses can be procured and used to test cross reactivity.

Evaluation method:

Resolution of discrepant results:

test. test.

Page 95 of 459

under evaluation.

be as follows:
comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).
10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

Page 96 of 459

moderate and weak positive samples) and 3 negative samples should be tested 5 (qualitative).

(qualitative).

(qualitative).

Page 97 of 459

  1. Acceptance criteria Sensitivity: ≥90% Specificity: ≥95%
    To achieve at least the performance characteristics outlined in the acceptance criteria, ≥140 positive samples and ≥80 negative samples should be used for evaluation. author(s). acceptable.
    VI. References:

  2. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Pelegrino JL, Vázquez S, Artsob H, Drebot M, Gubler DJ, Halstead SB, Guzmán MG, Margolis HS, Nathanson CM, Rizzo Lic NR, Bessoff KE, Kliks S, Peeling RW. Evaluation of commercially available anti-Dengue virus immunoglobulin M tests. Emerg Infect Dis. 2009 Mar;15(3):436-40. doi: 10.3201/eid1503.080923.

  3. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Vazquez S, for the detection of Dengue virus NS1 antigen and anti-Dengue virus IgM antibody. PLoSNegl Trop Dis. 2014 Oct 16;8(10):e3171. doi: 10.1371/journal.pntd.0003171.

  4. WHO, Evaluation of commercially available anti-Dengue virus immunoglobulin M tests. (Diagnostics evaluation series, 3). ISBN 978 92 4 159775 3.

  5. Central Drugs Standard Control Organization. Guidance on Performance Evaluation of In- vitro Diagnostic Medical Devices.

Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/medical device/guidanceperformanceivd.pdf 5. Central Drugs Standard Control Organization. In-Vitro Diagnostic (IVD) Medical Devices Frequently Asked Questions. 2022. Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf- documents/IVD/FAQs/CDSCO-IVD-FAQ-03-2022-.pdf

Page 98 of 459

  1. U.S. Food and Drug Administration. Dengue Virus Serological Reagents - Class II Special Controls Guideline for Industry and Food and Drug Administration Staff. 2014. Available at: https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation- emitting-products/Dengue-virus-serological-reagents-class-ii-special-controls-guideline- industry-and-food-and-drug
  2. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
8. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

*The validation protocols need to be revisited after introduction of Dengue vaccines and the acceptance criteria needs revisiting every year so as to enable the availability of best diagnostic kits.

Page 99 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE IgM ELISA KIT site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type

Results:

Positive Negative Total Name of Dengue IgM ELISA kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity: o

(Sensitivity and specificity have been assessed in controlled lab setting using kits provided by the manufacturer from the batch mentioned above using ….. sample. Results should not be extrapolated to other sample types.)

Page 100 of 459

Disclaimers Note: This report is exclusively for ………………………Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal………………………………………

Page 101 of 459

Performance evaluation protocol for Dengue NS1/IgM combo RDT kits I. Background:
II. Purpose: To evaluate the performance characteristics of Dengue NS1/IgM combo RDT kits in the diagnosis of Dengue infection. III. Requirements:
V. Procedure: through
Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC ⮚ Handling of Dengue NS1/IgM combo IVD kits received for performance evaluation (Verification/Storage/Unpacking etc).

Page 102 of 459

Preparation of Dengue RDT IVD kit evaluation panel: Well characterised Dengue RDT IVD kit evaluation panel is a critical requirement for should be available from Dengue confirmed cases. Further characterised for Dengue NS1 and IgM positivity by using approved reference kits having high sensitivity and specificity. Dengue NS1/IgM performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples. 4. Reference assay:
Anti-DENV IgM detection ELISA WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved kit AND/OR DENV NS1 ELISA WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved kit Serotype status to be assessed using a combination of CDC and/or NIV real-time PCR serotyping protocols.
All positive samples need confirmation reference NS1/IgM ELISA assay and real-time PCR assay. 5. samples of Dengue against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For require usage of the sample size outlined for 85% specificity.

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

Page 103 of 459

Positive samples: Samples which are positive for IgM or NS1 or both by the reference assays will be considered as true positive samples. There should be representation of samples positive for all four serotypes.

Negative samples: These should include samples negative by all the reference assays and real- time PCR using CDC and/or NIV serotyping protocol (True negatives). Table 1. Sample sizes and panel composition of positive Dengue samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off for balanced allocation] # 99%# 16 28 *NS1 positive and IgM negative: 8 *NS1 and IgM positive: 12 *NS1 negative and IgM positive: 8 95% 77 84 *NS1 positive and IgM negative: 24 *NS1 and IgM positive: 36 *NS1 negative and IgM positive: 24 90% 145 160 *NS1 positive and IgM negative: 44 *NS1 and IgM positive: 72 *NS1 negative and IgM positive: 44 85% 206 220 *NS1 positive and IgM negative: 60 *NS1 and IgM positive: 100 *NS1 negative and IgM positive: 60 80% 258 260 *NS1 positive and IgM negative: 72 *NS1 and IgM positive: 116 *NS1 negative and IgM positive: 72

  • all 4 serotypes shall be represented

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

Page 104 of 459

Note:
In the absence of natural samples, spiked samples may be used as per details provided below:

Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. Recombinant NS1 antigen of cross reactive flaviviruses (Zika, West Nile and Japanese Encephalitis viruses) expressed in mammalian cells can be obtained commercially and reconstituted in serum samples (100 ng -1 µg/ml) and diluted in the ratio of 1:2 and used accordingly (at least five dilutions for each virus specific NS1). Before used for evaluation, flavivirus NS1 reconstituted in serum samples needs to be tested by the dengue NS1 reference assay, and dilutions which are negative for dengue should be used for evaluation. The serum samples used for reconstitution should be negative for Dengue NS1, RNA and IgM antibody.

Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off for balanced allocation]# 99%# 16 30 For antibody:  Chikungunya IgM positive: 1 @  Japanese Encephalitis IgM positive: 1 @  West Nile Virus IgM positive: 1 *  Zika Virus IgM positive: 1 *

For antigen:  Chikungunya PCR positive: 1 @  Japanese Encephalitis NS1/PCR positive: 1 @  West Nile Virus NS1/PCR positive: 1 *  Zika Virus NS1/PCR positive: 1 *

disease cases: 5 3.aAcute febrile cases: 12
4.bHealthy subjects from endemic regions: 5
95% 77 90 For antibody:  Chikungunya IgM positive: 3 @

Page 105 of 459

 Japanese Encephalitis IgM positive: 3 @  West Nile Virus IgM positive: 3 *  Zika Virus IgM positive: 3 *

For antigen:  Chikungunya PCR positive: 3 @  Japanese Encephalitis NS1/PCR positive: 3 @  West Nile Virus NS1/PCR positive: 3 *  Zika Virus NS1/PCR positive: 3 *

disease cases: 15 3.aAcute febrile cases: 36
4.bHealthy subjects from endemic regions: 15
90% 145 160 For antibody:  Chikungunya IgM positive: 5 @  Japanese Encephalitis IgM positive: 5 @  West Nile Virus IgM positive: 5 *  Zika Virus IgM positive: 5 *

For antigen:  Chikungunya PCR positive: 5 @  Japanese Encephalitis NS1/PCR positive: 5 @  West Nile Virus NS1/PCR positive: 5 *  Zika Virus NS1/PCR positive: 5 *

disease cases: 30
3.aAcute febrile cases: 60
4.bHealthy subjects from endemic regions: 30
85% 206 220 For antibody:  Chikungunya IgM positive: 7 @  Japanese Encephalitis IgM positive: 7 @  West Nile Virus IgM positive: 7 *  Zika Virus IgM positive: 7 *

For antigen:  Chikungunya PCR positive: 7 @  Japanese Encephalitis NS1/PCR positive: 7 @  West Nile Virus NS1/PCR positive: 7 *

Page 106 of 459

 Zika Virus NS1/PCR positive: 7 *

disease cases: 34 3.aAcute febrile cases: 90
4.bHealthy subjects from endemic regions: 40
80% 258 260 For antibody:  Chikungunya IgM positive: 8 @  Japanese Encephalitis IgM positive: 8 @  West Nile Virus IgM positive: 8 *  Zika Virus IgM positive: 8 *

For antigen:  Chikungunya PCR positive: 8 @  Japanese Encephalitis NS1/PCR positive: 8 @  West Nile Virus NS1/PCR positive: 8 *  Zika Virus NS1/PCR positive: 8 *

disease cases: 45 3.aAcute febrile cases: 106
4.bHealthy subjects from endemic regions: 45
a Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR) b Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, RNA)

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

Note: Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available.
If IgM positive samples for cross reactive flaviviruses are not available, commercially available IgM sera panel for different viruses can be procured and used to test cross reactivity.

Before used for evaluation, the NS1 reconstituted in serum samples needs to be tested by the reference assay and dilution which are positive only should be used for evaluation.

Page 107 of 459

The serum sample used for spiking or reconstitution should be negative for Dengue NS1, RNA and IgM antibody.

Evaluation method:

  1. Repeatability and Reproducibility Assessment:

This should be done to assess the repeatability of the detection of target analyte(s) using the kit under evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples for each analyte should be tested 5 times in independent runs.

be as follows:
comprising 10 low positive AND 5 moderate/ strong positive samples per target analyte, and 10 negative samples).
10 low positive AND 5 moderate/ strong positive samples per target analyte, and 10 negative samples).

Page 108 of 459

two different operators, keeping all other testing parameters undisturbed. Within- run and between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples for each analyte should be tested 5 times in independent runs.
(qualitative).

and 3 negative samples for each analyte should be tested 5 times in independent runs.
(qualitative).

samples for each analyte should be tested 5 times in independent runs.
(qualitative).

Page 109 of 459

  1. Interpretation of results: Since the kits have been provided in combo format, concordance has to be calculated separately for NS1 and IgM, and the overall sensitivity and specificity have to be calculated based on the combined results of NS1 and IgM. If the sample is positive for any one or both analytes (NS1 or IgM or both), then the sample is considered positive. Refer the table below for interpretation: NS1 Reference test result IgM reference test result Final Reference test result NS1 Index test result IgM Index test result Final index test result Interpretation for dengue positivity

Positive +

Positive True Positive + + Positive

Positive True Positive + + Positive

Negative False Negative + + Positive + + Positive True Positive +

Positive +

Positive True Positive +

Positive

Positive True Positive +

Positive

Negative False Negative

Positive +

Positive True Positive

Positive

Positive True Positive

Positive

Negative False Negative

Negative

Positive False Positive

Negative +

Positive False Positive Consider one-on-one results for analyte-specific analysis. 10. Resolution of discrepant results:

test. test.

  1. Blinding of laboratory staff:

Page 110 of 459

  1. Acceptance criteria: A minimum concordance of 80% for NS1 and 80% for IgM should be achieved with the reference assay, and an overall combined sensitivity* and specificity$ of ≥90% each.
  • Samples which are positive for NS1 or IgM or both by the kit under evaluation (irrespective of the reference assay results) will be considered as positive for dengue and used for overall sensitivity calculation. To achieve at least the overall combined performance characteristics outlined in the acceptance criteria, ≥160 positive samples and ≥160 negative samples should be used for evaluation. $ Sample which are negative for both NS1 and IgM by kit under evaluation (irrespective of the reference assay results) will be considered negative for dengue and used for overall specificity calculation
  1. Publication Rights: author(s). acceptable.

Page 111 of 459

VI. References:

  1. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Vazquez S, for the detection of Dengue virus NS1 antigen and anti-Dengue virus IgM antibody. PLoSNegl Trop Dis. 2014 Oct 16;8(10):e3171. doi: 10.1371/journal.pntd.0003171.
  2. Hermann LL, Thaisomboonsuk B, Poolpanichupatam Y, Jarman RG, Kalayanarooj S, Nisalak A, Yoon IK, Fernandez S. Evaluation of a Dengue NS1 antigen detection assay sensitivity and specificity for the diagnosis of acute Dengue virus infection. PLoSNegl Trop Dis. 2014 Oct 2;8(10):e3193. doi: 10.1371/journal.pntd.0003193.
  3. Yow KS, Aik J, Tan EY, Ng LC, Lai YL. Rapid diagnostic tests for the detection of recent Dengue infections: An evaluation of six kits on clinical specimens. PLoS One. 2021 Apr 1;16(4):e0249602. doi: 10.1371/journal.pone.0249602.
  4. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
5. WHO, Evaluation of commercially available anti-Dengue virus immunoglobulin M tests. (Diagnostics evaluation series, 3). ISBN 978 92 4 159775 3.
6. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability and diagnostic performance: a go-to review. European radiology experimental, 8(1), 79. https://doi.org/10.1186/s41747-024-00474-w

Page 112 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE NS1 and IgM COMBO RDT KIT

site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to:

Application No. Sample Panel Sample type

Results:

Positive Negative Total Name of Dengue NS1 and IgM combo RDT kit Positive

Negative

Total

Prepare the above 2x2 table individually for each analyte, as well as for overall performance characteristics

Page 113 of 459

Estimate (%) 95% CI Combined Sensitivity

Combined Specificity

Dengue NS1 standalone sensitivity

Dengue NS1 standalone specificity

Dengue IgM standalone sensitivity

Dengue IgM standalone specificity

o Cross-reactivity: ● Conclusions: o Disclaimers

Note: This report is exclusively for ………….. Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ………………………………

Page 114 of 459

Field evaluation protocol for Dengue NS1 and IgM combo RDT kits

I. Background:
II. Purpose: To evaluate the performance characteristics of Dengue NS1/IgM RDT combo kits in the diagnosis of Dengue infection in individuals with unknown disease status. III. Requirements:

The study will be initiated after approval from the institutional human ethics committee.
V. Procedure:

  1. Study design/type: Cross-sectional study through
    Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC ⮚ Handling of Dengue NS1 RDT/IgM RDT IVD kits received for performance evaluation (Verification/Storage/Unpacking etc).

Page 115 of 459

  1. Sample size for performance evaluation: Sample sizes of positive and negative samples of Dengue against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. It is further assumed that 30% of the individuals attending the health care facilities for acute febrile illness and suspected for Dengue will be positive for Dengue. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.

𝑑2 x P x (1 −IR) 𝑑2 x P x (1 −IR)

· n (se) is the minimum number of number of individuals to be enrolled to obtain the requisite number of positive samples. · n (sp) is the minimum number of number of individuals to be enrolled to obtain the requisite number of negative samples.

P is the prevalence of the disease · d is the predetermined marginal error (5%).

Sample size has to be calculated based on both the sensitivity and the specificity. The final sample size will be the maximum of the two. For example, at 95% sensitivity and 95% specificity, the sample size required will be 260 (maximum of 260 and 110). It is desirable to cover at least one Dengue season so that adequate samples are available for evaluation.

Page 116 of 459

Table 1. Sample sizes for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size
Minimum no. of individuals to be enrolled* off] # 99% 53 60 95% 255 260 90% 484 490 85% 686 690 80% 861 870

  • Individuals attending the health care facilities for

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity Table 2. Sample sizes for different values of specificity claimed by the manufacturer. Specificity Calculated sample size Minimum no. of individuals to be enrolled* off] # 99%# 23 30 95% 109 110 90% 207 210 85% 294 300 80% 369 370

  • Individuals attending the health care facilities for

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity Recruitment of cases shall be halted once desired number of positive and negative samples are reached. 4. Inclusion criteria: Patient with Dengue like illness (A patient with acute febrile illness of 1-14 days with two or more manifestations: Head ache, retro-orbital pain, myalgia, arthralgia, rash, hemorrhagic manifestations etc. The 1-14 days disease duration shall cover viraemic as well as convalescent phase of Dengue infection, so that both Dengue NS1 and IgM positive cases are enrolled.)

Page 117 of 459

  1. Exclusion criteria: Individuals with already known positive history for other pathogens

  2. Reference assay:
    Anti-DENV IgM detection ELISA WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved kit AND/OR DENV NS1 ELISA WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved kit Serotype status to be assessed using a combination of CDC and/or NIV real-time PCR serotyping protocols.

  3. Study implementation: The individuals with Dengue like illness will be recruited into the study and five ml of whole blood will be collected in vacutainer tubes and the serum will be separated by centrifugation and used for the study.
    It needs to be ensured that the samples are tested by reference tests and index test simultaneously.

  4. Evaluation method:

  5. Interpretation of results:

  6. Positive samples: Samples which are positive for IgM or NS1 or both by the reference assays will be considered as true positive samples.

  7. Negative samples:
    Samples which are negative by the reference assay will be considered as negative. N.B.:
    test. test.

Page 118 of 459

A. Cross reactivity (other flavivirus infections): A.1 NS1: Clinical samples or commercially available NS1 antigens from other flaviviruses will be used to test cross reactivity of the NS1 component of index test. i. Chikungunya PCR positive: 5 samples @ ii. Japanese Encephalitis PCR/antigen positive: 5 samples @ iii. West Nile Virus PCR/antigen: 5 samples * iv. Zika Virus PCR/antigen: 5 samples * reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the

In the absence of natural samples, spiked samples may be used, as per details provided in the note below. Note:
Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. Recombinant NS1 antigen of cross reactive flaviviruses (Zika, West Nile and Japanese Encephalitis viruses) expressed in mammalian cells can be obtained commercially and reconstituted in serum samples (100 ng -1 µg/ml) and diluted in the ratio of 1:2 and used accordingly (at least five dilutions for each virus specific NS1). Before used for evaluation, NS1 reconstituted in serum samples needs to be tested by the reference assay and dilution which are positive only should be used for evaluation. The serum samples used for reconstitution should be negative for Dengue NS1, RNA and IgM antibody.

A.2 IgM: Clinical samples positive for IgM for other flaviviruses will be used to test cross reactivity of the IgM component of index test. i. Chikungunya IgM positive: 5 samples @ ii. Japanese Encephalitis IgM positive: 5 samples @ iii. West Nile Virus IgM positive: 5 samples * iv. Zika Virus IgM positive: 5 samples *

Page 119 of 459

reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the Note: Depending on the availability of IgM positive samples for cross reactive flaviviruses, the requirement of samples for each virus may be increased or decreased accordingly to reach the total number of samples. If IgM positive samples for cross reactive flaviviruses are not available, commercially available IgM sera panel for different viruses can be procured and used to test cross reactivity. 12. Statistical analysis:
Concordance will be calculated separately for Dengue NS1 and IgM. Combined sensitivity and specificity will also be calculated. Interim analysis of data shall be conducted on completing evaluation of 25%, 50% and 75% of samples. If, at any point, the performance of the assay is found to be not satisfactory, the assay shall not be evaluated further. Evaluation fee shall be charged accordingly.

This should be done to assess the repeatability of the detection of target analyte(s) using the kit under evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples for each analyte should be tested 5 times in independent runs.

be as follows:
comprising 10 low positive AND 5 moderate/ strong positive samples per target analyte, and 10 negative samples).
10 low positive AND 5 moderate/ strong positive samples per target analyte, and 10 negative samples).

Page 120 of 459

two different operators, keeping all other testing parameters undisturbed. Within- run and between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples for each analyte should be tested 5 times in independent runs.
(qualitative).

and 3 negative samples for each analyte should be tested 5 times in independent runs.
(qualitative).

samples for each analyte should be tested 5 times in independent runs.
(qualitative).

Page 121 of 459

  1. Interpretation of results:
    Since the kits have been provided in a combo format, the sensitivity and specificity has to be calculated based on the combined results of the NS1 and IgM. If the sample is positive for any one or both analytes (NS1 or IgM or both), then the sample is considered positive. Refer the table below: NS1 Reference test result IgM reference test result Final Reference test result NS1 Index test result IgM Index test result Final index test result Interpretation for dengue positivity

Positive +

Positive True Positive + + Positive

Positive True Positive + + Positive

Negative False Negative + + Positive + + Positive True Positive +

Positive +

Positive True Positive +

Positive

Positive True Positive +

Positive

Negative False Negative

Positive +

Positive True Positive

Positive

Positive True Positive

Positive

Negative False Negative

Negative

Positive False Positive

Negative +

Positive False Positive Consider one-on-one results for analyte-specific analysis. 15. Resolution of discrepant results:

test. test.

  1. Blinding of laboratory staff

be blinded to the results of the reference assay. The PI of the evaluation exercise should remain unblinded, i.e., privy to the results of the reference test. Another senior laboratory staff selected by the PI may remain unblinded for overseeing the activity and maintaining the database of results, and the PI should maintain confidentiality of data. The data should be analyzed only by the PI of the evaluating lab.
17. Acceptance criteria: A minimum concordance of 80% for NS1 and 80% for IgM should be achieved with the reference assay, and an overall combined sensitivity* and specificity$ of ≥90% each.

Page 122 of 459

  • Samples which are positive for NS1 or IgM or both by the kit under evaluation (index test) irrespective of the reference assay results will be considered positive for dengue and used for overall sensitivity calculation $ Samples which are negative for both NS1 and IgM by kit under evaluation will be considered as negative for dengue and used for overall specificity calculation To achieve at least the overall performance characteristics outlined in the acceptance criteria (≥90% overall sensitivity and ≥90% overall specificity), ≥490 individuals satisfying the case definition need to be enrolled to obtain the requisite number of positive samples. This sample size is sufficient for required number of negative samples. Recruitment should be terminated once the desired number of positive cases is enrolled and tested.
  1. Publication Rights: author(s). acceptable.

VI. References:

  1. Hunsperger EA, Yoksan S, Buchy P, Nguyen VC, Sekaran SD, Enria DA, Vazquez S, for the detection of Dengue virus NS1 antigen and anti-Dengue virus IgM antibody. PLoSNegl Trop Dis. 2014 Oct 16;8(10):e3171. doi: 10.1371/journal.pntd.0003171.
  2. Hermann LL, Thaisomboonsuk B, Poolpanichupatam Y, Jarman RG, Kalayanarooj S, Nisalak A, Yoon IK, Fernandez S. Evaluation of a Dengue NS1 antigen detection assay sensitivity and specificity for the diagnosis of acute Dengue virus infection. PLoSNegl Trop Dis. 2014 Oct 2;8(10):e3193. doi: 10.1371/journal.pntd.0003193.
  3. Ganeshkumar P, Murhekar MV, Poornima V, Saravanakumar V, Sukumaran K, Anandaselvasankar A, John D, Mehendale SM. Dengue infection in India: A systematic review and meta-analysis. PLoSNegl Trop Dis. 2018 Jul 16;12(7):e0006618. doi: 10.1371/journal.pntd.0006618.
  4. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1

Page 123 of 459

  1. WHO, Evaluation of commercially available anti-Dengue virus immunoglobulin M tests. (Diagnostics evaluation series, 3). ISBN 978 92 4 159775 3.
  2. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

Page 124 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE NS1 and IgM COMBO RDT KIT site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type Positive samples: Not applicable, may categorize cases as per duration of illness Negative samples (may categorize as per duration of illness, must include cross reactivity panel) Results

Positive Negative Total Name of NS1 and IgM combo RDT kit Positive

Negative

Total

Prepare the above 2x2 table individually for each analyte, as well as for overall performance characteristics

Estimate (%) 95% CI Combined Sensitivity

Combined Specificity

Denue NS1 standalone sensitivity

Dengue NS1 standalone specificity

Dengue IgM standalone sensitivity

Dengue IgM standalone specificity

Page 125 of 459

o Cross-reactivity: ● Conclusions: o (Sensitivity and specificity have been assessed in using kits provided by the manufacturer from the batch mentioned above using ….. sample in ……. (field/controlled lab). Results should not be extrapolated to other sample types.) Disclaimers Note: This report is exclusively for …………………..NS1 and IgM combo Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In charge …………………… Seal ………………………………

Page 126 of 459

Performance evaluation protocol for Dengue real-time PCR kit I. Background:
This recommendation focuses on the laboratory performance evaluation of Dengue virus molecular diagnostic test. All clinical samples tested in the study should be evaluated in accordance with the candidate test’s instructions for use.
II. Purpose: To evaluate the performance characteristics of Dengue real-time PCR kits in the diagnosis of Dengue infection using irreversibly de-identified leftover archived/ spiked clinical samples. III. Requirements:
Performance evaluation activities using irreversibly de-identified leftover clinical samples are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024.
V. Procedure: Identified IVD kit evaluation laboratories should establish their proficiency through
Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory.

Page 127 of 459

⮚ Handling of Dengue RT-PCR kits received for performance evaluation 3. Preparation of Dengue RNA evaluation panel: Well characterised Dengue serum/plasma panel positive for RNA by RT-PCR is a critical requirement for performance evaluation of IVD kits utilizing genome detection. Hence statistically significant number of sera/plasma samples should be available from Dengue PCR confirmed cases.
4. RNA extraction: RNA extraction should be performed as per reference test and index test Instructions for Use (IFU). 5. Real-Time PCR System: 6. Internal control/Extraction control: The test under evaluation should have an internal control or extraction control (RNA added before extraction to a sample). 7. Reference assay: Any WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Dengue PCR assay or CDC/NIV protocol for detection of Dengue virus RNA should be used as the reference assay. All positive samples should be confirmed positive for at least one serotype by real-time PCR assay using CDC/NIV protocol.
All negative samples should be negative for all the markers of Dengue infection (NS1, IgM, and RNA).

Page 128 of 459

samples and sample panel composition against different values of sensitivity and 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

Positive samples: These include samples positive by the reference real-time PCR assay (True positives) and representative of all four serotypes.

Negative samples: All negative samples should be negative by reference real-time PCR assay, WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved NS1 antigen ELISA kit-and WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved IgM Capture ELISA.

Page 129 of 459

Table 1. Sample sizes and panel composition of positive Dengue samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off] # 99% 16 20 Strong positive (Ct value <25): 6 Moderate positive (Ct value between 25-30): 7
Weak positive (Ct value >30 to 34): 7
95% 77 80 24
between 25-30): 28
34): 28 90% 145 150 44
between 25-30): 53
34): 53 85% 206 210 62
between 25-30): 74
34): 74 80% 258 260 78
between 25-30): 91
34): 91 Note: If clinical samples positive for a particular serotype is not available, tissue culture fluid (5-10 different isolates with a plaque forming unit of 105-6/ml) (Heat-inactivated) from reference laboratories can be obtained, spiked in serum samples (15 µl isolate + 150 µl) and can be further diluted in the ratio of 1:10, frozen at -80°C, and tested by the reference assay when needed and the positive samples can be used for evaluation.
The serum used for spiking isolate should be negative for Dengue virus RNA, and NS1.

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

Page 130 of 459

Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off]

99% 16 20 1.Cross-reactivity panel:  Chikungunya PCR positive: 4 @  Japanese Encephalitis PCR positive: 1 @  Zika Virus PCR positive: 1 *  West Nile Virus PCR positive: 1 * 2.aAcute febrile cases: 8
3.bHealthy subjects from endemic regions: 5
95% 77 80  Chikungunya PCR positive: 15 @  Japanese PCR Encephalitis positive: 5 @  Zika Virus PCR positive: 5 *  West Nile Virus PCR positive: 5 * 2.aAcute febrile cases: 30
3.bHealthy subjects from endemic regions: 20
90% 145 150  Chikungunya PCR positive: 28 @  Japanese Encephalitis PCR positive: 9 @  Zika Virus PCR positive: 9 *  West Nile Virus PCR positive: 9 * 2.aAcute febrile cases: 57
3.bHealthy subjects from endemic regions: 38
85% 206 210  Chikungunya PCR positive: 39 @  Japanese Encephalitis PCR positive: 13 @  Zika Virus PCR positive: 13 *  West Nile Virus PCR positive: 13 * 2.aAcute febrile cases: 79
3.bHealthy subjects from endemic regions: 53
80% 258 260  Chikungunya PCR positive: 49 @  Japanese Encephalitis PCR positive: 16 @

Page 131 of 459

 Zika Virus PCR positive: 16 *  West Nile Virus PCR positive: 16 * 2.aAcute febrile cases: 98
3.bHealthy subjects from endemic regions: 65
a Acute febrile cases negative for all markers of Dengue (NS1 & IgM & IgG & RNA) b Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, nucleic acid)

  • Note:
    Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. If PCR positive samples for cross reactive flaviviruses not available, commercially available RNA panels should be used to test cross reactivity.

#It is recommended to calculate the sample size as per manufacturer's claims of

It is recommended to test for all pathogens listed in the cross-reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

  1. Evaluation method:

  2. Interpretation of results:

  3. Resolution of discrepant results:

test. test.

Page 132 of 459

  1. Repeatability and Reproducibility Assessment:

evaluation.

be as follows:
 First lot of the kit: should be tested on statistically significant number of positive and negative samples as calculated in the protocol.
 Second lot of the kit: should be tested on 25 samples (15 positive samples comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).
10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

Page 133 of 459

moderate and weak positive samples) and 3 negative samples should be tested 5 (qualitative).

(qualitative).

(qualitative).

recommended). 13. Blinding of laboratory staff:

Page 134 of 459

  1. Acceptance Criteria: Sensitivity: ≥95% Specificity: ≥98% To achieve at least the performance characteristics outlined in the acceptance criteria, ≥80 positive samples and ≥80 negative samples should be used for evaluation.
  2. Publication Rights: author(s). acceptable.

Page 135 of 459

VI. References:

  1. Santiago, G.A., Vázquez, J., Courtney, S. et al. Performance of the Trioplex real-time RT- PCR assay for detection of Zika, Dengue, and Chikungunya viruses. Nat Commun 9, 1391 (2018). https://doi.org/10.1038/s41467-018-03772-1
  2. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
3. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

Page 136 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE REAL-TIME PCR KITS site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: clinical/ spiked, strong, moderate, weak/simulated samples)
Negative samples (provide details clinical/ spiked,

Results

Positive Negative Total Name of Dengue real- time PCR kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity:

Page 137 of 459

Disclaimers Note: This report is exclusively for Dengue………….. Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ………………………….

Page 138 of 459

Field evaluation protocol for Dengue real-time PCR kits I. Background:
II. Purpose: To evaluate the performance characteristics of Dengue real-time PCR kits in the diagnosis of Dengue infection in individuals with unknown disease status. III. Requirements:
The study will be initiated after approval from the institutional human ethics committee.
V. Procedure:

  1. Study design/type: Cross-sectional study through
    Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory. ⮚ Handling of Dengue RT-PCR kits received for performance evaluation

Sample size for performance evaluation: Sample sizes of positive and negative samples of Dengue against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. It is further assumed that 30% of the

Page 139 of 459

individuals attending the health care facilities for acute febrile illness and suspected for Dengue will be positive for Dengue. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For require usage of the sample size outlined for 85% specificity.

𝑑2 x P x (1 −IR) 𝑑2 x P x (1 −IR)

· n (se) is the minimum number of number of individuals to be enrolled to obtain the requisite number of positive samples. · n (sp) is the minimum number of number of individuals to be enrolled to obtain the requisite number of negative samples.

P is the prevalence of the disease · d is the predetermined marginal error (5%).

Sample size has to be determined based on both the sensitivity and the specificity. The required sample size will be the maximum of the two. For example, at 95% sensitivity and 95% specificity, the sample size required will be 260 (maximum of 260 and 110).

Table 1. Sample sizes for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size
Minimum no. of individuals to be enrolled* off] #

Page 140 of 459

99% 53 60 95% 255 260 90% 484 490 85% 686 690 80% 861 870

  • Individuals attending the health care facilities for

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

Table 2. Sample sizes for different values of specificity claimed by the manufacturer. Specificity Calculated sample size Minimum no. of individuals to be enrolled* off] # 99% 23 30 95% 109 110 90% 207 210 85% 294 300 80% 369 370

  • Individuals attending the health care facilities for

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity Recruitment of cases shall be halted once desired number of positive and negative samples are reached. 4. Inclusion criteria: Individuals with Dengue like illness (A patient with acute febrile illness of 2-7 days with two or more manifestations: Head ache, retro-orbital pain, myalgia, arthralgia, rash, hemorrhagic manifestations) 5. Exclusion criteria: Individuals with already known positive history for other pathogens 6. RNA extraction RNA extraction should be performed as per reference test and index test Instructions for Use (IFU).

Page 141 of 459

Real-Time PCR System: 8. Internal control/Extraction control: The test under evaluation should have an internal control or extraction control (RNA added before extraction to a sample). 9. Reference assay:
Any WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Dengue PCR assay or CDC/NIV protocol for detection of Dengue RNA should be used as the reference assay. All positive samples should be confirmed positive for at least one serotype by real-time PCR assay using CDC/NIV protocol.
All negative samples should be negative for all the markers of Dengue infection (NS1 & IgM & IgG and RNA).
10. Study implementation: The individuals with Dengue like illness will be recruited into the study and five ml of whole blood will be collected in vacutainer tubes and the serum will be separated by centrifugation and used for the study.
It needs to be ensured that the samples are tested by reference tests and index test simultaneously. 11. Evaluation method:

  1. Interpretation of results:
  2. Positive samples: Samples which are positive by reference real-time PCR assay will be considered as true positive sample.

Page 142 of 459

  1. Negative samples:
    Samples which are negative by the reference assay will be considered as negative. N.B.:
    test. test. A. Cross reactivity: Clinical samples or commercially available validated viral RNA genome of other flaviviruses that are accepted by accreditation agencies/RNA from sequence confirmed virus isolates will be used to test cross reactivity of the index test. a. Chikungunya PCR positive: 5 samples @ b. Japanese Encephalitis PCR positive: 5 samples @ c. West Nile Virus PCR positive: 5 samples * d. Zika Virus PCR positive: 5 samples * Alternatively, tissue culture fluid of cross reactive flaviviruses (with a plaque forming unit of 105-6/ml)(Heat inactivated) from reference laboratories can be obtained, spiked in serum samples (15 µl isolate + 150 µl) and can be further diluted in the ratio of 1:10, tested by the reference assay and the negative samples can be used for evaluation. The serum used for spiking isolate should be negative for Dengue virus RNA, and NS1. reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the
  2. Statistical analysis:
    Sensitivity and specificity will be calculated. Interim analysis of data shall be conducted on completing evaluation of 25%, 50% and 75% of samples. If, at any point, the performance of the assay is found to be not satisfactory, the assay shall not be evaluated further. Evaluation fee shall be charged accordingly.
  3. Repeatability and Reproducibility Assessment:
    evaluation.

Page 143 of 459

be as follows:
comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).
10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

Page 144 of 459

moderate and weak positive samples) and 3 negative samples should be tested 5 (qualitative).

(qualitative).

(qualitative).

recommended).

  1. Resolution of discrepant results:

test. test.

  1. Blinding of laboratory staff

be blinded to the results of the reference assay. The PI of the evaluation exercise should remain unblinded, i.e., privy to the results of the reference test. Another senior laboratory staff selected by the PI may remain unblinded for overseeing the activity and maintaining the database of results, and the PI should maintain confidentiality of data. The data should be analyzed only by the PI of the evaluating lab.
19. Acceptance Criteria Sensitivity: ≥95%

Page 145 of 459

Specificity: ≥98% To achieve at least the performance characteristics outlined in the acceptance criteria, ≥260 individuals satisfying the case definition need to be enrolled to obtain the requisite number of positive samples. This sample size is sufficient for required number of negative samples. Recruitment should be terminated once the desired number of positive cases is enrolled and tested. 20. Publication Rights: author(s). acceptable.
VI. References:

  1. Santiago, G.A., Vázquez, J., Courtney, S. et al. Performance of the Trioplex real-time RT- PCR assay for detection of Zika, Dengue, and Chikungunya viruses. Nat Commun 9, 1391(2018). https://doi.org/10.1038/s41467-018-03772-1
  2. Ganeshkumar P, Murhekar MV, Poornima V, Saravanakumar V, Sukumaran K, Anandaselvasankar A, John D, Mehendale SM. Dengue infection in India: A systematic review and meta-analysis. PLoSNegl Trop Dis. 2018 Jul 16;12(7):e0006618. doi: 10.1371/journal.pntd.0006618.
  3. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
4. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

Page 146 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE REAL-TIME PCR KITS site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type Positive samples: Not applicable, may categorize cases as per duration of illness Negative samples (may categorize as per duration of illness, must include cross reactivity panel)

Results

Positive Negative Total Name of Dengue real- time PCR kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity:

Page 147 of 459

Disclaimers

Note: This report is exclusively for Dengue………….. Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ………………………………..

Page 148 of 459

Performance evaluation protocol for Real-time PCR tests for Zika virus I. Background:
CDSCO and ICMR, New Delhi, aimed at facilitating the evaluation and deployment of Quality-Assured Diagnostics kits appropriate for use in India. Hence the following guidelines shall establish the uniformity in performance evaluation of in-vitro diagnostic kits (IVD). The performance evaluation is to independently verify the manufacturer’s claim regarding in-vitro diagnostic kit (IVD) performance. This recommendation focuses on the laboratory performance evaluation of Zika virus molecular diagnostic test. All clinical samples tested in the study should be evaluated in accordance with the candidate test’s proposed diagnostic algorithm (i.e., tested using the procedure in the instructions for use), including retesting when appropriate.
II. Purpose:
To evaluate the performance characteristics of Zika virus RT-PCR test for diagnosis of Zika infection using irreversibly de-identified leftover archived/ spiked clinical samples.
III. Requirements:

  1. Supply of kits under evaluation (along with batch/lot No. Expiry & required details)
  2. Evaluation site/laboratory should be equipped with necessary equipment and supplies for molecular testing. Any essential equipment and consumables for closed system to be supplied and maintained from the manufacturer, during the period of evaluation.
  3. Characterized evaluation panel IV. Ethics approval:

V. Procedure:

  1. Study design: Diagnostic accuracy study using irreversibly de-identified leftover

Page 149 of 459

  1. Preparation of Evaluation site/laboratory: Performance evaluation performance and report to be issued only from designated reference testing laboratory/ ISO accredited laboratory, as specified by state or central licensing authority.

  2. Identified IVD kit evaluation laboratories should establish their proficiency through: Laboratory/ Calibration Laboratory (ISO/IEC 17025), Medical Laboratory (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory. B. Staff training: All the staff involved in the IVD kit evaluation should undergo hands on training and competency testing on following
    ⮚ Preparation & characterization of evaluation panel
    ⮚ Handling of Zika molecular diagnostic kits received for performance evaluation (Verification/Storage/Unpacking etc.) ⮚ Testing, interpretation, recording of results & reporting

  3. Preparation of Zika reference evaluation panel: Well characterized Zika molecular evaluation panel is a critical requirement for performance evaluation of IVD kits. Hence, statistically significant number of clinical samples should be used for evaluation. ● Frozen samples (≤-70°C) may be used, if stored appropriately and analytical data demonstrate that accuracy of test results is not affected.
    ● Samples that previously tested positive by WHO Pre-Qualified/US FDA/ATAGI
    Australia/ PMDA Japan approved PCR and/or CDC/NIV approved protocols may be used. ● In the absence of natural samples, spiked clinical samples may be used.

RNA extraction: RNA extraction should be performed as per reference test and index test Instructions for Use (IFU). 6. Real-Time PCR System:

Page 150 of 459

Internal control/Extraction control: The test under evaluation should have an internal control or extraction control (RNA added before extraction to a sample). 8. Reference assay:
Any WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Zika PCR assay or CDC/NIV protocol for detection of Zika RNA should be used as the reference assay. Evaluations with the reference test should be conducted as per the manufacturer’s instructions for use.

Positive and negative samples should be subjected to both the reference test and test under evaluation. 9. samples and panel composition against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For require usage of the sample size outlined for 85% specificity.

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

Page 151 of 459

Positive Samples:
 Clinical positive samples: Sample tested positive by Zika virus molecular reference assay from clinically suspect cases.  Contrived positive samples: In absence of reference clinical samples, a contrived positive sample may be used.
Contrived positive samples should be prepared using spiking of diluted Zika virus culture isolate in unique negative samples, as per the note below: Table 1. Sample sizes and panel composition of positive Zika virus samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off] # 99%# 16 20 6 between 25-30): 7
34): 7
95% 77 80 24
between 25-30): 28
34): 28 90% 145 150 44
between 25-30): 53
34): 53 85% 206 210 62
between 25-30): 74
34): 74 80% 258 260 78
between 25-30): 91
34): 91

Page 152 of 459

Note 1: Representative positive samples from genotype (African, Asian/American) may be included, if feasible.
Note 2: Contrived positive samples – In absence of reference clinical samples, a contrived positive sample may be used.
Contrived positive samples should be prepared using spiking of diluted Zika virus culture isolate in unique negative samples, as follows: Tissue culture fluid (3-5 different isolates with a plaque forming unit of 105-6/ml) (Heat inactivated) from reference laboratories can be obtained, spiked in serum samples (15 µl isolate + 150 µl) and can be further diluted in the ratio of 1:10, tested by the reference assay and the positive samples can be used for evaluation. The serum used for spiking isolate should be negative for Dengue virus RNA, and NS1.

#It is recommended to calculate the sample size as per manufacturer's claims of

Table 2. Sample sizes and panel composition of negative Zika virus samples for different values of specificity claimed by the manufacturer. Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off]

99%# 16 20  Dengue PCR positive: 4 @
 Chikungunya PCR positive: 1 @  Japanese Encephalitis positive: 1 *  West Nile Virus positive: 1 * 2.aAcute febrile cases: 10
3.Healthy subjects from endemic regions: 3
95% 77 80  Dengue PCR positive: 15 @
 Chikungunya PCR positive: 5 @  Japanese Encephalitis positive: 5 *  West Nile Virus positive: 5 * 2.aAcute febrile cases: 40
3.Healthy subjects from endemic regions: 10
90% 145 150  Dengue PCR positive: 28 @
 Chikungunya PCR positive: 9 @  Japanese Encephalitis positive: 9 *  West Nile Virus positive: 9 *

Page 153 of 459

2.aAcute febrile cases: 76
3.Healthy subjects from endemic regions: 19
85% 206 210  Dengue PCR positive: 40 @
 Chikungunya PCR positive: 13 @  Japanese Encephalitis positive: 13 *  West Nile Virus positive: 13 * 2.aAcute febrile cases: 105
3.Healthy subjects from endemic regions: 26
80% 258 260  Dengue PCR positive: 49 @
 Chikungunya PCR positive: 16 @  Japanese Encephalitis positive: 16 *  West Nile Virus positive: 16 * 2.aAcute febrile cases: 130
3.Healthy subjects from endemic regions: 33
a Acute febrile cases negative by Zika virus molecular reference assay

Positive samples / samples spiked with culture filtrate of viruses mentioned in the cross- reactivity panel may be used.

#It is recommended to calculate the sample size as per manufacturer's claims of

It is recommended to test for all pathogens listed in the cross-reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing/dropping only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

Note:
Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required analyte level is not available. If PCR positive samples for cross reactive flaviviruses are not available, commercially available RNA panels/RNA from virus isolates should be used to test cross reactivity.

  1. Evaluation method:

  2. Interpretation of results:

Page 154 of 459

  1. Resolution of discrepant results:

test. test.

evaluation.

be as follows:
comprising 10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).
10 low positive AND 5 moderate/ strong positive samples, and 10 negative samples).

Page 155 of 459

moderate and weak positive samples) and 3 negative samples should be tested 5 (qualitative).

(qualitative).

(qualitative).

Page 156 of 459

recommended). 14. Blinding of laboratory staff:

  1. Acceptance criteria: Sensitivity: ≥95% Specificity: ≥98%

Page 157 of 459

Agreement between sample types– Candidate tests meant for testing multiple sample matrices should demonstrate a minimum of 95% positive percent agreement (PPA) and negative percent agreement (NPA) for all specimen types.

To achieve at least the performance characteristics outlined in the acceptance criteria, ≥80 positive samples and ≥80 negative samples should be used for evaluation.

  1. Publication Rights: author(s).

acceptable.

VI. References:

  1. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
2. Santiago GA, Vázquez J, Courtney S, Matías KY, Andersen LE, Colón C, Butler AE, Roulo R, Bowzard J, Villanueva JM, Muñoz-Jordan JL. Performance of the Trioplex real-time RT- PCR assay for detection of Zika, Dengue, and Chikungunya viruses. Nat Commun. 2018 Apr 11;9(1):1391. doi: 10.1038/s41467-018-03772-1. 3. Stone M, Bakkour S, Grebe E, Emperador DM, Escadafal C, Deng X, Dave H, Kelly-Cirino C, Lackritz E, Rojas DP, Simmons G, Rabe IB, Busch MP. Standardized evaluation of Zika nucleic acid tests used in clinical settings and blood screening. PLoS Negl Trop Dis. 2023 Mar 17;17(3):e0011157. 4. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

Page 158 of 459

PERFORMANCE EVALUATION REPORT FOR ZIKA REAL-TIME PCR KIT site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: clinical/spiked, strong, moderate, weak/simulated samples)
Negative samples (provide details clinical/spiked, Results

Positive Negative Total Name of Zika virus real-time PCR kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity:

FINAL CONCLUSION

Page 159 of 459

Disclaimers Note: This report is exclusively for ………….. Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal……………………………….

Page 160 of 459

DENGUE IgG BASED ASSAYS

Page 161 of 459

Performance evaluation protocol for Dengue IgG RDT kits I. Background:
CDSCO/ICMR, New Delhi, have aimed at facilitating the availability of Quality-Assured evaluation is to independently verify the manufacturer’s claim regarding in-vitro diagnostic kit (IVD) performance. II. Purpose: To evaluate the performance characteristics of Dengue IgG RDT kits in the diagnosis of primary and secondary dengue infections using irreversibly de-identified leftover archived clinical samples. III. Requirements:
V. Procedure:

  1. Study design/type: Diagnostic accuracy study using irreversibly de-identified archived/ spiked leftover clinical samples
    through ALL of the following: A. Accreditation for at least one of the Quality management systems (accreditation for Testing Lab / Calibration Lab (ISO/IEC: 17025), Medical Lab (ISO: 15189), PT provider (ISO/IEC: 17043) or CDSCO approved Reference laboratory.

Page 162 of 459

⮚ Handling of Dengue IgG Rapid IVD kits received for performance evaluation

  1. Preparation of Dengue IgG Rapid IVD kit evaluation panel: Well characterised Dengue IVD kit evaluation panel is a critical requirement for should be collected from Dengue NS1/PCR/IgM confirmed cases. Further characterised for Dengue IgG positivity by using approved reference kits having high sensitivity and specificity. Dengue IgG performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples.
  2. Reference assay:
    Positive and negative samples should be characterized using composite reference standard of Dengue IgG AND one additional marker of Dengue (NS1 or IgM or PCR). The following kits should be used for characterization of the sample panel:  Panbio Dengue IgG capture ELISA kit/ WHO Pre-Qualified/ US-FDA/ ATAGI Australia/ PMDA Japan approved Dengue IgG ELISA kit  WHO Pre-Qualified/ US-FDA/ ATAGI Australia/ PMDA Japan approved Dengue IgM ELISA kit  NS1 antigen status to be assessed using WHO Pre-Qualified/ US-FDA/ ATAGI Australia/ PMDA Japan approved NS1 ELISA kit  Serotype status to be assessed using a combination of CDC/NIV real-time PCR serotyping protocols.
  3. Sample size for performance evaluation:

Sample sizes of positive and negative samples and sample panel composition against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5% using the following formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

Page 163 of 459

Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.
Positive samples: The panel of positive samples should include samples positive for IgG by the reference assay. The samples should also be positive for either dengue NS1 antigen or dengue IgM antibodies.
It is recommended (but not mandatory) to have IgG positive samples from primary and secondary dengue cases.

Negative samples: Samples which are negative by reference dengue IgG test should form the negative sample panel.
Table 1. Sample sizes and panel composition of positive dengue IgG samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off] # 99% 16 20 Strong Positive: 6 Weak Positive: 7 95% 77 80 Strong Positive: 24 Weak Positive: 28 90% 145 150 Strong Positive: 44 Moderate Positive: 53 Weak Positive: 53 85% 206 210 Strong Positive: 62

Page 164 of 459

Moderate Positive: 74 Weak Positive: 74 80% 258 260 Strong Positive: 78 Moderate Positive: 91 Weak Positive: 91 The samples need to be classified as strong, moderate and weak positives based on ELISA units of the reference assay.

#It is recommended to calculate the sample size as per manufacturer's claims of

Table 2. Sample sizes and panel composition of negative dengue IgG samples for different values of specificity claimed by the manufacturer. Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off]

99%# 16 20 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 7
2.Acute febrile illness cases: 8
 Chikungunya positive samples:2  Dengue (NS1 & IgM & IgG & PCR) negative samples:6 3.Samples from other flavivirus disease cases (cross- reactive panel): 3  Japanese Encephalitis IgM/IgG positive: 1 @  West Nile Virus IgM/IgG positive: 1 *  Zika Virus IgM/IgG positive: 1 * 4. aHealthy subjects from endemic regions: 2
95% 77 80 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 27
2.Acute febrile illness cases: 32
 Chikungunya positive samples:8 negative samples:24 3.Samples from other flavivirus disease cases(cross- reactive panel): 9  Japanese Encephalitis IgM/IgG positive: 3 @  West Nile Virus IgM/IgG positive: 3 *  Zika Virus IgM/IgG positive: 3 * 4. aHealthy subjects from endemic regions: 12
90% 145 150 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 50
2.Acute febrile illness cases: 60

Page 165 of 459

 Chikungunya positive samples:15 negative samples:45 3.Samples from other flavivirus disease cases(cross- reactive panel): 15  Japanese Encephalitis IgM/IgG positive: 5 @  West Nile Virus IgM/IgG positive: 5 *  Zika Virus IgM/IgG positive: 5 * 4. aHealthy subjects from endemic regions: 25
85% 206 210 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 70
2.Acute febrile illness cases: 84
 Chikungunya positive samples:21 negative samples:63 3.Samples from other flavivirus disease cases(cross- reactive panel): 21  Japanese Encephalitis IgM/IgG positive: 7 @  West Nile Virus IgM/IgG positive: 7 *  Zika Virus IgM/IgG positive: 7 * 4. aHealthy subjects from endemic regions: 35
80% 258 260 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 85
2.Acute febrile illness cases: 104
 Chikungunya positive samples:26 negative samples:78 3.Samples from other flavivirus disease cases(cross- reactive panel): 27  Japanese Encephalitis IgM/IgG positive: 9 @  West Nile Virus IgM/IgG positive: 9 *  Zika Virus IgM/IgG positive: 9 * 4. aHealthy subjects from endemic regions: 44
a Samples from healthy subjects from endemic regions negative for all dengue markers (NS1, IgM, IgG, RNA)

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics.

there is an acute shortfall or non-availability of clinical samples, one may consider reducing only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

Page 166 of 459

Note: If IgM/IgG positive samples for cross reactive flaviviruses are not available, commercially available validated standard panels that are accepted by accreditation agencies.

  1. Evaluation method:

The index test and the reference assay should be run simultaneously on the sample panel, and

  1. Interpretation of results:
  2. Resolution of discrepant results:

test. test.

under evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples

be as follows:
negative samples).
samples).

Page 167 of 459

two different operators, keeping all other testing parameters undisturbed. Within- run and between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples (qualitative).

(qualitative).

3 positive samples (strong, moderate and weak positive samples) and 3 negative (qualitative).

Page 168 of 459

  1. Acceptance Criteria: Sensitivity: ≥80% Specificity: ≥90% Cross reactivity: Minimal To achieve at least the performance characteristics outlined in the acceptance criteria, ≥260 positive samples and ≥150 negative samples should be used for evaluation.

Page 169 of 459

The PI (s) of the evaluating labs shall retain publication rights of the evaluation as lead author(s). acceptable.
VI. References:

  1. Vazquez S, Hafner G, Ruiz D, Calzada N, Guzman MG. Evaluation of immunoglobulin M and G capture enzyme-linked immunosorbent assay Panbio kits for diagnostic dengue infections. J Clin Virol. 2007 Jul;39(3):194-8. doi: 10.1016/j.jcv.2007.04.003..
  2. WHO, Evaluation of commercially available anti-Dengue virus immunoglobulin M tests. (Diagnostics evaluation series, 3). ISBN 978 92 4 159775 3.
  3. Central Drugs Standard Control Organization. Guidance on Performance Evaluation of In-vitro Diagnostic Medical Devices.

Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/medical device/guidanceperformanceivd.pdf 4. Central Drugs Standard Control Organization. In-Vitro Diagnostic (IVD) Medical Devices Frequently Asked Questions. 2022. Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/IVD/FAQs/CDSCO- IVD-FAQ-03-2022-.pdf
5. U.S. Food and Drug Administration. Dengue Virus Serological Reagents - Class II Special Controls Guideline for Industry and Food and Drug Administration Staff. 2014. Available at: https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation- emitting-products/Dengue-virus-serological-reagents-class-ii-special-controls-guideline-industry- and-food-and-drug
6. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3. 2017. Available at: eng.pdf;sequence=1
7. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability and diagnostic performance: a go-to review. European radiology experimental, 8(1), 79. *The validation protocols need to be revisited after introduction of Dengue vaccines and the acceptance criteria needs revisiting every year so as to enable the availability of best diagnostic kits.
VII.

Page 170 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE IgG RDT KIT site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type

Results:

Positive Negative Total Name of Dengue IgG antibody - based RDT kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity: o Invalid test rate:

Page 171 of 459

Disclaimers Note: This report is exclusively for ………………………Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ……………………….

Page 172 of 459

Performance evaluation protocol for Dengue IgM and IgG RDT combo kits I. Background:
CDSCO/ICMR, New Delhi, have aimed at facilitating the availability of Quality-Assured evaluation is to independently verify the manufacturer’s claim regarding in-vitro diagnostic kit (IVD) performance. II. Purpose: To evaluate the performance characteristics of Dengue IgM and IgG RDT combo kits in the diagnosis of dengue and discriminating primary and secondary dengue infections using irreversibly de-identified leftover archived clinical samples. III. Requirements:
V. Procedure:

  1. Study design/type: Diagnostic accuracy study using irreversibly de-identified archived/ spiked leftover clinical samples through ALL of the following: A. Accreditation for at least one of the Quality management systems (accreditation for Testing Lab / Calibration Lab (ISO/IEC: 17025), Medical Lab (ISO: 15189), PT provider (ISO/IEC: 17043) or CDSCO approved Reference laboratory.

Page 173 of 459

⮚ Handling of Dengue IgM and IgG Rapid IVD kits received for performance evaluation (Verification/Storage/Unpacking etc).

  1. Preparation of Dengue IgM and IgG Rapid IVD kit evaluation panel: Well characterised Dengue IVD kit evaluation panel is a critical requirement for should be collected from Dengue NS1/PCR/IgM confirmed cases. Further characterised for Dengue IgG positivity by using approved reference kits having high sensitivity and specificity. Dengue IgG performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples.
  2. Reference assay:
    Positive and negative samples should be characterized using reference standard for Dengue IgG (and one additional marker of Dengue - NS1 or IgM or PCR) AND IgM. The following kits should be used for characterization of the sample panel:  Panbio Dengue IgG capture ELISA kit/ WHO Pre-Qualified/ US-FDA/ ATAGI Australia/ PMDA Japan approved Dengue IgG ELISA kit  WHO Pre-Qualified/ US-FDA/ ATAGI Australia/ PMDA Japan approved Dengue IgM ELISA kit  NS1 antigen status to be assessed using WHO Pre-Qualified/ US-FDA/ ATAGI Australia/ PMDA Japan approved NS1 ELISA kit  Serotype status to be assessed using a combination of CDC/NIV real-time PCR serotyping protocols.
    samples of Dengue against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5% using the following formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

Page 174 of 459

Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.

Positive samples: The samples should be positive for dengue IgM antibodies. The panel of positive samples should include 50% of samples positive for IgG by the reference assay. Samples should be representative of varying degrees of positivity. It is recommended (but not mandatory) to have IgG positive samples from primary and secondary dengue cases.

Negative samples: These should include samples negative by all the reference assays (True negatives).

Table 1. Sample sizes and panel composition of positive Dengue samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off for balanced allocation] # 99% 16 20

  1. 10 samples positive for Dengue IgM  Strong positive:3  Moderate positive: 3  Weak positive: 4
  2. 10 samples positive for both Dengue IgM and IgG  Strong positive IgG:3  Moderate positive IgG: 3  Weak positive IgG: 4

Page 175 of 459

95% 77 80 40 samples positive for Dengue IgM  Strong positive:12  Moderate positive: 14  Weak positive: 14

40 samples positive for both Dengue IgM and IgG  Strong positive IgG:12  Moderate positive IgG: 14  Weak positive IgG: 14 90% 145 150 75 samples positive for Dengue IgM  Strong positive:23  Moderate positive: 26  Weak positive: 26

75 samples positive for both Dengue IgM and IgG  Strong positive IgG: 23  Moderate positive IgG: 26  Weak positive IgG: 26 85% 206 210 105 samples positive for Dengue IgM  Strong positive:31  Moderate positive: 37  Weak positive: 37

105 samples positive for both Dengue IgM and IgG  Strong positive IgG: 31  Moderate positive IgG: 37  Weak positive IgG: 37 80% 258 260 130 samples positive for Dengue IgM  Strong positive:38  Moderate positive: 46  Weak positive: 46

130 samples positive for both Dengue IgM and IgG  Strong positive IgG: 38  Moderate positive IgG: 46  Weak positive IgG: 46

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity

Page 176 of 459

Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off for balanced allocation] # 99% 16 20 1.aSamples from acute febrile illness cases negative for dengue: 9

 Samples positive for chikungunya: 2
 Other Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR):7

2.Samples from other flavivirus disease cases (cross-reactive panel): 3  Japanese Encephalitis IgM/IgG positive: 1@  West Nile Virus IgM/IgG positive:1*  Zika Virus IgM/IgG positive: 1 *

  1. bHealthy subjects from endemic regions: 8
    95% 77 80 1.aSamples from acute febrile illness cases negative for dengue: 44

 Samples positive for chikungunya: 8
 Other Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR):36

2.Samples from other flavivirus disease cases (cross-reactive panel): 6 positive: 2@  West Nile Virus IgM/IgG positive:2*  Zika Virus IgM/IgG positive: 2 *

30 90% 145 150 1.aSamples from acute febrile illness cases negative for dengue: 80

 Samples positive for chikungunya: 15

Page 177 of 459

 Other Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR):65

2.Samples from other flavivirus disease cases (cross-reactive panel): 15 positive: 5 @  West Nile Virus IgM/IgG positive:5*  Zika Virus IgM/IgG positive: 5*

55 85% 206 210 1.aSamples from acute febrile illness cases negative for dengue: 110

 Samples positive for chikungunya: 21
 Other Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR):89

2.Samples from other flavivirus disease cases (cross-reactive panel): 24 positive: 8 @  West Nile Virus IgM/IgG positive:8*  Zika Virus IgM/IgG positive: 8*

76 80% 258 260 1.aSamples from acute febrile illness cases negative for dengue: 138

 Samples positive for chikungunya: 26  Other Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR):112

2.Samples from other flavivirus disease cases (cross-reactive panel): 27 positive: 9 @  West Nile Virus IgM/IgG positive:9*  Zika Virus IgM/IgG positive: 9*

95

Page 178 of 459

a Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR) b Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, RNA)

#It is recommended to calculate the sample size as per manufacturer's claims of

It is recommended to test for all pathogens listed in the cross reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross reactive sample panel remain the same by compensating with the

Note: If IgM/IgG positive samples for cross reactive flaviviruses are not available, commercially available validated standard panels that are accepted by accreditation agencies can also be used.

  1. Evaluation method:

The index test and the reference assay should be run simultaneously on the sample panel, and

  1. Resolution of discrepant results:

test. test.

  1. Repeatability and Reproducibility Assessment:

under evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples for each analyte should be tested 5 times in independent runs.

Page 179 of 459

be as follows:
comprising 10 low positive AND 5 moderate/strong positive samples per target analyte, and 10 negative samples).
10 low positive AND 5 moderate/strong positive samples per target analyte, and 10 negative samples).

two different operators, keeping all other testing parameters undisturbed. Within- run and between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples for each analyte should be tested 5 times in independent runs.
(qualitative).

Page 180 of 459

and 3 negative samples for each analyte should be tested 5 times in independent runs.
(qualitative).

samples for each analyte should be tested 5 times in independent runs.
(qualitative).

  1. Blinding of laboratory staff:

Page 181 of 459

  1. Acceptance Criteria: Sensitivity for each analyte: ≥80% Specificity for each analyte: ≥90% Cross-reactivity: Minimal To achieve at least the performance characteristics outlined in the acceptance criteria, ≥260 positive samples and ≥150 negative samples for each analyte should be used for evaluation.
  2. Publication Rights: author(s). acceptable.

VI. References:

  1. Vazquez S, Hafner G, Ruiz D, Calzada N, Guzman MG. Evaluation of immunoglobulin M and G capture enzyme-linked immunosorbent assay Panbio kits for diagnostic dengue infections. J Clin Virol. 2007 Jul;39(3):194-8. doi: 10.1016/j.jcv.2007.04.003..
  2. WHO, Evaluation of commercially available anti-Dengue virus immunoglobulin M tests. (Diagnostics evaluation series, 3). ISBN 978 92 4 159775 3.
  3. Central Drugs Standard Control Organization. Guidance on Performance Evaluation of In-vitro Diagnostic Medical Devices.

Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/medical device/guidanceperformanceivd.pdf 4. Central Drugs Standard Control Organization. In-Vitro Diagnostic (IVD) Medical Devices Frequently Asked Questions. 2022. Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/IVD/FAQs/CDSCO- IVD-FAQ-03-2022-.pdf
5. U.S. Food and Drug Administration. Dengue Virus Serological Reagents - Class II Special Controls Guideline for Industry and Food and Drug Administration Staff. 2014. Available at: https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation-

Page 182 of 459

emitting-products/Dengue-virus-serological-reagents-class-ii-special-controls-guideline- industry-and-food-and-drug
6. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3. 2017. Available at: eng.pdf;sequence=1
7. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability and diagnostic performance: a go-to review. European radiology experimental, 8(1), 79.

*The validation protocols need to be revisited after introduction of Dengue vaccines and the acceptance criteria needs revisiting every year so as to enable the availability of best diagnostic kits.
VII.

Page 183 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE IgM and IgG COMBO RDT KIT

site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type

Positive Negative Total Name of Dengue antibody combo RDT kit Positive

Negative

Total

Prepare the above 2x2 table for each analyte and for overall performance characteristics

● Details of cross reactivity:
● Invalid test rate:

Page 184 of 459

● Conclusions: o Sensitivity, specificity for dengue IgM: o Sensitivity, specificity for dengue IgG: o Invalid test rate: o Performance:
▪ Satisfactory / Not satisfactory for Dengue IgM ▪ Satisfactory / Not satisfactory for Dengue IgG

Disclaimers

Note: This report is exclusively for ………….. Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ………………………………

Page 185 of 459

Performance evaluation protocol for Dengue IgG ELISA kits I. Background:
CDSCO/ICMR, New Delhi, have aimed at facilitating the availability of Quality-Assured evaluation is to independently verify the manufacturer’s claim regarding in-vitro diagnostic kit (IVD) performance. II. Purpose: To evaluate the performance characteristics of Dengue IgG ELISA kits in the diagnosis of Dengue infection using irreversibly de-identified leftover archived clinical samples. III. Requirements:
V. Procedure:

  1. Study design/type: Diagnostic accuracy study using irreversibly de-identified archived/ spiked leftover clinical samples through ALL of the following: A. Accreditation for at least one of the Quality management systems (accreditation for Testing Lab / Calibration Lab (ISO/IEC: 17025), Medical Lab (ISO: 15189), PT provider (ISO/IEC: 17043) or CDSCO approved Reference laboratory.

Page 186 of 459

⮚ Handling of Dengue IgG ELISA IVD kits received for performance evaluation 3. Preparation of Dengue IgG ELISA IVD kit evaluation panel: Well characterised Dengue IVD kit evaluation panel is a critical requirement for should be collected from Dengue NS1/PCR/IgG confirmed cases. Further characterised for Dengue IgM positivity by using approved reference kits having high sensitivity and specificity. Dengue IgG performance evaluation panel need to be tested again by the reference assays at the time of evaluating a particular index test to confirm the positive and negative status of the samples. 4. Reference assay:
Positive and negative samples should be characterized using composite reference standard of Dengue IgG AND one additional marker of Dengue (NS1 or IgM or PCR). The following kits should be used for characterization of the sample panel:  Panbio Dengue IgG capture ELISA kit/ WHO Pre-Qualified/ US-FDA/ ATAGI Australia/ PMDA Japan approved Dengue IgG ELISA kit  WHO Pre-Qualified/ US-FDA/ ATAGI Australia/ PMDA Japan approved Dengue IgM ELISA kit  NS1 antigen status to be assessed using WHO Pre-Qualified/ US-FDA/ ATAGI Australia/ PMDA Japan approved NS1 ELISA kit  Serotype status to be assessed using a combination of CDC/NIV real-time PCR serotyping protocols.
5. Sample size for performance evaluation: Sample sizes of positive and negative samples and sample panel composition against different values of sensitivity and specificity are provided in Tables 1 and 2. Sample sizes have been calculated assuming 95% level of significance, and an absolute precision of 5% using the following formulae:

𝑑2

𝑑2

Page 187 of 459

Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.

Positive samples: The panel of positive samples should include samples positive for IgG by the reference assay. The samples should also be positive for either dengue NS1 antigen or dengue IgM antibodies. Samples should be representative of varying degrees of positivity. It is recommended (but not mandatory) to have IgG positive samples from primary and secondary dengue cases.

Negative samples: These should include samples negative by the reference assays for dengue IgG.

Table 1. Sample sizes and panel composition of positive Dengue samples for different values of sensitivity claimed by the manufacturer. Sensitivity Calculated sample size Minimum no. of Positive Samples required off] # 99% 15 20 Strong Positive: 6 Weak Positive: 7 95% 73 80 Strong Positive: 24 Weak Positive: 28 90% 138 140 Strong Positive: 42 Moderate Positive: 49 Weak Positive: 49 85% 196 200 Strong Positive: 60 Moderate Positive: 70 Weak Positive: 70 80% 246 250 Strong Positive: 75 Moderate Positive: 87 Weak Positive: 88

Page 188 of 459

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. Specificity Calculated sample size Minimum no. of Negative Samples required [Sample size rounded off]# 99% 15 20 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 7
2.Acute febrile illness cases: 8
 Chikungunya positive samples:2 negative samples:6 3.Samples from other flavivirus disease cases (cross-reactive panel): 3 positive: 1 @  West Nile Virus IgM/IgG positive: 1 *  Zika Virus IgM/IgG positive: 1 * 4. aHealthy subjects from endemic regions: 2
95% 73 80 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 27
2.Acute febrile illness cases: 32
 Chikungunya positive samples:8 negative samples:24 3.Samples from other flavivirus disease cases(cross-reactive panel): 9 positive: 3 @  West Nile Virus IgM/IgG positive: 3 *  Zika Virus IgM/IgG positive: 3 * 4. aHealthy subjects from endemic regions: 12
90% 138 140 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 45
2.Acute febrile illness cases: 60
 Chikungunya positive samples:15 negative samples:45

Page 189 of 459

3.Samples from other flavivirus disease cases(cross-reactive panel): 15 positive: 5 @  West Nile Virus IgM/IgG positive: 5 *  Zika Virus IgM/IgG positive: 5 * 4. aHealthy subjects from endemic regions: 20
85% 196 200 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 65
2.Acute febrile illness cases: 84
 Chikungunya positive samples:21 negative samples:63 3.Samples from other flavivirus disease cases(cross-reactive panel): 21 positive: 7 @  West Nile Virus IgM/IgG positive: 7 *  Zika Virus IgM/IgG positive: 7 * 4. aHealthy subjects from endemic regions: 30
80% 246 250 1.Samples positive for dengue IgM/NS1/RNA but negative for IgG: 80
2.Acute febrile illness cases: 104
 Chikungunya positive samples:26 negative samples:78 3.Samples from other flavivirus disease cases(cross-reactive panel): 27 positive: 9 @  West Nile Virus IgM/IgG positive: 9 *  Zika Virus IgM/IgG positive: 9 * 4. aHealthy subjects from endemic regions: 39
a Acute febrile cases negative for Dengue (NS1 & IgM & IgG & PCR) b Samples from healthy subjects from endemic regions negative for all Dengue markers (NS1, IgM, IgG, RNA)

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. It is recommended to test for all pathogens listed in the cross reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross reactive sample panel remain the same by compensating with the

Page 190 of 459

Note: If IgM/IgG positive samples for cross reactive flaviviruses are not available, commercially available validated standard panels that are accepted by accreditation agencies can also be used.

  1. Evaluation method:

The index test and the reference assay should be run simultaneously on the sample panel, and results should be recorded.

  1. Interpretation of results:
  2. Resolution of discrepant results:

False positive samples: These are samples negative by reference assay and positive by index test. False negative samples: These are samples positive by reference assay and negative by index test.

This should be done to assess the repeatability of the detection of target analyte(s) using the kit under evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples

be as follows:
negative samples).

Page 191 of 459

samples).

moderate and weak positive samples) and 3 negative samples should be tested 5 (qualitative).

(qualitative).

Page 192 of 459

(qualitative).

  1. Acceptance criteria: Sensitivity: ≥90% Specificity: ≥95%

Page 193 of 459

To achieve at least the performance characteristics outlined in the acceptance criteria, ≥140 positive samples and ≥80 negative samples should be used for evaluation. author(s). acceptable.

VI. References:

  1. Vazquez S, Hafner G, Ruiz D, Calzada N, Guzman MG. Evaluation of immunoglobulin M and G capture enzyme-linked immunosorbent assay Panbio kits for diagnostic dengue infections. J Clin Virol. 2007 Jul;39(3):194-8. doi: 10.1016/j.jcv.2007.04.003..
  2. WHO, Evaluation of commercially available anti-Dengue virus immunoglobulin M tests. (Diagnostics evaluation series, 3). ISBN 978 92 4 159775 3.
  3. Central Drugs Standard Control Organization. Guidance on Performance Evaluation of In- vitro Diagnostic Medical Devices.

Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/medical device/guidanceperformanceivd.pdf 4. Central Drugs Standard Control Organization. In-Vitro Diagnostic (IVD) Medical Devices Frequently Asked Questions. 2022. Available at: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf- documents/IVD/FAQs/CDSCO-IVD-FAQ-03-2022-.pdf
5. U.S. Food and Drug Administration. Dengue Virus Serological Reagents - Class II Special Controls Guideline for Industry and Food and Drug Administration Staff. 2014. Available at: https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation- emitting-products/Dengue-virus-serological-reagents-class-ii-special-controls-guideline- industry-and-food-and-drug
6. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3. 2017. Available at: eng.pdf;sequence=1 7. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

Page 194 of 459

*The validation protocols need to be revisited after introduction of Dengue vaccines and the acceptance criteria needs revisiting every year so as to enable the availability of best diagnostic kits.

VII.

Page 195 of 459

PERFORMANCE EVALUATION REPORT FOR DENGUE IgG ELISA KIT site office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

date:

Valid Up to: Application No. Sample Panel Sample type

Results:

Positive Negative Total Name of Dengue IgG antibody -based ELISA kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity: o Invalid test rate: o (Sensitivity and specificity have been assessed in controlled lab setting using kits provided by the manufacturer from the batch mentioned above using ….. sample. Results should not be extrapolated to other sample types.) Disclaimers

Page 196 of 459

Note: This report is exclusively for ………………………Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal………………………………………

Page 197 of 459

Information on Operational and Test Performance Characteristics Required from Manufacturers for Dengue/Chikungunya/ Zika IVD The manufacturer should provide the following details about the IVD:

  1. Instructions for Use
  2. Scope of the IVD: to diagnose Dengue and/or/Chikungunya and/or Zika virus
  3. Intended Use Statement
  4. Principle of the assay
  5. Intended testing population(cases of acute febrile illness/suspected cases of Dengue and/or Chikungunya and/or Zika virus infection)
  6. Intended user(laboratory professional and/or health care worker at point-of-care)
  7. Detailed test protocol
  8. Lot/batch No.
  9. Date of manufacture
  10. Date of Expiry
  11. Information on operational Characteristics i. Configuration of the kit/device ii. Requirement of any additional equipment, device iii. Requirement of any additional reagents iv. Operation conditions v. Storage and stability before and after opening vi. Internal control provided or not vii. Quality control and batch testing data viii. Biosafety aspects- waste disposal requirements
  12. Information on Test Performance Characteristics i. Type of sample-serum/plasma/whole blood/other specimen (specify) ii. Volume of sample iii. Sample rejection criteria (if any) iv. Any additional sample processing required v. Any additional device/consumable like sample transfer device, pipette, tube, etc required vi. Name of analyte to be detected vii. Pathogens targeted by the kit

Page 198 of 459

viii. Time taken for testing ix. Time for result reading and interpretation x. Manual or automated(equipment)reading xi. Limit of detection/ Limit of Quantification and rage of detection xii. Diagnostic sensitivity xiii. Diagnostic specificity xiv. Stability and reproducibility (including data) xv. Training required for testing (if any) xvi. If yes, duration xvii. Details of Cut-off and /or Equivocal Zone for interpretation of test xviii. Details of cross reactivity, if any xix. Interpretation of invalid and indeterminate results to be provided xx. It is recommended to provide data demonstrating accuracy and precision

*Please mention “Not applicable” against sections not pertaining to the kit.

Page 199 of 459

INFLUENZA VIRUS, SARS-COV-2, RESPIRATORY SYNCYTIAL VIRUS IN-VITRO DIAGNOSTICS

Page 200 of 459

A. Working Group:

  1. Dr. Varsha Potdar, Scientist-E, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  2. Dr. Biswajyoti Borkakoty, Scientist-F, ICMR-Regional Medical Research Center, Dibrugarh, Department of Health Research, Ministry of Health and Family

  3. Dr. Shantala GB, Professor of Microbiology, Bangalore Medical College and Research Institute, Bengaluru

  4. Dr. Rizwan SA, Scientist-D, ICMR-National Institute of Epidemiology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  5. Dr. Labanya Mukhopadhyay, Scientist-C, Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  6. Dr. Mandeep Chadha, Former Scientist-G, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  7. Dr. Amita Jain, Former Professor and Head of the Department of Microbiology, King George’s Medical University, Lucknow, Uttar Pradesh

  8. Dr. Lalit Dar, Professor and Head of the Department of Microbiology, All India Institute of Medical Sciences, Delhi

  9. Dr. Manoj Murhekar, Scientist-G and Director, ICMR-National Institute of Epidemiology, Department of Health Research, Ministry of Health and Family

  10. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control Organization, Directorate General of Health Services, Ministry of Health and Family Welfare, Government of India

  11. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control Organization, Directorate General of Health Services, Ministry of Health and Family Welfare, Government of India

  12. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central

  13. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

Page 201 of 459

General Guidelines Protocols for performance evaluation of in vitro molecular diagnostic kits for detection and differentiation of Influenza virus and/or SARS-CoV-2 and/or RSV

  1. Introduction: This document provides a framework for evaluating the performance characteristics of in vitro diagnostic (IVD) kits used in identifying and distinguishing various strains of Influenza viruses and/or SARS-CoV-2 and/or RSV, aligning with international standards to ensure reliability and accuracy in diagnosis. The coronavirus diseases 2019 (COVID-19) pandemic, caused by the SARS-CoV-2 virus, has necessitated the rapid development and validation of in vitro molecular diagnostic kits. These kits are crucial for the timely detection and differentiation of major respiratory viruses (influenza virus and/or its strains/SARS-CoV- 2/RSV) to control their spread. This protocol outlines a systematic approach for validating these diagnostic kits to ensure their accuracy, sensitivity, specificity, and reliability. Although SARS-CoV -2 is no longer a public health emergency globally, it is prudent to implement integrated surveillance for Influenza, SARS-CoV-2 and other respiratory viruses, making differential diagnosis for these viruses essential. Additionally, timely diagnosis of other respiratory viruses, particularly Respiratory syncytial virus (RSV), is crucial for providing effective clinical management to pediatric cases.
    This document provides guidance for single plex or multiplex assays for the differential diagnosis of Influenza and/or its strains) and/or SARS-CoV-2 and/or RSV. It outlines the evaluation of IVD kits intended for the detection and differentiation of influenza virus strains and/or detection of SARS-CoV-2 and/or detection and differentiation of RSV using nucleic acid detection methods as outlined in the scope below. This includes IVD kits that detect and differentiate between influenza virus types (Influenza A or B), subtypes (A (H1N1) pdm09 or A (H3N2)), and/or multiple influenza virus types/subtypes; kits that identify only SARS-CoV- 2, as well as kits that only detect and/or differentiate RSV. Additionally, this protocol may be used for multiplex IVD kits designed to simultaneously detect Influenza A & B (with or without subtyping), and/or SARS-CoV-2, and/or RSV. This document outlines the following aspects of performance evaluation of IVD kits as per the scope outlined in the document: 1.1 The procedure for validating entities to determine operational parameters of IVD kits that detect influenza virus gene segment(s). 1.2 T procedure for validating entities to determine operational parameters of IVD kits that detect SARS-CoV-2 gene segment(s).
    1.3 The procedure for validating entities to determine operational parameters of IVD kits that detect RSV gene segment(s).
    1.4 The techniques for identifying influenza virus/SARS-CoV-2/RSV nucleic acid targets in single-plex or multiplex formats (using appropriate protocols listed in the document).

Page 202 of 459

1.5 This document does not cover performance evaluation of serological assays for detection of antigen and antibody for influenza viruses/SARS-CoV-2/RSV. The IVD kit to be validated is henceforth known as the “Kit under Evaluation.” 2. Objective:
This document aims to offer a comprehensive set of instructions for evaluating the performance of molecular IVD assays mentioned in the scope below for detecting Influenza A and Influenza B viruses with/without subtyping, and other common respiratory viruses such as SARS-CoV-2 and RSV. This evaluation will focus on measuring the analytical sensitivity and specificity, cross-reactivity, repeatability, and reproducibility as compared against a reference assay using clinical sample panel. In brief, the objectives are as follows: 2.1 To validate the performance characteristics of in vitro molecular diagnostic kits for detecting Influenza A & B (with/without subtyping)/ SARS-CoV-2/ RSV. 2.2 To ensure the kits under evaluation meet the necessary standards for sensitivity, specificity, repeatability, and reproducibility. 2.3 To evaluate the cross-reactivity of the kits with other respiratory viruses. 3. Scope: This guideline is solely for the evaluation and establishment of the performance characteristics of IVD kits designed for the detection and subtyping of commonly circulating seasonal Influenza viruses (Influenza A(H1N1) pdm09, Influenza A(H3N2), Influenza B(Yamagata) and Influenza B(Victoria) subtypes) and/or other common respiratory viruses such as SARS- CoV-2 and RSV, using single or multiplex molecular assays (as outlined in the scope below) intended for human clinical samples. This document is a guide to assess: 3.1 The analytical assay performance characteristics with clinical specimens for the detection and/or differentiation of influenza viruses. (Protocol A)
3.2 The analytical assay performance characteristics with clinical specimens for the detection of SARS-CoV-2 (Protocol B)
3.3 The analytical assay performance characteristics with clinical specimens for the detection of RSV (Protocol C)
3.4 The analytical performance characteristics of multiplex assay for detection of two or more of these viruses by combining Protocols A, B & C as per the kit format.
3.5 Analytical performance characteristics which should include sensitivity, specificity, cross-reactivity, and lot-to-lot variation including functionality of devices that identify and/or differentiate influenza viruses, SARS-CoV-2 and/or RSV depending on the kit format. 3.6 The performance of the kit, only if the kit includes an internal control (preferably endogenous, or exogenous).
3.7 This document may also apply to forthcoming influenza, SARS-CoV-2 and RSV molecular diagnostic kits that do not fit within these current classifications.

Page 203 of 459

3.8 The document will serve as a reference for assessing kits based on Nucleic Acid Amplification Test (single plex or multiplex assays) as listed below: 3.8.1 Real-time Reverse Transcription Polymerase Chain Reaction format (rRT-PCR): including Real-time PCR probe-based assays or non-probe based assays 3.8.2 Other NAT testing platforms such as LAMP/RPA, and other closed system platforms such as TrueNat /cartridge-based assays

Note: This protocol is not suitable for the kits where amplicons are handled outside the amplification system.

  1. Requirements:
    4.1 Supply of kits under evaluation (Along with batch/lot No. Expiry & required details). supply the required equipment and consumables. 4.2 Evaluation sites/laboratories (With required equipment) 4.4 Reference test kits 4.4 Characterized clinical samples for evaluation panel 4.5 Laboratory supplies

  2. Ethical approvals:
    Laboratory validation of IVDs using irreversibly de-identified samples is exempted from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024. A self-declaration form as provided in ICMR guidelines to be submitted by the investigators to the institutional authorities and ethics committee for information (https://ethics.ncdirindia.org/asset/pdf/Guidance_on_Ethical_Requirements_for_Laborat ory_Validation_Testing.pdf )

  3. Procedure: 6.1 Study design/type: Diagnostic accuracy study using leftover irreversibly de- identified archived clinical samples.
    6.2 Evaluation site/laboratory considerations: Identified IVD kit evaluation laboratories should establish their proficiency through
    6.2.1 Accreditation for at least one of the Quality management systems (accreditation for Testing Lab / Calibration Lab (ISO/IEC: 17025), Medical Lab (ISO 15189), PT provider (ISO/IEC: 17043) or CDSCO approved Reference laboratory. 6.2.2 Have sufficient numbers of archived as well as contemporary (collected and tested within the preceding 1 year) clinical specimens positive for respiratory viruses targeted by the kit under evaluation (Influenza A(H1N1)pdm09, A(H3N2), B(Yamagata), B(Victoria), and/or SARS-CoV-2 and/or RSV A & B), with aliquots stored at -80 ℃ deep freezers or in lyophilized form.

Page 204 of 459

6.2.3 Virus strains should be well-characterized by ICMR approved or US FDA/ ATAGI Australia/PMDA Japan approved/WHO Pre-Qualified reference assay and/or by influenza virus HA gene/segment or gene-specific sequencing (for SARS-CoV-2 and RSV) or Next-Generation Sequencing. 6.2.4 Have a minimum BSL-2 level facility with trained manpower and at least two different Real Time platforms to perform molecular diagnostic assays for Influenza virus and other respiratory viruses. 6.2.5 Have a good record of External Quality Assurance programs for influenza, SARS-CoV-2, and other respiratory viruses. 6.2.6 Staff training: All the staff involved in IVD kit evaluation should undergo hands-on training and competency testing on the following: 6.2.6.1 Preparation & characterization of kit evaluation panel
6.2.6.2 Handling of respiratory virus PCR kits received for performance evaluation (Verification/Storage/Unpacking etc). 6.2.6.3 Testing, interpreting, recording of results & reporting

6.2.6.4 Data handling, data safety & confidentiality 6.3 Performance characteristics: To be assessed for all assay targets of influenza A/B, SARS-CoV-2 and RSV (single plex or multi-plex assays) 6.3.1 Analytical Sensitivity and specificity
6.3.2 Cross-reactivity
6.3.3 Repeatability 6.3.4 Reproducibility

Page 205 of 459

Protocol A Performance Evaluation of Molecular IVD Kit detecting influenza A & B viruses, and subtyping into A (H1N1) pdm 09, A(H3N2), B(Yamagata) & B(Victoria) in single plex or multiplex format

  1. Objective:
    1.1 To evaluate the performance of molecular IVD kit for detection and differentiation of Influenza viruses as per the scope outlined in this document.
    1.2 To ensure the kits under evaluation meet the necessary standards for sensitivity, specificity, repeatability, and reproducibility.
    1.3 To evaluate the cross-reactivity of the kits with other respiratory viruses.

  2. Evaluation of performance characteristics should be done for the following parameters: 2.1 Sensitivity and specificity 2.2 Cross-reactivity 2.3 Repeatability 2.4 Reproducibility

  3. Panel development: Clinical sample (archived/contemporary) panel for testing: 3.1 Contemporary (collected and tested within the preceding 1 year) leftover irreversibly de-identified clinical/archived respiratory samples (in VTM) for the panel should be irreversibly de-identified. 3.2 Samples to be used for panel preparation shall be stored properly at – 80 ℃ or lyophilized. 3.3 Unless the manufacturer has specific requirement of nucleic acid extraction kit, the validation laboratory can use WHO Pre-Qualified/ US FDA/ ATAGI Australia/ PMDA Japan approved/ICMR validated total RNA / viral RNA extraction kits for the evaluation.
    3.4 Clinical samples for evaluation should be characterized by a reference kit / Sequencing/NGS. 3.5 All positive samples should be confirmed positive for the target pathogens by the reference assay. 3.6 All negative samples should be confirmed negative for the target pathogens by the reference assay.

Page 206 of 459

  1. Sample size and sample panel composition for evaluation of performance characteristics: Sample sizes of positive and negative samples of the analyte/pathogen targeted by the kit against different values of sensitivity and specificity are provided in Table 1. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity. Sample sizes are calculated using the formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· n (se) is the number of positive samples. · n (sp) is the number of negative samples.

Sample sizes for positive samples and their composition for evaluating subtyping are provided in Table 2.

Table 1. Sample sizes per target pathogen for different values of sensitivity/ specificity claimed by the manufacturer.

Page 207 of 459

Sensitivity/ Specificity Sample size: Minimum number of positive samples¥ Composition of positive samples# Sample size: Minimum number of negative samples (rounded) ¥ Minimum number of cross reactive* samples among the negative samples 99% 16 (rounded to 20 for better distribution of samples) Strong positive = 06 Moderate positive = 07 Weak positive = 07 20 5 95% 77 (rounded to 80 for better Strong positive = 24 Moderate positive = 28 Weak positive = 28 80 20 90% 146 (rounded to 155 for better Strong positive = 45 Moderate positive = 55 Weak positive =55 150 38 85% 207 (rounded to 215 for better Strong positive = 63 Moderate positive = 76 Weak positive = 76 210 53 80% 259 (rounded to 260 for better Strong positive = 78 Moderate positive = 91 Weak positive = 91 260 65 #Strong positive: (Ct value <25) Moderate positive: (Ct value between 25-30) Weak positive: (Ct value >30 and ≤ 34)

¥ Equal distribution of positive nasopharyngeal and/or oropharyngeal swabs in virus transport medium (VTM) to be used

  • Samples positive for common respiratory viruses (such as SARS-CoV-2, Parainfluenza viruses, Adenoviruses, Rhinoviruses, Respiratory Syncytial Virus (including its types and subtypes), common human coronaviruses), other than the ones targeted by the kit under evaluation. Equal distribution of cross-reactive viruses is desirable. Commercially available validated standard panels that are accepted by accreditation agencies can also be used if clinical samples with required target is not available for cross-reactivity analysis. It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. Table 2. Sample sizes for positive samples and their composition for evaluating subtyping
    Sample size* (per target pathogen) Minimum total number of positive samples (rounded figures) Influenza A (H1N1) pdm09 Influenza A/H3N2 Influenza B (Victoria) Sensitivity Minimum number of nasopharyngeal swabs/ oropharyngeal swabs (rounded figures) Minimum number of nasopharyngeal swabs/ oropharyngeal swabs (rounded figures) Minimum number of nasopharyngeal swabs/ oropharyngeal swabs (rounded figures) 99% 20 20 20 60 95% 80 80 80 240 90% 150 150 150 450

Page 208 of 459

85% 210 210 210 630 80% 260 260 260 780 *Combination of strong, moderate and weak positive samples should be considered as per the information provided in Table 1. It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. Influenza B (Yamagata) is not mentioned in the table in view nil/minimal global and national circulation of the strain for the past few years. Samples positive for influenza B (Yamagata) should be used for performance evaluation of kits if the strain starts circulating in the country. 5. Methodology: 5.1 Samples should be tested in parallel with the Kit Under Evaluation and the reference assay. The ICMR-NIV RT-qPCR assay for Influenza/SARS-CoV-2 or WHO Pre- Qualified/ US FDA/ ATAGI Australia/ PMDA Japan approved kit will be considered as the reference assay for these parameters. 5.2 The validation laboratory can use the established total RNA / viral RNA extraction protocol for the evaluation.
5.3 The instruction for the assay setup and the interpretation of the results will be as per the protocol outlined by the manufacturer of the reference test and the kit under evaluation.
5.4 The results shall be compared with the reference assay for sensitivity and specificity calculations. 5.5 If there is a discrepancy observed in the results with the index test, this discrepancy should be taken as discordant. Repetition of the assay may introduce bias. If the reference kit itself has failed, then these samples with discrepancies should be discarded, and new well-characterized samples should be used instead.
True positive samples: These are samples positive by both reference assay and index test. True negative samples: These are samples negative by both reference assay and index test. False positive samples: These are samples negative by reference assay and positive by index test. False negative samples: These are samples positive by reference assay and negative by index test. 5.6 The interpretation for internal control (preferably endogenous, or exogenous) will be as per manufacturer's instruction.
5.7 PCR should be performed using IVD-approved machines. If any equipment(s) is specified in the IFU of the index test, it should be used for the evaluation, and it should be provided by the manufacturer if not available within the lab’s IVD evaluation scope.

Page 209 of 459

Real-time closed systems/devices awaiting evaluation should be provided by the manufacturer along with all necessary components, supplies and reagents. The details on the Real-time Equipment used for validation should be recorded, including calibration status.

  1. Cross-reactivity Analysis: 6.1 Objective:
    To assess the primer-probe set for true detection of influenza viruses and assess its cross-reactivity with other respiratory viruses. 6.2 Methodology: 6.2.1 Potential cross-reactivity of the kit shall be ruled out by testing other respiratory pathogen positive samples as part of the negative sample panel, which is outlined in Table 1.* 6.2.2 Cross-reactivity will be assessed by comparing the results of these samples using kit under evaluation and reference kit.
  • For multiplex assays targeting influenza, SARS-CoV-2, and RSV, samples positive for these viruses may be suitably interchanged for assessing cross-reactivity, apart from the ORV panel. (i.e. Influenza A positive samples may be used for detecting cross- reactivity against Influenza B)
  1. Acceptance criteria for the kit:
    Sensitivity for each pathogen/ type/ subtype: ≥95%
    Specificity for each pathogen/ type/ subtype: ≥99%

To achieve at least the performance characteristics outlined in the acceptance criteria, ≥80 positive samples and ≥20 negative samples should be tested for evaluation for each pathogen/ type/ subtype.

  1. Repeatability and Reproducibility Assessment:

evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples per target should be tested 5 times in independent runs.

Page 210 of 459

be as follows:
negative samples per target).
10 low positive AND 5 moderate/strong positive samples, and 10 negative samples per target).  There should be no lot-to-lot variability (qualitative).

moderate and weak positive samples) and 3 negative samples per target should be (qualitative).

and 3 negative samples per target should be tested 5 times in independent runs.
(qualitative).

samples per target should be tested 5 times in independent runs.
(qualitative).

recommended).

Page 211 of 459

  1. Internal Control Analysis: 9.1 Monitor the internal control (preferably RNaseP or other housekeeping gene) to ensure consistent extraction and amplification efficiency across samples and runs. 9.2 Ct-values of internal controls should be within the manufacturer’s prescribed limit. 9.3 Tests will be marked invalid if Ct-values are outside the prescribed limit.

  2. Blinding of Laboratory Staff:

PI of the evaluating lab, and the PI should maintain confidentiality of data. Refer to Fig. 1.

Fig.1: Blinding in evaluation exercise

  1. Publication Rights:
    author(s).

Page 212 of 459

  1. Conclusion: Based on the comprehensive evaluation conducted, the [Kit & Manufacturer’s Name] Influenza Virus molecular IVD kit has been found [Satisfactory/Not Satisfactory] for its intended in vitro diagnostic (IVD) use.
    The assay demonstrates [Strengths/Concerns] in terms of sensitivity, specificity, and performance characteristics compared to established reference IVD approved RT-PCR kits. acceptable.

  2. Performance evaluation report format The performance evaluation report format (page 34) is designed for multiplex assays with several targets. It should be modified and used accordingly for single plex assays/multiplex assays with fewer targets.

Page 213 of 459

Protocol B: Performance Evaluation of Molecular IVD Kit detecting SARS-CoV-2 in single plex or multiplex format

  1. Objective: 1.1. To validate the performance characteristics of in vitro molecular diagnostic kits for detecting SARS-CoV-2 as per the scope outlined in this document. 1.2. To ensure the kits under evaluation meet the necessary standards for sensitivity, specificity, repeatability, and reproducibility. 1.3. To evaluate the cross-reactivity of the kits with other respiratory viruses.

  2. Evaluation of Performance characteristic should be done for the following:
    2.1 Sensitivity and specificity 2.2 Cross-reactivity
    2.3 Repeatability 2.4 Reproducibility

  3. Panel development: Clinical sample (archived/ contemporary) panel for testing: 3.1 Contemporary (collected and tested within the preceding 1 year) leftover irreversibly de-identified clinical/archived respiratory samples in VTM for the panel should be irreversibly de-identified. 3.2 Samples to be used for panel preparation shall be stored properly at – 80 ℃ or lyophilized. 3.3 Unless the manufacturer has specific requirement of nucleic acid extraction kit, the MDTLs/ validation laboratory can use WHO Pre-Qualified/ US FDA/ ATAGI Australia/ PMDA Japan approved/ICMR validated an established total RNA / viral RNA extraction kits for the evaluation.
    3.4 Clinical samples for evaluation should be characterized by a reference kit / Sequencing/NGS. 3.5 All positive samples should be confirmed positive for the target pathogens by the reference assay. 3.6 All negative samples should be confirmed negative for the target pathogens by the reference assay.

  4. Sample size and sample panel composition for evaluation of performance characteristics:

Page 214 of 459

Sample sizes of positive and negative samples of SARS-CoV-2 against different values of sensitivity and specificity are provided in Table 3. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ a claim of 87% specificity would require usage of the sample size that is outlined for 85% specificity. Sample sizes are calculated using the formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· n (se) is the number of positive samples. · n (sp) is the number of negative samples.

Table 3. Sample sizes for different values of sensitivity/ specificity claimed by the manufacturer. Sensitivity/ Specificity Sample size: Minimum number of positive samples¥ Composition of positive samples# Sample size: Minimum number of negative samples (rounded)¥ Minimum number of cross reactive* samples among the negative samples 99% 16 (rounded to 20 for better Strong positive = 06 Moderate positive = 07 Weak positive = 07 20 5

Page 215 of 459

95% 77 (rounded to 80 for better Strong positive = 24 Moderate positive = 28 Weak positive = 28 80 20 90% 146 (rounded to 155 for better distribution of samples) Strong positive = 45 Moderate positive = 55 Weak positive =55 150 38 85% 207 (rounded to 215 for samples) Strong positive = 63 Moderate positive = 76 Weak positive = 76 210 53 80% 259 (rounded to 260 for samples) Strong positive = 78 Moderate positive = 91 Weak positive = 91 260 65 #Strong positive: (Ct value <25) Moderate positive: (Ct value between 25-30) Weak positive: (Ct value >30 and ≤ 34)

¥ Nasopharyngeal/ oropharyngeal swabs in virus transport medium (VTM) to be used *Samples positive for common respiratory viruses (such as Influenza (including its types and subtypes), Parainfluenza viruses, Adenoviruses, Rhinoviruses, Respiratory Syncytial Virus (including its types and subtypes), common human coronaviruses), other than the ones targeted by the kit under evaluation. Equal distribution of cross-reactive viruses is desirable. Commercially available validated standard panels that are accepted by accreditation agencies can also be used if clinical samples with required target is not available for cross-reactivity analysis. It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. 4.1 Repeatability testing will be performed on 3 positive (strong, moderate and weak positive) and 3 negative samples (within the selected positive and negative samples) per target pathogen 5 times (replicates of 5).
5. Methodology: 5.1 Samples should be tested in parallel with the Kit Under Evaluation and the reference assay. The ICMR-NIV RT-qPCR assay for Influenza/SARS-CoV-2 or WHO Pre- Qualified/ US FDA/ PMDA Japan/ ATAGI Australia approved kit will be considered as the reference assay for these parameters. 5.2 The validation laboratory can use established total RNA / viral RNA extraction protocol for the evaluation.
5.3 The instruction for the assay setup and the interpretation of the results will be as per the protocol outlined by the manufacturer of the reference test and the kit under evaluation. The results shall be compared with the reference assay for sensitivity and specificity calculations. 5.4 If there is a discrepancy observed in the results with the index test, this discrepancy should be taken as discordant. Repetition of the assay may introduce bias. If the reference kit itself has failed, then these samples with discrepancies should be discarded, and new well-characterized samples should be used instead.
True positive samples: These are samples positive by both reference assay and index test. True negative samples: These are samples negative by both reference assay and index test.

Page 216 of 459

False positive samples: These are samples negative by reference assay and positive by index test. False negative samples: These are samples positive by reference assay and negative by index test.

5.5 The interpretation for internal control (preferably endogenous, or exogenous) will be as per manufacturer's instruction.
5.6 PCR should be performed using IVD-approved machines. If any equipment(s) is specified in the IFU of the index test, it should be used for the evaluation, and it should be provided by the manufacturer if not available within the lab’s IVD evaluation scope.
The details on the Real-time Equipment used for validation should be recorded, including calibration status. The details on the Real-time Equipment used for validation should be recorded including calibration status.

  1. Cross-reactivity Analysis: 6.1 Objective:
    To assess the primer-probe set for true detection of SARS-CoV-2 and assess its cross- reactivity with other respiratory viruses. 6.2 Methodology:

6.2.1 Potential cross-reactivity of the kit shall be ruled out by testing other respiratory pathogen positive samples as part of the negative sample panel, as outlined in Table 3.*
6.2.2 Cross-reactivity will be assessed by comparing the results of these samples using kit under evaluation and reference kit.

  • For multiplex assays targeting influenza, SARS-CoV-2, and RSV detection, samples positive for these viruses may be suitably interchanged for assessing cross-reactivity
  1. Acceptance criteria for the kit:
    Sensitivity: ≥95%
    Specificity: ≥99%

To achieve at least the performance characteristics outlined in the acceptance criteria, ≥80 positive samples and ≥20 negative samples should be tested for evaluation for each pathogen/ type/ subtype.

Page 217 of 459

  1. Repeatability and Reproducibility Assessment:

This should be done to assess the repeatability of the detection of target using the kit under evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples per target should be tested 5 times in independent runs.

be as follows:
negative samples per target).
10 low positive AND 5 moderate/strong positive samples, and 10 negative samples per target).  There should be no lot-to-lot variability (qualitative).

moderate and weak positive samples) and 3 negative samples per target should be (qualitative).

and 3 negative samples per target should be tested 5 times in independent runs.
(qualitative).

samples per target should be tested 5 times in independent runs.
(qualitative).

Page 218 of 459

recommended).

  1. Internal Control Analysis: 9.1 Monitor the internal control (preferably RNaseP or other housekeeping gene) to ensure consistent extraction and amplification efficiency across samples and runs.

9.2 Ct-values of internal controls should be within the manufacturer’s prescribed limit.

9.3 Tests will be marked invalid if Ct-values are outside the prescribed limit.

  1. Blinding of Laboratory Staff:

To ensure rigor of the evaluation process, laboratory staff performing the evaluation should be blinded to the status of the clinical samples. The PI of the evaluation exercise should remain unblinded, i.e., privy to the status of the samples. Another senior laboratory staff selected by the PI may remain unblinded and carry out coding of samples and dispensing them into similar-looking vials to be used for testing, and maintaining the database of results. Staff performing the reference test and the test under evaluation, interpretation of the test result, and entering the results against the coded samples in the database, should remain blinded to the status of samples till the completion of evaluation. The data should be analyzed only by the PI of the evaluating lab, and the PI should maintain confidentiality of data. Refer to Fig. 1 in Protocol A.

  1. Conclusion: Based on the comprehensive evaluation conducted, the [Kit & Manufacturer’s Name] SARS-CoV-2 molecular IVD kit has been found [Satisfactory/Not Satisfactory] for its intended in vitro diagnostic (IVD) use.
    The assay demonstrates [Strengths/Concerns] in terms of sensitivity, specificity, and performance characteristics compared to established reference IVD approved RT-PCR kits. author(s). acceptable.

Page 219 of 459

  1. Performance evaluation report format: The performance evaluation report format (page 34) is designed for multiplex assays with several targets. It should be modified and used accordingly for single plex assays/multiplex assays with fewer targets.

Page 220 of 459

Protocol C: Performance Evaluation of Molecular IVD Kit detecting Respiratory Syncytial Virus (RSV /RSV A/RSV B) in single plex or multiplex format

  1. Objective: 1.1 To validate the performance characteristics of in vitro molecular diagnostic kits for detecting and/or differentiating RSV A/B as per the scope outlined in this document. 1.2 To ensure the kits under evaluation meet the necessary standards for sensitivity, specificity, repeatability, and reproducibility. 1.3 To evaluate the cross-reactivity of the kits with other respiratory viruses.
  2. Evaluation of Performance characteristic should be done for the following:
    2.1 Sensitivity and specificity 2.2 Cross-reactivity
    2.3 Repeatability 2.4 Reproducibility
  3. Panel development: Clinical sample (archived/ contemporary) panel for testing: 3.1 Contemporary (collected and tested within the preceding 1 year) leftover irreversibly de-identified clinical/archived respiratory samples in VTM for the panel should be irreversibly de-identified. 3.2 Samples to be used for panel preparation shall be stored properly at – 80 ℃ or lyophilized. 3.3 Unless the manufacturer has specific requirement of nucleic acid extraction kit, the MDTLs/ validation laboratory can use WHO Pre-Qualified/ US FDA/ ATAGI Australia/ PMDA Japan approved/ ICMR validated an established total RNA / viral RNA extraction kits for the evaluation.
    3.4 Clinical samples for evaluation should be characterized by a reference kit / Sequencing/NGS. 3.5 All positive samples should be confirmed positive for the target pathogens by the reference assay. 3.6 All negative samples should be confirmed negative for the target pathogens by the reference assay.
  4. Sample size and sample panel composition for evaluation of performance characteristics:
    Sample sizes of positive and negative samples of the RSV A/B against different values of sensitivity and specificity are provided in Table 4. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%.

Page 221 of 459

Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ a claim of 87% specificity would require usage of the sample size outlined for 85% specificity. Sample sizes for positive samples and their composition for evaluating subtyping (RSV A/B) are provided in Table 5. Sample sizes are calculated using the formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· n (se) is the number of positive samples. · n (sp) is the number of negative samples.

Table 4. Sample sizes per target pathogen (RSV A/B) for different values of sensitivity/ specificity claimed by the manufacturer. Sensitivity/ Specificity Sample size: Minimum number of positive samples¥ Composition of positive samples# Sample size: Minimum number of negative samples (rounded)¥ Minimum number of cross reactive* samples among the negative samples 99% 16 (rounded to 20 for better Strong positive = 06 Moderate positive = 07 Weak positive = 07 20 5 95% 77 (rounded to 80 for better Strong positive = 24 Moderate positive = 28 Weak positive = 28 80 20

Page 222 of 459

90% 146 (rounded to 155 for samples) Strong positive = 45 Moderate positive = 55 Weak positive =55 150 38 85% 207 (rounded to 215 for samples) Strong positive = 63 Moderate positive = 76 Weak positive = 76 210 53 80% 259 (rounded to 260 for samples) Strong positive = 78 Moderate positive = 91 Weak positive = 91 260 65 #Strong positive: (Ct value <25) Moderate positive: (Ct value between 25-30) Weak positive: (Ct value >30 and ≤ 34)

¥ Nasopharyngeal/ oropharyngeal swabs in virus transport medium (VTM) to be used Samples positive for common respiratory viruses (such as Influenza (including its types and subtypes), SARS-CoV-2, Parainfluenza viruses, Adenoviruses, Rhinoviruses, common human coronaviruses), other than the ones targeted by the kit under evaluation. Equal distribution of cross-reactive viruses is desirable. Commercially available validated standard panels that are accepted by accreditation agencies can also be used if clinical samples with required target is not available for cross-reactivity analysis. It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. Table 5. Sample sizes for positive samples and their composition for evaluating subtyping
Sensitivity Sample size
(per target pathogen) RSV A RSV B Minimum total positive samples Minimum number of nasopharyngeal swabs/ oropharyngeal swabs Minimum number of nasopharyngeal swabs/ oropharyngeal swabs 99% 20 20 20 40 95% 80 80 80 160 90% 150 150 150 300 85% 210 210 210 420 80% 260 260 260 520 *Combination of strong, moderate and weak positive samples should be considered as per the information provided in Table 4. It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. 4.1 Repeatability testing will be performed on 3 positive (strong, moderate and weak positive) and 3 negative samples (within the selected positive and negative samples) per target pathogen 5 times (replicates of 5).
5. Methodology: 5.1 Samples should be tested in parallel with the Kit Under Evaluation and the reference assay. The ICMR-NIV RT-qPCR assay for RSV or WHO Pre-Qualified/ US FDA/

Page 223 of 459

ATAGI Australia/ PMDA Japan approved kit will be considered as the reference assay for these parameters. 5.2 The validation laboratory can use established total RNA / viral RNA extraction protocol for the evaluation.
5.3 The instruction for the assay setup and the interpretation of the results will be as per the protocol outlined by the manufacturer of the Kit Under Evaluation.
5.4 The results shall be compared with the reference assay for sensitivity and specificity calculations. 5.5 If there is a discrepancy observed in the results with the index test, this discrepancy should be taken as discordant. Repetition of the assay may introduce bias. If the reference kit itself has failed, then these samples with discrepancies should be discarded, and new well-characterized samples should be used instead.
True positive samples: These are samples positive by both reference assay and index test. True negative samples: These are samples negative by both reference assay and index test. False positive samples: These are samples negative by reference assay and positive by index test. False negative samples: These are samples positive by reference assay and negative by index test. 5.6 The interpretation for internal control (preferably endogenous, or exogenous) will be as per manufacturer's instruction.
5.7 PCR should be performed using IVD-approved machines. If any equipment(s) is specified in the IFU of the index test, it should be used for the evaluation, and it should be provided by the manufacturer if not available within the lab’s IVD evaluation scope.
Real-time closed systems/devices awaiting evaluation should be provided by the manufacturer along with all necessary components, supplies and reagents. The details on the Real-time Equipment used for validation should be recorded, including calibration status. The details on the Real-time Equipment used for validation should be recorded including calibration status. 6. Cross-reactivity Analysis: 6.1 Objective:
To assess the primer-probe set for true detection of RSV and assess its cross-reactivity with other respiratory viruses. 6.2 Methodology: 6.2.1 Potential cross-reactivity of the kit shall be ruled out by testing other respiratory pathogen positive samples as part of the negative sample panel, as outlined in Table 4.* 6.2.2 Cross-reactivity will be assessed by comparing the results of these samples using kit under evaluation and reference kit.

Page 224 of 459

  • For multiplex assays targeting influenza, SARS-CoV-2, and RSV detection, samples positive for these viruses may be suitably interchanged for assessing cross-reactivity
  1. Acceptance criteria for the kit:

Sensitivity for each pathogen/ type/ subtype: ≥95%
Specificity for each pathogen/ type/ subtype: ≥99%
To achieve at least the performance characteristics outlined in the acceptance criteria, ≥80 positive samples and ≥20 negative samples should be tested for evaluation for each pathogen/ type/ subtype. 8. Repeatability and Reproducibility Assessment:

evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples per target should be tested 5 times in independent runs.

be as follows:
negative samples per target).
10 low positive AND 5 moderate/strong positive samples, and 10 negative samples per target).  There should be no lot-to-lot variability (qualitative).

Page 225 of 459

moderate and weak positive samples) and 3 negative samples per target should be (qualitative).

and 3 negative samples per target should be tested 5 times in independent runs.
(qualitative).

samples per target should be tested 5 times in independent runs.
(qualitative) should be 100% based on positive and negative test result (qualitative).

recommended).

  1. Internal Control Analysis: 9.1 Monitor the internal control (RNaseP or other endogenous housekeeping gene) to ensure consistent extraction and amplification efficiency across samples and runs. 9.2 Ct-values of internal controls should be within the manufacturer’s prescribed limit. 9.3 Tests will be marked invalid if Ct-values are outside the prescribed limit.

  2. Blinding of Laboratory Staff:

To ensure rigor of the evaluation process, laboratory staff performing the evaluation should be blinded to the status of the clinical samples. The PI of the evaluation exercise should remain unblinded, i.e., privy to the status of the samples. Another senior laboratory staff selected by the PI may remain unblinded and carry out coding of samples and dispensing them into similar-looking vials to be used for testing, and maintaining the database of results. Staff performing the reference test and the test under evaluation, interpretation of the test result, and entering the results against the coded samples in the database, should remain blinded to the status of samples till the completion of evaluation. The data should be analyzed only by the PI of the evaluating lab, and the PI should maintain confidentiality of data. Refer to Fig. 1 in Protocol A.

Page 226 of 459

  1. Conclusion: Based on the comprehensive evaluation conducted, the [Kit & Manufacturer’s Name] RSV molecular IVD kit has been found [Satisfactory/Not Satisfactory] for its intended in vitro diagnostic (IVD) use.
    The assay demonstrates [Strengths/Concerns] in terms of sensitivity, specificity, and performance characteristics compared to established reference IVD approved RT-PCR kits. author(s). acceptable.
  2. Performance evaluation report format: The performance evaluation report format (page 34) is designed for multiplex assays with several targets. It should be modified and used accordingly for single plex assays/multiplex assays with fewer targets.

Page 227 of 459

Protocol D: Performance Evaluation of Molecular IVD Kit detecting Influenza virus and SARS- CoV-2 in multiplex format

To assess the performance of multiplex assays, Protocols A and B can be used as per kit format to check the performance of each virus for its sensitivity and specificity assessment, including cross reactivity, repeatability, reproducibility and Lot to lot variation.
A comprehensive report can be generated which will include sensitivity and specificity for all targets.
Sample size for multiplex molecular assay (as per the scope outlined in the document) detecting Influenza virus and SARS-CoV-2 in multiplex format is given below. All other relevant performance criteria (such as repeatability/ reproducibility/ cross-reactivity) outlined in the single plex protocols (Protocols A and B) are to be essentially followed and met for each target pathogen/type/subtype.

  1. Sample size and sample panel composition for evaluation of performance characteristics:

Sample sizes of positive and negative samples against different values of sensitivity and specificity are provided in Table 6. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity. Sample sizes are calculated using the formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· n (se) is the number of positive samples. · n (sp) is the number of negative samples.

Page 228 of 459

Table 6. Sample sizes for different values of sensitivity/ specificity claimed by the manufacturer. Sensitivit y/ Specificit y Sample size for each of the 04 target pathogensa: Minimum number of positive samples¥ Composition of positive samples for each pathogen/type/subtype# Total number of positive samples (includin g all 04 pathogen s) Sample size: Minimum number of negative samples¥ Minimum number of cross reactive* samples among the negative samples 99% 16 (rounded to 20 for samples) Strong positive = 06 Moderate positive = 07 Weak positive = 07 80 20 5 95% 77 (rounded to 80 for samples) Strong positive = 24 Moderate positive = 28 Weak positive = 28 320 80 20 90% 146 (rounded to 155 for better distribution of samples) Strong positive = 45 Moderate positive = 55 Weak positive =55 620 150 38 85% 207 (rounded to 215 for better distribution of samples) Strong positive = 63 Moderate positive = 76 Weak positive = 76 860 210 53 80% 259 (rounded to 260 for better distribution of samples) Strong positive = 78 Moderate positive = 91 Weak positive = 91 1040 260 65 aInfluenza A: (H1N1) pdm09, Influenza A/H3N2, Influenza B (Victoria), and SARS CoV-2 #Strong positive: (Ct value <25) Moderate positive: (Ct value between 25-30) Weak positive: (Ct value >30 and ≤ 34)

¥ Nasopharyngeal or oropharyngeal swabs in virus transport medium (VTM) to be used *Samples positive for common respiratory viruses (such as Parainfluenza viruses, Adenoviruses, Rhinoviruses, common human coronaviruses, RSV), other than the ones targeted by the kit under evaluation. Equal distribution of cross-reactive viruses is desirable.

For multiplex assays targeting influenza and SARS-CoV-2, samples positive for these viruses may be suitably interchanged for assessing cross-reactivity. Influenza B (Yamagata) is not mentioned in the table in view nil/minimal global and national circulation of the strain for the past few years. Samples positive for influenza B (Yamagata) should be used for performance evaluation of kits if the strain starts circulating in the country.

It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics.

Page 229 of 459

  1. Acceptance Criteria for the kit:

Sensitivity for each pathogen/ type/ subtype: ≥95%
Specificity for each pathogen/ type/ subtype: ≥99%
Cross-reactivity: Minimal per pathogen/ type/ subtype To achieve at least the performance characteristics outlined in the acceptance criteria, ≥80 positive samples and ≥20 negative samples should be tested for evaluation for each pathogen/ type/ subtype. acceptable.

Page 230 of 459

Protocol E: Performance Evaluation of Molecular IVD Kit detecting Influenza virus, SARS-CoV-2 and Respiratory Syncytial Virus (RSV) in multiplex format To assess the performance of multiplex assays, Protocols A, B or C can be used as per kit format to check the performance of each virus for its sensitivity and specificity assessment, including cross reactivity, repeatability, reproducibility and Lot to lot variation.
A comprehensive report can be generated which will include sensitivity and specificity for all targets.
Sample size for multiplex molecular assay (as per the scope outlined in the document) detecting Influenza virus, SARS-CoV-2 and Respiratory Syncytial Virus (RSV) in multiplex format is given below. All other relevant performance criteria (such as repeatability/ reproducibility/ cross-reactivity) outlined in the single plex protocols (Protocol A, B and C) are to be essentially followed and met for each target pathogen/type/subtype.

  1. Sample size and sample panel composition for evaluation of performance characteristics:

Sample sizes of positive and negative samples against different values of sensitivity and specificity are provided in Table 7. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate ≤5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity. Sample sizes are calculated using the formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· n (se) is the number of positive samples. · n (sp) is the number of negative samples.

Page 231 of 459

Table 7. Sample sizes for different values of sensitivity/ specificity claimed by the manufacturer. Sensitivit y/ Specificit y Sample size for each of the 06 target pathogensa: Minimum number of positive samples¥ Composition of positive samples for each pathogen/type/subtype# Total number of positive samples (including all 06 pathogens) Sample size: Minimum number of negative samples¥ Minim um numbe r of cross reactiv e* sample s among the negativ e sample s 99% 16 (rounded to 20 for samples) Strong positive = 06 Moderate positive = 07 Weak positive = 07 120 20 5 95% 77 (rounded to 80 for samples) Strong positive = 24 Moderate positive = 28 Weak positive = 28 480 80 20 90% 146 (rounded to 155 for better distribution of samples) Strong positive = 45 Moderate positive = 55 Weak positive =55 930 150 38 85% 207 (rounded to 215 for better distribution of samples) Strong positive = 63 Moderate positive = 76 Weak positive = 76 1290 210 53 80% 259 (rounded to 260 for better distribution of samples) Strong positive = 78 Moderate positive = 91 Weak positive = 91 1560 260 65 aInfluenza A: (H1N1) pdm09, Influenza A/H3N2, Influenza B (Vivtoria), SARS CoV-2, RSV A, and RSV B #Strong positive: (Ct value <25) Moderate positive: (Ct value between 25-30) Weak positive: (Ct value >30 and ≤ 34) ¥ Nasopharyngeal/ oropharyngeal swabs in virus transport medium (VTM) to be used

  • Samples positive for common respiratory viruses (such as Parainfluenza viruses, Adenoviruses, Rhinoviruses, common human coronaviruses), other than the ones targeted by the kit under evaluation. Equal distribution of cross-reactive viruses is desirable.

For multiplex assays targeting influenza, SARS-CoV-2, and RSV, samples positive for these viruses may be suitably interchanged for assessing cross-reactivity. Influenza B (Yamagata) is not mentioned in the table in view nil/minimal global and national circulation of the strain for the past few years. Samples positive for influenza B (Yamagata) should be used for performance evaluation of kits if the strain starts circulating in the country. It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure

Page 232 of 459

adequate power of the study in case the kit falls short of claimed performance characteristics.

  1. Acceptance Criteria for the kit:

Sensitivity for each pathogen/ type/ subtype: ≥95%
Specificity for each pathogen/ type/ subtype: ≥99%

Cross-reactivity: Minimal per pathogen/ type/ subtype To achieve at least the performance characteristics outlined in the acceptance criteria, ≥80 positive samples and ≥20 negative samples should be tested for evaluation for each pathogen/ type/ subtype.

acceptable.

Reference for sample size calculation:

  1. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

Page 233 of 459

Performance Evaluation Report Format

Performance evaluation report for Respiratory Virus in-vitro molecular diagnostic kit

Name and address of the actual manufacturing site Name of supplier: Manufacturer/Importer/Port office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use License Number: Issue date:

Valid Up to: Application No. Sample Panel Sample type Negative samples (provide details, including cross reactivity panel/simulated samples)

i. Analytes/Pathogens targeted by the kit under evaluation: ………………………………………………… ii. ………………………………………………… iii. ………………………………………………… iv. ………………………………………………… v. ………………………………………………… vi. ………………………………………………… vii. …………………………………………………

RESULTS INTERPRETATION

SENSITIVITY AND SPECIFICITY FOR INDIVIDUAL VIRUS TARGETS

Page 234 of 459

  1. Sensitivity and specificity for Influenza A (H1N1) pdm09
    Name of the Kit Under Evaluation

Reference assay Positive Negative Total Positive

Negative

Total

Estimate (%) CI 95% Sensitivity

Specificity

  1. Sensitivity and specificity for Influenza A (H3N2)
    Name of the Kit Under Evaluation

Reference assay Positive Negative Total Positive

Negative

Total

Estimate (%) CI 95% Sensitivity

Specificity

  1. Sensitivity and specificity for Influenza B (Victoria)
    Name of the Kit Under Evaluation

Reference assay Positive Negative Total Positive

Negative

Total

Estimate (%) CI 95%

Page 235 of 459

Sensitivity

Specificity

  1. Sensitivity and specificity for Influenza B (Yamagata) –to be assessed if the strain circulates nationally and/or globally Name of the Kit Under Evaluation

Reference assay Positive Negative Total Positive

Negative

Total

Estimate (%) CI 95% Sensitivity

Specificity

  1. Sensitivity and specificity for SARS-CoV-2
    Name of the Kit Under Evaluation

Reference assay Positive Negative Total Positive

Negative

Total

Estimate (%) CI 95% Sensitivity

Specificity

  1. Sensitivity and specificity for RSV
    Name of the Kit Under Evaluation

Reference assay Positive Negative Total Positive

Negative

Page 236 of 459

Total

Estimate (%) CI 95% Sensitivity

Specificity

  1. Sensitivity and specificity for RSV A
    Name of the Kit Under Evaluation

Reference assay Positive Negative Total Positive

Negative

Total

Estimate (%) CI 95% Sensitivity

Specificity

  1. Sensitivity and specificity for RSV B
    Name of the Kit Under Evaluation

Reference assay Positive Negative Total Positive

Negative

Total

Estimate (%) CI 95% Sensitivity

Specificity

a) Cross-reactivity Analysis: b) Invalid test rate: c) Repeatability Assessment: d) Precision (Reproducibility):

Page 237 of 459

Lot to Lot Details of lots tested (3 lots to be tested):

  1. Lot No.:

Lot No:

Tested By: 2. Lot No.:

Lot No:

Tested By: 3. Lot No.:

Lot No:

Tested By:

▪ Lot-to-lot variation was observed / not observed.

a. Internal Control Analysis: Conclusion: Satisfactory / Not satisfactory

RECOMMENDATIONS: Suggestions for improvements or modifications (if applicable):



▪ ICMR-CDSCO guidelines were followed for kit performance evaluation.
This evaluation report is exclusively for____________________________________ In Vitro Molecular Diagnostic Kit manufactured by . Sensitivity and specificity have been assessed in controlled lab settings using the kits of the Lot number: i. Lot No., ii. Lot No., iii. Lot No._____, Provided by the manufacturer, using ……. samples. Results should not be extrapolated to other sample types. DISCLAIMER:

  1. This validation process does not approve/disapprove the Kit design.
  2. This validation process does not certify user friendliness of the Kit.
  3. Influenza and SARS-CoV-2 are continuously evolving viruses and therefore primer probe sequences of the assay may require periodic updates, which will amount to a changed version of the assay. Re-validation is required for changed version of the assay, and needs to be considered while issuing license

Page 238 of 459

Signature of the Lab Manager

Signature of the Lab Director

Signature of Head of the Institute

Seal of Head of the Institute

Page 239 of 459

Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers

  1. Instructions for Use
  2. Scope of the IVD: to diagnose influenza and/or SARS-CoV-/RSV.
  3. Intended Use Statement
  4. Principle of the assay
  5. Intended testing population (cases of ARI/ILI/SARI)
  6. Intended user (laboratory professional and/or health care worker at point-of-care)
  7. Lot/batch No.
  8. Date of manufacture
  9. Date of Expiry
  10. Information on operational Characteristics i. Configuration of the kit
  11. Information on Test Performance Characteristics i. Type of sample-NP/OP swab, other respiratory specimen iii. Any specific sample NOT to be tested v. Any additional device/consumable like sample transfer device, pipette, tube, etc required vii. Pathogens targeted by the kit

Page 240 of 459

x. Manual or automated(equipment)reading xi. Limit of detection/Limit of Quantification and range of detection xiv. Stability and reproducibility xv. Training required for testing xviii. Interpretation of invalid and indeterminate results to be provided xix. It is recommended to provide data demonstrating accuracy and precision xx. Limit of detection

Page 241 of 459

HUMAN METAPNEUMOVIRUS REAL- TIME PCR

Page 242 of 459

A. Working Group:

  1. Dr. Varsha Potdar, Scientist-E, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  2. Ms. Krittika Bhattacharyya, Statistical Officer (Planning), Directorate of Economics and Statistics, Government of National Capital Territory of Delhi

  3. Dr. Labanya Mukhopadhyay, Scientist-C, Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  4. Dr. Mandeep Chadha, Former Scientist-G, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  5. Dr. Shobha Broor, Former Professor and Head of Virology Division, All India Institute of Medical Sciences, Delhi, and current Emeritus Professor, SGT University, Gurgaon, Haryana

  6. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control Organization, Directorate General of Health Services, Ministry of Health and Family Welfare, Government of India

  7. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control Organization, Directorate General of Health Services, Ministry of Health and Family Welfare, Government of India

  8. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central

  9. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

Page 243 of 459

Performance evaluation protocol for Human Metapneumovirus real-time PCR kit I. Background:
evaluation is to independently verify the manufacturer’s claim regarding in-vitro diagnostic kit (IVD) performance. This recommendation focuses on the laboratory performance evaluation of Human Metapneumovirus (hMPV) virus real time PCR kit. All clinical samples tested in the study should be evaluated in accordance with the candidate test’s instructions for use.
II. Purpose: To evaluate the performance characteristics of hMPV real-time PCR kits in the diagnosis of hMPV infection/ disease using irreversibly de-identified leftover archived/ spiked clinical samples. III. Requirements:
Performance evaluation activities using irreversibly de-identified leftover clinical samples are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024.
V. Procedure:

Identified IVD kit evaluation laboratories should be well-equipped and establish their proficiency through ALL of the following: A. Accreditation for at least one Quality management system for at least one respiratory viral pathogen molecular testing (accreditation for Testing Lab /

Page 244 of 459

Calibration Lab as per ISO/IEC 17025, Medical Lab as per ISO 15189, PT provider as per ISO/IEC 17043), or CDSCO approved Reference laboratory.
B. Staff training: All the staff involved in hMPV virus IVD evaluation should undergo hands-on training and competency testing on following
⮚ Preparation & characterization of reference sample panel (at least 2 staff) ⮚ Handling of hMPV RT-PCR kits received for performance evaluation 3. Preparation of hMPV RNA evaluation panel A well characterised panel of hMPV positive human samples is a critical requirement for evaluation of these RT-PCR IVD kits. A statistically significant number of clinical samples should be used for the evaluation. The sample type for hMPV detection is nasopharyngeal/oropharyngeal swab. If a kit claims to detect hMPV across several sample types, attempt should be made to evaluate the assay across all the sample types. In case all the sample types mentioned in the IFU are not available with the lab, the performance evaluation report should clearly mention the sample type against which the kit is evaluated. There should be no ambiguity about the type of sample used for evaluation. 4. RNA extraction RNA extraction should be performed as per reference test and index test Instructions for Use (IFU). 5. Real-Time PCR System 6. Internal control/Extraction control The index test must have an internal control (housekeeping gene), with or without an extraction control (RNA added before extraction to a sample).

Page 245 of 459

  1. Reference assay:
    WHO Pre-Qualified/ US FDA/ ATAGI Austtralia/ PMDA Japan approved real-time PCR assay/ ICMR-NIV Pune in-house Real Time PCR Assay should be used as the Reference Assay.
    All positive samples should be confirmed positive by the reference assay. All negative samples should be confirmed negative by the reference assay.
  2. Sample size for performance evaluation: Sample size is calculated assuming 95% sensitivity and specificity of the index test, 95% confidence level, absolute precision of 5% and ≤5% invalid test rate. A minimum of 77 (rounded to 80) positive clinical samples and a minimum of 77 (rounded to 80) negative clinical samples are required for performance evaluation. However, for negative samples, a minimum of 115 specimens are suggested to account for a rigorous cross reactivity panel.

𝑛𝑠𝑒 ≥𝑍2 𝑥 𝑆𝑒 (1 −𝑆𝑒) 𝑑2 𝑥 (1 −𝐼𝑅)
𝑛𝑠𝑝 ≥𝑍2 𝑥 𝑆𝑝 (1 −𝑆𝑝) 𝑑2 𝑥 (1 −𝐼𝑅)

  1. Sample panel composition: A. Human samples A.1 Positive samples (Minimum n=80): Clinical samples positive by the reference real-time PCR assay
    A.1.1 Strong positive (Ct value <25) = 24 samples
    A.1.2. Moderate positive (Ct value between 25-30) = 28 samples A.1.3 Weak positive (Ct value >30-35) = 28 samples

Page 246 of 459

Note: If possible, attempt should be made to include all lineages of hMPV in the positive sample panel.

A.2 Negative samples: All negative samples should be negative by reference real- time PCR assay. Distribution of the negative samples should be as follows: A.2.1 NP/OP swab from individuals with respiratory infection that are negative for hMPV RNA = 30 samples A.2.2 NP/OP swab from apparently healthy individuals with no respiratory symptoms = 20 samples A.2.3 Cross reactivity panel (Table 1): Samples negative for hMPV RNA but positive for other common respiratory viruses = 65 samples Table 1: Cross reactivity panel for performance evaluation of HMPV real time PCR kit S.N. Pathogen Minimum no. of positive samples needed (n=65) Additional comments i. RSV A @ 5
In case adequate number of one RSV type is unavailable, supplement with the available RSV type ii. RSV B @ 5
iii. Measles @ 5

iv. Mumps @ 5
Buccal swab is the preferred sample type for Mumps, and the same (or throat swab) should be used for evaluation v. Seasonal Influenza A (H1N1pdm09 and H3N2) @ 10 (5 of each)

vi. Seasonal Influenza B (Victoria, with/without Yamagata) @ 5

vii. SARS-CoV-2 5

viii. Seasonal coronaviruses @ 3
OC43 AND 229E
ix. Respiratory Adenovirus @ 5
Representation from all respiratory types is desirable

Page 247 of 459

x. Human Respiroviruses 1 and 3, Human Rubulaviruses 2 and 4 @
5
Representation from all types is desirable xi. Rhinovirus @ 5
In case samples available with the lab are not typed into Rhinovirus and non- Rhinovirus Enteroviruses, please use 10 such samples to represent these 2 pathogens xii. Enterovirus * 5
xiii. Cytomegalovirus * 2
Lower respiratory specimen positive for CMV is acceptable

It is recommended to test for all pathogens listed in the cross-reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the available “essentially to be tested” samples. If available, samples positive for relevant bacterial pathogens and other relevant viruses (with which majority of the population is likely to be infected), should also be included in the cross-reactivity panel.
Commercially available validated standard panels that are accepted by accreditation agencies can also be used if clinical samples with required target for cross-reactivity analysis is not available. 10. Evaluation method: The index test and the reference assay should be run simultaneously on the sample panel, and

  1. Interpretation of results:
  2. Resolution of discrepant results: test.

Page 248 of 459

test.

evaluation. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples

be as follows:
 First lot of the kit: should be tested on statistically significant number of positive and negative samples as calculated in the protocol.
 Second lot of the kit: should be tested on 25 samples (15 positive samples comprising 10 low positive AND 5 moderate/strong positive samples, and 10 negative samples).
10 low positive AND 5 moderate/strong positive samples, and 10 negative samples).

Page 249 of 459

moderate and weak positive samples) and 3 negative samples should be tested 5 (qualitative).

(qualitative).

(qualitative).

recommended). 14. Blinding of laboratory staff:

Page 250 of 459

  1. Acceptance Criteria: Sensitivity: ≥95% Specificity: ≥98% Cross reactivity with other viruses as outlined in the negative sample panel: Minimal
  2. Publication Rights: author(s). acceptable.

Page 251 of 459

VI. References:

  1. U.S. Food and Drug Administration: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays - Class II Special Controls Guidance for Industry and FDA Staff. 2009. Available at: https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation-emitting- products/testing-human-metapneumovirus-hmpv-using-nucleic-acid-assays-class-ii-special-controls- guidance#3 [Accessed on January 11, 2025]
  2. Amarasinghe, G.K., Ayllón, M.A., Bào, Y. et al. Taxonomy of the order Mononegavirales: update
  3. Arch Virol 164, 1967–1980 (2019). https://doi.org/10.1007/s00705-019-04247-4

VII.

Page 252 of 459

PERFORMANCE EVALUATION REPORT FOR HUMAN METAPNEUMOVIRUS (HMPV) REAL-TIME PCR KITS

Name and address of the actual manufacturing site

office of

Lot No / Batch No.:

Type of Assay

Kit components

Manufacturing Date

Expiry Date

Intended Use

date:

Valid Up to:

Application No.

Sample Panel Sample type

Positive samples (provide details: clinical/spiked, strong, moderate, weak/simulated samples)

Negative samples (provide details (clinical/spiked),

Results

Positive Negative Total Name of hMPV virus real-time PCR Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

● Cross-reactivity ● Invalid test rate

Page 253 of 459

Disclaimers Note:
This report is exclusively for Human Metapneumovirus………….. Kit (Lot No……) manufactured by …………… (supplied by ……….) The kit has been validated against the pathogen (as a whole) with statistically significant sample size, and NOT against different lineages of the pathogen. Signature of Director/ Director-In-charge …………………… Seal ………………………….

Page 254 of 459

Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers

  1. Instructions for Use
  2. Scope of the IVD:
  3. Intended Use Statement
  4. Principle of the assay
  5. Intended testing population (cases of ARI/ILI/SARI)
  6. Intended user (laboratory professional and/or health care worker at point-of-care)
  7. Lot/batch No.
  8. Date of manufacture
  9. Date of Expiry
  10. Information on operational Characteristics i. Configuration of the kit/device
  11. Information on Test Performance Characteristics i. Type of sample-NP/OP swab, other respiratory specimen iii. Any specific sample NOT to be tested v. Any additional device/consumable like sample transfer device, pipette, tube, etc required vii. Pathogens targeted by the kit

Page 255 of 459

x. Manual or automated (equipment) reading xi. Limit of detection/Limit of Quantification and range of detection xiv. Stability and reproducibility xv. Training required for testing xviii. Interpretation of invalid and indeterminate results to be provided xix. It is recommended to provide data demonstrating accuracy and precision

*Please mention “Not applicable” against sections not pertaining to the kit.

Page 256 of 459

MULTIPLEX RESPIRATORY VIRUS REAL TIME PCR

Page 257 of 459

A. Working Group:

  1. Dr. Labanya Mukhopadhyay, Scientist-C, Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family

  2. Ms. Krittika Bhattacharyya, Statistical Officer (Planning), Directorate of Economics and Statistics, Government of National Capital Territory of Delhi

  3. Dr. Mandeep Chadha, Former Scientist-G, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  4. Dr. Shobha Broor, Former Professor and Head of Virology Division, All India Institute of Medical Sciences, Delhi, and current Emeritus Professor, SGT University, Gurgaon, Haryana

  5. Dr. Vasanthapuram Ravi, Former Dean Research and Head of Neurovirology, National Institute of Mental Health and Neuro-Sciences, Bengaluru, Karnataka

  6. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control

  7. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control

  8. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central

  9. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family

Page 258 of 459

Performance evaluation protocol for multiplex respiratory virus real-time PCR kit

I. Background:
diagnostic kits appropriate for use in India. Hence the following guidelines shall establish the This recommendation focuses on the laboratory performance evaluation of multiplex respiratory virus real time PCR kit. All clinical samples tested in the study should be evaluated in accordance with the candidate test’s instructions for use.
II. Purpose: To evaluate the performance characteristics of multiplex respiratory virus real-time PCR kits using irreversibly de-identified leftover archived clinical/spiked samples. III. Scope of the document: This document outlines performance evaluation protocol for multiplex real time PCR assays detecting the following respiratory viruses of utmost importance in human clinical specimens (Table 1), as determined by ICMR appointed working group and expert group of physicians and clinical microbiologists following extensive literature review and real-life experience. This pathogen list has been developed as part of the National One Health Mission. Table 1: List of respiratory viruses within the scope of this performance evaluation protocol

  1. Influenza virus A 
    
  2. Influenza virus B
  3. SARS Coronavirus-2
  4. Respiratory syncytial virus 
    
  5.  Adenovirus 
    
  6. Human Respiroviruses 1 and 3 and Human Rubulaviruses 2 and 4 (erstwhile Human 
    

Parainfluenzaviruses 1-4) 7. Human metapneumovirus
8. Measles virus 9. Rhinovirus 10. Human Bocavirus 11. Enterovirus 12. Cytomegalovirus

IV. Requirements:

Page 259 of 459

V. Ethical approvals:
VI. Procedure:

  1. Study design/type: Diagnostic accuracy study using irreversibly de-identified archived/ spiked clinical samples

Identified IVD kit evaluation laboratories should be well-equipped and establish their proficiency through ALL of the following:

A. Accreditation at least one of the Quality management systems for at least one respiratory viral pathogen molecular testing (accreditation for Testing Lab / Calibration Lab as per ISO/IEC 17025, Medical Lab as per ISO 15189, PT provider as per ISO/IEC 17043), or CDSCO approved Reference laboratory.
B. Staff training: All the staff involved in IVD evaluation should undergo hands-on training and competency testing on the following:
⮚ Preparation & characterization of reference sample panel ⮚ Handling of multiplex respiratory virus RT-PCR kits received for performance evaluation (Verification/Storage/Unpacking etc). ⮚ Testing

  1. Preparation of multiplex respiratory virus evaluation panel: A well characterised panel of positive and negative clinical samples is a critical requirement for evaluation of these RT-PCR IVD kits. Also, a statistically significant number of clinical samples should be used for the evaluation. The sample type for respiratory virus detection is usually nasopharyngeal/oropharyngeal swab. If a kit claims to detect these viruses across several sample types, attempt should be made to evaluate the assay across all the sample types. In case all the sample types mentioned in the IFU are not available with the lab, the performance evaluation report should clearly

Page 260 of 459

mention the sample type against which the kit is evaluated. There should be no ambiguity about the type of sample used for evaluation. 4. Nucleic acid extraction: Nucleic acid extraction should be performed as per reference test and index test Instructions for Use (IFU). *Caution is advised in the selection of a nucleic acid extraction kit since the target pathogens comprise both RNA and DNA viruses.
5. Real-Time PCR System: 6. Internal control/Extraction control: The test under evaluation (index test) must have an internal control (housekeeping gene), with or without an extraction control (nucleic acid added before extraction to a sample). 7. Reference assay:
The following points are to be noted: i. A WHO Pre-Qualified/ US FDA/ ATAGI Australia/ PMDA Japan approved single plex (for a particular target pathogen) or multiplex real-time PCR assay/ ICMR-NIV Pune in-house single plex (for a particular target pathogen) or multiplex Real Time PCR Assay should be used as the reference assay for each target pathogen/type/subtype. ii. Since the list of target pathogens is extensive, a combination of single plex and/or multiplex assays may be used as the reference assay(s), as long as these reference assays satisfy the criteria outlined in point 7(i).
All samples positive for a particular pathogen should be confirmed positive by the reference assay. All samples negative for a particular pathogen should be confirmed negative by the reference assay. 8. Sample size for performance evaluation: The 2009 FDA guidance document “Respiratory Viral Panel Multiplex Nucleic Acid Assay - Class II Special Controls Guidance for Industry and FDA Staff”, recommends including a sufficient number of prospectively collected samples for each specimen type to generate a result with at least 90% sensitivity with a lower bound of the two-sided 95% confidence interval (CI) greater than 80, and demonstrate specificity with a lower bound of the two-sided 95% CI greater

Page 261 of 459 than 90%. In accordance with these guidelines and for feasibility of evaluation of these extensive multiplex panels, sample size for each pathogen is calculated assuming ≥90% sensitivity and specificity of the index test, 95% confidence level, absolute precision of 7.5%, and ≤5% invalid test rate. A minimum of 65 positive clinical samples (rounded to 70) and a minimum of 65 negative clinical samples for each target pathogen are required for performance evaluation of the assay. However, 120 negative samples are recommended per pathogen to account for an extensive cross reactivity panel. Sample sizes are calculated using the formulae: 𝑑2 x (1 −IR) 𝑑2 x (1 −IR) · n (se) is the minimum number of positive samples. · n (sp) is the minimum number of negative samples. · Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96). · Se is the predetermined sensitivity. · Sp is the predetermined specificity. · d is the predetermined marginal error (5%) · IR is the invalid test rate The details of sample requirement are outlined in Table 2. Table 2: No. of samples required for performance evaluation: Pathogen Minimum no. of positive samples needed per pathogen Minimum no. of negative samples recommended per pathogen 1. Influenza virus A* 70 120 2. Influenza virus B* 70 120 3. SARS Coronavirus-2 70 120 4. Respiratory syncytial virus* 70 120 5. Adenovirus* 70 120 6. Human Respirovirus 1 and Human Respirovirus 3 and Human Rubulavirus 2 and Human Rubulavirus 4* 70 120 7. Human metapneumovirus * 70 120 8. Measles virus 70 120 9. Rhinovirus** 70 120

Page 262 of 459 10. Human Bocavirus 70 120 11. Enterovirus** 70 120 12. Cytomegalovirus 70 120 *If a kit claims to differentiate between virus types/subtypes, please use minimum 70 positive samples and minimum 120 negative samples for each virus type/subtype. If such type/subtype specific samples are not available (only for predicate device) or if the kit does not claim to differentiate between pathogen types/subtypes, and the kit is evaluated against the pathogen as a whole, the reports should be issued with a disclaimer that performance characteristics against pathogen types/subtypes have not been evaluated separately. However, in such a scenario, the evaluating centre should try to include all types/subtypes of the pathogen in the evaluation panel (even if the numbers are not statistically significant for each pathogen type).
**If clinical samples positive separately for Rhinovirus/Enterovirus are not available (only for predicate device), or if the kit does not differentiate between Enteroviruses and Rhinoviruses, please use minimum 70 samples positive for Rhinovirus/Enterovirus in the positive sample panel and issue the reports with a disclaimer that performance characteristics against Rhinovirus/Enterovirus have not been evaluated separately. Influenza virus, SARS Coronavirus 2, Respiratory Syncytial Virus and Human Metapneumovirus positive samples used for evaluation should have been collected within the past 1 year. Notes for Table 2:

  1. Samples positive for currently circulating virus strains should be used in the positive sample panel, with representation from all virus types/subtypes.
  2. Sample positive for a particular virus type and negative for the target pathogen being considered may be used in the negative sample panel for the target pathogen, e.g.: a sample positive for SARS-CoV-2 may be used as a negative sample for RSV.
  3. Sample panel composition: A. Human samples A.1 Positive samples for each pathogen/ type or subtype of pathogen (Minimum n=70): Clinical samples positive by the reference real-time PCR assay should be included, as per the following criteria
    A.1.1 Strong positive (Ct value <25) = 20 samples A.1.2. Moderate positive (Ct value between 25-30) = 25 samples A.1.3 Weak positive (Ct value >30-36) = 25 samples A.2 Negative samples for each pathogen/ type or subtype of pathogen (Minimum n=120): All negative samples should be negative for the target pathogen/ its type or subtype by the reference real-time PCR assay. Distribution of the negative samples A.2.1 NP/OP swab from individuals with respiratory infection that are negative for the target pathogen/its type or subtype = 35 samples **

Page 263 of 459

A.2.2 NP/OP swab from apparently healthy individuals with no respiratory symptoms = 23 samples ** A.2.3 Cross reactivity panel (Table 3): Samples negative for the target pathogen but
positive for other common respiratory viruses = 62 samples *** Archived frozen sample aliquots if used for the evaluation, should not be thawed more than once.
** If samples are available with the evaluating lab that satisfy these criteria and are negative for all the pathogens targeted by the kit, the same samples may be included in the negative sample panel for all target pathogens to prevent wastage of resources.
*** Same positive samples may be included in the cross-reactivity panel of several target pathogens to prevent wastage of resources e.g.: the same Influenza A virus positive sample may be included in the cross-reactivity panel for RSV, Human Metapneumovirus, SARS-CoV- 2 etc.

Page 264 of 459 Table 3: Cross reactivity panel for performance evaluation of multiplex respiratory virus real time PCR kit Target Pathoge n Virus-wise no. of samples needed for cross reactivity analysis Total no. of cross reacti ve sampl es per patho gen Inf lue nza vir us A * Inf lue nza vir us B* SAR S Coro navir us-2 * Res pirat ory sync ytial viru s * Ade novi rus @ Huma n Respir ovirus es 1 and 3 , Huma n Rubul avirus es 2 and 4

Human metapn eumovi rus@ Meas les virus * Rhi nov irus @ $ Huma n Bocav irus Ent ero viru s $ Cytom egalovi rusᴥ Seas onal coro navir uses * Rub ella

  1. Influenza virus A 0 5 5 5 5 5 5 5 5 5 5 5 5 2 62
  2. Influenza virus B 5 0 5 5 5 5 5 5 5 5 5 5 5 2 62

SARS Coronavirus -2 5 5 0 5 5 5 5 5 5 5 5 5 5 2 62 4. Respira tory syncytial virus 5 5 5 0 5 5 5 5 5 5 5 5 5 2 62 5. Adeno virus 5 5 5 5 0 5 5 5 5 5 5 5 5 2 62 6. Human Respiroviru ses 1 and 3, Human Rubulavirus es 2 and 4 5 5 5 5 5 0 5 5 5 5 5 5 5 2 62 7. Human metapneum ovirus 5 5 5 5 5 5 0 5 5 5 5 5 5 2 62 8. Measle s virus 5 5 5 5 5 5 5 0 5 5 5 5 5 2 62 9. Rhinovirus 5 5 5 5 5 5 5 5 0 5 5 5 5 2 62 10. Human Bocavirus 5 5 5 5 5 5 5 5 5 0 5 5 5 2 62 11. Enterovirus 5 5 5 5 5 5 5 5 5 5 0 5 5 2 62 12. Cytomegalo virus 5 5 5 5 5 5 5 5 5 5 5 0 5 2 62 *Include all currently circulating strains/types/subtypes @It is desirable to have representation from all types of the pathogen, since even approved assays may not always differentiate between pathogen types.

Include at least 1 of each

$ If clinical samples positive separately for Rhinovirus/Enterovirus are not available, please use total 10 samples positive for Rhinovirus/Enterovirus in the cross-reactivity panel for remaining pathogens. ᴥ Can use lower respiratory tract specimen

Page 265 of 459 If a kit claims to differentiate between virus types/subtypes, please use 5 positive samples for each virus type in the cross reactivity panel for other target pathogens. If such type specific samples are not available and the kit is evaluated against the pathogen as a whole, it should be clearly mentioned in the report. If available, samples positive for relevant bacterial pathogens and other relevant viruses (with which majority of the population is likely to be infected), should also be included in the cross- reactivity panel. Influenza virus, SARS Coronavirus 2, Respiratory Syncytial Virus and Human Metapneumovirus positive samples used for evaluation should have been collected within the past 1 year. B. Contrived samples: Contrived positive and negative samples may be used for evaluation in case of paucity/unavailability of human clinical samples. Positive contrived samples should be positive and negative contrived samples should be negative for the target pathogen/type/subtype using the reference assay. The number and distribution of positive and negative samples, including the cross reactivity panel, should remain the same. Contrived positive samples (as part of positive sample panel/ cross-reactivity panel) should be prepared by spiking a sample matrix negative for the pathogen with a pathogen-infected cell line, genomic DNA plasmids or RNA transcripts. It is recommended to demonstrate equivalence between contrived and clinical specimens. Serial dilutions of clinical sample and serial dilutions of contrived sample with targeted levels of analyte should be compared for demonstrating equivalence. 10. Evaluation method: The index test and the reference assay should be run simultaneously on the sample panel, and 11. Interpretation of results: 12. Resolution of discrepant results: test. test.

Page 266 of 459

under evaluation. samples per target should be tested 5 times in independent runs.

Three lots of an IVD kit shall be evaluated. Sample size for lot-to-lot reproducibility per target negative samples).
samples).

Page 267 of 459

moderate and weak positive samples) and 3 negative samples per target should be (qualitative).

and 3 negative samples per target should be tested 5 times in independent runs.

(qualitative).

positive samples (strong, moderate and weak positive samples) and 3 negative samples per target should be tested 5 times in independent runs.

Page 268 of 459

(qualitative).

recommended). 14. Blinding of laboratory staff:

To ensure rigor of the evaluation process, laboratory staff performing the evaluation should be blinded to the status of the clinical samples. The PI of the evaluation exercise should remain the PI may remain unblinded and carry out coding of samples and dispensing them into similar- looking vials to be used for testing, and maintaining the database of results. Staff performing the reference test and the test under evaluation, interpretation of the test result, and entering the results against the coded samples in the database, should remain blinded to the status of samples till the completion of evaluation. The data should be analyzed only by the PI of the evaluating lab, and the PI should maintain confidentiality of data. Refer to Fig. 2.

  1. Acceptance Criteria Sensitivity for each pathogen/type/subtype: ≥90% Specificity for each pathogen/type/subtype: ≥95%

Page 269 of 459

Cross reactivity with other viruses as outlined in the negative sample panel: Minimal

  1. Publication Rights: author(s). acceptable.
    VII. References:
  2. Food and Drug Administration. Respiratory Viral Panel Multiplex Nucleic Acid Assay - Class II Special Controls Guidance for Industry and FDA Staff. Available at: https://www.fda.gov/medical- devices/guidance-documents-medical-devices-and-radiation-emitting-products/respiratory-viral- panel-multiplex-nucleic-acid-assay-class-ii-special-controls-guidance-industry-and [Accessed on 22nd January, 2025].
  3. Food and Drug Administration. 510(k) Substantial Equivalence Determination Decision Summary, Biofire Diagnostics LLC, FilmArray Pneumonia Panel. Available at: https://www.accessdata.fda.gov/cdrh_docs/reviews/K180966.pdf [Accessed on 19th January 2025]
  4. Food and Drug Administration: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays - Class II Special Controls Guidance for Industry and FDA Staff. 2009. Available at: https://www.fda.gov/medical-devices/guidance-documents-medical-devices-and-radiation- emitting-products/testing-human-metapneumovirus-hmpv-using-nucleic-acid-assays-class-ii- special-controls-guidance#3 [Accessed on January 11, 2025]

VIII. Performance evaluation report format

Page 270 of 459

Performance evaluation report for multiplex respiratory virus real-time PCR kits

Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Pathogens detected by the assay Intended Use

Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: clinical/spiked, strong, moderate, weak/simulated samples)
Negative samples (provide details: clinical/spiked, including cross reactivity panel/simulated samples)

Results: Tables 1 and 2 should be made for each pathogen/type of pathogen targeted by the kit under evaluation

Table 1: 2x2 table for sensitivity and specificity calculation (prepare 1 table for each target pathogen /type/ subtype)

Positive Negative Total Name of kit and name of target pathogen Positive

Negative

Total

Table 2: Sensitivity and specificity (calculate for each target pathogen /type/ subtype)

Estimate (%) 95% CI

Page 271 of 459

Sensitivity

Specificity

o Cross reactivity o Invalid test rate Disclaimers Note:
This report is exclusively for Human Metapneumovirus………….. Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ………………………….

Page 272 of 459

Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers

Instructions for Use 3. Scope of the IVD:
4. Pathogens/type/subtype of pathogens targeted by the kit 5. Intended Use Statement 6. Principle of the assay 7. Intended testing population (cases of ARI/ILI/SARI) 8. Intended user (laboratory professional and/or health care worker at point-of-care) 9. Lot/batch No. 10. Date of manufacture 11. Date of Expiry 12. Information on operational Characteristics i. Configuration of the kit/device 13. Information on Test Performance Characteristics i. Type of sample-NP/OP swab, other respiratory specimen iii. Any specific sample NOT to be tested v. Any additional device/consumable like sample transfer device, pipette, tube, etc required vii. Pathogens targeted by the kit

Page 273 of 459

x. Manual or automated (equipment) reading xi. Limit of detection/Limit of Quantification and range of detection xiv. Stability and reproducibility xv. Training required for testing xviii. Interpretation of invalid and indeterminate results to be provided xix. It is recommended to provide data demonstrating accuracy and precision xx. Limit of detection *Please mention “Not applicable” against sections not pertaining to the kit.

Page 274 of 459

MALARIA IN-VITRO DIAGNOSTICS

Page 275 of 459

A. Working Group:

  1. Dr. Praveen K. Bharti, Scientist-F, ICMR-National Institute of Malaria Research, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  2. Dr. Gauri Misra, Scientist-I, Head Molecular Diagnostics and COVID-19 Kit Testing Laboratories, National Institute of Biologicals, Noida

  3. Dr Shafeeq K Shahul Hameed, Former Scientific Officer, Center for Brain Research, Indian Institute of Science, Bengaluru, and currently Senior Epidemiologist, Health Emergency Department, Ministry of Public Health, Qatar

  4. Dr. Vishal Deo, Scientist-C, ICMR- National Institute for Research in Digital Health and Data Science, Department of Health Research, Ministry of Health and Family

  5. Dr. Labanya Mukhopadhyay, Scientist-C, Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  6. Dr. Ashis Das, Senior Professor, Department of Biological Sciences, Birla Institute of Technology and Science, Pilani

  7. Dr. Pawan Malhotra, Department of Malaria Biology, International Center for Genetic Engineering & Biotechnology, New Delhi

  8. Dr. Lokesh Kori, Scientist-D, Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  9. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control

  10. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control

  11. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central

  12. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family

Page 276 of 459

Performance evaluation protocol for Malaria Rapid diagnostic test (RDT) kits

I. Background:
CDSCO/ICMR, New Delhi have aimed to facilitate the evaluation and supply of Quality- Assured in vitro Diagnostics (IVD) kits suitable for use in India. Hence, the following guidelines shall establish the uniformity during the performance evaluation of IVD kits The objective of performance evaluation is to independently validate the manufacturer’s claim regarding in-vitro diagnostic kit (IVD) performance. II. Purpose: To evaluate the performance characteristics of rapid diagnostic test kit for the diagnosis of malaria parasite using irreversibly de-identified leftover archived clinical samples. III. Requirements:
a) Instructions for use (IFU) b) Supply of RDT kits under evaluation (with batch no.; lot no.; manufacturing and expiry date and other required details).
c) Evaluation sites/laboratories (With required equipment) d) Reference test kits e) Characterised Evaluation panel f) Laboratory supplies
Investigators are required to submit a self-declaration form, as outlined in the ICMR guidelines, to the institutional authorities and ethics committee for information. V. Procedure: clinical/leftover samples. through:
a) Laboratory accreditation: Accreditation for at least one of the Quality management systems (accreditation for Testing Lab / Calibration Lab (ISO/IEC: 17025),

Page 277 of 459

Medical Lab (ISO: 15189), PT provider (ISO/IEC: 17043) or CDSCO approved Reference laboratory.
b) It is recommended that malaria Medical Device Testing Labs (MDTLs) participate in Quality Control exercises such as EQAP (External Quality Assurance Programme). c) Staff training: All the staff involved in IVD kit evaluation should undergo hands on training and competency testing on the following at suitable malaria labs before initiation of MDTL activity:
⮚ Preparation and characterization of evaluation panel for the respective IVD kit. ⮚ Management of RDT kits (specific for Plasmodium falciparum / Plasmodium vivax) received for performance evaluation (Verification/Storage/Unpacking etc.). ⮚ Perform tests interpretation and documentation of results, and reporting. ⮚ Data management and safety and confidentiality. 3. Preparation of sample panel for Malaria RDT kit evaluation
To evaluate the performance of IVD kit, a well characterized species-specific malaria antigen sample panel is required. Statistically significant number of blood samples as defined in this protocol should be used (as mentioned in Table 1). The panel should comprise positive and negative samples as described in section 5.
The reference sample panel should be stored in appropriate storage conditions (depending on the sample type and planned storage duration), and the quality of the panel should be checked periodically (at least once a year) with appropriate tests (e.g.: reference test/parasite culture/enzymatic activity/other relevant test). 4. Reference assay:
WHO Pre-Qualified/ US FDA/ ATAGI Australia/ PMDA Japan approved RDT should be used as reference standard. Parasitemia level should be determined by microscopy and/or other relevant test results. All positive samples should be confirmed positive by the reference assay. All negative samples should be confirmed negative by the reference assay.

  1. Sample size and sample panel composition for performance evaluation:
    Sample sizes of positive and negative samples of each species targeted by the kit against different values of sensitivity and specificity are provided in Tables 1 and 2, with recommended composition. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate of 5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in

Page 278 of 459

the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table).
For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity.

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96).

Table 1. Positive sample sizes (per species) and composition for different values of sensitivity claimed by the manufacturer for evaluation of Pf (single/combo RDT) or Pv (single/combo RDT) Sensitivity Sample size: Minimum number of positive samples # Composition of positive sample panel 99% 16 (rounded to 20)
Equal number of high and low parasitemic samples 95% 77 (rounded to 80)
90% 146 (rounded to 150) 85% 207 (rounded to 210) 80% 259 (rounded to 260) 75% 305 (rounded to 310)

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics.

Page 279 of 459

Table 2. Negative sample sizes and composition for different values of specificity claimed by the manufacturer for evaluation of Pf (single/combo RDT) or Pv (single/combo RDT) Specificity Sample size: Minimum number of negative samples # Composition of negative samples# 99% 16 (rounded to 20) Dengue NS1/IgM positive samples: 03 Chikungunya IgM positive samples:03 Serum reactive for RA factor – low positive and high positive:02 Serum reactive for TPHA/other specific test for syphilis:02 Healthy controls from endemic regions: 10

95% 77 (rounded to 80) Dengue NS1/IgM positive samples: 10 Chikungunya IgM positive samples:10 Serum reactive for RA factor – low positive and high positive:10 Serum reactive for TPHA/other specific test for syphilis:10 Healthy controls from endemic regions: 40

90% 146 (rounded to 150) Dengue NS1/IgM positive samples: 18 Chikungunya IgM positive samples:18 Serum reactive for RA factor – low positive and high positive:18 Serum reactive for TPHA/other specific test for syphilis:18 Healthy controls from endemic regions: 78

85% 207 (rounded to 210) Dengue NS1/IgM positive samples: 26 Chikungunya IgM positive samples:26 Serum reactive for RA factor – low positive and high positive:26 Serum reactive for TPHA/other specific test for syphilis:26 Healthy controls from endemic regions: 106

80% 259 (rounded to 260) Dengue NS1/IgM positive samples: 35 Chikungunya IgM positive samples:35 Serum reactive for RA factor – low positive and high positive:30 Serum reactive for TPHA/other specific test for syphilis:30 Healthy controls from endemic regions: 130

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics.

Sample panel composition:

Positive samples: Malaria RDT/microscopy positive samples should be obtained from health facilities (tertiary care centers and their linked hospitals, private clinics, field practice areas etc.) and confirmed using PCR (Standardized Snounou protocol/US FDA/ATAGI Australia/PMDA Japan approved /WHO Pre-Qualified assay).

Page 280 of 459

 RDT/Microscopy AND PCR confirmed positive malaria samples should be characterized further for parasite load on in-house calibrated equipment using blood smear microscopy (and/or other relevant test results).
 Range of Parasitemia: Panel members should have low (≤200 parasites per microliter) to high (≥2000 parasites per microliter) range of Plasmodium falciparum, P. vivax and/or other Plasmodium species, as obtained from microscopy and/or other relevant test results results. Characterized panels must contain equal number of samples of both low and high parasitemia. Note for additional characterization (not mandatory):
If ELISA is used for characterization of samples in addition to the above-mentioned mandatory tests (RDT/microscopy AND PCR), consistent ELISA quantification results should be obtained in ≥3 runs of ELISA experiments performed for each of the three common antigens (PfHRP2, LDH and aldolase), with the results obtained at the 200 p/µL and the 2,000 p/µL being consistent with each other as well (factor of roughly 10 between results). The limit of detection of PfHRP2 is 0.6-74 ng/ mL, PvLDH is 1.6-47.9 ng/ mL, PfLDH is 0.2-53.5 ng/mL, and Pf aldolase is 0-9.9 ng/mL.
** If the pool of samples available for testing is sufficiently large in numbers, then the antigen concentration range at the 200 p/µL dilution should be restricted to 5-9.5 ng/mL for PfHRP2, 15-47.9 ng/mL for PvLDH, 10.8-53.5 ng/mL for PfLDH, and 1.7- 15ng/mL for Pf aldolase. If ELISA for other antigen/antibody is carried out for further characterization (for which limits of detection have not been established), testing should be carried out on calibrated equipment in ≥3 runs of ELISA, leading to their classification as low and high parasitemic samples or strong/moderate/weakly reactive samples.
Negative samples: These are samples negative by RDT/microscopy AND PCR(Standardized Snounou protocol/US FDA/ATAGI Australia/PMDA Japan approved /WHO Pre-Qualified assay).

  1. Repeatability and Reproducibility Assessment:

This should be done to assess the repeatability of the detection of target analyte(s) using the kit under evaluation. 3 positive samples (comprising low parasitemic AND high parasitemic samples) and 3 negative samples per species should be tested 5 times in independent runs.

Page 281 of 459

Three lots of an IVD kit shall be evaluated. Sample size for lot-to-lot reproducibility should be as follows:
 Second lot of the kit: should be tested on 25 samples per species (15 positive samples comprising 10 low parasitemic AND 5 high parasitemic samples per species, and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples per species (15 positive samples comprising 10 low parasitemic AND 5 high parasitemic samples per species, and 10 negative samples).

imprecision (if applicable) should be measured. 3 positive samples (comprising low

Page 282 of 459

parasitemic AND high parasitemic samples) and 3 negative samples per species (qualitative).

consecutive days. 3 positive samples (comprising low parasitemic AND high parasitemic samples) and 3 negative samples per species should be tested 5 times in independent runs.
(qualitative).

positive samples (comprising low parasitemic AND high parasitemic samples) and 3 negative samples per species should be tested 5 times in independent runs.
(qualitative).

Note: Testing Methodology Read the instructions for use (IFU) thoroughly. Take out the required number of RDTs kits from the manufacturer-recommended storage conditions. Bring RDTs to room temperature (20°C - 30°C) and thaw the required number of sample aliquots for a minimum of 20 minutes to maximum 60 minutes before performing the test. Note that more than one aliquot may be needed for the testing of each sample. Record the results of the performance evaluation on the recommended report format. 7. Evaluation method: The reference assay and the index test should be run on the sample panel in parallel. 8. Interpretation of results:

Results should be interpreted as per the IFU of the reference assay and the index test.

  1. Resolution of discrepant results:

True positive samples: These are samples positive by both reference assay and index test. True negative samples: These are samples negative by both reference assay and index test. test. test.

Page 283 of 459

  1. Acceptance criteria1:

Sensitivity: ≥75% for P. vivax and ≥95% for P. falciparum Specificity: ≥90% for P. vivax and ≥95% for P. falciparum

To achieve at least the performance characteristics outlined in the acceptance criteria, ≥310 positive samples and ≥150 negative samples should be tested for P vivax, and ≥80 positive samples and ≥80 negative samples should be tested for P falciparum.

  1. Blinding of laboratory staff

lab, and the PI should maintain confidentiality of data. Refer to Fig. 2.

Page 284 of 459

  1. Publication Rights

The PI(s) of the evaluating labs shall retain publication rights to the evaluation as lead author(s). After following due procedure as defined in this document, once any kit is found to be Not of Standard Quality, thereafter, no request for repeat testing of the same kit will be acceptable. Any request of re-validation from the same manufacturer for the same test type will only be entertained if valid proof of change in the kit composition is submitted. acceptable.

VI. References:

  1. Ministry of Health and Family Welfare. Guidelines for Bivalent RDT. Available at: guidelines- for-bivalent-rdt.pdf (mohfw.gov.in)
  2. World Health Organization. Malaria Rapid Diagnostic Test Performance - Results of WHO product testing of malaria RDTs: round 8 (2016–2018): Available at: https://iris.who.int/bitstream/handle/10665/276190/9789241514965-eng.pdf?sequence=1
  3. Snounou G, Viriyakosol S, Zhu XP, Jarra W, Pinheiro L, Do Rosario VE, et al. High sensitivity of detection of human malaria parasites by the use of nested polymerase chain reaction. Molecular and Biochemical Parasitology. 1993;61:315–20.
  4. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability
  5. World Health Organization. Methods manual for laboratory quality control testing of malaria rapid diagnostic test. Version Ten; 2023

VII.

Page 285 of 459

REPORT FORMAT Name of the Laboratory Name of the Institute, (with station) Certificate of Analysis File No.: _________________________________________

Name of supplier: Manufacturer/Importer/Port office of

Lot No / Batch No.:

Type of Assay

Kit components

Manufacturing Date

Expiry Date

Intended Use

Valid Up to:

Application No.

Sample Panel Sample type

Positive samples (provide details: low/high parasitemia/simulated samples)

Page 286 of 459

Negative samples (provide details, including cross reactivity panel/simulated samples)

Results:

Reference assay ……….……………… (name)

Positive Negative Total Name of index malaria RDT
Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity o Invalid test rate o Performance: Satisfactory / Not Satisfactory (Sensitivity and specificity have been assessed in controlled lab setting on …. samples only, using kits provided by the manufacturer from the batch mentioned above. Results should not be extrapolated for any other sample type.) Disclaimers Note: This report is exclusively for ………….. Kit (Lot No……) manufactured by …………… (Supplied by ……….) Signature of Director/ Director-In-charge ……………………

Seal ………………………………………

Page 287 of 459

Performance evaluation protocol for Malaria ELISA kits I. Background:
CDSCO/ICMR, New Delhi, have aimed to facilitate the evaluation and supply of Quality- Assured In Vitro Diagnostics kits suitable for use in India. Hence, the following guidelines shall establish the uniformity during the performance evaluation of IVD kits . The objective of performance evaluation is to independently validate the manufacturer’s claim regarding in-vitro diagnostic kit (IVD) performance. II. Purpose:
To evaluate the performance characteristics of malaria ELISA kits for the diagnosis of malaria parasite infection using irreversibly de-identified leftover archived clinical samples. The malaria ELISA kits are designed to detect antigens (hrp2, LDH, aldolases) occurring in subjects infected with species specific (P. falciparum, P. vivax) and stage specific antibodies (MSP1, MSP3, CSP, EBA175 etc.- parasite markers for the purpose of sero-survey). III. Requirements:

  1. Instructions for use (IFU)
  2. Supply of ELISA kits under evaluation (with batch no./lot no. expiry date & required details). In case the kit to be evaluated is designed to work in a closed system format, the manufacturer needs to supply the required equipment.
  3. Evaluation sites/laboratories (With required equipment)
  4. Reference test kits
  5. Characterised Evaluation panel
  6. Laboratory supplies
    Performance evaluation activities using irreversibly de-identified leftover clinical samples are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024.
    V. Procedure:
  7. Study design/type: Diagnostic accuracy study using irreversibly de-identified
    leftover clinical samples. Identified ELISA kit evaluation laboratories should establish their proficiency through
    a) Laboratory accreditation: Accreditation for at least one of the Quality management systems (accreditation for Testing Lab / Calibration Lab (ISO/IEC: 17025),

Page 288 of 459

Medical Lab (ISO:15189), PT provider ISO/IEC: 17043 or CDSCO approved Reference laboratory. b) It is recommended that malaria Medical Device Testing Labs (MDTLs) participate in Quality Control exercises such as EQAP (External Quality Assurance Programme). c) Staff training: All the staff involved in ELISA kit evaluation should undergo hands on training and competency testing on the following at suitable malaria labs before initiation of MDTL activity:
 Preparation and characterization of evaluation panel for the respective ELISA kit.  Management of malaria ELISA kits received for performance evaluation  Perform tests, interpretation and documentation of results and reporting.  Data management and safety and confidentiality 3. Reference sample panel and sample panel characterization:
To evaluate the performance of ELISA kit a well characterised malaria stage specific antigen/species specific antibody ELISA evaluation sample panel is required. In the absence of WHO Pre-Qualified/US FDA/ ATAGI Australia/ PMDA Japan approved malaria ELISA assay, it is recommended that performance evaluation of ELISA assays be carried out on a rigorously well characterized panel of positive and negative samples.
WHO Pre-Qualified/US FDA/ ATAGI Australia/ PMDA Japan approved ELISA kits may be used as the reference assay as and when such kits become available.
A statistically significant number of blood/plasma/sera samples should be used. All samples should be further characterized. A. Sample panel for malaria antigen ELISA evaluation:

Positive samples: These are samples positive by RDT/microscopy AND PCR (standardized Snounou protocol/WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved) AND characterized further. These samples should be additionally characterized for parasite load and analyte level on in-house calibrated equipment using microscopy and other relevant test results.

a. Range of Parasitemia: Panel members should have low (≤200 parasites per microliter) to high (≥2000 parasites per microliter) range of Plasmodium falciparum, P. vivax and/or other Plasmodium species, as obtained from microscopy and/or other relevant test results. Characterized panels must contain equal number of samples of both low and high parasitemia.
b. Analyte Characterization by ELISA: Consistent ELISA quantification results should be obtained in ≥3 runs of ELISA experiments performed for each of the three antigens (PfHRP2, LDH and aldolase), with the results obtained at the 200

Page 289 of 459

p/µL and the 2,000 p/µL being consistent with each other as well (factor of roughly 10 between results). The limit of detection of PfHRP2 is 0.6-74 ng/ mL, PvLDH is 1.6-47.9 ng/ mL, PfLDH is 0.2-53.5 ng/mL, and Pf aldolase is 0-9.9 ng/mL.
** If the pool of samples available for testing is sufficiently large in numbers, then the antigen concentration range at the 200 p/µL dilution should be restricted to 5-9.5 ng/mL for PfHRP2, 15-47.9 ng/mL for PvLDH, 10.8-53.5 ng/mL for PfLDH, and 1.7- 15ng/mL for Pf aldolase. For those kits which have other antigen as target analyte (for which limits of detection have not been established), characterization of samples for that analyte should be performed on calibrated equipment in ≥3 runs of ELISA, leading to their classification as low/high parasitemic AND strong/moderate/weakly positive samples, which will then be used for performance evaluation of the assay.

B. Sample panel for malaria antibody ELISA evaluation: For those kits which have antibodies as target analytes (for which limits of detection have not been established), characterization of sample panel for that analyte should be performed on calibrated equipment(≥3 runs of ELISA), leading to their classification as strong, moderate and weakly reactive samples and negative samples, which will then be used for performance evaluation of the assay. Samples for antibody kit evaluation must have been collected from microscopy/RDT/PCR positive malaria cases at requisite time points (depending on the antibody target).
The above-mentioned activities should not be performed with spiked/contrived samples. Equal representation of samples positive for Plasmodium (P.falciparum /P.vivax) species is preferred. C. Negative panel for antigen based ELISA should constitute malaria RDT/microscopy negative samples (confirmed by PCR) as described in point 3A. Negative panel for antibody based ELISA should be negative for the analyte when performed on calibrated equipment (≥3 runs of ELISA). The reference sample panel should be stored in appropriate storage conditions (depending on the sample type and planned storage duration), and the quality of the panel should be checked periodically (at least once a year) with appropriate tests (e.g.: reference test/parasite culture/enzymatic activity/other relevant test). Malaria positive samples should be obtained from health facilities, including tertiary care centers and their linked hospitals, private clinics, field practice areas etc.

Page 290 of 459

  1. Sample size: Sample sizes of positive and negative samples of each species targeted by the kit against different values of sensitivity and specificity are provided in Table 1 and Table 2, with recommended composition. Sample sizes have been calculated assuming 95% level of significance and an absolute precision of 5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For require usage of the sample size outlined for 85% specificity. Sample sizes are calculated using the formulae:

𝑑2

𝑑2

· Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96).

Table 1. Positive sample sizes (per species) and composition for different values of sensitivity claimed by the manufacturer for evaluation of Pf (single/combo) or Pv (single/combo) antigen/antibody ELISA Sensitivity Sample size: Minimum number of positive samples# Composition of positive sample panel 99% 16 (rounded to 20) in each category for antigen based ELISA. It is recommended to include 30% strongly reactive samples, and 35% each across moderately and weakly reactive (depending on ELISA OD cut-offs) in each category for both antigen and antibody-based ELISA.

95% 73 (rounded to 80) 90% 139 (rounded to 140) 85% 196 (rounded to 200) 80% 246 (rounded to 250) 75% 289 (rounded to 290)

Page 291 of 459

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. Table 2. Negative sample sizes and composition for different values of specificity claimed by the manufacturer for evaluation of Pf (single/combo) or Pv (single/combo) ELISA Specificity Sample size: Minimum number of negative samples # Composition of negative samples 99% 16 (rounded to 20) Dengue NS1/IgM positive samples: 03 Chikungunya IgM positive samples:03 Serum reactive for RA factor – low positive and high positive:02 Serum reactive for TPHA/other specific test for syphilis:02 Healthy controls from endemic regions: 10

95% 73 (rounded to 80) Dengue NS1/IgM positive samples: 10 Chikungunya IgM positive samples:10 Serum reactive for RA factor – low positive and high positive:10 Serum reactive for TPHA/other specific test for syphilis:10 Healthy controls from endemic regions: 40

90% 139 (rounded to 140) Dengue NS1/IgM positive samples: 18 Chikungunya IgM positive samples:18 Serum reactive for RA factor – low positive and high positive:18 Serum reactive for TPHA/other specific test for syphilis:18 Healthy controls from endemic regions: 68

85% 196 (rounded to 200) Dengue NS1/IgM positive samples: 25 Chikungunya IgM positive samples:25 Serum reactive for RA factor – low positive and high positive:25 Serum reactive for TPHA/other specific test for syphilis:25 Healthy controls from endemic regions: 100

80% 246 (rounded to 250) Dengue NS1/IgM positive samples: 30 Chikungunya IgM positive samples:30 Serum reactive for RA factor – low positive and high positive:30 Serum reactive for TPHA/other specific test for syphilis:30 Healthy controls from endemic regions: 130

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics.

  1. Repeatability and Reproducibility Assessment:

Page 292 of 459

under evaluation. 3 positive samples (comprising low and high analyte/parasitemic samples) and 3 negative samples per species (if applicable) should be tested 5 times in independent runs.

as follows:
comprising 10 low and 5 high analyte/parasitemic samples and 10 negative samples) per species (as applicable) )
10 low and 5 high analyte/parasitemic samples and 10 negative samples) per species (as applicable) )

Page 293 of 459

imprecision (if applicable) should be measured. 3 positive samples (comprising low and high analyte/parasitemic samples) and 3 negative samples per species (as applicable) should be tested 5 times in independent runs.
(qualitative).

consecutive days. 3 positive samples (comprising low and high analyte/parasitemic samples) and 3 negative samples per species (as applicable) should be tested 5 times in independent runs.
(qualitative).

positive samples (comprising low and high analyte/parasitemic samples) and 3 negative samples per species (as applicable) should be tested 5 times in independent runs.
(qualitative).

  1. Evaluation Methodology:

The index test should be tested on a rigorously well-characterized panel of samples from confirmed malaria positive and negative cases, which are also tested for the presence of malaria parasite using standardized Snounou protocol/microscopy/RDT or PCR Pre-Qualified by WHO or approved by US FDA/ATAGI Australia/PMDA Japan.

  1. Interpretation of results:

Results should be interpreted as per the IFU of the reference assay.

  1. Resolution of discrepant results:

True positive samples: These are well-characterized samples from confirmed malaria positive cases, which are also positive by the index test. True negative samples: These are well-characterized samples from confirmed malaria negative cases, which are also negative by the index test. False positive samples: These are well-characterized samples from confirmed malaria negative cases, which are positive by the index test. False negative samples: These are well-characterized samples from confirmed malaria positive cases, which are negative by the index test.

Page 294 of 459

  1. Acceptance Criteria:

Type of assay Acceptance criteria Minimum no. of samples needed to achieve at least the performance characteristics outlined in the acceptance criteria

Malaria antibody ELISA Sensitivity: ≥90%
Specificity: ≥95% Minimum no. of Positive samples = 140 Minimum no. of Negative samples = 80 Pv ELISA Sensitivity: ≥75%
Specificity: ≥95% Minimum no. of Positive samples = 290 Minimum no. of Negative samples = 80 Pf ELISA Sensitivity: ≥90%
Specificity: ≥95% Minimum no. of Positive samples = 140 Minimum no. of Negative samples = 80

lab, and the PI should maintain confidentiality of data. Refer to Fig. 2.

  1. Publication Rights:

author(s).

Page 295 of 459

acceptable.

VI. References:

  1. Snounou G, Viriyakosol S, Zhu XP, Jarra W, Pinheiro L, Do Rosario VE, et al. High sensitivity of detection of human malaria parasites by the use of nested polymerase chain reaction. Molecular and Biochemical Parasitology. 1993;61:315–20.
  2. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability
  3. World Health Organization. Methods manual for laboratory quality control testing of malaria rapid diagnostic test. Version Ten; 2023

VII.

Page 296 of 459

REPORT FORMAT Name of the Laboratory Name of the Institute, (with station) Certificate of Analysis File No.: _________________________________________ Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: low/high parasitemia/simulated samples)
Negative samples (provide details, including cross reactivity panel/simulated samples)

Page 297 of 459

Results:

Samples with confirmed disease status

Positive Negative Total Name of malaria ELISA kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity o Invalid test rate ● Conclusions: o Sensitivity, specificity o Invalid test rate o Performance: Satisfactory / Not Satisfactory (Sensitivity and specificity have been assessed in controlled lab setting on ……. samples only, using kits provided by the manufacturer from the batch mentioned above. Results should not be extrapolated for any other sample type.) Disclaimers Note: This report is exclusively for ………….. Kit (Lot No……) manufactured by …………… (Supplied by ……….)

Signature of Director/ Director-In-charge ……………………

Seal ………………………………………

Page 298 of 459

Performance evaluation protocol for Malaria real-time PCR kits I. Background:
CDSCO/ICMR, New Delhi, have aimed to facilitate the evaluation and supply of Quality- Assured Diagnostics kits appropriate for use in India. Hence the following guidelines shall establish the uniformity in performance evaluation of in-vitro diagnostic kits (IVD). The performance evaluation is to independently verify the manufacturer’s claim regarding IVD kit performance. II. Purpose:
To evaluate the performance characteristics of Malaria real-time PCR (RT-PCR) kits using irreversibly de-identified leftover archived clinical samples. III. Requirements:

  1. Instructions for use (IFU)
  2. Supply of kits under evaluation (with batch no. and lot no. ; Manufacturing and Expiry and other required details). If the kit to be evaluated works in a closed system format, the manufacturer needs to supply the required equipment.
  3. Evaluation sites/laboratories (With required equipment)
  4. Reference test kits
  5. Characterised Evaluation panel
  6. Laboratory supplies

Performance evaluation activities using irreversibly de-identified leftover clinical samples are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024.

V. Procedure: clinical samples. through
a) Laboratory accreditation: Accreditation for at least one of the Quality management systems (accreditation for Testing Lab / Calibration Lab (ISO/IEC: 17025), Medical Lab (ISO: 15189), PT provider (ISO/IEC: 17043) or CDSCO approved Reference laboratory. b) It is recommended that malaria Medical Device Testing Labs (MDTLs) participate in Quality Control exercises such as EQAP (External Quality Assurance Programme).

Page 299 of 459

c) Staff training: All the staff involved in IVD kit evaluation should undergo hands-on training and competency testing on the following at suitable malaria labs before initiation of MDTL activity:
 Preparation and characterization of evaluation panel for the respective IVD kit.  Management of RDT kits (specific for Plasmodium falciparum / Plasmodium vivax) received for performance evaluation (Verification/Storage/Unpacking etc.).  Perform tests interpretation and documentation of results, and reporting.  Data management and safety and confidentiality. 3. Preparation of evaluation sample panel for Malaria To evaluate the performance of malaria RT-PCR IVD kit, a well characterized species specific malaria whole genome panel is required. Hence, statistically significant number of whole blood samples should be used. The panel should comprise positive and negative samples as described in section 8. The reference sample panel should be stored in appropriate storage conditions (depending on the sample type and planned storage duration), and the quality of the panel should be checked periodically (at least once a year) with appropriate tests (e.g.: reference test/parasite culture/enzymatic activity/other relevant test). Malaria positive samples should be obtained from health facilities, including tertiary care centers and their linked hospitals, private clinics, field practice areas etc.
4. DNA extraction DNA extraction should be performed using a standard protocol/kit as recommended by the manufacturer, or fully automated DNA extractor may be used (as per manufacturer’s instruction and compatible reagent kits).
Note: If the manufacturer of the index test recommends a specific DNA extraction kit, it needs to be provided by the manufacturer, if the evaluation lab is unable to procure the same. 5. Real-time PCR system: 6. Internal Control/Extraction Control: The index test must have an internal control (housekeeping gene), with or without an extraction control. 7. Reference assay:
Two WHO Pre-Qualified/ US FDA/ ATAGI Australia/ PMDA Japan-approved malaria RT- PCR assays (or one approved assay from the aforementioned regulatory authorities and

Page 300 of 459

standardized Snounou protocol) should be used as reference assays for the characterization of samples, with 100% agreement between their results.
All positive samples should be confirmed positive by the reference assay(s). All negative samples should be confirmed negative by the reference assay(s).

  1. Sample size and sample panel composition for performance evaluation:
    Sample sizes of positive and negative samples of each species targeted by the kit against different values of sensitivity and specificity are provided in Table 1 and Table 2, with recommended composition. Sample sizes have been calculated assuming 95% level of significance, an absolute precision of 5%, and invalid test rate of 5%. Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sensitivity/specificity is not present in the table, the manufacturer needs to consider the sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ specificity available in the table). For example, if a manufacturer claims a sensitivity of 93%, they are required to use a sample size mentioned against 90% sensitivity. Similarly, a claim of 87% specificity would require usage of the sample size outlined for 85% specificity. Sample sizes are calculated using the formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96).

Table 1. Positive sample sizes (per species) and composition for different values of sensitivity claimed by the manufacturer for evaluation of malaria real time PCR kit Sensitivity Sample size: Minimum number of positive samples per species# Composition of positive samples 99% 16 (rounded to 20)

Page 301 of 459

95% 77 (rounded to 80) 90% 146 (rounded to 150) 85% 207 (rounded to 210) 80% 259 (rounded to 260) 75% 304 (rounded to 310) #It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics. Table 2. Negative sample sizes and composition for different values of specificity claimed by the manufacturer for evaluation of malaria real time PCR kit Specificity Sample size: Minimum number of negative samples per species# Composition of negative samples 99% 16 (rounded to 20) Dengue NS1/IgM positive samples: 03 Chikungunya IgM positive samples:03 Serum reactive for RA factor – low positive and high positive:02 Serum reactive for TPHA/other specific test for syphilis:02 Healthy controls from endemic regions: 10

95% 77 (rounded to 80) Dengue NS1/IgM positive samples: 10 Chikungunya IgM positive samples:10 Serum reactive for RA factor – low positive and high positive:10 Serum reactive for TPHA/other specific test for syphilis:10 Healthy controls from endemic regions: 40

90% 146 (rounded to 150) Dengue NS1/IgM positive samples: 18 Chikungunya IgM positive samples:18 Serum reactive for RA factor – low positive and high positive:18 Serum reactive for TPHA/other specific test for syphilis:18 Healthy controls from endemic regions: 78

85% 207 (rounded to 210) Dengue NS1/IgM positive samples: 26 Chikungunya IgM positive samples:26 Serum reactive for RA factor – low positive and high positive:26 Serum reactive for TPHA/other specific test for syphilis:26 Healthy controls from endemic regions: 106

80% 259 (rounded to 260) Dengue NS1/IgM positive samples: 35 Chikungunya IgM positive samples:35 Serum reactive for RA factor – low positive and high positive:30 Serum reactive for TPHA/other specific test for syphilis:30 Healthy controls from endemic regions: 130

#It is recommended to calculate the sample size as per manufacturer's claims of sensitivity and specificity; however, a higher sample size is suggested to ensure adequate power of the study in case the kit falls short of claimed performance characteristics.

Page 302 of 459

Sample panel composition:

A. Positive samples: RDT/microscopy positive malaria samples should be obtained from health facilities and confirmed using two USFDA/ATAGI Australia/PMDA Japan approved/WHO Pre-Qualified PCR Kits (or one approved kit from any of the aforementioned authorities and standardized Snounou protocol). Once the positive samples are well-characterized with these two PCR assays (100% agreement between results), they should be classified as per their parasite load using microscopy and/or other relevant test format.
Range of Parasitemia: Panel members should have a low (≤200 parasites per microliter) to high (≥2000 parasites per microliter) range of Plasmodium falciparum, P. vivax, as obtained from microscopy and/or other relevant test results. Characterized panels must contain equal number of samples of both low and high parasitemia. Note for additional characterization (not mandatory):
If ELISA is used for characterization of samples in addition to the above-mentioned mandatory tests (RDT/microscopy AND PCR), consistent ELISA quantification results should be obtained in ≥3 runs of ELISA experiments performed for each of the three common antigens (PfHRP2, LDH and aldolase), with the results obtained at the 200 p/µL and the 2,000 p/µL being consistent with each other as well (factor of roughly 10 between results). The limit of detection of PfHRP2 is 0.6-74 ng/ mL, PvLDH is 1.6-47.9 ng/ mL, PfLDH is 0.2-53.5 ng/mL, and Pf aldolase is 0-9.9 ng/mL.
** If the pool of samples available for testing is sufficiently large in numbers, then the antigen concentration range at the 200 p/µL dilution should be restricted to 5-9.5 ng/mL for PfHRP2, 15-47.9 ng/mL for PvLDH, 10.8-53.5 ng/mL for PfLDH, and 1.7- 15ng/mL for Pf aldolase. If ELISA for other antigen is carried out for further characterization (for which limits of detection have not been established), testing should be carried out on calibrated equipment in ≥3 runs of ELISA, leading to their classification as low and high parasitemic samples or strong/moderate/weakly reactive samples.
Negative samples: These are samples negative by RDT/microscopy AND two USFDA/ATAGI Australia/PMDA Japan approved/WHO Pre-Qualified PCR Kits (or one approved kit from any of the aforementioned authorities and standardized Snounou protocol). The above-mentioned activities should not be performed with spiked/contrived samples. If the kit does not differentiate between Pf and Pv, performance characteristics may be evaluated for the kit as a whole with statistically significant sample size according to different performance metrics outlined in Tables 1 and 2, with equal representation of samples positive for P.falciparum and /P.vivax. It is desirable (but not mandatory) to evaluate the kit using some positive samples from mixed Pf/Pv infection cases.

Page 303 of 459

under evaluation. 3 positive samples (comprising low and high parasitemic samples) and 3 negative samples per target pathogen should be tested 5 times in independent runs.

Three lots of an IVD kit shall be evaluated. Sample size for lot-to-lot reproducibility

comprising 10 low parasitemic AND 5 high parasitemic samples, and 10 negative samples) per target pathogen.
10 low parasitemic AND 5 high parasitemic samples, and 10 negative samples) per target pathogen.

Page 304 of 459

imprecision (if applicable) should be measured. 3 positive samples (comprising low and high parasitemic samples) and 3 negative samples per target pathogen should be tested 5 times in independent runs.
(qualitative).

consecutive days. 3 positive samples (comprising low and high parasitemic samples) and 3 negative samples per target pathogen should be tested 5 times in independent runs.
(qualitative).

positive samples (comprising low and high parasitemic samples) and 3 negative samples per target pathogen should be tested 5 times in independent runs.
(qualitative).

Page 305 of 459

recommended). 10. Testing Methodology:

The reference assay and the index test should be run on the sample panel in parallel.

  1. Interpretation of results:

Results should be interpreted as per the IFU of the reference assay and the index test.

  1. Resolution of discrepant results:

True positive samples: These are samples positive by both reference assay and index test. True negative samples: These are samples negative by both reference assay and index test. test. test.

  1. Acceptance Criteria:

Target Plasmodium species
Acceptance criteria Minimum no. of samples needed to achieve at least the performance characteristics outlined in the acceptance criteria Pf PCR Sensitivity ≥98% Specificity ≥98% Limit of detection: 1 parasite/µl Minimum no. of Positive samples = 80

Minimum no. of Negative samples = 80 Pv PCR Sensitivity ≥95% Specificity ≥98% Limit of detection: 1-2 parasites/µl Minimum no. of Positive samples = 80

Minimum no. of Negative samples = 80 Multiplex PCR - Pf & Pv
For Pf:
 Sensitivity: ≥98%
 Specificity: ≥98%  Absolute precision 5%
 95% CI  Invalid test rate ≤5%  Limit of detection: 1 parasite/µl

For Pv:  Sensitivity: ≥95%
 Specificity: ≥98%  Absolute precision 5%
 95% CI  Invalid test rate ≤5% For Pf: Minimum no. of Positive samples = 80

Minimum no. of Negative samples = 80

For Pv: Minimum no. of Positive samples = 80

Page 306 of 459

 Limit of detection: 1-2 parasites/µl

Minimum no. of Negative samples = 80

  1. Blinding of laboratory staff:

lab, and the PI should maintain confidentiality of data. Refer to Fig. 2.

  1. Publication Rights

author(s).

Page 307 of 459

acceptable.

VI. References:

  1. Snounou G, Viriyakosol S, Zhu XP, Jarra W, Pinheiro L, Do Rosario VE, et al. High sensitivity of detection of human malaria parasites by the use of nested polymerase chain reaction. Molecular and Biochemical Parasitology. 1993;61:315–20.
  2. Ramírez AM, Tang THT, Suárez ML, Fernández AÁ, García CM, Hisam S, Rubio JM. Assessment of Commercial Real-Time PCR Assays for Detection of Malaria Infection in a Non- Endemic Setting. Am J Trop Med Hyg. 2021 Oct 12;105(6):1732-1737. doi: 10.4269/ajtmh.21-
  3. PMID: 34662870; PMCID: PMC8641344.
  4. Bouzayene, A., Zaffaroullah, R., Bailly, J. et al. Evaluation of two commercial kits and two laboratory-developed qPCR assays compared to LAMP for molecular diagnosis of malaria. Malar J 21, 204 (2022). https://doi.org/10.1186/s12936-022-04219-1
  5. Aschar M, Sanchez MCA, Costa-Nascimento MJ, Farinas MLRN, Hristov AD, Lima GFMC, Inoue J, Levi JE, Di Santi SM. Ultrasensitive molecular tests for Plasmodium detection: applicability in control and elimination programs and reference laboratories. Rev Panam Salud Publica. 2022 Mar 28;46:e11. doi: 10.26633/RPSP.2022.11. PMID: 35355692; PMCID: PMC8959250.
  6. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability
  7. World Health Organization. Methods manual for laboratory quality control testing of malaria rapid diagnostic test. Version Ten; 2023

VII.

Page 308 of 459

REPORT FORMAT Name of the Laboratory Name of the Institute, (with station) Certificate of Analysis File No.: _________________________________________ Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: low/high parasitemia/simulated samples)
Negative samples (provide details, including cross reactivity panel/simulated samples)

Page 309 of 459

Results:

Reference assay ……….……………… (name)

Positive Negative Total Name of malaria real time PCR kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity o Invalid test rate ● Conclusions: o Performance: Satisfactory / Not Satisfactory (Sensitivity and specificity have been assessed in controlled lab setting on ……. samples only, using kits provided by the manufacturer from the batch mentioned above. Results should not be extrapolated for any other sample type.) Disclaimers Note: This report is exclusively for ………….. Kit (Lot No……) manufactured by …………… (Supplied by ……….)

Signature of Director/ Director-In-charge ……………………

Seal ………………………………………

Page 310 of 459

Field evaluation protocol for combo Malaria Rapid Diagnostic Test (RDT) kits (detecting P vivax and P falciparum) I. Background:
CDSCO/ICMR, New Delhi, have aimed to facilitate the availability of Quality-Assured Diagnostics kits appropriate for use in India. Hence the following guidelines shall establish the evaluation is to independently verify the manufacturer’s claim regarding IVD kit performance. II. Purpose: To evaluate the performance characteristics of Malaria RDT kits (detecting P. vivax and/or P. falciparum) in the diagnosis of Malaria parasite infection in individuals with unknown disease status. III. Requirements:

  1. Supply of kits under evaluation (with batch no. and lot no. Manufacturing and Expiry dates other required details). If the kit to be evaluated works in a closed system format, the manufacturer needs to supply the required equipment. The study will be initiated after approval from the institutional human ethics committee.
    V. Procedure:
  2. Study design/type: Cross-sectional study Identified IVD kit evaluation laboratories should establish their proficiency through
    A. Laboratory accreditation: Accreditation for at least one of the Quality management systems (accreditation for Testing Lab / Calibration Lab (ISO/IEC: 17025), Medical Lab (ISO: 15189), PT provider (ISO/IEC: 17043) or CDSCO approved Reference laboratory. It is recommended that malaria Medical Device Testing Labs (MDTLs) participate in Quality Control exercises such as EQAP (External Quality Assurance Programme). training and competency testing on the following at suitable malaria labs before initiation of MDTL activity:
    ⮚ Preparation and characterization of evaluation panel for the respective IVD kit. ⮚ Management of RDT kits (specific for Plasmodium falciparum / Plasmodium vivax) received for performance evaluation (Verification/Storage/Unpacking etc.).

Page 311 of 459

⮚ Perform tests interpretation and documentation of results, and reporting. ⮚ Data management and safety and confidentiality. 3. Sample size for performance evaluation: Sample sizes of positive and negative samples against different values of sensitivity and 95% level of significance, an absolute precision of 5%, and invalid test rate 5%. It is further assumed that at least 5% of the individuals attending the health care facilities for acute febrile illness and suspected for Malaria will be positive for Malaria (P. vivax and P. falciparum). Appropriate sample size has to be chosen from the tables according to the values of sensitivity and specificity being claimed by the manufacturer. If a claimed sample size associated with the largest sensitivity/specificity provided in the table that is smaller to the claimed value (that is, as per the next smaller value of the sensitivity/ a claim of 87% specificity would require usage of the sample size outlined for 85% specificity. Sample sizes are calculated using the following formulae and assumption of 5% for prevalence of the disease:

𝑑2 x (1 −IR) x P 𝑑2 x (1 −IR) x P

· n (se) is the minimum number of individuals to be enrolled to obtain the requisite number of positive samples. · n (sp) is the minimum number of individuals to be enrolled to obtain the requisite number of negative samples. · Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96).

P is prevalence of the disease

Sample size has to be calculated based on both the sensitivity and the specificity. The final sample size will be the maximum of the two. For example, at 95% sensitivity and 95% specificity, the sample size required will be 1600 (maximum of 1600 and 84). Please note that since the prevalence is low, the final sample size is generally expected to be governed by the assumed sensitivity.

Page 312 of 459

Table 1. Sample sizes for different values of species-specific sensitivity being claimed
Sensitivity Minimum no. of positive samples required (rounded figure) # Minimum number of individuals to be enrolled in the study to obtain requisite number of positive samples 99% 20 400 95% 80 1600 90% 150 3000 85% 210 4200 80% 260 5200 75% 305 6100 #It is recommended to calculate the sample size as per manufacturer's claims of sensitivity Samples will be collected from individuals attending the health care facilities (tertiary care centers and their linked hospitals, private clinics, field practice areas etc.) for acute febrile illness in highly endemic areas.

The disease status of these cases will be unknown.

Table 2. Sample sizes for different values of species-specific specificity being claimed Specificity No. of negative samples required (rounded figure) Minimum number of individuals to be enrolled to obtain requisite number of negative samples 99% 20 21 95% 80 84 90% 150 158 85% 210 221 80% 260 274 75% 305 321 #It is recommended to calculate the sample size as per manufacturer's claims of sensitivity Samples will be collected from individuals attending the health care facilities (tertiary care centers and their linked hospitals, private clinics, field practice areas etc.) for acute febrile illness in highly endemic areas.

Since a large number of febrile cases have to be enrolled to obtain the requisite number of malaria positive samples, enrolling the number of cases mentioned in Table 1 will be sufficient to obtain the requisite number of negative samples.

Page 313 of 459

  1. Inclusion criteria: Individuals with the following clinical features may be enrolled in the study Fever and any 2 of the following: o Chills, sweating, headache, tiredness, nausea and vomiting, jaundice, splenomegaly

  2. Exclusion criteria  Individuals not satisfying inclusion criteria  Individuals with already known positive history for other pathogens

  3. Reference assay:
    WHO Pre-Qualified/ US FDA/ ATAGI Australia/ PMDA Japan approved Malaria PCR assay/ standardized Snounou protocol should be used as reference assay.

  4. Study implementation: The patients displaying Malaria like illness will be recruited into the study and five ml of whole blood will be collected in EDTA tubes. The whole blood sample will be subjected to the reference and the index test. The disease status of the enrolled cases will be unknown.

  5. Evaluation method:

  6. Interpretation of results:

  7. Positive samples: Samples positive by the reference assay will be considered as true positive samples.

  8. Negative samples:
    Samples negative by the reference assay will be considered as true negative samples. test. test.

  9. Cross reactivity: The RDT kit should have been evaluated against the following cross reactivity panel during the analytical performance evaluation:

Page 314 of 459

 Dengue NS1 positive samples (n=10 samples)  Chikungunya PCR positive samples (n=10 samples)  Healthy controls from endemic regions (n= 40 samples)  Serum reactive for RA factor – low positive and high positive (n=15 samples)  Serum reactive for TPHA/other specific test for syphilis (n= 10 samples) 13. Statistical analysis:
Sensitivity and specificity will be calculated. Interim analysis of data shall be conducted on completing evaluation of 25%, 50% and 75% of samples. If, at any point, the performance of the assay is found to be not satisfactory, the assay shall not be evaluated further. Evaluation fee shall be charged accordingly. 14. Repeatability and Reproducibility Assessment:

This should be done to assess the repeatability of the detection of target analyte using the kit under evaluation. 3 positive samples (comprising low parasitemic AND high parasitemic samples) and 3 negative samples per species should be tested 5 times in independent runs.

as follows:
comprising 10 low parasitemic AND 5 high parasitemic samples) per species, and 10 negative samples).
10 low parasitemic AND 5 high parasitemic samples) per species, and 10 negative samples).

Page 315 of 459

imprecision (if applicable) should be measured. 3 positive samples (comprising low parasitemic AND high parasitemic samples) and 3 negative samples per species (qualitative).

consecutive days. 3 positive samples (comprising low parasitemic AND high parasitemic samples) and 3 negative samples per species should be tested 5 times in independent runs.
(qualitative).

Page 316 of 459

positive samples (comprising low parasitemic AND high parasitemic samples) and 3 negative samples per species should be tested 5 times in independent runs.
(qualitative).

  1. Resolution of discrepant results:

True positive samples: These are samples positive by both reference assay and index test. True negative samples: These are samples negative by both reference assay and index test. test. test.

  1. Blinding of laboratory staff

blinded to the results of the reference assay. The PI of the evaluation exercise should remain unblinded, i.e., privy to the results of the reference test. Another senior laboratory staff selected by the PI may remain unblinded for overseeing the activity and maintaining the database of results. The data should be analyzed only by the PI of the evaluating lab, and the PI should maintain confidentiality of data.

  1. Acceptance criteria:

Sensitivity: ≥75% for P. vivax and ≥95% for P. falciparum Specificity: ≥90% for P. vivax and ≥95% for P. falciparum

To achieve at least the performance characteristics outlined in the acceptance criteria for P vivax, ≥6100 individuals satisfying the case definition need to be enrolled to obtain the requisite number of positive samples. This sample size is sufficient for requisite number of negative samples. To achieve at least the performance characteristics outlined in the acceptance criteria for P falciparum, ≥1600 individuals satisfying the case definition need to be enrolled to obtain the requisite number of positive samples. This sample size is sufficient for requisite number of negative samples.

Recruitment should be terminated once the desired number of positive cases is enrolled and tested.

  1. Publication Rights

author(s).

Page 317 of 459

acceptable.

VI. References:

  1. Ministry of Health and Family Welfare. Guidelines for Bivalent RDT. Available at: guidelines- for-bivalent-rdt.pdf (mohfw.gov.in)
  2. World Health Organization. Malaria Rapid Diagnostic Test Performance - Results of WHO product testing of malaria RDTs: round 8 (2016–2018): Available at: https://iris.who.int/bitstream/handle/10665/276190/9789241514965-eng.pdf?sequence=1
  3. Snounou G, Viriyakosol S, Zhu XP, Jarra W, Pinheiro L, Do Rosario VE, et al. High sensitivity of detection of human malaria parasites by the use of nested polymerase chain reaction. Molecular and Biochemical Parasitology. 1993;61:315–20.
  4. Integrated Disease Surveillance Project Training Manual For State & District Surveillance Officers - Case Definitions Of Diseases & Syndromes Under Surveillance (Module-5). Available at: https://idsp.mohfw.gov.in/WriteReadData/OldSite/2WkDSOSept08/Resources_files/DistrictS urvMan/Module5.pdf [Accessed on 25th June 2024]
  5. CDC. National Notifiable Diseases Surveillance System (NNDSS). Malaria (Plasmodium spp.) 2014 Case Definition. Available at: https://ndc.services.cdc.gov/case-definitions/malaria-2014/ [Accessed on 28th June, 2024]
  6. Kannambath R, Rajkumari N, Sivaradjy M. Prevalence of malaria: A 7-year trend analysis from a tertiary care center, Puducherry. Trop Parasitol. 2023 Jan-Jun;13(1):28-33. doi: 10.4103/tp.tp_41_22. Epub 2023 May 19. PMID: 37415756; PMCID: PMC10321582.
  7. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

Page 318 of 459

REPORT FORMAT Name of the Laboratory Name of the Institute, (with station) Certificate of Analysis File No.: _________________________________________ Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use Valid Up to: Application No. Sample Panel Sample type Positive samples: Not applicable, may categorize cases as per duration of illness Negative samples (may categorize as per duration of illness, must include cross reactivity panel)

Page 319 of 459

Results:

Reference assay ……….……………… (name)

Positive Negative Total Name of index malaria RDT Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity o Invalid test rate ● Conclusions: o Performance: Satisfactory / Not Satisfactory (Sensitivity and specificity have been assessed in field/controlled lab setting on……. samples only, using kits provided by the manufacturer from the batch mentioned above. Results should not be extrapolated for any other sample type.) Disclaimers Note: This report is exclusively for ………….. Kit (Lot No……) manufactured by …………… (Supplied by ……….)

Signature of Director/ Director-In-charge …………………… Seal ………………………………………

Page 320 of 459

Information on Operational and Test Performance Characteristics Required from Manufacturers for Malaria IVD

  1. Instructions for Use
  2. Scope of the IVD: to diagnose Malaria (Pf and/or Pv)
  3. Intended Use Statement
  4. Principle of the assay
  5. Intended testing population (cases of acute febrile illness/suspected cases of Malaria)
  6. Intended user(laboratory professional and/or health care worker at point-of-care)
  7. Detailed test protocol
  8. Lot/batch No.
  9. Date of manufacture
  10. Date of Expiry
  11. Information on operational Characteristics i. Configuration of the kit/device
  12. Information on Test Performance Characteristics i. Type of sample-serum/plasma/whole blood/other specimen (specify) iii. Sample rejection criteria (if any) v. Any additional device/consumable like sample transfer device, pipette, tube, etc required vii. Pathogens targeted by the kit

Page 321 of 459

x. Manual or automated (equipment)reading xi. Limit of detection/Limit of Quantification and range of detection xiv. Stability and reproducibility (including data) xv. Training required for testing (if any) xviii. Details of cross reactivity, if any xix. Interpretation of invalid and indeterminate results to be provided xx. It is recommended to provide data demonstrating accuracy and precision xxi. Limit of detection

*Please mention “Not applicable” against sections not pertaining to the kit.

Page 322 of 459

NIPAH VIRUS REAL TIME PCR KIT

Page 323 of 459

A. Working Group:

  1. Dr. Anita Shete-Aich, Scientist-E, ICMR National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  2. Dr. Labanya Mukhopadhyay, Scientist-C, Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  3. Ms. Krittika Bhattacharyya, Statistical Officer (Planning), Directorate of Economics and Statistics, Government of National Capital Territory of Delhi

  4. Dr. Vasanthapuram Ravi, Former Dean Research and Head of Neurovirology, National Institute of Mental Health and Neuro-Sciences, Bengaluru, Karnataka

  5. Dr. Pragya D Yadav, Scientist-F, ICMR National Institute of Virology and Director-in- Charge, National Institute of One Health, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  6. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control

  7. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control

  8. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central

  9. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family

Page 324 of 459

Performance evaluation protocol for Nipah virus real-time PCR kit

I. Background: Diagnostics kits appropriate for use in India. Hence the following guidelines shall establish the evaluation is to independently verify the manufacturer’s claim regarding in-vitro diagnostic kit (IVD) performance. This recommendation focuses on the laboratory performance evaluation of Nipah virus real time PCR kit. All clinical samples tested in the study should be evaluated in accordance with the candidate test’s instructions for use.
II. Purpose: To evaluate the performance characteristics of Nipah virus real-time PCR kits in the diagnosis of Nipah virus infection/ disease using irreversibly de-identified leftover archived/ spiked clinical samples. III. Requirements:
Investigators are required to submit a self-declaration form, as outlined in the ICMR guidelines, to the institutional authorities and ethics committee for information. V. Procedure:

  1. Study design/type: Diagnostic accuracy study using spiked/clinical samples (human specimens). Identified IVD kit evaluation laboratories should be well-equipped and establish their proficiency through ALL of the following:

Page 325 of 459

A. Availability of BSL-4 facility for handling of Nipah virus positive specimens B. Accreditation for at least one Quality management system for at least one respiratory viral pathogen molecular testing (accreditation for Testing Lab / Calibration Lab as per ISO/IEC 17025, Medical Lab as per ISO 15189, PT provider as per ISO/IEC 17043), or CDSCO approved Reference laboratory.
C. Staff training: All the staff involved in Nipah virus IVD evaluation should undergo hands on training and competency testing on following
⮚ BSL-4 practices
⮚ Nipah virus culture and handling ⮚ Preparation & characterization of reference sample panel ⮚ Handling of Nipah virus RT-PCR kits received for performance evaluation 3. Preparation of Nipah virus RNA evaluation panel This is a zoonotic disease, and well characterised Nipah virus positive human samples is a critical requirement for evaluation of RT-PCR IVD kits. A statistically significant number of clinical samples should be used for the evaluation. 4. RNA extraction RNA extraction should be performed as per reference test and index test Instructions for Use (IFU). 5. Real-Time PCR System 6. Internal control/Extraction control Assays must have an internal control (housekeeping gene), with or without an extraction control (RNA added before extraction to a sample).

Page 326 of 459

  1. Reference assay:
    The Nipah virus Real Time PCR Assay developed by ICMR-NIV Pune, or a WHO Pre- Qualified/ US FDA/ ATAGI Australia/ PMDA Japan approved real time PCR assay should be used as the Reference Standard.
    All positive samples should be confirmed positive by the reference assay. All negative samples should be confirmed negative by the reference assay.

  2. Sample size for performance evaluation: Sample size is calculated assuming 95%
    sensitivity and specificity of the index test, 95% confidence level, absolute precision of 5% and ≤5% invalid test rate. A minimum of 77 (rounded to 80) positive clinical samples and a minimum of 77 (rounded to 80) negative clinical samples are required. Sample sizes are calculated using the formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96).

Nipah virus is detectable from throat swab, urine, CSF. The assay should be validated with positive clinical/spiked samples, and negative samples for all the formats claimed by the manufacturer. However, if a particular sample matrix is used to evaluate the assay (as opposed to all the sample types claimed by the manufacturer), the performance evaluation report should clearly mention the performance characteristics of the assay against the sample type used for validation. There should be no ambiguity about the sample type used for assay validation.

  1. Sample panel composition: A. Human samples A.1 Positive samples (Minimum n=80 for each sample type): Clinical/ Spiked samples positive by the reference real-time PCR assay
    A.1.1 Strong positive (Ct value <25) = 24 samples

Page 327 of 459

A.1.2. Moderate positive (Ct value between 25-30) = 28 samples A.1.3 Weak positive (Ct value >30 to 34) = 28 samples
The sample type should be as per the index test IFU. If an assay claims to detect Nipah virus RNA in several sample types, attempt should be made to use 80 positive samples across each sample type, or at least the sample types available with the evaluating lab. This relaxation is provided since clinical samples are scarce and obtained only during outbreaks occurring every few years in India, which necessitates using spiked clinical samples. The latter is difficult since Nipah virus is a BSL-4 level pathogen and its handling requires sophisticated laboratory setup and trained manpower. In case the requisite number of specimens for a particular sample type are not available and a smaller number of samples are used for performance evaluation (i.e., sample size calculated assuming higher performance characteristics), it is necessary to ensure that the study has adequate power for acceptance of the evaluation results in case the assay falls short of the assumed performance characteristics. Note: If clinical samples positive for Nipah virus are not available, tissue culture fluid (Heat- inactivated) from reference laboratories can be used, spiked in serum/urine/Throat swab samples to obtain the panel with Ct value <25, 25-30 and >35 and tested by the reference assay, and the positive samples can be used for evaluation.
Confirmed negative samples would be used for spiking with Nipah virus.isolate.

A.2 Negative samples (number of samples will depend on sample type): All negative samples should be negative by reference real-time PCR assay. Distribution of the negative samples should be as follows

Page 328 of 459

Categories of samples as per the sample type Sample type
NP/TS (Minimum n= 80) Serum (Minimum n= 80) Urine (Minimum n=80) A.2.1 Samples from cases having similar illness/ spiked samples which are RT- PCR positive for common pathogens but negative for Nipah virus Samples from individuals presenting with ARI/ILI/SARI (n=45):

5 positive clinical/ spiked samples from each of the following diseases:

  1. Influenza A virus @
  2. Influenza B virus @
  3. SARS-CoV-2 @
  4. RSV A/B @
  5. HPIV @
  6. HMPV @
  7. Adenovirus @
  8. Seasonal Coronaviruses *
  9. Rhinovirus/Enterovirus*

Cross reactivity panel is arranged in descending order of priority. The pathogens marked @ are essentially to be tested. It is recommended to test for all pathogens listed in the cross reactivity panel. However, if there is an acute shortfall or non- availability of clinical samples, one may consider reducing only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross reactive sample panel remain the same by compensating with the available “essentially to be tested” samples. Samples from cases of AES (n=35):

5 positive clinical/ spiked samples from each of the following diseases:

  1. Japanese Encephalitis @
  2. Dengue @
  3. HSV @
  4. VZV @
  5. West Nile Virus *
  6. Chandipura virus *
  7. Rabies virus *

Cross reactivity panel is arranged in descending order of priority. The pathogens marked @ are essentially to be tested. It is recommended to test for all pathogens listed in the cross reactivity panel. However, if there is an acute shortfall or non-availability of clinical samples, one may consider reducing only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross reactive sample panel remain the same by compensating with the available “essentially to be tested” samples. 5 positive clinical/ spiked samples from each of the following diseases, presenting with respiratory and/or encephalitis symptoms (n=20):

  1. Measles
  2. Rubella
  3. Mumps
  4. SARS-CoV- 2

A.2.2 Samples from cases with acute respiratory disease/ acute encephalitis/ acute febrile illness and RT- PCR negative for the above- mentioned pathogens and Nipah virus 25 35 40

Page 329 of 459

A.2.3 Healthy/ asymptomatic cases

from endemic regions negative for Nipah virus 10
10 20 Serum/ throat swab/ urine samples collected from the same case may be used for evaluation. Commercially available validated standard panels that are accepted by accreditation agencies can also be used if the clinical sample with required target is not available for cross-reactivity analysis.

  1. Evaluation method:

  2. Interpretation of results:

  3. Resolution of discrepant results:

test. test.

under evaluation.

as follows:

Page 330 of 459

negative samples).
samples).

b. Inter-Operator variability: Testing should be conducted by two different operators, keeping all other testing parameters undisturbed. Within-run and between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

consecutive days. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

Page 331 of 459

IVD kit using two different manufacturer recommended platforms (if applicable). 3 positive samples (strong, moderate and weak positive samples)
(qualitative).

recommended). 14. Blinding of laboratory staff: lab, and the PI should maintain confidentiality of data. Refer to Fig. 2.

Page 332 of 459

  1. Acceptance Criteria Sensitivity: ≥95% Specificity: ≥98% Cross reactivity with other viruses as outlined in the negative sample panel: Minimal
  2. Publication Rights: author(s). acceptable.

VI. References:

  1. Yadav PD, Majumdar T, Gupta N, Kumar MA, Shete A, Pardeshi P, Sultana S, Sahay RR, Manoj MN, Patil S, Floura S, Gangakhedkar R, Mourya DT. Standardization & validation of Truenat™ point-of-care test for rapid diagnosis of Nipah. Indian J Med Res. 2021 Apr;154(4):645-649. doi: 10.4103/ijmr.IJMR_4717_20. PMID: 34854433; PMCID: PMC9205002.
  2. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification – Diagnostic Assessment TGS-3.

Available at: eng.pdf;sequence=1
3. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

VII.

Page 333 of 459

Performance evaluation report for Nipah virus real-time PCR kits site Name of supplier: Manufacturer/Importer/Port office of Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use

Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: clinical/spiked, strong, moderate, weak/simulated samples)
Negative samples (provide details (clinical/spiked,),

Results

Positive Negative Total Name of Nipah virus real-time PCR Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross reactivity o Invalid test rate

Page 334 of 459

Disclaimers Note: This report is exclusively for Nipah virus………….. Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal ………………………….

Page 335 of 459

Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers

  1. Instructions for Use
  2. Scope of the IVD:
  3. Intended Use Statement
  4. Principle of the assay
  5. Intended testing population (cases of AES/ARI/SARI)
  6. Intended user (laboratory professional and/or health care worker at point-of-care)
  7. Lot/batch No.
  8. Date of manufacture
  9. Date of Expiry
  10. Information on operational Characteristics i. Configuration of the kit/device
  11. Information on Test Performance Characteristics i. Type of sample- Nasopharyngeal swab/Throat swab/ CSF/Serum / Other specimen iii. Any specific sample NOT to be tested v. Any additional device/consumable like sample transfer device, pipette, tube, etc required vii. Pathogen(s) targeted by the kit

Page 336 of 459

x. Manual or automated(equipment)reading xi. Limit of detection/Limit of Quantification and range of detection xiv. Stability and reproducibility xv. Training required for testing xviii. Interpretation of invalid and indeterminate results to be provided xix. It is recommended to provide data demonstrating accuracy and precision

*Please mention “Not applicable” against sections not pertaining to the kit.

Page 337 of 459

CHANDIPURA VIRUS REAL TIME PCR KIT

Page 338 of 459

A. Working Group:

  1. Dr. Vijay P Bondre, Scientist-G, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  2. Ms. Krittika Bhattacharyya, Statistical Officer (Planning), Directorate of Economics and Statistics, Government of National Capital Territory of Delhi

  3. Dr. Labanya Mukhopadhyay, Scientist-C, Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  4. Dr. Vasanthapuram Ravi, Former Dean Research and Head of Neurovirology, National Institute of Mental Health and Neuro-Sciences, Bengaluru, Karnataka

  5. Dr. Reeta Mani, Professor and Head of Neurovirology, National Institute of Mental Health and Neuro-Sciences, Bengaluru, Karnataka

  6. Dr. Bhagirathi Dwibedi, Professor of Paediatrics, All India Institute of Medical Sciences, Bhubaneswar

  7. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control

  8. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control

  9. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central

  10. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family

Page 339 of 459

Performance evaluation protocol for Chandipura virus real-time PCR kits

I. Background
Diagnostics kits appropriate for use in India. Hence the following guidelines shall establish the evaluation is to independently verify the manufacturer’s claim regarding in-vitro diagnostic kit (IVD) performance. This recommendation focuses on the laboratory performance evaluation of Chandipura virus (CHPV) virus real time PCR kit. All clinical samples tested in the study should be evaluated in accordance with the candidate test’s instructions for use.
II. Purpose: To evaluate the performance characteristics of CHPV real-time PCR kits in the diagnosis of CHPV infection/ disease using irreversibly de-identified leftover archived/ spiked clinical samples.

III. Requirements:

  1. Kits Under Evaluation: Include detailed information such as batch number, lot number, expiry date, and other relevant specifications. For kits designed to operate within a closed system, manufacturers must provide the necessary equipment and consumables for testing.
  2. Evaluation Sites/Laboratories: Identify laboratories equipped with the required instruments and infrastructure to conduct the evaluation.
  3. Reference Test Kits: Use reference kits or in-house kits developed by the reference laboratory, which have been validated to demonstrate satisfactory performance.
  4. Evaluation Panel: Prepare a panel of well-characterised clinical samples from confirmed cases or spiked samples for a comprehensive evaluation.
  5. Laboratory Supplies: Ensure all necessary laboratory materials and supplies are available for the evaluation process.

IV. Ethical Approvals:
Performance evaluation activities using irreversibly de-identified clinical samples are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Investigators are required to submit a self-declaration form, as outlined in the ICMR guidelines, to the institutional authorities and ethics committee for information.

Page 340 of 459

V. Procedure:

  1. Study design/type: Diagnostic accuracy study using irreversibly de-identified archived clinical/spiked samples
    Identified IVD kit evaluation laboratories should establish their proficiency through the following: A) Accreditation for at least one of the Quality management systems, such as  Testing Laboratory or Calibration Laboratory (ISO/IEC 17025)  Medical Laboratory (ISO 15189)  Proficiency Testing Provider (ISO/IEC 17043) OR  CDSCO-approved reference laboratory B) Staff training: All staff involved in IVD kit evaluation process should undergo hands on training and competency assessment in the following areas:
     Preparation and characterization of kit evaluation panel
     Handling of Chandipura real-time PCR kits received for performance evaluation (verification/storage/unpacking etc.).
     Testing procedures, interpretation and recording of results, and reporting
     Data handling, data safety & confidentiality
  2. Preparation of Chandipura RNA evaluation panel:
    A well characterised panel of CHPV positive clinical samples is a critical requirement for evaluation of these RT-PCR IVD kits. A statistically significant number of clinical samples should be used for the evaluation. The sample type for CHPV detection is Cerebrospinal fluid (CSF) and serum. If a kit claims to detect CHPV in both sample types, attempt should be made to evaluate the assay across both serum and CSF using statistically significant sample size for each sample type. In case all the sample types mentioned in the IFU are not available with the lab, the performance evaluation report should clearly mention the sample type against which the kit is evaluated, ensuring statistical rigor. There should be no ambiguity about the type of sample used for evaluation.
  3. RNA extraction:
    RNA extraction should be performed as per reference test and index test Instructions for Use (IFU).

Page 341 of 459

  1. Real-time PCR system: PCR should be performed using IVD-approved machines. If any equipment(s) is specified in the IFU of the index test, it should be used for the evaluation, and it should be provided by the manufacturer if not available within the lab’s IVD evaluation scope.
    Real-time closed systems/devices awaiting evaluation should be provided by the manufacturer along with all necessary components, supplies and reagents.
  2. Internal Control/Extraction Control: The index test must have an internal control (housekeeping gene), with or without an extraction control (RNA added before extraction to a sample).
  3. Reference assay: A WHO Pre-Qualified/ US FDA/ ATAGI Australia/ PMDA Japan approved real time CHPV PCR assay/ ICMR-National Institute of Virology, Pune developed protocol for detection of Chandipura virus RNA will serve as the reference assay.
    All positive samples should be confirmed positive by the reference assay. All negative samples should be confirmed negative by the reference assay and CHPV IgM.
  4. Sample size for performance evaluation: Sample size is calculated assuming 95% sensitivity and specificity of the index test, 95% confidence level, absolute precision of 5% and ≤5% invalid test rate. A minimum of 77 (rounded to 80) positive clinical samples and a minimum of 77 (rounded to 80) negative clinical samples for each sample type are required for performance evaluation. Sample sizes are calculated using the formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96).

Page 342 of 459

  1. Sample panel composition: A) Positive samples (Minimum n=80 for each sample type): These samples should be clinical/spiked samples positive by reference real-time PCR assay and preferably represent all genetic variants. The distribution of samples should be as follows: Characteristic of positive sample Minimum no. of serum samples needed (for kits detecting CHPV in serum) Minimum no. of CSF samples needed (for kits detecting CHPV in CSF) A.1 Strong positive [Ct value ≤ 25] 24 24 A.2 Moderate positive [Ct value between >25 and ≤31] 28 28 A.3 Weak positive [Ct value

31 and ≤ 37] 28 28

For kits detecting CHPV in both serum and CSF, 80 positive serum samples and 80 positive CSF samples should be used for performance evaluation. One sample type should not be substituted by the other to reach the desired sample size in case there is paucity of samples. Note: Since such large number of positive clinical samples may NOT be available for Chandipura virus, pre-titrated and inactivated virus obtained from tissue culture fluid prepared in the laboratory will be used to spike serum and CSF samples [dilution factor: 1:10 to 1:1000 to generate samples with different intensities of positivity]. These spiked samples will be stored at -80°C, after being tested by the reference assay.
B) Negative samples (n=80 for each sample type): All negative samples should be negative by reference assay and CHPV IgM. Distribution of the negative samples should be as follows: Categories of samples as per the sample type Sample type Serum/plasma (Minimum n=80, (B.1 + B.2)) CSF (Minimum n=80, (B.1+B.2)) B.1 Samples from cases of AES/ spiked samples which are RT-PCR positive for known pathogens but negative for CHPV (CHPV RNA and serology) 30

5 positive clinical/ spiked samples from each of the following diseases (confirmed by PCR):

  1. Dengue virus @
  2. Japanese Encephalitis @
  3. HSV 1/2 *
  4. West Nile Virus*
  5. VSV *

35

  1. Seven (07) positive clinical/ spiked samples from each of the following diseases:

a) Japanese Encephalitis @ b) Dengue virus @ c) HSV 1/2 * d) West Nile Virus *

  1. Rabies virus (n=4)*
  2. VSV (n=3)*

Page 343 of 459

B.2 Samples from cases with acute encephalitis and RT- PCR negative for the above-mentioned pathogens and CHPV (CHPV RNA and serology) 50 45 B.3 Healthy/ asymptomatic cases
from endemic regions negative for CHPV (CHPV RNA and serology) 5 (desirable, not mandatory) 20 (desirable, not mandatory) Serum/plasma and CSF samples collected from the same case may be used for evaluation.

there is an acute shortfall or non-availability of clinical samples, one may consider reducing only the pathogens of lower priority marked by * , while ensuring that the actual numbers of cross-reactive sample panel remain the same by compensating with the available “essentially to be tested” samples.

Testing for Rabies and VSV is recommended since both the viruses belong to the same family as Chandipura virus (Rhabdoviridae). Spiked specimens/ synthetic transcripts may be used for these viruses. Commercially available validated standard panels that are accepted by accreditation agencies can also be used if clinical samplse with required target is not available.

  1. Evaluation method:
    The index test and reference tests should be conducted simultaneously on the sample panel to minimize the risk of false-negative results from the index test due to freeze-thaw cycles or sample degradation from prolonged storage.
  2. Interpretation of results:
  3. Resolution of discrepant results: True positive samples: These are samples positive by both the reference assay and index test. True negative samples: These are samples negative by both the reference assay and index test. test.

Page 344 of 459

test.

evaluation.

as follows:
negative samples).
samples).

Page 345 of 459

between-run imprecision (if applicable) should be measured. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be (qualitative).

consecutive days. 3 positive samples (strong, moderate and weak positive samples) and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

IVD kit using two different manufacturer recommended platforms (if applicable). 3 positive samples (strong, moderate and weak positive samples) (qualitative).

Page 346 of 459

recommended).

  1. Blinding of laboratory staff lab, and the PI should maintain confidentiality of data. Refer to Fig. 2.

  2. Acceptance criteria: Sensitivity: ≥ 95%
    Specificity: ≥ 98%
    Cross-reactivity with other rhabdoviruses: Minimal
    Invalid test rate ≤5%

Page 347 of 459

  1. Publication Rights:
    The PI(s) of the evaluating labs shall retain publication rights of the field evaluation as lead author(s).
    acceptable.

VI. References

  1. Sudeep AB, Gurav YK, Bondre VP. Changing clinical scenario in Chandipura virus infection. Indian J Med Res. 2016;143(6):712-721. doi:10.4103/0971-5916.191929.
  2. Sapkal GN, Sawant PM, Mourya DT. Chandipura Viral Encephalitis: A Brief Review. Open Virol J. 2018 Aug 31;12:44-51. doi: 10.2174/1874357901812010044. PMID: 30288194; PMCID: PMC6142667.
  3. World Health Organization. Technical Guidance Series (TGS) for WHO Prequalification-Diagnostic assessment TGS-3. 2017. Available at: https://iris.who.int/bitstream/handle/10665/258985/WHO- EMP-RHT-PQT-TGS3-2017.03-eng.pdf;sequence=1
  4. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability and diagnostic performance: a go-to review. European radiology experimental, 8(1),

VII. Performance Evaluation Report Format

Page 348 of 459 Performance evaluation report for Chandipura virus real-time PCR kits Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use Valid Up to: Application No. Sample Panel Positive samples (provide details: type,strong, moderate, Negative samples (provide details, type,including cross reactivity panel/simulated samples) Results Positive Negative Total Name of Chandipura real-time PCR kits Positive Negative Total Estimate (%) 95% CI Sensitivity Specificity o Cross reactivity with related viruses: o Invalid test rate: o (Sensitivity and specificity have been assessed in controlled lab setting using kits provided by the manufacturer from the batch mentioned above using ….. sample. Results should not be extrapolated to other sample types.) Disclaimers

Page 349 of 459

Note: This report is exclusively for Chandipura………….. Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In-charge …………………… Seal …………………………………

Page 350 of 459

Annexure-1: Information on Operational and Test Performance Characteristics Required from Manufacturers

  1. Instructions for Use
  2. Scope of the IVD:
  3. Intended Use Statement
  4. Principle of the assay
  5. Intended testing population (cases of Acute Febrile Illness/ AES)
  6. Intended user (laboratory professional and/or health care worker at point-of-care)
  7. Lot/batch No.
  8. Date of manufacture
  9. Date of Expiry
  10. Information on operational Characteristics i. Configuration of the kit/device
  11. Information on Test Performance Characteristics i. Type of sample-CSF/Serum/Other specimen iii. Any specific sample NOT to be tested v. Any additional device/consumable like sample transfer device, pipette, tube, etc required vii. Pathogen(s) targeted by the kit

Page 351 of 459

x. Manual or automated (equipment) reading xi. Limit of detection/Limit of Quantification and range of detection xiv. Stability and reproducibility xv. Training required for testing xviii. Interpretation of invalid and indeterminate results to be provided xix. It is recommended to provide data demonstrating accuracy and precision xx. Limit of detection

*Please mention “Not applicable” against sections not pertaining to the kit.

Page 352 of 459

TYPHOID IN-VITRO DIAGNOSTICS

Page 353 of 459

A. Working Group:

  1. Dr. Rajlakshmi Viswanathan, Scientist-E, ICMR-National Institute of Virology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  2. Dr. Labanya Mukhopadhyay, Scientist-C, Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  3. Ms. Krittika Bhattacharyya, Statistical Officer (Planning), Directorate of Economics and Statistics, Government of National Capital Territory of Delhi

  4. Dr. Suman Kanungo, Scientist-G and Director, ICMR-Regional Medical Research Center, Dibrugarh, Department of Health Research, Ministry of Health and Family

  5. Dr. Archana Angrup, Additional Professor, Department of Medical Microbiology, Postgraduate Institute of Medical Education and Research, Chandigarh

  6. Dr Shafeeq K Shahul Hameed, Former Scientific Officer, Center for Brain Research, Indian Institute of Science, Bengaluru, and currently Senior Epidemiologist, Health Emergency Department, Ministry of Public Health, Qatar

  7. Dr. Pratibha Kale, Additional Professor, Department of Clinical Microbiology, Institute of Liver and Biliary Sciences, New Delhi

  8. Dr Jeromie Wesley Vivian T, Scientist C, ICMR-National Institute of Epidemiology, Department of Health Research, Ministry of Health and Family Welfare, Government of India

  9. Dr. Vasanthapuram Ravi, Former Dean Research and Head of Neurovirology, National Institute of Mental Health and Neuro-Sciences, Bengaluru, Karnataka

  10. Dr. Arti Kapil, Professor & Head Department of Microbiology, North DMC Medical College and Hindu Rao Hospital, New Delhi & Ex-Professor and In-Charge of Bacteriology at All India Institute of Medical Sciences, Delhi

  11. Dr. V Balaji, Professor of Microbiology, Christian Medical College, Vellore India

  12. Dr. Jacob John, Professor of Community Health, Christian Medical College, Vellore India

  13. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control

  14. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control

  15. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central

  16. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family

Page 354 of 459

Field Evaluation Protocol for Typhoid Molecular IVDs

I. Background:
Diagnostics kits appropriate for use in India. Hence the following guidelines shall establish the II. Purpose:
To evaluate the performance characteristics of Typhoid molecular IVD kits in the diagnosis of Typhoid fever in individuals with unknown disease status. The protocol outlines field evaluation of IVD kits that detect S Typhi gene segments in blood/ serum/plasma samples. III. Requirements:

  1. Supply of kits under evaluation (Along with batch/lot No. Expiry & required details). If the kit to be evaluated works in a closed system format, the manufacturer needs to supply the required equipment. The study will be initiated after obtaining approval from the institutional human ethics committee.
    V. Procedure:
  2. Study design/type: Cross-sectional study through
    Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider (ISO/IEC 17043) or CDSCO approved Reference laboratory.  Preparation & characterization of kit evaluation panel
     Handling of Typhoid molecular IVD received for performance evaluation  Testing, interpreting, recording of results & reporting  Data handling, data safety & confidentiality

Page 355 of 459

  1. Sample size for performance evaluation: Sample size has been calculated assuming 95% level of significance, an absolute precision of 5%, sensitivity of ≥95%, specificity of ≥95%, invalid test rate ≤5%. Sample size is calculated using the formulae: 𝑑2 x (1 −IR)𝑥𝑃 𝑑2 x (1 −IR)𝑥𝑃

 n (se) is the minimum number of positive samples  n (sp) is the minimum number of negative samples  Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96)  Se is the predetermined sensitivity  Sp is the predetermined specificity  d is the predetermined marginal error (5%)  IR is the invalid test rate  P is the prevalence of the disease

A minimum of 2000 cases with acute febrile illness satisfying the case definition need to be enrolled in the study to achieve statistically significant minimum number of positive samples.

Explanation: A minimum of 77 positive cases and a minimum of 77 negative cases should be enrolled as per the aforementioned formulae. Rounding up, a minimum of 80 positive cases and a minimum of 80 negative cases should be enrolled for evaluation. Since approximately 4% of acute fever cases are positive for typhoid through blood culture (John J, 2023), approximately 2000 cases satisfying the case definition need to be enrolled to achieve the desired target of 80 positive cases. This sample size is adequate for the desired number of negative cases that need to be tested in the evaluation.

Recruitment of cases shall be halted once desired number of positive and negative samples are reached.

Note: Typhoid being a seasonal disease, attempts should be made to enrol more cases during seasons of high transmission. For multi-centric studies, clinicians at each participating site should be sensitized on case enrolment.

  1. Inclusion criteria: The case definition from India’s Integrated Disease Surveillance Programme (IDSP) is adapted to increase the probability of enrolment of blood culture positive typhoid fever,

Page 356 of 459

and ensure availability of adequate sample volumes (Ministry of Health & Family Welfare, 2024). Briefly, an individual aged 2-65 years from OPD/IPD settings will be included if they satisfy both the major and any one of the minor criteria mentioned below:
Major criteria: Documented fever of >38°C AND Duration of fever 4-10 days (preferably ≥3 consecutive days of fever) Minor criteria:  Toxic look  Coated tongue  Relative bradycardia  Splenomegaly

  1. Exclusion criteria:  Age <2 years or ≥65 years
     Pregnancy  Chronic diseases  Known or identified focus of infection
  2. Nucleic acid extraction Nucleic acid extraction should be performed as per reference test and index test (kit under evaluation) Instructions for Use (IFU). If the manufacturer of the index test recommends a specific nucleic acid extraction kit/system, the same needs to be provided by the manufacturer if the evaluating lab is unable to procure the same/it is not available within the lab’s IVD evaluation scope.
  3. IVD System (e.g.: Real-Time PCR System) The reference test and index test should be performed using IVD-approved machines. If any equipment(s) is specified in the IFU of the index test, it should be used for the evaluation, and it should be provided by the manufacturer if not available within the lab’s IVD evaluation scope.
    Closed systems/devices awaiting evaluation should be provided by the manufacturer
  4. Internal control/Extraction control The test under evaluation (index test) must have an internal control (housekeeping gene), with or without an extraction control (nucleic acid added before extraction to a sample).

Page 357 of 459

  1. Reference standard:
     A composite gold standard (blood culture AND/OR accredited molecular test/WHO Pre-Qualified/US FDA/ ATAGI Australia/ PMDA Japan approved molecular test – e.g.: for stag and/or vi genes) should be used as the reference standard. Use of composite gold standard is intended to increase the ability to detect a true-positive case. o Blood culture alone should be used as the reference standard till the time a well- performing reference molecular test (as outlined in this section) is available with the evaluating labs. o Automated blood culture and ID methods should be used, and these tests should be under laboratory accreditation scope. o Note: WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Typhoid molecular kit may be used as reference molecular test as and when these kits become available.

  2. Study implementation: The individuals satisfying the case definition will be recruited, and sample(s) will be collected as per the requirements of the study. The samples should be tested by the reference standard and the index test.

  3. Interpretation of results:
    Reference test and index test results will be interpreted as per kit IFU/SoP.

  4. Cross-reactivity: Cross-reactivity should be reported separately using the samples mentioned below:

    Cross-reactivity panel@:

-Acute phase samples from proven Paratyphoid A, B, NTS infections -Acute phase samples from confirmed cases of Brucellosis, Rickettsial infections, Enterobacteriaceae infections (e.g. sepsis), and Influenza Total 30 samples (described below)

  1. Paratyphoid samples (n=10)  Paratyphoid A (n=4)  Paratyphoid B (n=3)  Paratyphoid C (n=3)

  2. NTS samples (n=10)

One each for the following 10 NTS##:  Salmonella Typhimurium  Salmonella Enteritidis  Salmonella Kentucky  Salmonella Eastbourne  Salmonella Dublin  Salmonella Bareilly
 Salmonella Weltevreden

Page 358 of 459

 Salmonella Newport  Salmonella Infantis  Salmonella Agona

3.Cases of Brucellosis and Rickettsial infections (n=5)  Brucellosis (B. abortus and/or B. melitensis and/or B. suis) cases (n=2) *  Rickettsial (R. rickettsiae and/or R. conorii and/or R. typhi) infections (n=3) * 4. Enterobacteriaceae infection (n=3) Cases of Enterobacteriaceae infection (sepsis) – including E. coli, Klebsiella sp., Citrobacter sp. 5. Influenza cases (n=2) 2 cases of seasonal influenza virus infection (including influenza A and B) NTS = Non Typhoidal Salmonella

@ These samples should be characterized as per national guidelines/globally acceptable standards

In case of unavailability of samples positive for a cross-reacting pathogen, the gap can be met with

negative samples (negative for typhoid markers and other causes of acute febrile illness such as Dengue, malaria, chikungunya, leptospira, scrub typhus) spiked with requisite serovars. It is recommended to use well-characterized serovars from reputed and credible national/global sources or agencies.

  • Commercially available validated panels containing genomes of these pathogens, that are accepted by accreditation agencies, may be used in case of paucity of clinical samples.
  1. Statistical analysis:
    Sensitivity and specificity will be calculated.

Interim analysis of data shall be conducted on completing evaluation of 25%, 50% and 75% of samples. If, at any point, the performance of the assay is found to be not satisfactory, the assay shall not be evaluated further. Evaluation fee shall be charged accordingly.

  1. Repeatability and Reproducibility Assessment:

Page 359 of 459

under evaluation. 3 positive samples and 3 negative samples should be tested 5 times in independent runs.
Three lots of an IVD kit shall be evaluated. Sample size for lot-to-lot reproducibility  Second lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).

Page 360 of 459

between-run imprecision (if applicable) should be measured. 3 positive samples and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

consecutive days. 3 positive samples and 3 negative samples should be (qualitative).

Page 361 of 459

d. Machine-to-machine variability: It is desirable (not mandatory) to evaluate the IVD kit using two different manufacturer recommended platforms (if applicable). 3 positive samples and 3 negative samples should be tested 5 (qualitative).

It is desirable to report standard deviation of measured value across repeats (recommended ≤1 standard deviation variation). 15. Resolution of discrepant results: True positive samples: These are samples positive by reference standard and index test. True negative samples: These are samples negative by reference standard and index test. False positive samples: These are samples negative by reference standard and positive by index test. False negative samples: These are samples positive by reference standard and negative by index test.

  1. Blinding of laboratory staff be blinded to the results of the reference assay. The PI of the evaluation exercise should remain unblinded, i.e., privy to the results of the reference test. Another senior laboratory staff selected by the PI may remain unblinded for overseeing the activity and maintaining the database of results. The data should be analyzed only by the PI of the evaluating lab. The PI should ensure data confidentiality.

  2. Acceptance Criteria Sensitivity: ≥95% Specificity: ≥95%

  3. Publication Rights: author(s).

acceptable.

Page 362 of 459

References:

  1. John J, Bavdekar A, Rongsen-Chandola T, Dutta S, Gupta M, Kanungo S, et al. Burden of typhoid and paratyphoid fever in India. N Engl J Med 2023;388(16): 1491–500. https://doi.org/10.1056/NEJMoa2209449. PMID: 37075141; PMCID: PMC10116367.
  2. Ministry of Health and Family Welfare-Integrated Disease Surveillance Programme. Training manual for state & district surveillance officers case definitions of diseases & syndromes under surveillance module -5. https://idsp.mohfw.gov.in/WriteReadData/OldSite/2WkDSOSept 08/Resources_files/DistrictSurvMan/Module5.pdf. [Accessed 23 November 2024].
  3. Viswanathan, R., Mukhopadhyay, L., Angrup, A., Kale, P., Bhattacharyya, K., Kanungo, S., Jeromie Wesley, V. T., Shahul Hameed, S. K., Sella Senthil, M., Meshram, P., Kapil, A., Balaji, V., John, J., Ravi, V., & Gupta, N. (2025). Standard protocols for performance evaluation of typhoid fever in-vitro diagnostic assays in laboratory and field settings. Indian journal of medical microbiology, 55, 100857. https://doi.org/10.1016/j.ijmmb.2025.100857
  4. Mather RG, Hopkins H, Parry CM, Dittrich S. Redefining typhoid diagnosis: what would an improved test need to look like? BMJ Glob Health 2019;4(5):e001831. https://doi.org/10.1136/bmjgh-2019-001831. PMID: 31749999; PMCID: PMC6830052.
  5. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

Page 363 of 459

Performance Evaluation Protocol for Typhoid Molecular IVDs

I. Background:
diagnostic kits appropriate for use in India. Hence the following guidelines shall establish

II. Purpose: To evaluate the performance characteristics of Typhoid molecular IVD kits in the diagnosis of Typhoid fever using irreversibly de-identified leftover archived clinical samples.

The protocol outlines performance evaluation of IVD kits that detect S Typhi gene segments in blood/ serum/plasma samples.

Note: Performance evaluation with well-characterized leftover samples will be feasible once a well-performing reference molecular test (as outlined later in the document) and requisite clinical samples are available for use.

III. Requirements:

  1. Supply of kits under evaluation (Along with batch/lot No. Expiry & required details). If the kit to be evaluated works in a closed system format, the manufacturer needs to supply the required equipment.
  2. Evaluation sites/laboratories (With required equipment, and blood culture facilities)

Investigators are required to submit a self-declaration form, as outlined in the ICMR guidelines, to the institutional authorities and ethics committee for information.

V. Procedure: clinical samples. through

Page 364 of 459

Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider (ISO/IEC 17043) or CDSCO approved Reference laboratory.  Preparation & characterization of kit evaluation panel
 Handling of kits received for performance evaluation  Testing, interpreting, recording of results & reporting  Data handling, data safety & confidentiality 3. Preparation of Typhoid molecular IVD evaluation panel Well characterised sample panel is a critical requirement for performance evaluation of IVD kits. Hence statistically significant number of samples across relevant sample matrices should be available from cases with confirmed typhoid fever status.

  1. Nucleic acid extraction

Nucleic acid extraction should be performed as per reference test and index test (kit under evaluation) Instructions for Use (IFU). If the manufacturer of the index test recommends a specific nucleic acid extraction kit/system, the same needs to be provided by the manufacturer if the evaluating lab is unable to procure the same/it is not available within the lab’s IVD evaluation scope. 5. IVD System (e.g.: Real-Time PCR System)

The reference test and index test should be performed using IVD-approved machines. If any equipment(s) is specified in the IFU of the index test, it should be used for the evaluation, and it should be provided by the manufacturer if not available within the lab’s IVD evaluation scope.
Closed systems/devices awaiting evaluation should be provided by the manufacturer along with all necessary components, supplies and reagents. 6. Internal control/Extraction control The test under evaluation (index test) must have an internal control (housekeeping gene), with or without an extraction control (nucleic acid added before extraction to a sample). 7. Reference standard:
 A composite gold standard (blood culture AND/OR accredited well-performing molecular test/WHO Pre-Qualified/US FDA/ ATAGI Australia/ PMDA Japan approved molecular test – e.g.: for stag and/or vi genes) should be used to characterize the samples. Use of composite gold standard is intended to increase the ability to detect a true-positive sample. o Automated blood culture and ID methods should be used, and these tests should be under laboratory accreditation scope.

Page 365 of 459

It should be noted that the samples for blood culture and molecular tests may be different. It is recommended to build a biorepository of well-characterized positive and negative samples across relevant sample matrices from confirmed positive and negative cases of typhoid fever.
8. Sample size and sample panel composition: Sample size has been calculated assuming 95% level of significance, an absolute precision of 5%, sensitivity of ≥95%, specificity of ≥95%, and invalid test rate ≤5%. Sample sizes are calculated using the following formulae:

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

  ·      n (se) is the minimum number of positive samples 

· n (sp) is the minimum number of negative samples · Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96) · Se is the predetermined sensitivity · Sp is the predetermined specificity

A minimum of 77 positive samples and a minimum of 77 negative samples should be used. Rounding up, a minimum of 80 positive samples and a minimum of 80 negative samples should be used for evaluation.

A. Positive samples: These samples should be acute phase samples from confirmed cases of typhoid fever positive by composite gold standard, but negative for common pathogens causing febrile illness pathogens (Dengue, Malaria, Scrub Typhus, Leptospira, Chikungunya). B. Negative samples: These samples should be negative by both blood culture AND reference molecular test (once the requisite molecular test is available for use in India)
The sample size and sample panel composition are depicted in Table 1: Table 1: Sample size and panel composition for performance evaluation of molecular IVD kits: A. Minimum Positive samples in the panel (N= 80) Minimum Number of samples needed Composite gold standard positive acute phase samples
which are negative for common febrile illness pathogens in India@ (dengue, malaria, scrub typhus, leptospira, chikungunya) 80 B. Minimum Negative samples in the panel (N=80) (samples negative by both blood culture AND reference molecular test) Minimum Number of samples needed

Page 366 of 459

Samples from afebrile cases where blood culture and reference molecular test are negative for typhoid (e.g., healthy blood donors)

Afebrile: absence of fever for last 14 days 30 Acute phase samples from non-typhoid febrile cases– positive for dengue, malaria, scrub typhus, leptospira, chikungunya @ 20 (equal distribution for these diseases) Cross-reactivity panel@:

-Acute phase samples from proven Paratyphoid A, B, NTS infections -Acute phase samples from confirmed cases of Brucellosis, Rickettsial infections, Enterobacteriaceae infections (e.g. sepsis), and Influenza Total 30 samples (described below)

  1. Paratyphoid samples (n=10)  Paratyphoid A (n=4)  Paratyphoid B (n=3)  Paratyphoid C (n=3)

  2. NTS samples (n=10)
    One each for the following 10 NTS##:  Salmonella Typhimurium  Salmonella Enteritidis  Salmonella Kentucky  Salmonella Eastbourne  Salmonella Dublin  Salmonella Bareilly
     Salmonella Weltevreden  Salmonella Newport  Salmonella Infantis  Salmonella Agona 3.Cases of Brucellosis and Rickettsial infections (n=5)  Brucellosis (B. abortus and/or B. melitensis and/or B. suis) cases (n=2) *  Rickettsial (R. rickettsiae and/or R. conorii and/or R. typhi) infections (n=3) *

  3. Enterobacteriaceae infection (n=3) Cases of Enterobacteriaceae infection (sepsis) – including E. coli, Klebsiella sp., Citrobacter sp.

  4. Influenza cases (n=2) cases of seasonal influenza virus infection (including influenza A and B)

Page 367 of 459

NTS = Non Typhoidal Salmonella @ These samples should be characterized as per national guidelines/globally acceptable standards

In case of unavailability of samples positive for a cross-reacting pathogen, the gap can be met with negative samples (negative

for typhoid markers and other causes of acute febrile illness such as Dengue, malaria, chikungunya, leptospira, scrub typhus) spiked with requisite serovars. It is recommended to use well-characterized serovars from reputed and credible national/global sources or agencies.

  • Commercially available validated panels containing genomes of these pathogens, that are accepted by accreditation agencies, may be used in case of paucity of clinical samples.
  1. Evaluation method:

It is recommended to run the reference molecular test and the index test in parallel.

  1. Interpretation of results:

Reference test and index test results will be interpreted as per kit IFU/SoP.

  1. Resolution of discrepant results:

True positive samples: These are samples positive by reference test and index test. True negative samples: These are samples negative by reference test and index test. False positive samples: These are samples negative by reference test and positive by index test. False negative samples: These are samples positive by reference test and negative by index test.

  1. Repeatability and Reproducibility Assessment:

evaluation. 3 positive samples and 3 negative samples should be tested 5 times in independent runs.

as follows:

Page 368 of 459

 Second lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).

between-run imprecision (if applicable) should be measured. 3 positive samples and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

consecutive days. 3 positive samples and 3 negative samples should be (qualitative).

Page 369 of 459

d. Machine-to-machine variability: It is desirable (not mandatory) to evaluate the IVD kit using two different manufacturer recommended platforms (if applicable). 3 positive samples and 3 negative samples should be tested 5 (qualitative).

It is desirable to report standard deviation of measured value across repeats (recommended ≤1 standard deviation variation).

  1. Blinding of laboratory staff

lab, and the PI should ensure confidentiality of the data. Refer to Fig. 1. Fig.1: Blinding in evaluation exercise

  1. Acceptance Criteria Sensitivity: ≥95% Specificity: ≥95%

  2. Publication Rights:

Page 370 of 459

author(s).

acceptable.

References:

  1. John J, Bavdekar A, Rongsen-Chandola T, Dutta S, Gupta M, Kanungo S, et al. Burden of typhoid and paratyphoid fever in India. N Engl J Med 2023;388(16): 1491–500. https://doi.org/10.1056/NEJMoa2209449. PMID: 37075141; PMCID: PMC10116367.
  2. Ministry of Health and Family Welfare-Integrated Disease Surveillance Programme. Training manual for state & district surveillance officers case definitions of diseases & syndromes under surveillance module -5. https://idsp.mohfw.gov.in/WriteReadData/OldSite/2WkDSOSept 08/Resources_files/DistrictSurvMan/Module5.pdf. [Accessed 23 November 2024].
  3. Viswanathan, R., Mukhopadhyay, L., Angrup, A., Kale, P., Bhattacharyya, K., Kanungo, S., Jeromie Wesley, V. T., Shahul Hameed, S. K., Sella Senthil, M., Meshram, P., Kapil, A., Balaji, V., John, J., Ravi, V., & Gupta, N. (2025). Standard protocols for performance evaluation of typhoid fever in-vitro diagnostic assays in laboratory and field settings. Indian journal of medical microbiology, 55, 100857. https://doi.org/10.1016/j.ijmmb.2025.100857
  4. Mather RG, Hopkins H, Parry CM, Dittrich S. Redefining typhoid diagnosis: what would an improved test need to look like? BMJ Glob Health 2019;4(5):e001831. https://doi.org/10.1136/bmjgh-2019-001831. PMID: 31749999; PMCID: PMC6830052.
  5. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

Page 371 of 459

Field Evaluation Protocol for Typhoid Antigen-based IVDs

I. Background:
Diagnostics kits appropriate for use in India. Hence the following guidelines shall establish the

II. Purpose:
To evaluate the performance characteristics of Typhoid antigen-based kits in the diagnosis of Typhoid fever in individuals with unknown disease status.

The protocol outlines field evaluation of IVD kits that detect S Typhi antigens in blood/ serum/plasma samples.

III. Requirements:

  1. Supply of kits under evaluation (Along with batch/lot No. Expiry & required details). If the kit to be evaluated works in a closed system format, the manufacturer needs to supply the required equipment. The study will be initiated after obtaining approval from the institutional human ethics committee.
    V. Procedure: 1.Study design/type: Cross-sectional study 2.Preparation of Evaluation sites/laboratories:
    Identified IVD kit evaluation laboratories should establish their proficiency through
    Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider (ISO/IEC
  1. or CDSCO approved Reference laboratory. ⮚ Preparation & characterization of kit evaluation panel
    ⮚ Handling of Typhoid antigen-based IVD received for performance evaluation

Page 372 of 459

⮚ Testing, interpreting, recording of results & reporting ⮚ Data handling, data safety & confidentiality

  1. Sample size for performance evaluation: Sample size has been calculated assuming 95% level of significance, an absolute precision of 5%, sensitivity of ≥90%, specificity of ≥95%, and invalid test rate ≤5% (where applicable). Sample size is calculated using the formulae:

𝑑2 x (1 −IR)𝑥𝑃 𝑑2 x (1 −IR)𝑥𝑃

· n (se) is the minimum number of positive samples · n (sp) is the minimum number of negative samples ·
Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96) · Se is the predetermined sensitivity · Sp is the predetermined specificity

P is the prevalence of the disease

A minimum of 3750 cases with acute febrile illness satisfying the case definition need to be enrolled in the study to achieve statistically significant minimum number of positive samples.

Explanation: A minimum of 146 positive cases (139 samples for assays without the provision of invalid test) and a minimum of 77 negative cases (73 samples for assays without the provision of invalid test) should be enrolled as per the aforementioned formulae. Rounding up, a minimum of 150 positive cases and a minimum of 80 negative cases should be enrolled for evaluation. Since approximately 4% of acute fever cases are positive for

Page 373 of 459

typhoid through blood culture (John J, 2023), approximately 3750 cases satisfying the case definition need to be enrolled to achieve the desired target of 150 positive cases. This sample size is adequate for the desired number of negative cases that need to be tested in the evaluation. Recruitment of cases shall be halted once desired number of positive and negative samples are reached. Note: Typhoid being a seasonal disease, attempts should be made to enrol more cases during seasons of high transmission. For multi-centric studies, clinicians at each participating site should be sensitized on case enrolment. 4. Inclusion criteria: The case definition from India’s Integrated Disease Surveillance Programme (IDSP) was adapted to increase the probability of enrolment of typhoid blood culture positive, and ensure availability of adequate sample volumes (Ministry of Health & Family Welfare, 2024). Briefly, an individual aged 2-65 years from OPD/IPD settings will be included if they satisfy both the major and any one of the minor criteria mentioned below:
Major criteria: Documented fever of >38°C AND Duration of fever 4-10 days (preferably ≥3 consecutive days of fever) Minor criteria:  Toxic look  Coated tongue  Relative bradycardia  Splenomegaly 5. Exclusion criteria:

Age <2 years or ≥65 years

Pregnancy

Chronic diseases

Known or identified focus of infection 6. Reference standard:
 A composite gold standard (blood culture AND/OR accredited molecular test/WHO Pre-Qualified/US FDA/ ATAGI Australia/ PMDA Japan approved molecular test – e.g.: for stag and/or vi genes) should be used as the reference standard. Use of composite gold standard is intended to increase the ability to detect a true-positive case. o Blood culture alone should be used as the reference standard till the time a well- performing reference molecular test (as outlined in this section) is available with the evaluating labs. o Automated blood culture and ID methods should be used, and these tests should be under laboratory accreditation scope.

Page 374 of 459

o Note: WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Typhoid molecular kit may be used as reference molecular test as and when these kits become available. 7. Study implementation: The individuals satisfying the case definition will be recruited, and sample(s) will be collected as per the requirements of the study. The requisite samples should be tested by the reference standard and the index test. 8. Interpretation of results:
Reference test and index test results will be interpreted as per kit IFU/SoP.

Page 375 of 459

  1. Cross-reactivity: Cross-reactivity should be reported separately using the samples mentioned below:

    Cross-reactivity panel@:  
    

-Acute phase samples from proven Paratyphoid A, B, NTS

-Acute phase samples from confirmed cases of Brucellosis, Rickettsial infections, Enterobacteriaceae infections (e.g. sepsis), and Influenza Total 30 samples (described below)

  1. Paratyphoid samples (n=10)  Paratyphoid A (n=4)  Paratyphoid B (n=3)  Paratyphoid C (n=3)
  2. NTS samples (n=10)
    One each for the following 10 NTS##:  Salmonella Typhimurium  Salmonella Enteritidis  Salmonella Kentucky  Salmonella Eastbourne  Salmonella Dublin  Salmonella Bareilly
     Salmonella Weltevreden  Salmonella Newport  Salmonella Infantis  Salmonella Agona 3.Cases of Brucellosis and Rickettsial infections (n=5)  Brucellosis (B. abortus and/or B. melitensis and/or B. suis) cases (n=2)

 Rickettsial (R. rickettsiae and/or R. conorii and/or R. typhi) infections (n=3) * Proven cross-reactive samples@:

  • Acute phase samples from proven Paratyphoid A, B, NTS infections -Acute phase samples collected from confirmed cases of Brucellosis, Rickettsial infections, Enterobacteriaceae infections (e.g. sepsis), and Influenza 30## -4 Paratyphoid A, 3 Paratyphoid B, 3 Paratyphoid C -1 each for the following NTS:
  1. Salmonella Typhimurium
  2. ⁠ Salmonella Enteritidis
  3. Salmonella Kentucky
  4. Salmonella Eastbourne
  5. Salmonella Dublin
  6. Salmonella Bareilly
  7. Salmonella Weltevreden
  8. ⁠ Salmonella Newport
  9. ⁠ Salmonella Infantis
  10. ⁠ Salmonella Agona

Page 376 of 459

  1. Enterobacteriaceae infection (n=3) Cases of Enterobacteriaceae infection (sepsis) – including E. coli, Klebsiella sp., Citrobacter sp.
  2. Influenza cases (n=2) cases of seasonal influenza virus infection (including influenza A and B) -2 cases of Brucellosis (B. abortus and/or B. melitensis and/or B. suis) -3 cases of Rickettsial infections (R. rickettsiae and/or R. conorii and/or R. typhi)* -3 cases of Enterobacteriaceae infection (sepsis) – including E. coli, Klebsiella sp., Citrobacter sp.
  • 2 cases of influenza virus infections (influenza A and B) NTS = Non Typhoidal Salmonella

@ These samples should be characterized as per national guidelines/globally acceptable standards

In case of unavailability of samples positive for a cross-reacting pathogen, the gap

can be met with negative samples (negative for typhoid markers and other causes of acute febrile illness such as Dengue, malaria, chikungunya, leptospira, scrub typhus) spiked with requisite serovars. It is recommended to use well-characterized serovars from reputed and credible national/global sources or agencies.

AFI = Acute Febrile Illness NTS = Non Typhoidal Salmonella @ These samples should be characterized per national guidelines/globally acceptable standards

In case of unavailability of samples positive for a

cross-reacting pathogen, the gap can be met with negative samples (negative for typhoid markers and

Page 377 of 459

  • Commercially available validated panels containing genomes of these pathogens, that are accepted by accreditation agencies, may be used in case of paucity of clinical samples. other causes of acute febrile illness as listed in the table) spiked with requisite serovars.

  • In case of unavailability, compensate with other samples outlined in the cross-reactivity panel

  1. Statistical analysis:
    Sensitivity and specificity will be calculated. Interim analysis of data shall be conducted on completing evaluation of 25%, 50% and 75% of samples. If, at any point, the performance of the assay is found to be not satisfactory, the assay shall not be evaluated further. Evaluation fee shall be charged accordingly.
  2. Repeatability and Reproducibility Assessment:

This should be done to assess the repeatability of the detection of target using the kit under evaluation.

Page 378 of 459

3 positive samples and 3 negative samples should be tested 5 times in independent runs.

be as follows:

 Second lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).

Page 379 of 459

between-run imprecision (if applicable) should be measured. 3 positive samples (qualitative).

consecutive days. 3 positive samples and 3 negative samples should be tested 5 (qualitative).

IVD kit using two different manufacturer recommended platforms (if applicable). 3 positive samples and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

Page 380 of 459

  1. Resolution of discrepant results:

True positive samples: These are samples positive by reference standard and index test. True negative samples: These are samples negative by reference standard and index test. False positive samples: These are samples negative by reference standard and positive by index test. False negative samples: These are samples positive by reference standard and negative by index test.

  1. Blinding of laboratory staff

To ensure rigor of the evaluation process, laboratory staff performing the evaluation should be blinded to the results of the reference assay. The PI of the evaluation exercise should remain unblinded, i.e., privy to the results of the reference test. Another senior laboratory staff selected by the PI may remain unblinded for overseeing the activity and maintaining the database of results.. The data should be analyzed only by the PI of the evaluating lab. The PI should ensure data confidentiality.

  1. Acceptance Criteria Sensitivity: ≥90% Specificity: ≥95% Invalid test rate (if applicable): ≤5%

  2. Publication Rights: author(s).

acceptable.

Page 381 of 459

References:

  1. John J, Bavdekar A, Rongsen-Chandola T, Dutta S, Gupta M, Kanungo S, et al. Burden of typhoid and paratyphoid fever in India. N Engl J Med 2023;388(16): 1491–500. https://doi.org/10.1056/NEJMoa2209449. PMID: 37075141; PMCID: PMC10116367.
  2. Ministry of Health and Family Welfare-Integrated Disease Surveillance Programme. Training manual for state & district surveillance officers case definitions of diseases & syndromes under surveillance module -5. https://idsp.mohfw.gov.in/WriteReadData/OldSite/2WkDSOSept 08/Resources_files/DistrictSurvMan/Module5.pdf. [Accessed 23 November 2024].
  3. Viswanathan, R., Mukhopadhyay, L., Angrup, A., Kale, P., Bhattacharyya, K., Kanungo, S., Jeromie Wesley, V. T., Shahul Hameed, S. K., Sella Senthil, M., Meshram, P., Kapil, A., Balaji, V., John, J., Ravi, V., & Gupta, N. (2025). Standard protocols for performance evaluation of typhoid fever in-vitro diagnostic assays in laboratory and field settings. Indian journal of medical microbiology, 55, 100857. https://doi.org/10.1016/j.ijmmb.2025.100857
  4. Mather RG, Hopkins H, Parry CM, Dittrich S. Redefining typhoid diagnosis: what would an improved test need to look like? BMJ Glob Health 2019;4(5):e001831. https://doi.org/10.1136/bmjgh-2019-001831. PMID: 31749999; PMCID: PMC6830052.
  5. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

Page 382 of 459

Performance Evaluation Protocol for Typhoid Antigen-based IVDs

I. Background:
diagnostic kits appropriate for use in India. Hence the following guidelines shall establish

II. Purpose: To evaluate the performance characteristics of Typhoid antigen-based kits in the diagnosis of Typhoid fever using irreversibly de-identified leftover archived/ spiked clinical samples. The protocol outlines performance evaluation of IVD kits that detect S Typhi antigens in blood/ serum/plasma samples. Note: Performance evaluation with well-characterized leftover samples will be feasible once a well-performing reference molecular test and requisite clinical samples are available for use.

III. Requirements:
6. Supply of kits under evaluation (Along with batch/lot No. Expiry & required details). 7. Evaluation sites/laboratories (With required equipment, and blood culture facilities) 8. Reference test kits 9. Characterised Evaluation panel 10. Laboratory supplies

Investigators are required to submit a self-declaration form, as outlined in the ICMR guidelines, to the institutional authorities and ethics committee for information. V. Procedure:

Page 383 of 459

through
A. ccreditation for at least one Quality management system (accreditation for Testing Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO 15189), PT provider (ISO/IEC 17043) or CDSCO approved Reference laboratory. B. Staff training: All the staff involved in IVD kit evaluation should undergo hands on training and competency testing on following
⮚ Handling of kits received for performance evaluation 3. Preparation of Typhoid antigen-based IVD evaluation panel Well characterised sample panel is a critical requirement for performance evaluation of IVD kits. Hence statistically significant number of samples should be available from cases with confirmed typhoid fever status.

  1. Reference standard:
     A composite gold standard (blood culture AND/OR accredited well-performing molecular test/WHO Pre-Qualified/US FDA/ ATAGI Australia/ PMDA Japan approved molecular test – e.g.: for stag and/or vi genes) should be used to characterize the samples. Use of composite gold standard is intended to increase the ability to detect a true-positive sample. o Automated blood culture and ID methods should be used, and these tests should be under laboratory accreditation scope. It should be noted that the samples for blood culture and molecular tests may be different. It is recommended to build a biorepository of well-characterized positive and negative samples across relevant sample matrices from confirmed positive and negative cases of typhoid fever.
  2. Sample size and sample panel composition: Sample size has been calculated assuming 95% level of significance, an absolute precision of 5%, sensitivity of ≥90%, specificity of ≥95%, and invalid test rate ≤5% (where applicable). Sample size are calculated using the formulae:

Page 384 of 459

𝑑2 x (1 −IR)
𝑑2 x (1 −IR)

· n (se) is the minimum number of positive samples · n (sp) is the minimum number of negative samples ·
Z2 is the critical value from the standard normal distribution corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96) · Se is the predetermined sensitivity · Sp is the predetermined specificity

A minimum of 146 positive samples (139 samples for assays without the provision of invalid test) and a minimum of 77 negative samples (73 samples for assays without the provision of invalid test) should be used. Rounding up, a minimum of 150 positive samples and a minimum of 80 negative samples should be used for evaluation.

A. Positive samples: These samples should be acute phase samples from confirmed cases of typhoid fever positive by composite gold standard, but negative for common pathogens causing febrile illness in India (Dengue, Malaria, Scrub Typhus, Leptospira, Chikungunya).

B. Negative samples: These samples should be negative by both blood culture AND reference molecular test.

The sample size and sample panel composition are depicted in Table 1:

Table 1: Sample size and panel composition for performance evaluation of diagnostic antigen based kits:

A. Minimum Positive samples in the panel (N= 150) Minimum Number of samples needed

Page 385 of 459

Composite gold standard positive acute phase samples which are negative for common febrile illness pathogens in India@ (dengue, malaria, scrub typhus, leptospira, chikungunya) 150 B. Minimum Negative samples in the panel (N=80)# (samples negative by both blood culture AND reference molecular test) Minimum Number of samples needed Samples from afebrile cases where blood culture and reference molecular test are negative for typhoid (e.g.: healthy blood donors) Afebrile: absence of fever for last 14 days 30 Acute phase samples from non-typhoid febrile cases– positive for dengue, malaria, scrub typhus, leptospira, chikungunya @ 20 (equal distribution for these diseases) Cross-reactivity panel@:

-Acute phase samples from proven Paratyphoid A, B, NTS infections -Acute phase samples from confirmed cases of Brucellosis, Rickettsial infections, Enterobacteriaceae infections (e.g. sepsis), and Influenza Total 30 samples (described below)

  1. Paratyphoid samples (n=10)  Paratyphoid A (n=4)  Paratyphoid B (n=3)  Paratyphoid C (n=3)
  2. NTS samples (n=10)
    One each for the following 10 NTS##:  Salmonella Typhimurium  Salmonella Enteritidis  Salmonella Kentucky  Salmonella Eastbourne  Salmonella Dublin  Salmonella Bareilly
     Salmonella Weltevreden  Salmonella Newport

Page 386 of 459

 Salmonella Infantis  Salmonella Agona 3.Cases of Brucellosis and Rickettsial infections (n=5)  Brucellosis (B. abortus and/or B. melitensis and/or B. suis) cases (n=2) *  Rickettsial (R. rickettsiae and/or R. conorii and/or R. typhi) infections (n=3) * 4. Enterobacteriaceae infection (n=3) Cases of Enterobacteriaceae infection (sepsis) – including E. coli, Klebsiella sp., Citrobacter sp. 5. Influenza cases (n=2) cases of seasonal influenza virus infection (including influenza A and B) NTS = Non Typhoidal Salmonella These samples should be characterized as per national guidelines/globally acceptable standards

In case of unavailability of samples positive for a cross-reacting pathogen, the gap can be met with negative samples

(negative for typhoid markers and other causes of acute febrile illness such as Dengue, malaria, chikungunya, leptospira, scrub typhus) spiked with requisite serovars. It is recommended to use well-characterized serovars from reputed and credible national/global sources or agencies.

  • Commercially available validated panels containing genomes of these pathogens, that are accepted by accreditation agencies, may be used in case of paucity of clinical samples.
  1. Evaluation method: It is recommended to run the reference molecular test and the index test in parallel.

  2. Interpretation of results:
    Reference test and index test results will be interpreted as per kit IFU/SoP.

  3. Resolution of discrepant results: True positive samples: These are samples positive by reference test and index test. True negative samples: These are samples negative by reference test and index test. False positive samples: These are samples negative by reference test and positive by index test.

Page 387 of 459

False negative samples: These are samples positive by reference test and negative by index test.

under evaluation. 3 positive samples and 3 negative samples should be tested 5 times in independent runs.

as follows:
 Second lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).

Page 388 of 459

between-run imprecision (if applicable) should be measured. 3 positive samples and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

consecutive days. 3 positive samples and 3 negative samples should be (qualitative).

d. Machine-to-machine variability: It is desirable (not mandatory) to evaluate the IVD kit using two different manufacturer recommended platforms (if

Page 389 of 459

applicable). 3 positive samples and 3 negative samples should be tested 5 (qualitative).

  1. Blinding of laboratory staff To ensure rigor of the evaluation process, laboratory staff performing the evaluation should be blinded to the status of the clinical samples. The PI of the evaluation exercise should remain unblinded, i.e., privy to the status of the samples. Another senior laboratory staff selected by the PI may remain unblinded and carry out coding of samples and dispensing them into similar-looking vials to be used for testing, and maintaining the database of results. Staff performing the reference test and the test under evaluation, interpretation of the test result, and entering the results against the coded samples in the database, should remain blinded to the status of samples till the completion of evaluation. The data should be analyzed only by the PI of the evaluating lab, and the PI should ensure confidentiality of data. Refer to Fig. 1.

Fig.1: Blinding in evaluation exercise

  1. Acceptance Criteria Sensitivity: ≥90% Specificity: ≥95%

Page 390 of 459

Invalid test rate(if applicable): ≤5% author(s). acceptable.

References:

  1. John J, Bavdekar A, Rongsen-Chandola T, Dutta S, Gupta M, Kanungo S, et al. Burden of typhoid and paratyphoid fever in India. N Engl J Med 2023;388(16): 1491–500. https://doi.org/10.1056/NEJMoa2209449. PMID: 37075141; PMCID: PMC10116367.
  2. Ministry of Health and Family Welfare-Integrated Disease Surveillance Programme. Training manual for state & district surveillance officers case definitions of diseases & syndromes under surveillance module -5. https://idsp.mohfw.gov.in/WriteReadData/OldSite/2WkDSOSept 08/Resources_files/DistrictSurvMan/Module5.pdf. [Accessed 23 November 2024].
  3. Viswanathan, R., Mukhopadhyay, L., Angrup, A., Kale, P., Bhattacharyya, K., Kanungo, S., Jeromie Wesley, V. T., Shahul Hameed, S. K., Sella Senthil, M., Meshram, P., Kapil, A., Balaji, V., John, J., Ravi, V., & Gupta, N. (2025). Standard protocols for performance evaluation of typhoid fever in-vitro diagnostic assays in laboratory and field settings. Indian journal of medical microbiology, 55, 100857. https://doi.org/10.1016/j.ijmmb.2025.100857
  4. Mather RG, Hopkins H, Parry CM, Dittrich S. Redefining typhoid diagnosis: what would an improved test need to look like? BMJ Glob Health 2019;4(5):e001831. https://doi.org/10.1136/bmjgh-2019-001831. PMID: 31749999; PMCID: PMC6830052.
  5. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

Page 391 of 459

Field Evaluation Protocol for Typhoid Antibody-based IVDs

I. Background:
Diagnostic kits appropriate for use in India. Hence the following guidelines shall establish the

II. Purpose:

To evaluate the performance characteristics of Typhoid antibody-based kits in the diagnosis of Typhoid fever in individuals with unknown disease status.

The protocol outlines field evaluation of IVD kits that detect S Typhi antibodies in blood/ serum/plasma samples.

III. Requirements:

  1. Supply of kits under evaluation (Along with batch/lot No. Expiry & required details). If the kit to be evaluated works in a closed system format, the manufacturer needs to supply the required equipment. The study will be initiated after obtaining approval from the institutional human ethics committee.
    V. Procedure:
  2. Study design/type: Cross-sectional study Identified IVD kit evaluation laboratories should establish their proficiency through
    Lab / Calibration Lab (ISO/IEC 17025), Medical Lab (ISO/IEC 15189), PT provider (ISO/IEC 17043) or CDSCO approved Reference laboratory.  Preparation & characterization of kit evaluation panel
     Handling of Typhoid antibody-based IVD received for performance evaluation  Testing, interpreting, recording of results & reporting  Data handling, data safety & confidentiality

Page 392 of 459

  1. Sample size for performance evaluation: Sample size has been calculated assuming 95% level of significance, an absolute precision of 5%, sensitivity ≥80%, specificity ≥90%, and invalid test rate ≤5% (where applicable). Sample size is calculated using the formulae:

𝑑2 x (1 −IR)𝑥𝑃 𝑑2 x (1 −IR)𝑥𝑃

                               ·      n (se) is the minimum number of positive samples 

· n (sp) is the minimum number of negative samples · Z2 is the critical value from the standard normal distribution corresponding to the
desired confidence level (95% CI corresponds to Z2 =1.96) · Se is the predetermined sensitivity · Sp is the predetermined specificity

P is the prevalence of the disease

A minimum of 6500 cases with acute febrile illness satisfying the case definition need to be enrolled in the study to achieve statistically significant minimum number of positive samples.

Explanation: A minimum of 259 positive cases (246 cases for assays without the provision of invalid test) and a minimum of 146 negative cases (139 cases for assays without the provision of invalid test) should be enrolled as per the aforementioned formulae. Rounding up, a minimum of 260 positive cases and a minimum of 150 negative cases should be enrolled for evaluation. Since approximately 4% of acute fever cases are positive for typhoid through blood culture (John J, 2023), minimum 6500 cases satisfying the case definition need to be enrolled to achieve the desired target of minimum 260 positive cases. This sample size is adequate for the desired number of negative cases that need to be tested in the evaluation. Recruitment of cases shall be halted once desired number of positive and negative samples are reached. Note: Typhoid being a seasonal disease, attempts should be made to enrol more cases during seasons of high transmission. For multi-centric studies, clinicians at each participating site should be sensitized on case enrolment. 4. Inclusion criteria: The case definition from India’s Integrated Disease Surveillance Programme (IDSP) has been adapted to increase the probability of enrolment of typhoid blood culture positive, and ensure availability of adequate sample volumes (Ministry of Health & Family Welfare, 2024). Briefly, an individual aged 2-65 years from OPD/IPD settings will be included if they satisfy both the major and any one of the minor criteria mentioned below:

Page 393 of 459

Major criteria: Documented fever of >38°C AND Duration of fever 4-10 days (preferably ≥3 consecutive days of fever) Minor criteria:  Toxic look  Coated tongue  Relative bradycardia  Splenomegaly 5. Exclusion criteria:  Age <2 years or ≥65 years
 Pregnancy  Chronic diseases  Known or identified focus of infection

  1. Reference standard and sample characterization:
    a. Sample collection: There is no predicate IVD kit/device available for use as diagnostic typhoid antibody reference standard, and cross-reactivity and anamnestic responses significantly impact the specificity of these assays. Therefore, it is important to ensure characterization of the sample panel to be used for evaluation. The study should enrol acutely febrile cases satisfying the case definition, and collect blood samples for different tests for identifying true positive typhoid cases at various time points as outlined in Table 1.

Table 1: Sample type and Timelines for Collection:

Age of participants Sample/Timing Antibody detecting IVDs Adults Sample type and timing of collection
1.Blood in 2 automated blood culture bottles (10-20 ml blood in each bottle) from 2 different sampling sites Timing: on case enrolment (acute febrile phase) 2. 3-4 ml blood in EDTA vacutainer (for reference molecular test) Timing: on case enrolment (acute febrile phase)

  1. 3-4 ml blood in plain vacutainer (for accredited Widal test-baseline antibody titers) Timing: on case enrolment (acute febrile phase)

Page 394 of 459

  1. Follow up Sample: 3-4 ml blood in plain vacutainer (for accredited Widal test and index test) Timing: 3-4 weeks after onset of illness (which should also in accordance with the duration of illness mentioned by the manufacturer)

Children Sample type and timing of collection 1.Blood in 1 automated blood culture bottle (3-5 ml of blood for children aged 2-5 years and
5-10 ml of blood for children aged >5-13 years) from 1 sampling site Timing: on case enrolment (febrile phase, in accordance with the duration of illness mentioned by the manufacturer)

  1. 3-4 ml blood in EDTA vacutainer (for reference molecular test) Timing: on case enrolment (febrile phase, in accordance with the duration of illness mentioned by the manufacturer)

  2. 3-4 ml blood in plain vacutainer (for accredited Widal test-baseline antibody titers) Timing: on case enrolment (febrile phase, in accordance with the duration of illness mentioned by the manufacturer)

  3. Follow up Sample: 3-4 ml blood in plain vacutainer (for accredited Widal test and index test) Timing: 3-4 weeks after onset of illness (if needed, also in accordance with the duration of illness mentioned by the manufacturer)

If an antibody-based IVD kit claims to detect anti-typhoid antibodies in samples collected earlier in the course of the disease (less than 3-4 weeks), samples should be available from these time points as well for evaluation. If these samples are not available, the performance evaluation report should clearly indicate the timing of follow up sample collection. b. Confirmation of typhoid fever status and obtaining typhoid antibody-positive samples from convalescent cases:
i. On enrolment: Typhoid fever status should be confirmed for all cases on enrolment through a composite reference standard of blood culture AND/OR standardized well-performing rigorously validated molecular test (under laboratory accreditation scope/WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan

Page 395 of 459

approved assays). Use of composite gold standard is intended to increase the ability to detect a true-positive case. o Blood culture alone should be used for determining typhoid fever status till the time a well-performing reference molecular test (as outlined in this section) is available with the evaluating labs. o Automated blood culture and ID methods should be used, and these tests should be under laboratory accreditation scope. o Note: WHO Pre-Qualified/US FDA/ATAGI Australia/ PMDA Japan approved Typhoid molecular kit may be used as reference molecular test as and when these kits become available. o An accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available) should be performed on the accompanying serum sample collected on enrolment to establish baseline anti-typhoid antibody titer. ii. Follow-up sample: A follow-up serum sample should be collected 3- 4 weeks after the onset of illness to confirm the presence of anti- typhoid antibodies through Widal test.

Positive cases: Confirmed typhoid fever cases should demonstrate 4-fold rise in Widal TO/TH titers (>320) in samples collected after 3-4 weeks; this sample panel should comprise anti- typhoid antibody positive samples and should be used for evaluation of the index test.
Negative cases: Sera samples collected 3-4 weeks after onset of illness from cases without typhoid fever and not showing 4-fold rise in Widal TO/TH titers (>320), should be considered negative samples, and should be used for evaluation of the index test.
Note: Blood culture/molecular test negative typhoid cases: For a blood culture AND reference molecular test negative (as per the conditions outlined in Section 6) but baseline Widal positive (TH titer > 320) case to be considered a true positive case of typhoid, the case should demonstrate at least 4-fold rise in antibody titers when follow- up samples are collected prospectively 3-4 weeks after onset of illness, in the absence of another confirmed cause of fever that may lead to anamnestic O antigen-related responses. Such samples may also be included in the positive sample panel for performance evaluation.
Note:  All Widal tests performed for sample characterization as per Section 6 should be accredited as per ISO 15189 or ISO/IEC 17025 or ISO/IEC 17043.
 WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test may be used as and when they become available.

  1. Study implementation: The individuals satisfying the case definition will be recruited, and samples will be collected and tested as per the requirements of the study.

Page 396 of 459

The requisite samples collected 3-4 weeks after onset of illness as defined in Section 6 should be tested by accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available) and the index test simultaneously.

  1. Interpretation of results:
    All test results (blood culture, reference molecular test, Widal test, index test) will be interpreted as per respective kit IFU/SoP.
  2. Resolution of discrepant results:

 True Positive samples: Rigorously well-characterized sera samples collected 3-4 weeks after onset of illness from confirmed typhoid cases (as per Section 6) and showing TO/TH titers >320 with accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available), and positive results with the index test, will be considered as true positive sample.
For semi-quantitative antibody-based tests, results should closely mimic Widal test results.

 True Negative samples:
confirmed non-typhoid cases (as per Section 6) and testing non-reactive (or) for anti- typhoid antibodies with accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available), as well as the index test, will be considered as true negative sample.

 False positive samples:

confirmed non-typhoid cases (as per Section 6) and testing non-reactive for anti-typhoid antibodies with accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available), but reactive for anti-typhoid antibodies with the index test, will be considered as false positive samples.

 False negative samples: confirmed typhoid cases (as per Section 6) and showing TO/TH titers >320 with accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available), but non-reactive for anti-typhoid antibodies with the index test, will be considered false negative samples.

Page 397 of 459

  1. Cross-reactivity: Cross-reactivity should be reported separately using the samples mentioned below:

Proven cross-reactive samples:

  • Sera collected 2-4 weeks after onset of illness in proven Paratyphoid A, B, and NTS infections (blood culture/PCR confirmed cases) @

-Sera collected 2-4 weeks after onset of illness from confirmed cases of Brucellosis, Rickettsial infections, Enterobacteriaceae infections (e.g. sepsis), and Influenza Recommended: 30 samples Minimum: 25 samples (described below)

  1. Paratyphoid samples (n=10)  Paratyphoid A (n=4)  Paratyphoid B (n=3)  Paratyphoid C (n=3)
  2. NTS samples (n=10)
    One each for the following 10 NTS##:  Salmonella Typhimurium  Salmonella Enteritidis  Salmonella Kentucky  Salmonella Eastbourne  Salmonella Dublin  Salmonella Bareilly
     Salmonella Weltevreden  Salmonella Newport  Salmonella Infantis  Salmonella Agona 3.Cases of Brucellosis and Rickettsial infections (n=5)  Brucellosis (B. abortus and/or B. melitensis and/or B. suis) cases (n=2) *  Rickettsial (R. rickettsiae and/or R. conorii and/or R. typhi) infections (n=3) * ‡
  3. Enterobacteriaceae infection (n=3) Cases of Enterobacteriaceae infection (sepsis) – including E. coli, Klebsiella sp., Citrobacter sp.
  4. Influenza cases (n=2) cases of seasonal influenza virus infection (including influenza A and B) NTS = Non Typhoidal Salmonella

In case of unavailability of samples positive for a cross-reacting pathogen, attempt should be

made to meet the gap with samples from clinical cases of infection with other serovars. If deficit persists, the gap should be met with samples from Paratyphoid fever cases. If no sample positive

Page 398 of 459

from confirmed cases of these infections is available, test 5 more paratyphoid fever positive samples (in addition to the 10 samples outlined in Point No. 1); this will bring down the size of the cross-reactivity panel to 25 samples. Commercially available serology panels for these pathogens, that are accepted by accreditation agencies, may also be used.

  • Commercially available serology panels for these pathogens, that are accepted by accreditation agencies, may also be used in case of paucity of clinical samples. ‡ In case of unavailability/paucity of Rickettsial serology samples, the gap should be met with scrub typhus IgM antibody positive samples.
  1. Statistical analysis:
    Sensitivity and specificity will be calculated. Interim analysis of data shall be conducted on completing evaluation of 25%, 50% and 75% of samples. If, at any point, the performance of the assay is found to be not satisfactory, the assay shall not be evaluated further. Evaluation fee shall be charged accordingly.
  2. Repeatability and Reproducibility Assessment:

under evaluation. 3 positive samples and 3 negative samples should be tested 5 times in independent runs.
as follows:

Page 399 of 459

 Second lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).

between-run imprecision (if applicable) should be measured. 3 positive samples (qualitative).

consecutive days. 3 positive samples and 3 negative samples should be tested 5 (qualitative).

IVD kit using two different manufacturer recommended platforms (if

Page 400 of 459

applicable). 3 positive samples and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

  1. Blinding of laboratory staff blinded to the results of the reference assay. The PI of the evaluation exercise should remain unblinded, i.e., privy to the results of the reference test. Another senior laboratory staff selected by the PI may remain unblinded for overseeing the activity and maintaining the database of results. The data should be analyzed only by the PI of the evaluating lab. Data confidentiality should be ensured by the PI.

  2. Acceptance Criteria Sensitivity: ≥80% Specificity: ≥90% Invalid test rate (if applicable): ≤5%

  3. Publication Rights: author(s). acceptable.
    References:

  4. John J, Bavdekar A, Rongsen-Chandola T, Dutta S, Gupta M, Kanungo S, et al. Burden of typhoid and paratyphoid fever in India. N Engl J Med 2023;388(16): 1491–500. https://doi.org/10.1056/NEJMoa2209449. PMID: 37075141; PMCID: PMC10116367.

  5. Ministry of Health and Family Welfare-Integrated Disease Surveillance Programme. Training manual for state & district surveillance officers case definitions of diseases & syndromes under surveillance module -5. https://idsp.mohfw.gov.in/WriteReadData/OldSite/2WkDSOSept 08/Resources_files/DistrictSurvMan/Module5.pdf. [Accessed 23 November 2024].

Page 401 of 459

  1. Viswanathan, R., Mukhopadhyay, L., Angrup, A., Kale, P., Bhattacharyya, K., Kanungo, S., Jeromie Wesley, V. T., Shahul Hameed, S. K., Sella Senthil, M., Meshram, P., Kapil, A., Balaji, V., John, J., Ravi, V., & Gupta, N. (2025). Standard protocols for performance evaluation of typhoid fever in-vitro diagnostic assays in laboratory and field settings. Indian journal of medical microbiology, 55, 100857. https://doi.org/10.1016/j.ijmmb.2025.100857
  2. Mather RG, Hopkins H, Parry CM, Dittrich S. Redefining typhoid diagnosis: what would an improved test need to look like? BMJ Glob Health 2019;4(5):e001831. https://doi.org/10.1136/bmjgh-2019-001831. PMID: 31749999; PMCID: PMC6830052.
  3. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data

Page 402 of 459

Performance Evaluation Protocol for Typhoid Antibody-based IVDs

I. Background:
diagnostic kits appropriate for use in India. Hence the following guidelines shall establish

II. Purpose:

To evaluate the performance characteristics of Typhoid antibody-based kits in the diagnosis of Typhoid fever using irreversibly de-identified leftover archived clinical samples.

The protocol outlines performance evaluation of IVD kits that detect S Typhi antibodies in blood/ serum/plasma samples.

Note: Performance evaluation with leftover clinical samples will be feasible only when well- characterized samples, as outlined in the document, are available for use.

III. Requirements:

  1. Evaluation sites/laboratories (With required equipment, and blood culture facilities) Investigators are required to submit a self-declaration form, as outlined in the ICMR guidelines, to the institutional authorities and ethics committee for information.

V. Procedure: through

Page 403 of 459

Lab / Calibration Lab (ISO/IES 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory.  Preparation & characterization of kit evaluation panel
 Handling of kits received for performance evaluation (Verification/Storage/Unpacking etc).  Testing, interpreting, recording of results & reporting  Data handling, data safety & confidentiality 3. Preparation of Typhoid antibody-based IVD evaluation panel Well characterised sample panel is a critical requirement for performance evaluation of IVD kits. Hence statistically significant number of well-characterized samples should be available from confirmed Typhoid fever cases.

  1. Sample size and sample panel composition: Sample size has been calculated assuming 95% level of significance, an absolute precision of 5%, sensitivity of ≥80%, specificity of ≥90%, and invalid test rate ≤5% (where applicable). Sample size are calculated using the formulae: 𝑑2 x (1 −IR)
    𝑑2 x (1 −IR)

                            ·       n (se) is the minimum number of positive samples 
    

· n (sp) is the minimum number of negative samples · Z2 is the critical value from the standard normal distribution
corresponding to the desired confidence level (95% CI corresponds to Z2 =1.96) · Se is the predetermined sensitivity · Sp is the predetermined specificity

A minimum of 259 positive samples (246 samples for assays without the provision of invalid test) and a minimum of 146 negative samples (139 samples for assays without the provision of invalid test). Rounding up, a minimum of 260 positive samples and a minimum of 150 negative samples should be used for evaluation.

  1. Sample panel characterization and Reference standard:
    True positive and negative samples should be characterized as per the “Field Evaluation Protocol for Typhoid Antibody-based IVDs” (Sections 6 and 8 of the document). The sample size and sample panel composition are depicted in Table 1:

Page 404 of 459

Table 1: Sample size and panel composition for analytical evaluation of diagnostic antibody-based assays:
A. Positive samples in the panel (N= 260) Minimum Number of samples needed True positive samples as described in “Field Evaluation Protocol for Typhoid Antibody-based IVDs” (Sections 6 and 8 of the document).

260 B. Negative samples in the panel (minimum N=150) True negative samples should satisfy the conditions described in “Field Evaluation Protocol for Typhoid Antibody-based IVDs” (Sections 6 and 8 of the document). Minimum Number of samples needed Sera from confirmed afebrile non-typhoid cases AND negative for anti-typhoid antibodies (e.g.: healthy blood donors) 80 Sera collected 2-4 weeks after onset of illness in acutely febrile confirmed non-typhoid cases (samples positive for dengue, malaria, scrub typhus, leptospira, chikungunya)@ 40 Dengue: 14 (desirable to have representation from all 4 serotypes) Malaria: 5 Scrub typhus: 5 Leptospira: 8 Chikungunya: 8 Proven cross-reactive samples:

  • Sera collected 2-4 weeks after onset of illness in proven Paratyphoid A, B, and NTS infections (blood culture/PCR confirmed cases) @

-Sera collected 2-4 weeks after onset of illness from confirmed cases of Brucellosis, Rickettsial infections, Enterobacteriaceae infections (e.g. sepsis), and Influenza Recommended: 35 samples Minimum: 30 samples (described below)

  1. Paratyphoid samples (n=10)  Paratyphoid A (n=4)  Paratyphoid B (n=3)  Paratyphoid C (n=3)
  2. NTS samples (n=10)
    One each for the following 10 NTS##:  Salmonella Typhimurium  Salmonella Enteritidis  Salmonella Kentucky  Salmonella Eastbourne  Salmonella Dublin  Salmonella Bareilly
     Salmonella Weltevreden  Salmonella Newport  Salmonella Infantis  Salmonella Agona 3.Cases of Brucellosis and Rickettsial infections (n=8)  Brucellosis (B. abortus and/or B.

Page 405 of 459

melitensis and/or B. suis) cases (n=3) *  Rickettsial (R. rickettsiae and/or R. conorii and/or R. typhi) infections (n=5) *‡ 4. Enterobacteriaceae infection (n=5) Cases of Enterobacteriaceae infection (sepsis) – including E. coli, Klebsiella sp., Citrobacter sp. 6. Influenza cases (n=2) cases of seasonal influenza virus infection (including influenza A and B) NTS = Non Typhoidal Salmonella @ These samples should be characterized as per national guidelines/globally acceptable standards.

In case of unavailability of samples positive for a cross-reacting pathogen, attempt should

be made to meet the gap with samples from clinical cases of infection with other serovars. If deficit persists, the gap should be met with samples from Paratyphoid fever cases. If no sample positive from confirmed cases of these infections is available, test 5 more paratyphoid fever positive samples (in addition to the 10 samples outlined in Point No. 1); this will bring down the size of the cross-reactivity panel to 30 samples. Commercially available serology panels for these pathogens, that are accepted by accreditation agencies, may also be used.

  • Commercially available serology panels for these pathogens, that are accepted by accreditation agencies, may also be used in case of paucity of clinical samples.

‡ In case of unavailability/paucity of Rickettsial serology samples, the gap should be met with scrub typhus IgM antibody positive samples.

Well-characterized true positive and true negative samples as outlined in Table 1 should be subjected to an accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available) to ensure sample integrity at the time of evaluation.

  1. Evaluation method:

The accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available) and the index test should be run simultaneously on the sample panel.

  1. Interpretation of results:
    All test results will be interpreted as per respective kit IFU/SoP.

Page 406 of 459

  1. Resolution of discrepant results:  True Positive samples: confirmed typhoid cases (see Note below) and showing TO/TH titers >320 with accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available), and positive results with the index test, will be considered as true positive sample.
    For semi-quantitative antibody-based tests, results should closely mimic Widal test results.  True Negative samples:
    confirmed non-typhoid cases (see Note below) and testing non-reactive for anti-typhoid antibodies with accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available), as well as the index test, will be considered as true negative sample.
     False positive samples:

confirmed non-typhoid cases (see Note below) and testing non-reactive for anti-typhoid antibodies with accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available), but reactive for anti-typhoid antibodies with the index test, will be considered as false positive samples.

 False negative samples: confirmed typhoid cases (see Note below) and showing TO/TH titers >320 with accredited Widal test (or WHO Pre-Qualified/US FDA/ATAGI Australia/PMDA Japan approved Widal test as and when they become available), but non-reactive for anti-typhoid antibodies with the index test, will be considered false negative samples. Note: The sample panel should be characterized as per the “Field Evaluation Protocol for Typhoid Antibody-based IVDs” (Sections 6 and 8 of the document).

  1. Repeatability and Reproducibility Assessment:

evaluation. 3 positive samples and 3 negative samples should be tested 5 times in independent runs.

Page 407 of 459

Three lots of an IVD kit shall be evaluated. Sample size for lot-to-lot reproducibility  Second lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).
 Third lot of the kit: should be tested on 25 samples (15 positive samples and 10 negative samples).

between-run imprecision (if applicable) should be measured. 3 positive samples

Page 408 of 459

(qualitative).

consecutive days. 3 positive samples and 3 negative samples should be tested 5 (qualitative).

positive samples and 3 negative samples should be tested 5 times in independent runs.
(qualitative).

lab. The PI should maintain confidentiality of the data. Refer to Fig. 1.

Fig.1: Blinding in evaluation exercise

Page 409 of 459

  1. Acceptance Criteria Sensitivity: ≥80% Specificity: ≥90% Invalid test rate (if applicable): ≤5%

  2. Publication Rights: author(s).

acceptable.

References:

  1. John J, Bavdekar A, Rongsen-Chandola T, Dutta S, Gupta M, Kanungo S, et al. Burden of typhoid and paratyphoid fever in India. N Engl J Med 2023;388(16): 1491–500. https://doi.org/10.1056/NEJMoa2209449. PMID: 37075141; PMCID: PMC10116367.
  2. Ministry of Health and Family Welfare-Integrated Disease Surveillance Programme. Training manual for state & district surveillance officers case definitions of diseases & syndromes under surveillance module -5. https://idsp.mohfw.gov.in/WriteReadData/OldSite/2WkDSOSept 08/Resources_files/DistrictSurvMan/Module5.pdf. [Accessed 23 November 2024].
  3. Viswanathan, R., Mukhopadhyay, L., Angrup, A., Kale, P., Bhattacharyya, K., Kanungo, S., Jeromie Wesley, V. T., Shahul Hameed, S. K., Sella Senthil, M., Meshram, P., Kapil, A., Balaji, V., John, J., Ravi, V., & Gupta, N. (2025). Standard protocols for performance evaluation of typhoid fever in-vitro diagnostic assays in laboratory and field settings. Indian journal of medical microbiology, 55, 100857. https://doi.org/10.1016/j.ijmmb.2025.100857
  4. Mather RG, Hopkins H, Parry CM, Dittrich S. Redefining typhoid diagnosis: what would an improved test need to look like? BMJ Glob Health 2019;4(5):e001831. https://doi.org/10.1136/bmjgh-2019-001831. PMID: 31749999; PMCID: PMC6830052.
  5. Monti, C. B., Ambrogi, F., & Sardanelli, F. (2024). Sample size calculation for data reliability

Page 410 of 459

Performance evaluation report format for Typhoid IVDs

Lot No / Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use Valid Up to: Application No. Sample Panel Sample type Positive samples (provide details: clinical/spiked samples)
If the index test claims to detect the disease during a period of illness for which samples are not available, clearly mention the disease period covered by the sample panel used for evaluation.

Page 411 of 459

Negative samples (provide details samples) Results

Positive Negative Total Name of index Typhoid IVD kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

o Cross-reactivity:

FINAL CONCLUSION (Sensitivity and specificity have been assessed in controlled lab setting using kits provided by the manufacturer from the batch mentioned above using ….. sample. Results should not be extrapolated to other sample types.) Disclaimers Note: This report is exclusively for ………….. Kit (Lot No……) manufactured by …………… (supplied by ……….) Signature of Director/ Director-In-charge ……………………
Seal……………………………….

Page 412 of 459

Information on Operational and Test Performance Characteristics Required from Manufacturers

  1. Instructions for Use
  2. Scope of the IVD:
  3. Intended Use Statement
  4. Principle of the assay
  5. Intended testing population
  6. Intended user (laboratory professional and/or health care worker at point-of-care)
  7. Detailed test protocol
  8. Lot/batch No.
  9. Date of manufacture
  10. Date of Expiry
  11. Information on operational Characteristics i. Configuration of the kit/device
  12. Information on Test Performance Characteristics i. Type of sample-serum/plasma/whole blood/other specimen (specify) iii. Sample rejection criteria (if any) v. Any additional device/consumable like sample transfer device, pipette, tube, etc required vii. Pathogens targeted by the kit

Page 413 of 459

x. Manual or automated(equipment)reading xi. Limit of detection/Limit of Quantification and range of detection xiv. Stability and reproducibility (including data) xv. Training required for testing (if any) xviii. Details of cross reactivity, if any xix. Interpretation of invalid and indeterminate results to be provided xx. It is recommended to provide data demonstrating accuracy and precision

*Please mention “Not applicable” against sections not pertaining to the kit.

Page 414 of 459

ANALYTICAL PERFORMANCE EVALUATION OF IN-VITRO DIAGNOSTICS FOR PULMONARY TUBERCULOSIS

Page 415 of 459

LIST of Contributors

A. Working Group:

  1. Dr Sivakumar, Scientist E and Head, Department of Bacteriology, ICMR-NIRT, Chennai

  2. Dr Madhumathi J, Scientist D, Division of Communicable Diseases (CD), ICMR, New Delhi

  3. Dr Amita Jain, Professor and Head, Department of Microbiology, KGMU

  4. Dr RM Pandey, ICMR- Dr A.S. Paintal Distinguished Scientist Chair

  5. Dr. Camilla Rodrigues, Senior Consultant, P.D. Hinduja Hospital, Mumbai

  6. Dr Gita Nataraj, Professor Emeritus, Microbiology, Seth GS Medical College and KEM Hospital, Mumbai

  7. Dr Ashutosh Aggarwal, Professor and Head, Pulmonary Medicine, PGIMER

  8. Dr Venkataraghava Mohan, Professor & Head, Dept. of Community Health and Development, CMC Vellore

  9. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control

  10. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control

  11. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central

  12. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

Page 416 of 459

Analytical Performance Evaluation of IVD for Pulmonary Tuberculosis

I. Background diagnostic kits appropriate for use in India. This protocol gives the methods to be used for evaluating the analytical performance characteristics of the in-vitro diagnostic test in detecting pulmonary tuberculosis and drug-resistant tuberculosis. Note: According to CDSCO guidelines, "performance evaluation" refers to "analytical validation" required for obtaining "test license", while “field evaluation” refers to "clinical validation” performed in clinical samples in real world setting.

II. Purpose To evaluate the performance characteristics of nucleic acid amplification tests (NAAT) forthe diagnosis of pulmonary Mycobacterium tuberculosis (MTB) using irreversibly de-identified leftover archivedor spiked sputum samples.

III. Study Design Analytical validation of IVD using irreversibly de-identified leftover clinical/spiked samples.

IV. Ethical Considerations

  1. Leftover sputum specimens collected for routine diagnostic evaluation from patients who are suspected of having TB shall be used. No additional specimens should be requested.
  2. The probability of harm or discomfort anticipated in the research is nil or not expected.
  3. Performance evaluation activities using irreversibly de-identified leftover clinical samples are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024.
  4. Investigators are required to submit a self-declaration form, as outlined in the ICMR
  5. The protection of privacy of participants should be ensured by using de-identified samples and encrypting the patient identifiers.
  6. Respect for the dignity of participants shall be prioritized.

Page 417 of 459

V. Blinding of Laboratory Staff

To ensure the rigor of the evaluation process, laboratory staff performing the evaluation should unblinded, i.e., privy to the status of the samples. Another senior laboratory staff member selected by the PI may remain unblinded and carry out coding of samples and dispensing them into similar-looking vials to be used for testing, and maintain the database of results. Staff status of samples till the completion of evaluation. The data should be analyzed only by the PI of the evaluating lab. Refer to Fig. 1.

Figure 1 Blinding in evaluation exercise

Page 418 of 459

VI. Procedure

  1. Preparation of Evaluation sites/laboratories C. The laboratory must be approved by the National TB Elimination Program (NTEP). D. Accreditation for at least one Quality management system [accreditation for Testing Lab / Calibration Lab (ISO/IES 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory].

  2. Exclusion  Extra-pulmonary samples  Specimens with > 1 freeze-thaw cycle (or according to IFU, if specified)  Any exclusion criteria stated in the product IFU

  3. Reference tests  For detection of MTB: Mycobacterium Growth Indicator Tubes (MGIT) liquid culture.  For MDR-TB: MGIT drug sensitivity testing (DST)

  4. Preparation of samples  For LOD studies - MTBC-negative sputum: smear-negative and NAAT-negative sputum should be used for the spiking analytic studies  For analytical sensitivity and specificity: Well characterized archived samples (sputum or processed sputum); MTB positives, MTB negatives and Non-Mycobacterium tuberculosis (NTM) samples confirmed by liquid MGIT culture
     For drug sensitivity: MTB and NTM clinical isolatesthoroughly characterized through MGIT DST and sequencingshould be used.
     For inclusivity/exclusivity, resistance detection, and cross-contamination, mycobacterial strains should be diluted into 7H9 medium at the required concentrations.
     The concentrations (cfu/mL) should be estimated by adjusting the bacterial suspension density to the McFarland standards.

  5. Reference Strains The National Institute for Biological Standards and Control (NIBSC) internal reference standard for Mycobacterium tuberculosis (H37Rv) DNA for Nucleic Acid Amplification Test (NAAT) based assays (NIBSC code: 20/152) will be used for the LOD assay. It was established as the 1st WHO International Standard for Mycobacterium tuberculosis (H37Rv) DNA for NAAT-based assays in

  6. The intended uses of this material are for calibration of secondary or in-house reference

Page 419 of 459

materials used in the assays for the molecular detection of M. tuberculosis DNA.It may also be used for assay validation and monitoring the limit of detection of rapid diagnostic tests. This preparation contains an arbitrary unitage of 6.3 log10 (or 2 million) IU per vial.

  1. Sample size and sample panel composition With an anticipated sensitivity of 90% and relative precision of 7%, a minimum of 87 confirmed MTB positive samples by MGIT culture will be required for testing analytical sensitivity. With an anticipated specificity of 95% with 5% relative precision, the minimum sample size required for analytical specificity is 81 confirmed MTB negative samples by MGIT culture. To rule out NTM detection, with an assumed sensitivity of 90% and relative precision of 10%, around 50 confirmed NTM samples may be included to evaluate the index test kit. Hence, approximately 100 confirmed MTB positives, 100 confirmed MTB negatives and 50 NTM samples will be used for pre-validation studies.

The proposed evaluation study will be done using Sputum/MTB isolates stored at the biobank facility of the National TB reference laboratories (NRLs) or the pre-validation labs. The stored sputum/MTB isolate/processed sample/DNA samples will be of the following categories and sub-categories. Category 1: Positive for MTB by MGIT culture (N = 100) [MTB positives should include equal number of low, medium and high bacterial load samples demonstrated by Xpert MTB]

Category 2: Negative for MTB by MGIT culture (N = 150)
Within the MTB negative group, we propose the following two sub-categories: i. Negative for all Mycobacteria (N = 100) ii. Positive for Non-Tuberculous Mycobacterium (N = 50)

Equivalence studies:

If a new specimen or format needs to be evaluated for an already recommended IVD, an equivalence study may be performed with 50 MTB positive and 50 MTB negative specimens to establish relationship between IVD performance in the newly claimed specimen type or new format of same technology. However, if the technology used for the IVD design is different, it shall go through the full process of validation.

Category 3: If resistance detection has to be carried out, within the MTB positive group, we propose to use the following sub-categories:

Page 420 of 459

i. Sensitive to Rifampicin and Isoniazid, individually and combined (N =100) confirmed by Drug susceptibility testing on MGIT liquid culture. ii. Resistance to both Rifampicin and Isoniazid (N = 100) as detected by Drug susceptibility testing on MGIT liquid culture. iii. Isoniazid mono-resistance (N =45) as detected by DST on MGIT liquid culture. iv. Fluroquinolone resistance (N=45) (if applicable for the index test) as confirmed by DST on MGIT liquid culture.

Table 1: Sample size calculation with 95% confidence level Anticipated Sensitivity Relative Precision Sample size 90% 5% 171 90% 10% 43 90% 7% 87 95% 5% 81 95% 10% 20 95% 7% 41

Page 421 of 459

Analytical sensitivity and specificity:

Figure 2. Flowchart for Analytical Performance Evaluation for detection of; a) MTB detection, b) MDR-TB

  1. Limit of Detection (LOD) Assay The 95% LOD is defined as the minimum concentration of bacterium, expressed as CFU/ml or genomic DNA copy numbers/mL, in a sample volume that can be detected in 95% of tests. Finalize the LOD at least one concentration with a hit rate above 95% and two concentrations with hit rates between 10% and 90%. LOD should be always done with NIBSC H37Rv (20/152) standard and only reported in IU/ml or CFU/ml.

a) b)

Page 422 of 459

Preparation of samples for LOD

  1. The volume of sputum required for LOD is based on the IFU (Instruction for use) from the index test manufacturer, which generally varies between 1-2 ml of sputum.
  2. A minimum of 200ml of NAAT negative sputum is required for the full LOD studies for a single index test.
  3. Sputum samples which are negative by Xpert MTB will be stored at -20oC and once the required amount is obtained the samples will be pooled and tested for MTB using molecular and phenotypic test to prove no growth of MTB in the pooled samples.
  4. To perform the assay it may take two weeks to one month based on the multiplicities of test suggested in the IFU after the required volume of sputum is collected. Spiking of sputum samples
  5. The spiked sputum will be used to determine the LOD of the test kit. About 1.8 ml of negative sputum specimen will be spiked with 200 ul of the respective diluted suspension series of M. tuberculosis H37Rv.
  6. These dilutions will be added to the sputum to get the final concentration (10000, 1000, 100, and 10 IU/ml). Before spiking, the culture for CFU will be set up for the different dilutions.
  7. NIBSC reference standard will be reconstituted as directed by NIBSC using 1 mL nuclease free molecular biology grade purified water (MBGPW). From this stock 100 µL will be diluted ½ to get 10,00,000 IU/ml and serially diluted to give 100000, 10000, 1000 and 100 IU/ml with MBGW.
  8. Each dilution of the WHO International Standard, will be tested 24 times. The 24 replicates will be performed over at least three days by at least two users and, for low- throughput instruments, on at least three different instruments, or sets of instruments if applicable (e.g., DNA preparation and amplification instruments). For low through-put instruments, the number of testing days may be increased.
  9. When M.tuberculosis (H37Rv) is used, the 24 replicate tests shall comprise (8 replicate tests on each day for 3 days) of a minimum 8-member 0.5 log 10 dilution panel of a suitable biological reference material (e.g., WHO International Standard) (WHO TSS- 17).
  10. Each lot shall comprise different production (or manufacturing, purification, etc.) runs of critical reagents. Inter-lot variation must be evaluated by appropriate statistical means.

Page 423 of 459

  1. Lowest dilution at which the test detects M.tb will be determined a LOD, the corresponding CFU will also be counted and reported in terms of CFU/per ml. The LOD will be presented as IU/mL for each dilution.
  2. Analytical sensitivity shall be estimated by determining the 95% LOD with 95% confidence intervals (e.g., by probit analysis).
  3. If there are more than four invalid results with the same specimen (i.e. dilution) overall, then the specimen will be retested to get at least 20 valid results for each dilution. For tests that include a claim for drug resistance testing, at least 20 valid results (i.e., sensitive or resistant) for each of the claimed drugs should be obtained for each dilution.
  4. To arrive at the LOD a probit analysis should be performed, Probit analysis is defined as a specialized form of regression analysis applied to binomial response variables, transforming a concentration-response curve into a straight line for analysis through methods like least squares or maximum likelihood regression. It is primarily used in molecular biology measurement procedures, such as PCR, to determine the detection probability of analytes at various concentrations.
    LOD for detection of drug resistance
  5. To test the drug resistant MTB strains, well-characterized MTBC strains of known concentration (expressed as CFU/mL) shall be spiked into each claimed MTBC negative specimen type. DR strains shall be characterized by sequencing.
  6. Relevant DR strains (as mentioned in table below) shall be spiked into each claimed MTBC-negative specimen type (e.g., raw and/or processed sputum, and each claimed extra-pulmonary specimen).
  7. If the assay detects resistance to more than 1 target drug, the LOD for each target drug in addition to a composite resistance LOD, defined as the highest LOD among the tested target, shall be reported.
  8. Analytical sensitivity for resistance detection shall be estimated as the lowest number of colony-forming units (CFU) per specimen that can be reproducibly distinguished from negative specimens with 95% confidence.
  9. The determination shall comprise 24 replicate tests (8 replicate tests on each of 3 days) of a minimum 8 8-member 0.5log10 dilution panel. The replicate testing shall be conducted on three different days using 2 lots, and at least 2 dilution series shall be tested. Table 2: Anti-mycobacterial drugs and common mutations

Page 424 of 459

S.No Drugs Resistance mutation of strains to be tested
1 Isoniazid katG_S315T and fabG1_c-15t 2 Rifampicin rpoB_S450L; rpoB_D435V; rpoB_H445Y; rpoB_H445D; rpoB_D435Y; rpoB_S450W; rpoB_L452P; rpoB_H445L; rpoB_S450F; rpoB_L430P; rpoB_H445R; one rpoC mutation 3 Levofloxacin (CC) LFX2,3
gyrA_A90V, gyrA_D94G, gyrA_D94H, gyrA_D94N, gyrA_D94Y, gyrA_S91P 4 Moxifloxacin (CC and CB) gyrA_A90V, gyrA_D94G, gyrA_D94H, gyrA_D94N, gyrA_D94Y, gyrA_S91P 5 Bedaquiline Rv0678_LoF, pepQ_LoF, atpE_p.Ala63Pro 6 Linezolid rplC_p.Cys154Arg, rrl_n.2814G>T
7 Ethambutol embB_M306L, embB_M306V, embB_Q497R 8 Delamanid ddn_LoF, ddn_p.Leu49Pro, fbiC_LoF 9 Pyrazinamide pncA_ V139A, pncA_ V139G 10 Amikacin rrs_ A1401G, rrs_ A1401G, rrs_G1484T, eis /promoter_ C-12T, eis /promoter_C-14T
11 Kanamycin rrs_ A1401G, rrs_ A1401G, rrs_G1484T, eis /promoter_ C-12T, eis /promoter_C-14T 12 Capreomycin rrs_ A1401G, rrs_ A1401G, rrs_G1484T, eis /promoter_ C-12T, eis /promoter_C-14T 13 Ethionamide fabG1_c-15t, inhA_S94A, fabG1_ T-8C 14 Pretomanid# ddn_LoF, ddn_p.Leu49Pro, fbiC_LoF 15 Cycloserine Alr_C-8T, alr_M319T, alr_Y364D, ald_T-32C, ddlA T365A 16 PAS thyA T22A, folC I43T, folC R49W 8. Reproducibility Three lots of a test shall be evaluated for lot-to-lot reproducibility. Each lot will comprise different production (or manufacturing, purification, etc.) runs of critical reagents.  First lot of the assay: should be tested on statistically significant number of positive and negative samples as calculated in the protocol.
 Second lot of the assay: should be tested on 20 samples (10 MGIT positive samples and 10 MGIT negative samples).
 Third lot of the assay: should be tested on 20 samples (10 MGIT positive samples and 25 MGIT negative samples).

Within-run (same operator, same measuring system, same operating conditions, and same location), Between-run, -lot, -day, -site, -operator.

  1. Three specimens will be used; MTB sensitive (H37Rv), MTB resistant and MTB

Page 425 of 459

negative. 2. The effect of operator-to-operator variation on IVD performance will be included as part of the precision studies.
3. The nucleic extraction/purification component will also be considered for estimating precision.
4. Contrived specimens will be used (i.e., MTBC strains with specific/most common mutations in the target genes spiked into a clinical matrix claimed in the IFU) for repeatability and reproducibility studies. 5. DR specimens at the concentrations specified for each DRTB (i.e. RR-TB, Hr-TB, MDR-TB, TB resistant to fluoroquinolones) as described in the table on resistance detection.
6. If there are two or more invalid results for the same specimen in the same run, then the run should be repeated for this specimen. Invalid results should be reported. 7. Results will be statistically analyzed by ANOVA or other methods to identify and isolate the sources and extent of any variance.
8. Furthermore, the percentage of correctly identified, incorrectly identified, and invalid results will be compiled for each specimen and separately categorized by site, lot, and other factors. 9. Within-run and within-laboratory reproducibility will be assessed by measuring eight replicates. At least two operators will test a total of 40 positives and 40 negatives with two batches of the kit over a period of 5 days. 9. Inclusivity and exclusivity

  1. Inclusivity MTBC stains: For a claim of MTBC detection, the following strains shall be tested: M. tuberculosis, M. bovis, M. bovis BCG, M. africanum, M.microti and M.caprae
  2. Exclusivity NTM strains: M. avium, M. kansasii, M. intracellulare
  3. Representative MTBC and non-tuberculosis mycobacteria (NTM) strains will be tested in triplicate forinclusivity and exclusivity verification.
  4. Resistance detection: For assays with a claim for detection of drug resistance, the applicable specimens from the resistancedetection panelwill be tested in triplicate.
  5. The concentration of MTBC isolates used in inclusivity studies will be at levels at or near the specific LOD and will be confirmed by plating/ counting bacterial CFUs (estimated using Truenat).

Page 426 of 459

  1. The selection of specific MTBC strains with relevant genetic variations linked to DR will be made to support the claims in the IFU.

  2. This will involve testing strains that carry the most common mutations, including associated or interim resistance mutations, covering at least 80% of the resistance mechanisms observed globally for each of the assay target drugs (as shown in table 2).

  3. Cross-contamination/carry-over

  4. The experiment will allow the determination of the well-to-well or vial-to-vial cross- contamination rate ofhigh-throughput platforms or potential carryover in low- throughput instruments.

  5. This will be assessed byalternating one high-positivespecimen with one negative specimen and repeating thissequence twenty times.

  6. For high-throughput assays, this will be achieved by alternating high-positive andhigh- negative specimens in the same plate/run.

  7. For low-throughput assays, each sequence of highly positivespecimens followed by negative specimens should be done on the same instrument.

  8. If more than one instrument isused, each run (i.e same instrument and same day) should include a minimum of 2 sets of alternating high-positive and negative specimens.

  9. Contrived specimens prepared by spiking MTBC strains into MTBC negative clinical sputum will be used for these studies.

  10. Resolution of discrepancy:

 The results of MGIT culture should be used to resolve any discrepancy in detection of MTB  Results of phenotypic DST and sequencing should be used to resolve discrepancy in detection of MDR-TB.

VII. Statistical Analysis Plan

  1. The index molecular test should be evaluated for its analytical sensitivity and analytical specificity.
    Note: The strains used for assessment of reproducibility, inclusivity/exclusivity, resistance detection, and carry-over may be commercially acquired or locally prepared, well-characterized strains (by phenotypic DST and sequencing).

Page 427 of 459

  1. 95% Confidence interval should be calculated for each of the parameters. % Sensitivity = Positives by index test

x 100 = [a/a+c] * 100

         Confirmed positives by MGIT culture   

% Specificity= Negatives by index test

x 100 = [d/b+d] * 100 Confirmed negatives by MGIT culture

VIII. Acceptance Criteria

Acceptance criteria for Diagnostic tests:
Expected sensitivity: ≥ 90% Expected specificity: ≥ 95% Sample Size: ~ 100 confirmed MTB positives (by MGIT culture), ~ 100 confirmed MTB negatives (by MGIT culture) and ~ 50 NTM samples (confirmed by culture and identification)

Acceptance criteria for Screening tests:
Test Type Minimal Accuracy Optimal accuracy High Sensitivity high specificity screening test 90% sensitivity
80% specificity 95% sensitivity
95% specificity High Sensitivity screening test 90% sensitivity
60% specificity 95% sensitivity
85% specificity High specificity screening test 60% Sensitivity
98% specificity 70% sensitivity
98% specificity Source: WHO TPP 2025

IX. Publication Rights author(s).

IMPORTANT NOTE Once a kit is determined to be “Not of Standard Quality”, following the procedure outlined in this document, no further requests for repeat testing of that kit will be accepted. Any request of re-validation from the same manufacturer for the same test type will only be entertained after a minimum of 3 months and only if a high-level technical

Page 428 of 459

summary of modifications or functional improvements to the kit design is submitted, without explicit disclosure of proprietary information.

References

  1. Technical Specifications Series for Submission to WHO Prequalification – Diagnostic Assessment. TSS 17 In vitro diagnostic medical devices used for the qualitative detection of Mycobacterium tuberculosis complex DNA and mutations associated with drug-resistant tuberculosis.https://iris.who.int/bitstream/handle/10665/366068/9789240055865-eng.pdf.
  2. Miotto, P., Tessema, B., Tagliani, E., Chindelevitch, L., Starks, A. M., Emerson, C., Hanna, D., Kim, P. S., Liwski, R., Zignol, M., Gilpin, C., Niemann, S., Denkinger, C. M., Fleming, J., Warren, R. M., Crook, D., Posey, J., Gagneux, S., Hoffner, S., Rodrigues, C., … Rodwell, T. C. (2017). A standardised method for interpreting the association between mutations and phenotypic drug resistance in Mycobacterium tuberculosis. The European respiratory journal, 50(6), 1701354. https://doi.org/10.1183/13993003.01354-2017.
  3. Indian Catalogue of Mycobacterium tuberculosis Mutations and their Association with Drug Resistance, Version 2.0 2024, ICMR-NIRT. https://www.nirt.res.in/pdf/mutation_catalogue_v2.pdf.
  4. Target product profiles for tuberculosis screening tests. Geneva: World Health Organization; 2025. Licence: CC BY-NC-SA 3.0 IGO. Available at: https://iris.who.int/bitstream/handle/10665/382179/9789240113572-eng.pdf?sequence=1.
  5. WHO manual for the preparation of secondary reference materials for in vitro diagnostic assays designed for infectious disease nucleic acid or antigen detection: calibration to WHO International Standards. Available at: https://www.who.int/publications/m/item/annex-6-trs-no-1004

Page 429 of 459

PERFORMANCE EVALUATION REPORT FORMAT

Performance Evaluation Report For MTB/MDR-TB Kit

Name of the product (Brand/generic) Lot No /Batch No.: Type of Assay Kit components Manufacturing Date Expiry Date Intended Use Import license / Manufacturing license/ Test license

License Number: Issue date: Valid Upto: Application No. Sample Panel Sample type Positive samples (provide details: strong, moderate, weak)
Negative samples (provide detail: clinical/spiked, including cross reactivity panel)

Results:

Reference assay ……….…………… (MGIT/MGIT DST for RIF/INH/FQ/others)

Positive Negative Total Name of MTB or MDR-TB kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

Conclusions:

o Sensitivity, Specificity

Page 430 of 459

DISCLAIMERS

Note: This report is exclusively for ………………………Kit (Lot No……), version …………with the gene targets …………………...manufactured by …………… (Supplied by ……….).

Signature of Director/ Director-In-charge …………………… Seal …………………………

Page 431 of 459

FIELDPERFORMANCE EVALUATION OF IN- VITRO DIAGNOSTICS FOR PULMONARY TUBERCULOSIS

Page 432 of 459

List of Contributors

A. Working Group: 1. Dr Gomathi N. Sivaramakrishnan, Former Scientist, ICMR-NIRT, Chennai 2. Dr Sivakumar, Scientist E and Head, Department of Bacteriology, ICMR-NIRT, Chennai 3. Dr Madhumathi J, Scientist D, Division of Communicable Diseases (CD), ICMR, New Delhi 4. Dr Hansraj Choudhary, Scientist C, Division of Communicable Diseases (CD), ICMR, New Delhi

  1. Dr Amita Jain, Professor and Head, Department of Microbiology, KGMU
  2. Dr RM Pandey, ICMR- Dr A.S. Paintal Distinguished Scientist Chair
  3. Dr. Camilla Rodrigues, Senior Consultant, P.D. Hinduja Hospital, Mumbai
  4. Dr Gita Nataraj, Professor Emeritus, Microbiology, Seth GS Medical College and KEM Hospital, Mumbai
  5. Dr Ashutosh Aggarwal, Professor and Head, Pulmonary Medicine, PGIMER
  6. Dr Venkataraghava Mohan, Professor & Head, Dept. of Community Health and Development, CMC Vellore
  7. Mr. Pramod Meshram, Deputy Drugs Controller, Central Drugs Standard Control
  8. Dr. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control
  9. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central
  10. Dr. Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family Welfare, Government of India

Page 433 of 459

Field Performance Evaluation of IVD for Pulmonary Tuberculosis I. Background diagnostic kits appropriate for use in India. This protocol gives the methods to be used for evaluating the clinical performance characteristics of nucleic acid amplification based in-vitro diagnostic test in detecting pulmonary tuberculosis. Note: According to CDSCO guidelines, "performance evaluation" refers to "analytical validation" required for obtaining "test license", while “field evaluation” refers to "clinical validation” performed in clinical samples in real world setting.
II. Purpose To evaluate the clinical performance characteristics of nucleic acid amplification tests (NAAT) for diagnosis of pulmonary Mycobacterium Tuberculosis (MTB) using prospectively collected sputum samples in clinical setting. III. Study Design Cross-sectional prospective multi-centric diagnostic accuracy study of IVD for detection of pulmonary TB using Mycobacterium Growth Indicator Tube (MGIT) liquid culture as the microbiological reference standard. IV. Ethical Considerations

  1. The study should be compliant to the ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024. Performance evaluation activities using irreversibly de-identified leftover clinical samples are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024. Investigators are required to submit a self-declaration form, as outlined in the ICMR guidelines, to the institutional authorities and ethics committee for information.
  2. Sputum specimens should be collected, as required for routine diagnostic evaluation, from patients who are suspected of having pulmonary TB as per algorithm. Probability of harm or discomfort anticipated in the research is nil or not expected.
  3. Enrolment of subjects should be continued till the sample size is met or till the project duration is completed.
  4. If additional sputum sample is obtained, written consent must be obtained as per the ICMR National Ethical Guidelines for Biomedical and Health Research Involving Human Participants. The institutional ethics committee of each participating site should be intimated about the study for necessary approval prior to initiating the study. Assent form

Page 434 of 459

should be collected in addition to Informed Consent in case of adolescents (13 to 16 yrs). For children between 7 and 12 years old, oral assent should be obtained in presence of parent or legal guardian. For children under 7 years old, written informed consent should be obtained from parent or legal guardian. 5. The protection of privacy of research participants will be ensured by encrypting the patient identifiers. 6. Patients shall receive the best possible diagnostic work-up as per the routine practice and the National Tuberculosis Elimination Program (NTEP) guidelines. There should not be delay in sending report due to the study.
7. TB treatment decisions should not be made based on the result of the index test under evaluation, but on the basis of the routine clinical and laboratory methods (smear, solid / liquid culture, standard NAAT results, and clinical work-up).
8. Respect for the dignity of research participants should be prioritized. 9. No compensation shall be provided to the participants since there is no additional cost or travel involved in sample collection for the study. Patients should be compensated for travel and time only if they are asked to pay additional visits exclusively for the sake of the study and not during regular treatment visits. 10. Follow-up visits may be required for a very limited number of discrepant patients to exclude TB. 11. Leftover sputum samples and deposits should be stored for resolving discrepancies. One positive culture and two DNA samples per patient should be stored at -80oC for use later. 12. All the sites should follow up with all study participants till the final diagnosis is made and the patient should be initiated on appropriate treatment as per NTEP norms. Those found to be M. tuberculosis complex (MTB)positive by standard NAAT test should be started on anti-tuberculosis treatment (ATT) by medical officer of the study site as per NTEP guidelines.
13. The findings of the study should be made accessible through reports.

V. Blinding of Laboratory Staff looking vials to be used for testing, and maintaining the database of results.

Page 435 of 459

Staff performing the reference test and the test under evaluation (index test), interpretation of the test result, and entering the results against the coded samples in the database, should remain blinded to the status of samples till the completion of evaluation.

Operators conducting routine laboratory tests (smear, Xpert MTB/RIF, MGIT culture etc) will not participate in the index test evaluation. Instead, dedicated operators, who are not involved in routine testing and are blinded to the routine test results, will perform the index test.The results will be recorded independently for each test without any patient identifiers. The result sheets will be shared with the investigator for result analysis.The data should be analyzed only by the PI of the evaluating lab (Fig. 1).

Figure 1 Blinding in evaluation exercise

VI. Procedure

  1. Preparation of Evaluation sites/laboratories  Laboratory must be approved by the National TB Elimination Program (NTEP).
     Accreditation for at least one Quality management system [accreditation for Testing Lab / Calibration Lab (ISO/IES 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory].

Page 436 of 459

 Three or more sites from different geographical regions should perform clinical validation for representation of population in real world setting.

  1. Study Participants Individuals with symptoms of presumptive pulmonary TB attending hospital OPDs/Chest clinics/district microscopy centers (DMCs) and Directly Observed Therapy Short Course (DOTS) centers. All such consecutive cases willing to provide consent will be enrolled in the study.

  2. Eligibility of Participants Inclusion Criteria

  3. Individuals positive for TB by smear or any approved NAAT test (Xpert® MTB/RIF)

  4. Individuals willing to give consent

  5. Individuals who are able and willing to give two good quality mucopurulent sputum samples of ≥ 3 ml Exclusion criteria

  6. Individuals on TB treatment for >96 hrs

  7. Individuals not consenting for the study

  8. Individuals unable to produce two sputum samples of ≥ 3 ml

  9. Reference and Index tests
    Reference test: Mycobacterium Growth Indicator Tubes (MGIT) liquid culture Comparator: NTEP approved NAAT test (Xpert® MTB/RIF)

  10. Sample size
    The anticipated sensitivity of an index test is 90 % and with absolute 5 % precision, while the anticipated specificity is 99 per cent with 1 % precision. A higher precision for specificity Definition of Presumptive PTB: Patients with any of the following symptoms regardless of duration will be considered to have
    ‘presumptive TB’: cough for two weeks or more, fever for two weeks or more, night sweats, unintentional weight loss, hemoptysis, chest pain or loss of appetite, with any abnormality in chest radiograph (one or more of the following findings by standardized interpretative criteria: cavitary lesion(s), apical infiltrates, hilar lymphadenopathy, new infiltrates and other suggestive radiological findings).

Page 437 of 459

would be required to minimize false positivity. The minimum sample size requirement has been calculated as ~150 positives and ~470 negatives for MTB by the gold standard culture.
With a prevalence of 24 % culture positives among presumptive cases in hospital setting (Penn-Nicholson et al., 2021) and a 5 % loss due to indeterminate results, approximately 610 consecutive cases meeting the inclusion and exclusion criteria would be required to be enrolled for the detection of MTB (Jayaprakasam et al., 2024). Enrolment would be continued till the required number of participants is covered.

The formula for calculating sample size for determining sensitivity/specificity of the index test:

NSe = [Z (1-α/2)]2 (Se)(1-Se)] d2 or

NSp = [Z (1-α/2)]2 (Sp)(1-Sp)] d2 NSe: Sample size for estimating sensitivity,
Se: Anticipated sensitivity with reference to culture DST
Sp: Anticipated specificity with reference to culture DST
Z (1-α/2):1.96 for confidence level of 95% d: Absolute precision

Page 438 of 459

  1. Implementation Plan

Figure 2 Flowchart for evaluating NAAT test for detection of Mycobacterium Tuberculosis (MTB) among individuals with presumptive pulmonary TB (PTB)

Page 439 of 459

  1. Sample collection, processing and storage

  2. Two sputum samples each of minimum 3 ml should be collected (one spot and one morning specimen) and sent to laboratory.

  3. Approximately 1 ml of sample should be taken from each sample and pooled under sterile conditions (total of 2 ml).

  4. Around 1 ml of pooled sample should be tested by the standard NAAT (Xpert MTB/RIF®) and remaining sample used for index test(s).

  5. The remaining portion of each sputum sample should be subjected to direct smear and decontamination by NaLC-NaOH method individually.

  6. The resultant deposit should be used for inoculation into two MGIT960 tubes.

  7. All positive cultures should be identified using rapid Immuno-chromatography test (ICT). (Ideally, positive MGIT tubes are tested within 5 days of instrument positivity. Interpretation of the result should be done within 15 minutes).

  8. All sputum samples should be stored at -20oC for later use. Decontaminated sediments and one positive culture per patient should be stored at -80oC, if necessary for later use.

  9. Two DNA samples per patient should be stored at -20oC till the end of the study for resolution of discrepant results.

  10. The index tests should be carried out as per the algorithm (figure 2) and as per the manufacturers’ instructions in the instructions for use (IFU).

All conventional test procedures for smear, culture (solid and liquid) and Xpert MTB will be performed as per NTEP national laboratory guidelines (CTD, 2016; RNTCP 2009) and laboratory manual of ICMR-NIRT (NIRT, 2010). Standard operating procedures for index test(s) will be provided by the manufacturer(s) including use of positive and negative controls. All procedures for preparation of media, reagents, washing, decontamination, disposal and storage will be performed according to the standard operating procedures (SOP) of ICMR- NIRT (NIRT, 2010) and WHO, (WHO, 2022).

  1. Laboratory Tests

i. Smear microscopy: Two direct sputum smear
ii. MGIT culture (decontaminated with 1-1.5% final NaOH); Two MGIT tubes (one per specimen) for each patient iii. Speciation of culture: Rapid immune-chromatographic test (ICT) of MGIT culture

Page 440 of 459

iv. Xpert MTB/RIF (one test per patient)

  1. Data Analysis and resolution of discrepancy i. If the index test produces error or indeterminate results, then only one repeat is allowed. The results of first test and repeat test should be recorded separately.
    ii. All Invalids/Indeterminates/errors should be recorded and reported. iii. A subgroup analysis may be carried out for pediatric population.

  2. Quality Control (QC) measures All sites should ensure high quality of laboratory procedures, data recording and documentation. There should be no deviation from the protocol. All the sites should participate in internal quality control (IQC) and external quality assurance (EQA) for all methods as per the standard manuals of Global Laboratory Initiative (GLI, 2014). Culture: Positive (Reference strain H37Rv or H37Ra) and negative controls for MGIT and LJ cultures would be tested as per NTEP guidelines. MGIT Time to detection QC for MTB reference strain would be performed every month/new lot of reagents/machine service. Sterility and performance testing of culture media would be performed with every new batch or lot. Smear: Smear QC should be performed as per NTEP guidelines at regular intervals and with new lot of reagents.
    ICT Identification of MTB complex: Culture of M. tuberculosis reference strain in MGIT broth should be used as positive control. Culture of Mycobacteria other than tuberculosis (e.g., a well characterized strain of M. avium complex/M.kansasii) in MGIT broth should be used as negative control. QC for ICT should be performed every 3 months. Molecular diagnostics: For molecular diagnostics internal quality control includes control supplied by the manufacturer and control prepared by the lab from the previous testing. The internal control should be used whenever batch of test kit changes, machine is serviced, and newly trained person is introduced into the system.
    Avoiding Cross-contamination: Unidirectional workflow: The workflow of a molecular lab should be in one direction only. PCR master mix reagents and samples that may contain templates for PCR should be prepared in the pre-PCR room only. Tubes that have undergone amplification in the post-PCR room contain amplicons and will not be opened or introduced in the pre-PCR room. Consumables and PPE (lab coats, gloves, goggles, etc.) that have been used in the post-PCR room should not be placed back in the pre-PCR room without thorough

Page 441 of 459

decontamination. Aerosol resistant pipettes will be used for all procedures and standard aseptic cleaning technique should be carried out before and after PCR for work surface, bench top and equipment.

VII. Statistical Analysis Plan

i. The performance of the diagnostic kits should be evaluated by calculating the sensitivity, specificity, positive predictive value, negative predictive value and accuracy with reference to the gold standard. 95% Confidence interval should be calculated for each of the parameters. ii. The index molecular test should be evaluated for its performance with reference to the MGIT culture. iii. Similarly, the performance of standard molecular test (Xpert MTB/RIF) should be estimated with reference to MGIT culture. iv. The sensitivity and specificity of index test vs MGIT culture should be compared with that of Xpert® MTB/RIF Vs MGIT culture. v. The agreement between the index test and standard NAAT test (Xpert MTB/RIF) should be calculated with kappa statistic.

VIII. Acceptance Criteria for diagnostic tests Expected sensitivity: ≥85 ± 2% Expected specificity: ≥95 ± 2% Sample size: ~150 MTB positives and ~470 MTB negatives by MGIT culture For screening tests the acceptability criteria will be as per WHO TPP 2025
Acceptance criteria for Screening tests:
Test Type Minimal Accuracy Optimal accuracy High Sensitivity high specificity screening test 90% sensitivity
80% specificity 95% sensitivity
95% specificity High Sensitivity screening test 90% sensitivity
60% specificity 95% sensitivity
85% specificity High specificity screening test 60% Sensitivity
98% specificity 70% sensitivity
98% specificity Source: WHO TPP 2025

Page 442 of 459

IMPORTANT NOTE acceptable.

Atleast two different lots or batches should be used for the field validation of any new molecular test.

References

  1. Penn-Nicholson, A., Gomathi, S. N., Ugarte-Gil, C., Meaza, A., Lavu, E., Patel, P., Choudhury, B., Rodrigues, C., Chadha, S., Kazi, M., Macé, A., Nabeta, P., Boehme, C., Gangakhedkar, R. R., Sarin, S., Tesfaye, E., Gotuzzo, E., du Cros, P., Tripathy, S., Ruhwald, M., … Members of the Truenat Trial Consortium: (2021). A prospective multicentre diagnostic accuracy study for the Truenat tuberculosis assays. The European respiratory journal, 58(5), 2100526.
  2. Jayaprakasam, M., Pandey, R. M., Choudhary, H., Shanmugam, S., Sivaramakrishnan, G. N., & Gupta, N. (2024). Evaluation of molecular diagnostic test for detection of adult pulmonary tuberculosis: A generic protocol. The Indian journal of medical research, 159(2), 246–253.
  3. Technical and operational guidelines for tuberculosis control in India 2016. Central TB Division.
  4. RNTCP Standard Operating Procedures for Tuberculosis lab for culture and DST, 2009.
  5. Standard Operating Procedures (SOP) for Mycobacteriology laboratory, ICMR-NIRT,
  6. Practical manual on tuberculosis laboratory strengthening, 2022 update. Geneva: World Health Organization; 2022. Licence: CC BY-NC-SA 3.0 IGO.
  7. Mycobacteriology laboratory manual, Global laboratory initiative, First edition, April 2014, Stop TB Partnership.

Page 443 of 459

PERFORMANCE EVALUATION REPORT FORMAT

Performance Evaluation Report ForMTB Kit

Name of the product (Brand/generic) Name and address of the actual manufacturing site office of Lot No /Batch No.: Product Reference No/Catalogue No Type of Assay Kit components Manufacturing Date Expiry Date Intended Use Import license / Manufacturing license/ Test license License Number: Issue date: Valid Upto: Application No. Sample Panel Sample type weak)
Negative samples (provide detail: clinical/spiked, including cross reactivity panel)

Results:

Test Number of samples tested Positive Negative Invalids/ Indeterminates/Error/ Contamination (culture) Smear

MGIT culture

Xpert MTB/RIF

New MTB kit

Reference assay ……….……………

Page 444 of 459

(MGIT culture)

Positive Negative Total Name of MTB kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

Conclusions:

o Sensitivity, specificity

DISCLAIMERS

Note: This report is exclusively for ………………………Kit (Lot No……), version …………with the gene targets …………………...manufactured by …………… (Supplied by ……….).

Signature of Director/ Director-In-charge …………………… Seal …………………………

Page 445 of 459

FIELD PERFORMANCE EVALUATION OF IN- VITRO DIAGNOSTICS FOR PULMONARY DRUG RESISTANT TUBERCULOSIS

Page 446 of 459

List of Contributors

A. Working Group:

  1. Dr Sivakumar, Scientist E and Head, Department of Bacteriology, ICMR-NIRT, Chennai

  2. Dr Joy Sarojini Michael, Professor, CMC, Vellore

  3. Dr Shubhada Shenai, Senior Scientist, FIND

  4. Dr Madhumathi J, Scientist D, Division of Communicable Diseases (CD), ICMR, New Delhi

  5. Dr Amita Jain, Professor and Head, Department of Microbiology, KGMU

  6. Dr RM Pandey, ICMR- Dr A.S. Paintal Distinguished Scientist Chair

  7. Dr. Camilla Rodrigues, Senior Consultant, P.D. Hinduja Hospital, Mumbai

  8. Dr Gita Nataraj, Professor Emeritus, Microbiology, Seth GS Medical College and KEM Hospital, Mumbai

  9. Dr Ashutosh Aggarwal, Professor and Head, Pulmonary Medicine, PGIMER

  10. Dr Venkataraghava Mohan, Professor & Head, Dept. of Community Health and Development, CMC Vellore

  11. Mr PramodMeshram, Deputy Drugs Controller, Central Drugs Standard Control

  12. D. Sella Senthil, Assistant Drugs Controller, Central Drugs Standard Control

  13. Dr. Md Omair Anwar, Drugs Inspector (Medical Devices), IVD Division, Central Drugs Standard Control Organization, Government of India

  14. Dr Nivedita Gupta, Scientist-G and Head of the Division of Communicable Diseases, ICMR Headquarters, Department of Health Research, Ministry of Health and Family

Page 447 of 459

Field Performance Evaluation of IVD for Pulmonary DR-TB

I. Background

diagnostic kits appropriate for use in India. This protocol gives the methods to be used for evaluating the clinical performance characteristics of the in-vitro diagnostic test in detecting pulmonary drug resistant tuberculosis (DR-TB). Note: According to CDSCO guidelines, "performance evaluation" refers to "analytical validation" required for obtaining "test license", while “field evaluation” refers to "clinical validation” performed in clinical samples in real world setting.

II. Purpose To evaluate the clinical performance characteristics of nucleic acid amplification tests (NAAT) fordiagnosis of pulmonary drug resistant tuberculosis (DR-TB) using prospectively collected sputum samples in clinical settings. Primary Objectives

  1. To determine the diagnostic accuracy of new multi-drug resistant (MDR) NAAT test against culture based drug sensitivity testing (DST) in detecting first line drug resistance [Rifampicin (RIF), Isoniazid (INH)] among the microbiologically confirmed TB patients (positive by smear or NAAT test).
  2. To determine the diagnostic accuracy of new NAAT test against culture-based drug sensitivity testing (DST) in detecting fluroquinolone drug resistance (FQ) among the microbiologically confirmed TB patients (positive by smear or NAAT test).

III. Study Design Cross-sectional prospective multi-centric diagnostic accuracy study of IVD for detection of pulmonary drug resistant TB,using Mycobacterium Growth Indicator Tube culture and drug sensitivity testing (MGIT-DST) as the microbiological reference standard.

IV. Ethical Considerations

  1. The study should be compliant to the ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing, 2024. Performance evaluation activities using irreversibly de-identified leftover clinical samples are exempt from ethics approval as per ICMR’s Guidance on Ethical Requirements for Laboratory Validation Testing,

Page 448 of 459

  1. Investigators are required to submit a self-declaration form, as outlined in the ICMR guidelines, to the institutional authorities and ethics committee for information.
  2. Sputum specimens should be collected, as required for routine diagnostic evaluation, from patients who are suspected of having pulmonary TB as per algorithm. Probability of harm or discomfort anticipated in the research is nil or not expected.
  3. Enrolment of subjects should be continued till the sample size is met or till the project duration is completed.
  4. If additional sputum sample is obtained, written consent must be obtained as per the ICMR National Ethical Guidelines for Biomedical and Health Research Involving Human Participants. The institutional ethics committee of each participating site should be intimated about the study for necessary approval prior to initiating the study. Assent form should be collected in addition to informed consent in case of adolescents (13 to 16 yrs). For children between 7 and 12 years old, oral assent should be obtained in presence of parent or legal guardian. For children under 7 years old, written informed consent should be obtained from parent or legal guardian.
  5. The protection of privacy of research participants will be ensured by encrypting the patient identifiers.
  6. Patients shall receive the best possible diagnostic work-up as per the routine practice and the National Tuberculosis Elimination Program (NTEP) guidelines. There should not be delay in sending report due to the study.
  7. TB treatment decisions should not be made based on the result of the index test under evaluation, but on the basis of the routine clinical and laboratory methods (smear, solid / liquid culture, standard NAAT results, and clinical work-up).
  8. Respect for the dignity of research participants should be prioritized.
  9. No compensation shall be provided to the participants since there is no additional cost or travel involved in sample collection for the study. Patients should be compensated for travel and time only if they are asked to pay additional visits exclusively for the sake of the study and not during regular treatment visits.
  10. Follow-up visits may be required for a very limited number of discrepant patients to exclude TB.
  11. Leftover sputum samples and deposits should be stored for resolving discrepancies. One positive culture and two DNA samples per patient should be stored at -80oC for use later.
  12. All the sites should follow up with all study participants till the final diagnosis is made

Page 449 of 459

and the patient should be initiated on appropriate treatment as per NTEP norms. Those found to be M. tuberculosis complex (MTB)positive by standard NAAT test should be started on anti-tuberculosis treatment (ATT) by medical officer of the study site as per NTEP guidelines.
13. The findings of the study should be made accessible through reports.

V. Blinding of Laboratory Staff looking vials to be used for testing, and maintaining the database of results.

Staff performing the reference test and the test under evaluation (index test), interpretation of the test result, and entering the results against the coded samples in the database, should remain blinded to the status of samples till the completion of evaluation.

Operators conducting routine laboratory tests (GeneXpert MTB/RIF, MGIT DST, LPA etc.) will not participate in the index test evaluation. Instead, dedicated operators, who are not involved in routine testing and are blinded to the routine test results, will perform the index test. The results will be recorded independently for each test without any patient identifiers. The result sheets will be shared with the investigator for result analysis.The evaluation study data should be analyzed only by the PI of the evaluating lab (Fig. 1).

Page 450 of 459

Figure 1 Blinding in evaluation exercise

VI. Procedure

  1. Preparation of Evaluation sites/laboratories  Laboratory must be approved by the NTEP.
     Accreditation for at least one Quality management system [accreditation for Testing Lab / Calibration Lab (ISO/IES 17025), Medical Lab (ISO 15189), PT provider ISO/IEC 17043 or CDSCO approved Reference laboratory].  Three or more sites from different geographical regions should perform clinical validation for representation of population in real world setting.

  2. Study Participants People with microbiologically confirmed pulmonary TB by smear and/or NTEP approved NAAT test attending hospital OPDs/Chest clinics/district microscopy centers (DMCs) and Directly Observed Therapy Short Course (DOTS) centers. All such consecutive cases (not currently receiving ATT) and willing to provide consent should be enrolled in the study.

  3. Eligibility of Participants Inclusion criteria for testing First Line Drugs i. Individuals positive for TB by smear or any approved NAAT test (Xpert®

Page 451 of 459

MTB/RIF) and not receiving ATT
ii. Individuals willing to give consent iii. Individuals who are able and willing to give two good quality mucopurulent sputum samples of ≥ 3 ml

Exclusion criteria i. Individuals on TB treatment for >10 days ii. Individuals not consenting for the study
iii. Individuals unable to produce two sputum samples of ≥ 3 ml

  1. Reference and Index tests

Index test Reference Test Comparator First Line Drug Resistance New NAAT test for RIF/INH MGIT Culture DST for RIF and INH FL-LPA: GenoType MTBDRplus Second Line Drug Resistance New NAAT test for FQ MGIT Culture DST for Moxifloxacin (0.25, 1 mg) and Levofloxacin (1 mg) SL-LPA: GenoType MTBDRsl

  1. Sample size
    Sample size for RIF and INH resistance among TB patients

The expected sensitivity of the index test is about 90% with 5 % precision and the expected specificity is 95% with 5% precision. With a confidence interval of 95 % and assuming 10 % loss due to indeterminate results, the sample size required is estimated to be approximately 200patient’s positive each for INH and RIF resistance either alone or in combination. The average prevalence of Isoniazid and Rifampicin are ~18 % and 7.3 % respectively, among the new and previously treated TB patients combined together(Report of drug resistance survey, 2014-16). The number needed to screen to obtain 200 drug resistant cases will be approximately 1111 for INH resistance and 2857 for RIF resistance. The participants will be enrolled till the required sample size is achieved for INH and RIF resistance.

The expected sensitivity of the index test for detecting FQ resistance is 85 % with 7 % precision and the expected specificity is 95 % with 5 % precision. Assuming 10 % loss, the sample size required is 111 FQ resistant cases. The prevalence of FQ resistance among TB patients is ~3

Page 452 of 459

% (Report of drug resistance survey, 2014-16). Hence the number needed to screen will be approximately 3333. The participants will be enrolled till the required sample size is achieved for FQ resistance. Table 1 shows sample sizes required for RIF, INH and FQ drug resistance.

Table 1. Sample sizes for RIF, INH and FQ Drug Resistance

Assumptio ns for Sensitivity Assumptio ns for Specificity Sensitivity/Specificity of the new test (%) 90 95 Relative precision (d) (%)
5 5 Desired confidence level (1- alpha) % 95 95 Number of drug resistance (INH and RIF) cases required 178 84 Number of drug resistant cases required with 10 % loss due to indeterminate results
~200 ~93 Number needed to be screened assuming a combined weighted average prevalence of ~18 % for INH resistance among the new and previously treated TB patients 1111 517 Number needed to be screened assuming a combined weighted average prevalence of ~7 % for RIF resistance among the new and previously treated TB patients 2857 1329

Other disease controls (to check cross-reactivity in real patients) Include people with common alternative diagnoses to mirror programmatic reality and probe false positives. This subset helps characterize clinical exclusivity beyond simple “TB- negative” status: i. Non-Tuberculous Mycobacteria (Culture or PCR confirmed): ~30 ii. Other respiratory diseases [e.g., bacterial pneumonia, chronic obstructive pulmonary disease (COPD), lung cancer, chronic fungal (like Histoplasmosis or Aspergillosis)]: ~30 patients combined.

Page 453 of 459

  1. Implementation Plan The samples will be collected and tested as per the routine practice for smear, Xpert MTB/RIF®, LPA, MGIT culture and DST. The samples with positive result for MTB either in smear or NAAT test should be tested for first line and second line drug resistance (RIF, INH and FQ).

Figure 2. Flowchart for evaluating IVDs for testing drug resistance to RIF, INH and FQ among pulmonary TB (PTB) patients

  1. Sample collection, processing and storage

  2. Two sputum samples each of minimum 3 ml should be collected (one spot and one morning specimen) and sent to laboratory.

  3. Approximately 1 ml of sample should be taken from each sample and pooled under sterile conditions (total of 2 ml).

  4. Around 1 ml of pooled sample should be tested by the standard NAAT (Xpert MTB/RIF®) and remaining sample used for index test(s).

  5. The remaining portion of each sputum sample should be subjected to direct smear and

Page 454 of 459

decontamination by NaLC-NaOH method individually.
5. All smear positive or NAAT positive samples will be tested by Line Probe Assay (LPA). 6. The resultant deposit should be used for inoculation into two MGIT960 tubes.
7. All positive cultures should be identified using rapid Immuno-chromatography test (ICT). (Ideally, positive MGIT tubes are tested within 5 days of instrument positivity. Interpretation of the result should be done within 15 minutes). 8. The positive cultures should be tested for drug sensitivity. 9. All sputum samples should be stored at -20oC for later use. Decontaminated sediments and one positive culture per patient should be stored at -80oC, if necessary for later use. 10. Two DNA samples (one DNA sample extracted for index test and one for LPA) per patient should be stored at -20oC till the end of the study for resolution of discrepant results.
11. The index tests should be carried out as per the algorithm (figure 2) and as per the manufacturers’ instructions in the instructions for use (IFU).

All conventional test procedures for smear, culture (solid and liquid) and Xpert MTB will be performed as per NTEP national laboratory guidelines (CTD, 2016; RNTCP 2009) and laboratory manual of ICMR-NIRT (NIRT, 2010). Standard operating procedures for index test(s) will be provided by the manufacturer(s) including use of positive and negative controls. All procedures for preparation of media, reagents, washing, decontamination, disposal and storage will be performed according to the standard operating procedures (SOP) of ICMR- NIRT (NIRT, 2010) and WHO, (WHO, 2022).

  1. Laboratory Tests

v. Smear microscopy: Two direct sputum smear
vi. MGIT culture (decontaminated with 1-1.5% final NaOH); Two MGIT tubes (one per specimen) for each patient vii. MGIT drug sensitivity testing (DST) for Rif, INH: Drug sensitivity testing will be carried out from any one positive MGIT culture. viii. MGIT drug sensitivity testing for moxifloxicin (0.25 mg and 1 mg) and levofloxacin (1 mg). Drug sensitivity testing should be carried out in from any one positive MGIT culture. ix. Speciation of culture: Rapid immunochromatographic test (ICT) of MGIT culture
x. LPA: LPA shall be carried out as per routine practice and as per NTEP guidelines. Direct LPA should be carried out from any one smear positive sample. If the sample is smear

Page 455 of 459

negative and culture positive, indirect LPA should be carried out from culture. First line LPA (FL-LPA) will be carried out (Rif and INH resistance) xi. Xpert MTB/RIF (one test per patient)

  1. Index test i. Index test will be performed as per manufacturer’s instructions following blinded study protocols. ii. At least 2 different lots of reagents should be tested across the study population to demonstrate consistency of test performance and minimize lot-related bias. iii. The results of the index test will not be disclosed to study participants or clinicians and will not be used to guide treatment decisions.

10 . Data Analysis and resolution of discrepancy iv. If the index test produces error or indeterminate results, then only one repeat is allowed. The results of first test and repeat test should be recorded separately. All Invalids/Indeterminates/errors should be recorded and reported. v. Results for new patients and previously treated patients should be entered separately. Result analysis will be carried out for these two populations separately as well as combined. vi. A subgroup analysis may be carried out for pediatric population.

  1. Quality Control (QC) measures All sites should ensure high quality laboratory procedures, data recording and documentation. There should be no deviation from the protocol. All the sites should participate in internal quality control (IQC) and external quality assurance (EQA) for all methods as per the standard manuals of Global Laboratory Initiative (GLI, 2014). Culture: Positive (Reference strain H37Rv or H37Ra) and negative controls for MGIT and LJ cultures would be tested as per NTEP guidelines. MGIT Time to detection QC for MTB reference strain would be performed every month/new lot of reagents/machine service. Sterility and performance testing of culture media would be performed with every new batch or lot. Drug sensitivity testing (DST): Standard ATCC strains should be used for each drug as reference control. QC should be performed whenever a new batch of drugs is prepared, after servicing of the instrument and after long gap of setting up DST. Molecular diagnostics: For molecular diagnostics internal quality control includes control

Page 456 of 459

supplied by the manufacturer and control prepared by the lab from the previous testing. The internal control should be used whenever batch of test kit changes, machine is serviced, and newly trained person is introduced into the system.

VII. Statistical Analysis Plan i. The performance of the diagnostic kits should be evaluated by calculating the sensitivity, specificity, positive predictive value, negative predictive value and accuracy with reference to the gold standard. 95% Confidence interval should be calculated for each of the parameters. ii. The index molecular test will be evaluated for its performance with reference to MGIT DST (for RIF/INH/FQ).
iii. Similarly, the performance of NTEP approved molecular test (Xpert MTB/RIF and LPA) should be estimated with reference to MGIT DST. iv. The agreement between the index test and molecular test for drug resistance (LPA) should be calculated using kappa statistic.

VIII. Acceptance Criteria Expected minimal sensitivity for MTB and Drug Resistant TB: ≥85 ± 2% Expected minimal specificityfor MTB and Drug Resistant TB: ≥95 ± 2% Sample size: ~200 positives for each drug resistance (RIF or INH or FQ etc) (either alone or in combination) and ~ 100 negatives for each drug resistance (RIF or INH or FQ etc).

IMPORTANT NOTE Once a kit is determined to be “Not of Standard Quality”, following the procedure outlined in this document, no further requests for repeat testing of that kit will be accepted. Any request of re-validation from the same manufacturer for the same test type will only be entertained after a minimum of 3 months and only if a high-level technical summary of modifications or functional improvements to the kit design is submitted, without explicit disclosure of proprietary information.

Atleast two different lots or batches should be used for the field validation of any new molecular test.

Page 457 of 459

References

  1. Report of the first national anti-tuberculosis drug resistance survey India, 2014-2016.
  2. Technical and operational guidelines for tuberculosis control in India 2016. Central TB Division.
  3. RNTCP Standard Operating Procedures for Tuberculosis lab for culture and DST, 2009.
  4. Standard Operating Procedures (SOP) for Mycobacteriology laboratory, ICMR-NIRT,
  1. Practical manual on tuberculosis laboratory strengthening, 2022 update. Geneva: World Health Organization; 2022. Licence: CC BY-NC-SA 3.0 IGO.
  2. Mycobacteriology laboratory manual, Global laboratory initiative, First edition, April 2014, Stop TB Partnership.

Page 458 of 459

PERFORMANCE EVALUATION REPORT FORMAT Performance Evaluation Report For MDR-TB Kit

Name of the product (Brand/generic) Lot No /Batch No.: Product Reference No/Catalogue No Type of Assay Kit components Manufacturing Date Expiry Date Intended Use Import license / Manufacturing license/ Test license

License Number: Issue date: Valid Upto: Application No. Sample Panel Sample type

Positive samples (provide details: strong, moderate, weak)
Negative samples (provide detail: clinical/spiked, including cross reactivity panel)

Results:

Test Number of samples tested Positive Negative Invalids/Indeterminates/ Error/Contamination (culture) Smear

MGIT culture

Xpert MTB/RIF

Number of samples tested Sensitive Resistant

FL LPA – RIF

FL LPA - INH

SL LPA- FQ

MGIT-DST- RIF

MGIT-DST-INH

Page 459 of 459

MGIT-DST-FQ

New IVD- RIF

New IVD-INH

New IVD-FQ

Reference assay ……….…………… (MGITDST – RIF/INH/FQ)*

Positive Negative Total Name of MDR-TB kit Positive

Negative

Total

Estimate (%) 95% CI Sensitivity

Specificity

*Report RIF/INH/FQ as separate tables

Conclusions:

o Sensitivity, specificity

DISCLAIMERS

Note: This report is exclusively for ………………………Kit (Lot Nos.……), version …………with the gene targets …………………...manufactured by …………… (Supplied by ……….).

Signature of Director/ Director-In-charge …………………… Seal …………………………

******End of the Document