GUIDANCE DOCUMENT VALIDATION OF RAPID DIAGNOSTICS FOR PATHOGEN IDENTIFICATION AND ANTIMICROBIAL SUSCEPTIBILITY TESTING (AST) 2025 Division of Descriptive Research, Indian Council of Medical Research
GUIDANCE DOCUMENT
VALIDATION OF RAPID DIAGNOSTICS
FOR PATHOGEN IDENTIFICATION
AND ANTIMICROBIAL SUSCEPTIBILITY TESTING (AST)
2025
Division of Descriptive Research In-Vitro Diagnostics Division
Standard Validation Protocol /ICMR
ICMR, New Delhi
GUIDANCE DOCUMENT: VALIDATION OF RAPID DIAGNOSTICS FOR PATHOGEN IDENTIFICATION AND ANTIMICROBIAL SUSCEPTIBILITY TESTING (AST), 2025.
1st edition.
© 2025, Indian Council of Medical Research. All rights reserved
Publications of the ICMR can be obtained from ICMR-HQ,
Division of Descriptive Research,
Indian Council of Medical Research
Ansari Nagar, New Delhi, India
Requests for permission to reproduce or translate ICMR publications – whether for sale or for non-commercial distribution – should be addressed to ICMR-Hq at the above address.
The mention of specific companies or of certain manufacturers’ products does not imply that they are endorsed or recommended by the ICMR in preference to others of a similar nature that are not mentioned. Errors and omissions excepted. All reasonable precautions have been taken by the ICMR to verify the information contained in this publication. However, the published material is being distributed without warranty of any kind, either expressed or implied. The responsibility for the interpretation and use of the material lies with the reader. In no event shall ICMR be liable for damages arising from its use.
PREFACE
This document has been designed to provide assistance to the innovators and testing laboratories in
validating diagnostics meant for pathogen identification and antimicrobial susceptibility testing.
This validation protocol lays out a comprehensive framework to systematically evaluate and confirm
the diagnostic performance of a given test, ensuring its reliability and its utility for clinical decision-
making. This document describes evaluation criteria for methods to detect, identify and quantify
pathogenic micro-organisms or the genetic materials i.e. DNA, RNA, toxins, antigens, or any other
product of these organisms as well as methods for antimicrobial susceptibility and illustrates the
steps to validate these diagnostic tests. This guidance document encompasses the assessment of
precision, accuracy, reproducibility, and the ability to correctly identify the target pathogen and/or
antimicrobial susceptibility. It will also help the innovators and developers understand the kind of
evidence required to be generated before approaching the validating laboratory and prepare for the
validation process. This document delineates the steps in the pathway of regulatory approval for a
test. This is an evolving document that aims to align with emerging technologies and novel
methodologies.
This document has been prepared by the AMR co-ordination unit of ICMR and is the third version
of the draft protocol, having undergone two rounds of inputs from experts and Central Drugs
Standard Control Organization (CDSCO) through e-mails and stakeholder consultations held at
ICMR headquarters at New Delhi.
CONTENTS
Abbreviations………………………………………………………. 1-2
Introduction……………………………………………………… 3-8
Regulatory process followed by CDSCO…………………………
General Considerations for Validation of Diagnostics………
9-12 3. Assay development and validation criteria for new rapid diagnostics for pathogen identification and antimicrobial susceptibility testing (AST) ………………………………………
13-31
Phase 1: Assay Development Step 1: Define the purpose of validation for diagnostics………… 1.1. Pathogen identification 1.2. Antimicrobial susceptibility testing (AST)
14-15
Step 2: Optimisation of the reagents and protocol……………… 15
Step 3. Quality control (QC) and Quality assurance (QA)………… 15-16
Phase 2: Assay Validation Step 4: Analytical performance: Standard Operating Procedures (SOPs) for validation…… 4.1. Study design 4.2. Sample size 4.3. Reference or gold standard method 4.4. Specimen type 4.5. Test method 4.6. Shelf Life 4.7. Statistical evaluation of results
16-19
Step 5: Analytical Performance: Testing of performance and/or operational parameters…… 5.1. Pathogen identification 5.1.1 Accuracy 5.1.2. Analytical Sensitivity 5.1.3. Analytical Specificity 5.1.4. Reproducibility 5.1.5. Precision 5.1.6. Reportable range 5.1.7. Reference interval 5.2. Antimicrobial susceptibility testing 5.2.1. Accuracy 5.2.2. Error rates 5.2.3. Precision 5.2.4. Reproducibility 5.2.5. Address errors and bias 5.2.6. Address void testing bias 19-27
Step 6: Clinical Validation: Testing of clinical characteristics
6.1. Study design
6.1.1. Selection of target population
6.1.2. Exclusion and inclusion criteria
6.2. Study settings and number of sites
6.3. Test Purpose
6.4. Specimen collection and handling
6.5. Ethical considerations for Clinical performance studies
6.6. Clinical Performance Evaluation Criteria
6.6.1. Clinical Validity
6.6.2. Clinical Utility
6.6.3. Clinical Sensitivity
6.6.4. Clinical Specificity
27-30
Phase 3: Assay Maintenance
Step 7. Maintenance and extension validation criteria………. 30
Step 8: Dossier of validation (technical report)………………. 31
Step 9: Outcome and Decision…………………………………
31 4 Annexures 32-46
Annexure I: Instructions
Annexure II: Checklist
Annexure III: Definitions
Annexure IV: Process for submission of hard copy of application
Annexure V: Process for application through online SUGAM portal
Annexure VI: Validation report summary format
Annexure VII: CDSCO approved AMR predicate IVDs
5 References………………………………………………………. 47-50
Acknowledgement 51
ABBREVIATIONS
AD
Agar Dilution
AMR
Antimicrobial resistance
AST
Antimicrobial Susceptibility Testing
ATCC
American Type Culture Collection
BMD
Broth Microdilution
CA
Categorical Agreement
CDSCO
Central Drugs Standard Control Organization
CLIA
Clinical Laboratory Improvement Amendments
CLSI
Clinical and Laboratory Standards Institute
CPE
Clinical Performance Evaluation
CRS
Composite Reference Standard
DD
Disk Diffusion
DNA
Deoxyribonucleic Acid
EA
Essential Agreement
ER
Error Rates
EUCAST
European Committee on Antimicrobial Susceptibility Testing
FDA
Food and Drug Administration
ID
Infectious Diseases
ISO
International Organization for Standardization
IVD
In Vitro Diagnostics
IVDMD
In Vitro Diagnostic Medical Device
LLOQ
Lower limit of quantification
LOD
Limit of Detection
MDR
Multiple Drug Resistance
MEs
Major Errors
mEs
Minor Errors
2
MIC
Minimum Inhibitory Concentration
MTCC
Microbial Type Culture Collection & Gene Bank
NABL
National Accreditation Board for Testing and Calibration Laboratories
NPV
Negative Predictive Value
PHC
Primary Health Care
PPV
Positive Predictive Value
QA
Quality Assurance
QC
Quality Control
RNA
Ribonucleic Acid
SDD
Susceptible Dose Dependent
S/I/R
Susceptible/Intermediate resistance/Resistance
SOP
Standard Operating Procedure
S/R
Susceptible/Resistance
UTI
Urinary Tract Infection
VMEs
Very Major Error
3
Introduction
4
Chapter 1 Introduction
Antimicrobial resistance (AMR) poses a critical global health concern, particularly in countries
like India with dense population, a significant burden of infectious diseases and diverse healthcare
practices which may lead to antibiotic consumption. Accurate diagnosis of the infection using a
quality assured diagnostic test is essential for the timely initiation of treatment and for reducing
the misuse of antimicrobials. A number of innovators are working towards the development of
new indigenous diagnostics to address the healthcare needs of Indian population. These tests can
be helpful in early identification of the pathogen and rapid testing of antimicrobial
susceptibility phenotype or detection of antimicrobial resistance genes or markers. The impact
of these tests on clinical decisions relies on their accuracy, which is established by comparing
them to reference standard methods. However, systematic validation of these diagnostics remains
challenging, impacting their adoption in healthcare (Sharma et al., 2021). Despite the availability
of enormous literature on the validation of diagnostic tests, the steps that should be followed to
undertake validation needs to be clearly defined for the diagnostic tests meant for AMR
containment.
Validation of a diagnostic test is a systematic evaluation of the test which has been developed,
standardized and optimized , to determine its fitness for the specific intended use (OIE, 2019).
The World Health Organization (WHO) defines validation as ‘the action (or process) of proving
that a procedure, process, system equipment or method used works as expected and achieves the
intended result’ (WHO, 1995). The process of validation includes assessment of the analytical
and diagnostic performance characteristics of a test. The parameters that should be tested are
often considered to be based upon the type of test and are not uniform across all the tests
or assays.
One of the major bottlenecks in undertaking a validation study is the lack of defined uniform
protocols and absence of clarity on the parameters that need to be tested to establish the fitness-
of-purpose of a new indigenous test. Guidelines for study design, interpretation of the results,
execution in appropriate settings and performance evaluation remains ambiguous which
decelerate the processes of development as well as regulatory approvals for indigenous AMR
diagnostic tests. The developer derives information for test validation from various standard
5 documents that are available from Clinical and Laboratory Standards Institute (CLSI), European Committee on Antimicrobial Susceptibility Testing (EUCAST), Clinical Laboratory Improvement Amendments (CLIA), the International Organization for Standardization (ISO), Bureau of Indian Standards (BIS), United States’ Food and Drug Administration (FDA) and WHO.
Figure 1: The process and the steps required in the validation of an indigenously developed diagnostic test In India, various processes related to approvals on In Vitro Diagnostics (IVD) are regulated by CDSCO and a recently created online platform of MedTech Mitra (ICMR initiative in collaboration with the CDSCO and under the guidance of NITI Aayog) has been created by GoI to provide strategic support to MedTech innovators for clinical evaluation, regulatory facilitation
6
and the adoption of new products. Also, the Manthan digital platform, an initiative of the Office
of the Principal Scientific Adviser, aims to bridge the gap between industry demand and the
academic and startup ecosystem.
This document attempts to elaborate on the requisites, steps and process-flow for undertaking the
validation of indigenous diagnostic tests for AMR. It supplements the existing CDSCO guidelines
(CDSCO/IVD/GD/Stability/01/2022, CDSCO/IVD/GD/PER/01/2022) by bridging information
gaps and providing guidance on test validation for innovators and developers in the country. This
document establishes evaluation criteria for methods to detect, identify and quantify pathogenic
micro-organisms or the genetic material i.e. DNA, RNA, toxins, antigens, or any other product
of these organisms and testing method of antimicrobial susceptibility and illustrates the steps to
validate these diagnostic tests. Figure1 illustrates the components and steps of validation study.
This document provides predetermined acceptance criteria to ensure the safety and performance
of IVD devices, in line with Medical Device Rules, 2017 and CDSCO requirements. Developers
can evaluate their application using the guidance provided before submitting the application
dossier to CDSCO (see Annexure 1-VI).
Regulatory process followed by CDSCO
The validation/performance evaluation of in-vitro diagnostics is crucial for verifying the
performance and operational characteristics of IVDs during pre-qualification. Performance
evaluation is a key part of pre-qualification assessments, conducted by specified CDSCO
collaborating centres or designated laboratories. An IVD validation study consists of two
important aspects: Analytical performance studies and Clinical performance studies. The clinical
performance studies need to be conducted as per the standard ISO 20916:2019 in vitro diagnostic
medical devices, in compliance with good study practice and with the national or regional
requirements for ethics committee approval (ISO 20916, 2019). ISO 15189:2022 & ISO
20916:2019 emphasises on establishing the performance specifications as well as quality
assessment measures as part of the validation (Álvarez & Andreu, 2011; ISO 15189, 2022; ISO
20916, 2019).
For a new In-vitro Diagnostic Medical Device (where no predicate device is available in the
country), the applicant must first obtain a Test License in form MD-13 (for domestic
manufacture) to produce test batches (Figure 2). These batches are necessary for generating in-
house quality/validation data or for evaluation in an external laboratory, as applicable.
Subsequently, to generate clinical performance data for the device, the applicant must secure
permission to conduct CPE by obtaining MD-25 (for protocol approval by the IVD experts
7 committee constituted by CDSCO). Upon receiving MD-25, the applicant can proceed with the CPE. The study results and generated data, must be submitted for approval to manufacture the new IVD, using form MD-29 (for clinical data approval by the IVD experts committee constituted by CDSCO). Following this, the applicant must obtain the appropriate manufacturing license, either MD-5/MD-6 (approved by the State Licensing Authority) or MD-9/MD-10 (approved by the Central Licensing Authority, DCGI) for sale or distribution. The details for submitting applications are outlined in Annexures IV and V.
Fig 2: Regulatory licenses required for evaluation and approvals For an In-vitro Diagnostic Medical Device with an existing predicate device in the country, the applicant must first obtain a Test License in form MD-13 (for domestic manufacture) to produce test batches. These batches are necessary for generating in-house quality/validation data or for evaluation in an external laboratory, as applicable. Following this, the applicant must secure the appropriate manufacturing license for sale or distribution. This involves obtaining MD-5/MD-6 (approved by the State Licensing Authority) or MD-9/MD-10 (approved by the Central Licensing Authority, DCGI). The classification of in vitro diagnostic (IVD) medical devices is determined by the level of risk associated with their use, ranging from Class A (lowest risk) to Class D (highest risk). For any new IVD devices classified as Class B, C, or D, a CPE is mandatory. Table 1 provides a curated
8 list of selected examples of CDSCO-approved IVDs that are crucial for diagnosing pathogens and/or antimicrobial resistance, with further details available in Annexure VII (source: https://cdscomdonline.gov.in/NewMedDev/ListOfIvdMdApprovedDevices).
Table 1: Examples of CDSCO approved antimicrobial resistant IVDs (August 2024)*
Class Risk Level Example A Low Risk WASPLab® System, Microplate ELISA Reader B Low Moderate Risk ETEST Azithromycin, ETEST Cefixime, Antimicrobial Susceptibility System(ASS)- HiMic Plate Kit, Mueller Hinton 2 agar + 5% sheep blood C Moderate High Risk AMR Direct Flow Chip Kit (Manual and Auto), NG-Test CARBA 5, UTI Antimicrobial Susceptibility Testing PCR Kit D High Risk
*Table 1 and Annexure VII provide examples of CDSCO approved IVDs purely for reference purposes only. None of the examples indicate any endorsement through this document.
9
General considerations for validation of diagnostics
10
Chapter 2 General considerations for validation of diagnostics
I. Sites: In an ideal situation, the validation studies must preferably be carried out at multiple sites. However, in cases where this is not feasible, validation should be carried out in a minimum of two different sites representing the country’s geographic variations to test the validity and usefulness of the diagnostic. The performance evaluation sites should be chosen from the CDSCO approved list and must comply with ISO15189 standards. (CDSCO, 2023).
II. Sample: The type of specimen to be used for testing must be mentioned, such as invasive (CSF, blood) or non -invasive (sputum, urine) etc. Isolates from sterile sites like blood or CSF are more clinically significant than those from non-sterile sites like urine or sputum, which may indicate colonization. The sample size must be determined with the help of statistical expert to ensure that sufficient number of specimens are tested to provide statistically reliable results taking into consideration the prevalence of the condition/infection. These are required to justify any claim and to provide reasonable estimates of uncertainty. Sample collection and handling must be performed by trained personnel. Also the sample storage must be done at optimum conditions to preserve integrity of the samples.
III. Test methods: Specify the type of diagnostic, whether it is a method, assay, kit or a device and detail its necessary components. Indicate the method involved e.g. biochemical assay, molecular assay or mass-spectrometry based assay. While most test methods yield numeric, quantitative results, some assays only provide qualitative outcomes.
IV. Standards: Reference material (standard strains, toxins, analytes, antimicrobials etc) for validating the test must be of certified quality; if not available, then the molecule with highest possible purity must be used to ensure the test results are reliable (GUIDELINE 2.1., 2012). Strains used to measure test performance should be obtained from reference centre collections (e.g., ATCC, etc), academic government reference laboratories, or other collections that are available to the scientific community. The characteristics of the reference material conferring
11
it a particular trait must be ascertained while procuring the material. For validation of the test
performance characteristics, the target strain should be associated historically with the
specimen, or an outbreak. These should be the first strains of choice for conducting the
validation study (Caliendo et al., 2013).
Standards for interpretation of results and susceptibilities:
• Breakpoints: These are the values used by clinical microbiology laboratories to interpret the
results of antimicrobial susceptibility testing (AST) and classify isolates as susceptible or
resistant. Clinical and Laboratory Standards Institute (CLSI) and the European Committee
on Antimicrobial Susceptibility Testing (EUCAST) set clinical breakpoints. Indian
laboratories majorly use CLSI breakpoints. For a given antibiotic, the breakpoint may differ
for different sites (e.g., urine versus tissues or skin) and for different infecting organisms.
• Minimum Inhibitory Concentration (MIC): It is defined as the lowest concentration of an
antibiotic which prevents visible growth of a bacterium. MICs are determined and then
interpreted according to the breakpoint – if the MIC is less than the susceptible breakpoint,
the organism is considered as susceptible and can be successfully treated with that particular
antibiotic, whilst if the MIC is higher than the susceptible breakpoint, it is considered as
non-susceptible (intermediate resistant or resistant).
V.
Reference method: The reference method is defined as a method/test by which the
performance of an alternate method is measured or evaluated. The reference method is
usually the gold standard for the organism under study. Validation studies must include
comparison to a recognized reference method to demonstrate equivalence of performance,
the significance of which must be determined statistically. The performance of the test under
question must be comparable to the gold standard. In case there is no gold standard for an
organism then the next best test available or composite reference standard (CRS) is used. A
composite reference standard is a fixed rule used to make a final diagnosis based on the results
of two or more tests, referred to as component tests. For each possible pattern of component
test results (test profiles), a decision is made about whether it reflects presence or absence of
the target disease. Though it is simple and easy to interpret, it can lead to seriously biased
estimates of accuracy indices (sensitivity and specificity) and should be avoided whenever
possible.
VI. Media: Bacteriological media that ISO standards should be utilised for antimicrobial susceptibility testing. All the media must be tested for physical and chemical parameters
12 and growth promotion parameters. The composition of the culture media used and the name and catalogue number of the manufacturer needs to be mentioned. This is important as the accuracy of the test results is dependent on the quality of media used for susceptibility testing.
VII. Supplies: The test kits, reagents, chemicals, equipment etc required to conduct validation of the test will be provided by the developer of the test. The test kits and reagents must be supplied at temperature that is suitable to preserve the test efficacy. Also, the storage conditions must be clearly stated by the developer and must be followed as indicated in the instructions for use / labels. Some products may not need refrigeration. If refrigerated storage space is inadequate to store the entire test kit, they may be divided such that labile reagents can be refrigerated separately from the non-labile supplies. Calibrated thermometers or other environmental monitoring devices must be placed at each location where reagents and specimens are stored, i.e. ambient, refrigerator and freezer; and temperatures must be recorded daily.
13
Criteria for development and validation of assays for new rapid diagnostics in pathogen identification and antimicrobial susceptibility testing (AST)
14
Chapter 3 Criteria for development and validation of assays for new rapid diagnostics in pathogen identification and antimicrobial susceptibility testing (AST)
Phase 1: Assay Development
Step 1: Define the purpose of validation for diagnostics
The intended goals and applications of the diagnostic, such as the identification of pathogens,
testing of antimicrobial susceptibility, or detection of antimicrobial resistance markers, should be
outlined. The type of diagnostic, whether it is a method, assay, kit, or device, must be specified,
along with a detailed description of its necessary components. It should be indicated whether the
test detects the whole organism, toxin, analyte, DNA/RNA, etc. The following considerations
should be taken into account when planning the validation:
1.1. Pathogen identification
• The disease or condition to be diagnosed.
• Whether the test can provide a qualitative or quantitative result for optimal clinical utility.
• Whether a screening or confirmatory test is required. Screening tests determine the status of a
disease, disorder or other physiological state in an asymptomatic individual whereas
confirmatory test establish the presence (or absence) of infection for treatment decisions in
symptomatic or screen positive individuals.
• The test purpose will directly influence the subject sample size (N) and selection criteria
(including inclusion and exclusion) when planning and designing the study. For example, if the
prevalence is low and test is meant to screen asymptomatic individuals, specimens from a large
number of subjects may be required to provide sufficient evidence of optimal performance.
However, if the test is to be used for diagnosis in symptomatic individuals, specimens from a
smaller number of subjects may be adequate.
15 1.2. Antimicrobial susceptibility testing (AST) • Whether genotypic or phenotypic based AST. • Whether test can differentiate between susceptible(S)/resistance (R) or susceptible/intermediate resistance/resistance (S/I/R) for a particular antibiotic. • Whether a single test or a diagnostic algorithm is required. • Whether test can provide a qualitative or quantitative result for optimal clinical utility. • Whether a screening or confirmatory test is required.
Step 2: Optimization of the reagents and protocol The reagents and protocol must be optimized to ensure that consistency of the assay is maintained, considering factors such as temporal, chemical and physical variables. These include test operating conditions, storage conditions during transportation, temperature, pH, dilutions, buffers and other relevant factors during laboratory usage (Jennings et al., 2009). A kit should be self- sufficient to allow testing and analysis e.g., brochure/software to facilitate training of the laboratory personnel, instructions for storage, processing of samples and analysis and reporting of the data (Crowther et al., 2006). Shelf life (expiry date) and the lot number of the reagents and kit should be clearly labelled. A certificate of analysis for reagents in the kit can be provided. ‘Single test’ device or consumables not to be reused should be preferred. Optimal quality of materials and media should be included for the study i.e. as certificate of analysis. The sample profile should be clearly detailed for precise validation (CLSI/NCCLS, 2003). Step 3. Quality control (QC) and Quality assurance (QA) Before validation is initiated, the test and reference method should have passed defined quality control criteria. QC is necessary to ensure the performance of the test method under evaluation and the reproducibility of the results. All QC parameters should be in range before validation. If any result is obscured, then additional replicates should be used until 95% of results are in the expected range. QC measures and strains to be used for the test method should be defined by the manufacturer and at least three replicates should be recommended. For the reference method, QC protocols and strains should follow CLSI guidelines (CLSI, 2018c) and up to nine replicates should be tested, including at least one QC organism (Humphries et al., 2018). Microbial reference stains should be obtained from standard sources (e.g. ATCC, MTCC (IMTECH), MCC (NCCS)) and should be well characterized, including stable defined antimicrobial susceptibility phenotypes. Isolates should be stored under optimal conditions to
16 ensure that their phenotypic and genotypic characteristics are retained, for e.g., resistance mechanisms especially plasmid-borne resistance can be lost under suboptimal conditions. The common storage conditions recommended for the isolates are -70°C to -80°C, in 20% glycerol or any another suitable medium such as dimethyl sulfoxide (Humphries et al., 2018). Quality assurance implements monitoring and evaluating the records, calibration and maintenance of equipment, training and QC and helps to ensure that materials and processes consistently give quality results (OIE, 2019). If a diagnostic is a machine or uses auxiliary equipment, it must be serviced, calibrated, maintained and monitored appropriately to ensure the reproducibility of the test conditions in assay performance.
Phase 2: Assay Validation
Step 4: Analytical validation: Standard Operating Procedures (SOPs) for validation
Analytical performance describes a device's capacity to precisely detect or measure a specific
analyte. These studies are generally designed following CLSI guidelines, with the study approach
and plan recorded in an analytical performance study. Standard Operating Procedures (SOPs)
should be prepared as ready-to-use working drafts and updated as needed throughout the
validation process. Upon completion of the validation study, these SOPs will serve as controlled
documents.
4.1. Study design and site selection
The design of a study for diagnostic validation requires meticulous planning, sample size
calculation, an appropriate study design, specimen type, considerations for multiple matrix
specimen and establishment of a clear reference standard. Additional considerations includes
standardization of data collection, planning statistical analyses, obtaining ethical approval and
adherence to reporting guidelines.
4.2. Sample size
A defined number of adequate samples are required to ensure that precise inferences can be made
from the statistical analysis and level of confidence can be achieved for accurate results (Hajian-
Tilaki, 2014). Sample size and data points for different parameters are mentioned in Table 2.
4.2.1. Pathogen identification- A representative number of samples including positives and
negatives should be tested in parallel to check the performance parameters of diagnostic test
(Rabenau et al., 2007).
17
4.2.2. Antimicrobial susceptibility testing- A representative number of samples positive for
clinically relevant pathogens indicative of susceptibility, intermediate resistance and resistance
(S/I/R) for antimicrobials must be tested in parallel in the study. A minimum of 100 isolates
recovered from clinical samples are recommended for each group of micro-organisms (CLSI,
2018c) to determine the relevance of antimicrobials for infectious disease management.
Susceptibility testing will be performed for a panel of antibiotics based on the drug-bug profile
combination, including the recently approved novel antibiotics using the reference broth
microdilution method. However, a greater number of samples will always help in improving the
confidence in test results and minimize the error or bias in process.
4.3. Reference or gold standard method
The reference or gold standard should be used as per the type of diagnostic test for the validation.
Standard and updated guidelines should be followed for the testing protocol. In case there is no
gold standard for an organism then the next best test available or composite reference standard
(CRS) is used. Another approach to analytical validation entails comparing the performance and
specifications of the IVD with an existing predicate device. Demonstrating that the new
diagnostic is equivalent or superior to the predicate in all aspects supports a claim of substantial
equivalence.
4.3.1. Pathogen identification- Standard reference materials/isolates recommended by CLSI in
their recent guidelines should be used. These standards perfectly discriminate between
participants with or without the disease and provide unbiased results for the diagnostic accuracy
measure of index test. Ideally, a gold standard with 100% sensitivity and specificity should be
used as reference. The storage conditions of the reference materials/ isolates should be defined in
accordance with standards such as those defined by CLSI, ATCC etc., to preserve the genotypic
and phenotypic stability.
4.3.2. Antimicrobial susceptibility testing- The standard methods and guidelines on S/I/R as
recommended by CLSI should be used for diagnostic evaluation. For AST, ideally a gold standard
with 100% category agreement and essential agreement should be used as reference method. The
reference method for AST is MIC determination by the broth microdilution (BMD) method
(CLSI, 2018a). Whereas other reference methods such as agar dilution (AD) and disk diffusion
(DD) are acceptable substitutes in case of non-availability of BMD (CLSI, 2018b). In such cases,
exceptions or modifications as suggested by CLSI must be considered as per requirement
while evaluating the diagnostic test. For example, inducible clindamycin resistance is determined
by the D-zone test, a reference DD method. Another exception is for fosfomycin resistance or
18
testing for Neisseria gonorrhoeae, because of complications with BMD, the AD is used as a
reference method (CLSI, 2018c). Other exception include the case of methicillin resistant
Staphylococci (detection of mecA or mecC) and vancomycin resistant Enterococci and
Staphylococci, where molecular detection of a resistance gene is used as the gold standard
method. However, the resistance in other species is multifactorial and not defined by single
molecular target, especially in gram-negative bacteria. Problems such as bias, errors and
inadequate results can occur in absence of reference/gold standard or in using imperfect reference
standards. Therefore, multiple imperfect reference standards should be considered, but only with
expert informed opinion.
4.4. Specimen type
The capability of the diagnostic to analyse one sample-type or multiple sample-types (e.g. urine,
blood, cerebrospinal fluid, etc.) should be clearly outlined. If diagnostic is capable of utilizing
more than one sample type (e.g. urine, serum, spinal fluid, etc.) for microbial detection, all the
analytical and performance studies should be independently repeated. A linearity study should be
carried out for each specimen type to omit the effect of matrix background on results.
4.5. Shelf life (Stability) of the diagnostic
The shelf-life of a diagnostic, whether an assay or device, must be assessed to ensure its viability
in maintaining acceptable performance characteristics over a defined time interval under specified
storage conditions. The shelf-life cannot be measured directly and it is assessed from the
accuracy and performance characteristics. Stability of diagnostic includes reagents, media,
controls, reconstituted lyophilized materials, working solutions, matrix and calibrators etc. when
used, stored and transported under specific conditions (CDSCO/IVD/GD/Stability/01/2022,
2022). EN 13640 and CLSI EP25- A describe the need for different types of product stability
testing(CLSI EP25, 2024).
The shelf life of a diagnostic can be assessed by performing real-time or accelerated stabilities
studies, component stability studies, reconstituted stability testing, transport simulated stability
assessment, open or in-use stability evaluation. The selection of lot/batches and the number of
samples used affect the stability assessment. A minimum of three different batches to verify shelf
life in real-time are recommended (Marimuthu et al., 2019; Medical Devices Rules, 2017, n.d.).
4.6. Statistical evaluation of results
A thorough statistical analysis of the assay's test results should be conducted to ensure its accuracy
and all findings and inferences should be documented with precision. Additionally, the method's
19 characteristics should be compared with those of a reference standard or a previously validated method to assess its reliability. Establish cut-off values or assign numerical parameters to further evaluate the method's effectiveness. Cut-off values should be established, or numerical parameters should be assigned, for further evaluation of the method's effectiveness.
Step 5: Analytical validation: Testing the performance and/or operational parameters
The performance and/or operational parameters that need to be validated for a particular
test/method should be identified and the formulas and worksheets and should be defined. Each
analytical parameter should be considered on a case by case basis and if any are not applicable a
justification should be provided (e.g. linearity is not applicable to qualitative devices). The
number of parameters that should be tested can vary slightly based on the type of diagnostic i.e.
whether test is biochemical, molecular or microbiological etc., or its format i.e. whether it is a
method, device and/or a test kit.
A diagnostic test can be qualitative (reported as positive or negative or indeterminate) or
quantitative. Assessment of minimum three determinants of validity i.e. accuracy, reproducibility
and precision have been recommended for both types of diagnostic test outputs (Cleophas &
Zwinderman, 2009).
While CLIA lists the performance specifications that must be established, it does not specify the
scientific methodology or data analysis tools to be used (Burd, 2010). Guidelines to assist in
establishing these performance specifications have been published by the CLSI and ISO in several
documents (CLSI, 2018b).
5.1. Pathogen identification
For diagnostic aimed at rapid identification of pathogen, performance parameters should be
assessed as mentioned in Table 2.
5.1.1. Accuracy: The test should accurately identify individuals with the disease and provide
insight into its severity.
5.1.2. Analytical Sensitivity: The test should be able to distinguish between two close
concentrations of the analyte.
5.1.3. Analytical Specificity: The test’s ability to accurately quantify the analyte in the presence
of potentially interfering substances must be thoroughly evaluated. This evaluation should take
20 into account potential interference from both endogenous and exogenous sources, with appropriate testing conducted to ensure reliable performance. 5.1.4. Reproducibility: Consistency is ensured when the second test yields the same result as the first when a subject is tested twice. 5.1.5. Precision: A test method is said to be precise when repeated determinations (analyses) on the same sample give similar results. When a test method is precise, the amount of random variation is small. The test method can be trusted because results are reliably reproduced time after time. 5.1.6. Reportable range: The range of analyte concentrations for which the test system can accurately report results. 5.1.7. Reference range: The range of IVD output values that correspond to normal or healthy populations.
Table 2. Evaluation of performance parameters for diagnostic for rapid pathogen identification S. No Performance characteristics/parameters Data point and minimum sample testing requirements Determina nt of validity Statistical test for Qualitative diagnostic Quantitative diagnostic 1 Accuracy • Sensitivity • Specificity • Overall Accuracy Receiver operated curves (ROC) • Barnett’s test; • Intraclass correlation vs gold test, • Bland-Altman test
Analytical
Sensitivity
(limit-of-
detection study):
• 60 data points (e.g. 8-12 replicates
from 4-5 samples in the range of the
expected detection limit);
• conduct the study over 5 days;
• probit regression analysis (or standard
deviation with confidence limits if
limit of blank studies are used)
Analytical Specificity (interference
studies): • No minimum no. of samples recommended; • test sample-related interfering substances (hemolysis etc.) and genetically similar organisms or organisms found in same sample sites with same clinical presentation; • spike with low concentration of analyte; • paired-difference (t test) statistics
21
Accuracy (comparison-of-methods
study): • Test in duplicate by both the comparative and test procedures over at least 5 operating days; • typically 40 or more specimens; • xy scatter plot with regression statistics; • Bland-Altman difference plot with determination of bias; % agreement with kappa statistics and/or Lin’s concordance correlation coefficient
Calculate the Diagnostic sensitivity
(TP/TP+FN) and diagnostic specificity (TN/ TN+FP).
Calculate the Positive Predictive Value
(PPV), Negative predictive value ( NPV), Likelihood Ratio (LR+ and LR-) of a positive and negative test. 2 Reproducibilit y Cohen’s kappas; � Duplicate standard errors, � Repeatability coefficients, � Intraclass correlations vs duplicate test • At least 10 and preferably 20 runs of the assay to give estimates of these parameters 3 Precision (replication study) Confidence intervals Confidence intervals
For qualitative test:
� minimum of 3 concentrations (LOD, 20% above LOD, 20% below LOD) and obtain 40 data points; � test in duplicate over 15 days (include data from analytical sensitivity runs to provide data over 20 days)
For quantitative test:
� minimum of 3 concentrations (high, low, LOD) and test in duplicate 1-2 times/day over 20 days (include data from reportable range study as day 1 to provide data over 20 days); � calculate SD and/or CV within run, between run, day to day, total variation 4 Reportable Range or Analytical Measurement Range (linearity study) NA (Qualitative tests do not require linearity, Analytical Measurement Range a linearity experiment to determine reportable range and lower limit of quantification (LLOQ)
For quantitative assays:
• 7-11 concentrations across anticipated measuring range (or 20- 30% beyond to ascertain widest possible range); (CLSI/NCCLS, 2003)
22 Source: (Burd, 2010; CLSI, 2008; CLSI and IFCC, 2008; CLSI/NCCLS, 2003; Humphries et al., 2018, 2023; Patel et al., 2013; Rabenau et al., 2007; van Belkum et al., 2019)
5.2. Antimicrobial susceptibility testing For detection of the S/I/R criteria, assessment of accuracy, error rates, precision and reproducibility of diagnostics for AST has been recommended for qualitative and quantitative types of diagnostics (Humphries et al., 2018; Patel et al., 2013). Data points and interpretive criteria for DD and MICs methods as described by CLSI guidelines M02 and M23 documents (CLSI, 2008, 2018b) should be used for testing so as to minimize the number of category errors during testing. 5.2.1. Accuracy: It shows the closeness of the result under evaluation to the true value (i.e., agreement with the reference standard or predicate test) that can be measured by two ways, categorical agreement (CA) and essential agreement (EA). • Categorical agreement: Percentage of isolates tested producing the same category result i.e. susceptible, intermediate, susceptible dose dependent, resistant, or non-susceptible as compared to the reference standard method. Susceptible-dose dependent (SDD) category implies that susceptibility of an isolate is dependent on the dosing regimen that is used in the patient. The dosing regimens (i.e., higher doses, more frequent doses, or both) used to set the reference range studies) • 2-4 replicates at each concentration on same day; • polynomial regression analysis 5 Reference Interval (reference value study) NA (reference range is typically negative or not- detected and reference interval studies do not need to be performed if target is always absent in a healthy individual)
For quantitative assays:
• reference interval will be reported as below the LOD or LLOQ; • for some analytes, the reference interval may be a clinical decision limit; • if the intended use of the test is limited to patients known to be positive for the analyte being assayed, a reference interval may not be applicable • it can be verified by testing 20 known normal samples; if no more than 2 results fall outside the manufacturer/published range then that reference range can be considered verified.
23
SDD interpretive criterions are provided in Appendix E in M100 (CLSI, 2018b). Non-
susceptible implies that only a susceptible interpretive criterion has been designated in these
isolates because of the absence or rare occurrence of resistant strains. An isolate that is
interpreted as non-susceptible does not necessarily mean that the isolate has a resistance
mechanism.
• Essential agreement: Percentage of isolates tested producing MICs that are within 1 log2
dilution (±1 doubling dilution) of the reference BMD MIC value. EA is applicable only to those
AST methods that determine MIC values.
5.2.2. Error rates (ER): ER shows the CA discrepancies of the test from the reference method. These are divided into following three types of errors: • Minor errors (mEs): the test shows minor discrepancies/errors between susceptible vs. intermediate and intermediate vs. resistant when compared to reference method. It can have the least detrimental influence on therapeutic decision. • Major errors (MEs): the test displays ‘resistant’ results while the result is ‘susceptible’ by the reference method. These errors limits the therapeutic options and tends to overuse last resort of antibiotics. • Very major error (VMEs): the test expresses ‘susceptible’ result while the result is ‘resistant’ by the reference method. This is a critical error that leads to use of an ineffective therapeutic agent for treatment against an infection and are associated with high mortality rate.
5.2.3. Precision: It is the closeness of agreement of the test to give same value when the same isolate is tested repeatedly under specified conditions. It can be determined within a run (repeatability); across several runs in one day; or across multiple runs across multiple days (reproducibility). 5.2.4. Reproducibility: It can be determined by testing the QC organisms using different personnel/operators, different test batches, different lots, different laboratories and after different time intervals. For diagnostic aimed at AST of pathogen, performance parameters should be assessed as mentioned in Table 3.
24 Table 3. Evaluation of performance parameters for diagnostic for rapid AST S. No. Performance parameters Calculations Calculation terms Acceptance criteria* 1 Accuracy Categorical Agreement (CA)
NCA/NT *100
NCA= number of isolates with an AST result with the same categorical interpretation as reference method ; NT= number of isolates tested
≥90% CA
Essential Agreement (EA)
Sensitivity Specificity NEA/NT * 100
TP/TP+FN TN/TN+FP NEA= number of isolates with the same or within one doubling dilution MIC value as the reference method; NT= number of isolates tested;
TP=True positive TN=True negative FP=False positive FN=False negative ≥90% EA (comparison-of- methods study) Test in duplicate at least 5 operating days; n=≥40; xy scatter plot with regression statistics; Bland-Altman difference plot with determination of bias; % agreement with kappa statistics 2 Error rates (ER) Minor Errors (mEs)
Major Errors (MEs)
Very Major Errors (VMEs)
NmE/NT * 100
NME/NRefS *100
NVME/NRef R * 100
NmE= number of isolates having minor errors NT= number of isolates tested;
NME= number of isolates that yielded false- resistant results; NRefS= number of isolates susceptible by the reference method
NVME= number of isolates that tested false-susceptible results NRefR= number of isolates resistant by the reference method
≤10% mE
<3% ME
<3% VME (FDA uses <1.5% VME)
25
3
Precision
% CV=
SD/mean
*100
Calculate in terms of
Imprecision (random
error): standard
deviation (SD) &
coefficient of variation
(CV)
Sample testing in
duplicate over 20 days
4
Reproducibilit
y
NA
More number of
variables can
strengthen the results
≥95%
5
Reportable
Range
NA (for
quantitative
analysis)
A linearity experiment
to determine reportable
range and lower limit of
quantification (LLOQ)
At least 10 and
preferably 20 runs of
the assay;
7-11 concentrations
across anticipated
measuring range (or
20- 30% beyond to
ascertain widest
possible range);
2-4 replicates at each
conc. on same day;
polynomial regression
analysis
6
Reference
range
NA
n=20 representative of the population [if the population is different, n=60 (minimum, 40)]; Out of the 20 samples, if no more than 2 results fall outside the published range then the reference range can be considered to be verified *Wherever applicable, percent (%) values are as recommended by ISO 20776-2:2007 and FDA. Source: (CLSI, 2008; CLSI and IFCC, 2008; CLSI/NCCLS, 2003; Humphries et al., 2018, 2023; Patel et al., 2013; Rabenau et al., 2007; van Belkum et al., 2019) 5.2.5. Address errors and bias Errors and variables can be assessed using various methods. Due to the inherent variation in MIC end points, the error rate will be directly proportional to the percentage of isolates with antimicrobial agent MICs in the range of one two-fold concentration above the intermediate MIC (I + 1) and two-fold concentration below the MIC (I-1)(Humphries et al., 2018). Thus, when an entire population is used as the denominator for calculating error rates, the rate will be
26
determined largely by the population of MICs in the I + 1 to I − 1 range. For example, when
90% of the isolates have highly susceptible drug MICs (as it is common with newer
antimicrobial agents), the error rate will be considerably less than that of a population in which
40% of the MICs fall in the I + 1 to I − 1 range. Using the total I + 1 to I − 1 subpopulation as
the denominator for calculating discrepancies provides a more accurate assessment of the
discrepancy by accounting for normal technical variability in the testing method. For
antimicrobial agents for which clinical use will primarily be for organisms with specific types
of resistance mechanisms, scatterplots and error rates are evaluated and presented separately for
these types of organisms.
Table 3 : Acceptance performance rates for ASTs by error-rate bound method for antimicrobials
with an intermediate category
Reference MIC range for isolates to include in
denominator of error calculations
Acceptable Error Rates
1-dilution
intermediate range
2-dilution
intermediate range
mE
ME
VME
≥ I + 2
≥ Ihigh + 2
< 2%
ND
< 5%
I+1 to I-1
Ihigh +1 to Ilow -1
<
40%
<
10%
< 10%
≤ I-2
≤ Ilow -2
ND
< 2%
< 5%
MIC=minimal inhibitory concentration; mE=minor error; ME=major error; VME=very major error;
I=Intermediate MIC value; Ihigh=high end of the MIC range for intermediate category; Ilow=low end of
the MIC range for intermediate category; ND=not determined
Error-rate-bound method (Brunden et al., 1992; FDA, 2009): It can be used for following
conditions:
• To evaluate MIC distribution in the isolates that differs significantly from the normal
distribution of MICs.
• To calculates the breakpoints, if >20% of the isolates tested are within 1log2 dilution.
• To analyse linear regression when the population of bacteria tested is enriched with a non-
wild-type population
• Traditionally used to evaluate disk breakpoints when the bacterial population is not binomial
Model based approach (DePalma et al., 2017): Determine Disk diffusion breakpoints
interpretive criteria, whereby a fitted model is used to take into account the proportion of
isolates at each
27 Table 4: Acceptance performance rates for ASTs by error-rate bound method for antimicrobials when no intermediate category exists
MIC=minimal inhibitory concentration; mE=minor error; ME=major error; VME=very major error; R=resistant MIC value; S=Susceptible MIC value; ND=not determined 5.2.6. Address void testing bias: Define possible bias that can be introduced in testing e.g., constant bias, proportional bias etc. Avoid discrepant analysis techniques (McAdam, 2000) as it has tendency to overestimate the sensitivity and specificity and PPV of a test. It has been emphasized that these should be avoided especially for diagnostic tests (like therapeutics). Re- testing of both concordant and discrepant samples should be done to avoid test bias.
Step 6: Clinical validation: Testing of clinical characteristics
Clinical performance evaluation (CPE) is the systematic evaluation of a new in vitro diagnostic
device using the specimens collected from human participants to assess its performance. It
evaluates the accuracy and reliability of a diagnostic test in a clinical setting. These studies often
involve testing the diagnostic test on patients with known disease or condition and comparing
the results to a gold standard or reference standard (Baumfeld Andre et al., 2022).
In India, clinical performance evaluation is required for all the class B, C and D (classification
as per Chapter II, Rule 4, Sub-rule (2) of MDR 2017) (Medical Devices Rules, 2017, n.d.) in
vitro diagnostic devices that are new. The IVD may be exempted from clinical performance
evaluation if it is being marketed for at most two years in one of the countries like United States
of America, Australia, Canada and Japan. The manufacturer or importer is required to take
permission to conduct clinical performance evaluation from the Central licensing authority
(CLA). Before beginning the enrolment process, the clinical performance evaluation must be
recorded in the Clinical Trial Registry of India. The annual status report for every clinical
Reference MIC range for isolates to
include in denominator of error
calculations
Acceptable Error Rates
mE
ME
VME
≥ R + 1
< 2%
ND
5%
R + S
< 40%
< 10%
< 10%
≤ S - 1
ND
< 2%
< 5%
28 performance evaluation concerning whether evaluation is ongoing, complete, or ended should be sent to CDSCO.
6.1. Study design
CPE studies should be designed to maximize the value of the data while minimizing bias.
Evaluating both diagnostic accuracy and clinical utility ensures robust validation studies,
producing reliable results for diagnostic tests. IVD medical device performance evaluation can
be designed as observation or intervention. An observational study is the one in which the results
obtained during the study are not used in the treatment of the patient and do not affect treatment
decisions. An interventional study is a study in which the results obtained from the study can
influence patients' decisions and be used to guide treatment. The design of the study must be
decided by taking opinion from experts in the field.
6.1.1. Selection of target population: The condition or disease targeted by the diagnostic
determines selection of appropriate population (for e.g. whether diagnostic aimed at neonatal
sepsis or sepsis etc.), the inclusion and exclusion criteria for enrolling participants and selecting
appropriate study settings. The letter of consent should be signed by the participant (or their
legal guardian) for enrolment in the study.
6.1.2. Inclusion and Exclusion criteria: The inclusion and exclusion criteria for enrolment of
the suitable target population should be outlined to ensure reliable results. Inclusion criteria
might include specific age ranges, clinical setting (in-patient or out-patient), severity of illness
and ability to provide informed consent. Exclusion criteria could encompass those outside the
age range, individuals with co-morbid conditions that could confound results, previous
participants in similar studies, known allergies or contraindications, pregnant or breastfeeding
women, history of non-compliance, acute or chronic diseases and recent surgery or trauma.
These criteria ensure a homogeneous population, enhancing the study's validity and reliability
(Clodi-Seitz et al., 2024).
6.2. Study settings and number of sites Site settings for diagnostic validation should be clearly defined, preferably using accredited laboratories (e.g., NABL). For testing at PHCs or in the field, training requirements for operating protocols and proficiency testing must be established to ensure competency. A documented criterion for the training and retraining of personnel, along with competency assessment records,
29 should be maintained. Personnel performing tests should have access to diagnosis and treatment in case of exposure and the anticipated number of operators should be specified without affecting test results. Requirements for the installation and maintenance of equipment, including electrical needs, should be outlined, with minimal needs for sites like PHCs or bedside locations. Validation should involve multiple sites to avoid bias and ensure accuracy and the validating site must be accredited by a competent authority.
6.3. Test purpose IVD medical devices can be designed for a variety of intended uses like diagnosis, screening, monitoring. The purpose of the test will directly influence the study's sample size (N) and sample selection (including inclusion and exclusion) when planning and designing the evaluation plan. For example, if the prevalence of disease/infection is low and the purpose of the test is to examine the asymptomatic population, a sample of larger subjects would be required to provide sufficient evidence. However, if the test is used for diagnostic purposes in patients, a small sample of subjects will be sufficient. An expert’s opinion must be taken while deciding the sample size.
6.4. Specimen Collection and Handling Samples used in clinical performance studies can be derived from specimens which might have been obtained from different sources, including purposefully-collected specimens, leftover specimens, or archived specimens. The samples must be collected, transported and stored appropriately to preserve the integrity. In case where leftover, or archived specimens are used, there must be sufficient information available necessary to perform data analysis.
6.5. Ethical Considerations for Clinical Performance Studies As a general principle, the rights, safety and well-being of subjects participating in IVD medical device clinical performance studies shall be protected. Approval of ethical committees, informed consents wherever applicable must be taken.
6.6. Clinical Performance Evaluation Criteria Clinical validation evaluates the clinical validity and utility of a test based on the disease or marker being tested. Data can be sourced from laboratory studies, peer-reviewed literature, or other reliable sources. CLIA mandates that laboratories have a qualified director responsible for ensuring the clinical utility of the tests performed.
30
6.6.1. Clinical Validity: The ability of a test to detect or predict the associated disorder
(phenotype)
6.6.2. Clinical Utility: The usefulness of the test in the diagnosis or treatment of patients. The
purpose of test (screening, diagnostic, predictive, etc.) must be clearly defined. Documented via
literature review and/or independent evaluation by the laboratory
6.6.3. Clinical Sensitivity: The proportion of patients with the mutation/disease who have a
positive test result, or the likelihood that a positive result correctly determines that the patient has
the condition being tested. Positive Predictive value = True positive results ÷ (True positive +
False positive). Predictive values take into account the prevalence of the disease in the population
being tested [e.g., the higher the prevalence, the higher the likelihood that a positive result is a
true positive
6.6.4. Clinical Specificity: The proportion of patients who lack the mutation/disease who have a
negative test result, or the likelihood that a negative result correctly determines that the patient
does not have the condition being tested. Negative Predictive Value = True negative results ÷
(True negative + False negative)
Phase 3: Assay Maintenance
Step 7. Maintenance and extension of validation criteria Validation ensures an understanding of the reproducibility, strengths, accuracy and limitations of the study. A validated assay needs constant monitoring and maintenance to retain its designation. Once the assay is used into routine, internal quality control is accomplished by consistently monitoring the assay for assessment of repeatability and accuracy (Cembrowski & Sullivan, n.d.). assay. Reproducibility is assessed through external quality control programmes such as proficiency testing. Continuous monitoring and timely review of assay performance is needed to: • monitor accuracy post validation e.g. the test’s performance in routine usage and to ensure that the expected performance is maintained throughout the life of the test; • monitor reproducibility and limitations; • determine need of calibration and control procedures • perform risk assessments which need to reviewed or written
31 Step 8: Dossier of validation (technical report) Documentation of all validation and verification experiments must be kept by the laboratory for as long as the test is in use but for no less than 2 years (US Federal code; Indian law/BIS/NABL requirement). It should contain following: • Record of each sample testing and method modification, result interpretation and analysis; • Deficiencies if any which could not be resolved in assay validation or in method; • Justification for inadequate data and for each possible scenario. Annexure VI provides the format for the validation report summary. Interpretation of results should be detailed, easy and simple for the personnel performing the test. For instance, whether interpretation of results will be based on visual color change, (or) it will be generated automatically with the inbuilt database (or) results can be printable etc. Instructions for use (brochure)/software should be provided which include details like whether test results will be qualitative (positive/negative or present/absent) or quantitative. Uncertain test results must be reported and a protocol for repeat testing should be provided. Results of validation from third-party (independent) evaluation of the test may be considered for objective assessment and ensuring diagnostic conformity to the stated purpose and claimed outcome.
Step 9: Outcome and Decision Decision on the outcome of validation process of diagnostic will depend on the practicability of the method, assay usability, performance parameters data and user feedback. Approximate cost per test may have decisive impact on the usability of diagnostic which could be determined using cost-effectiveness studies. The barriers to implementation and final recommendations should be made based upon the validation exercise.
32
Annexures
33
Chapter 5 Annexures
S.NO. TITLE Annexure I. INSTRUCTIONS Annexure II. CHECKLIST Annexure III. DEFINITIONS Annexure IV. PROCESS FOR SUBMISSION OF HARD COPY OF APPLICATION Annexure V. PROCESS FOR APPLICATION ON ONLINE SUGAM PORTAL Annexure VI. VALIDATION REPORT SUMMARY FORMAT Annexure VII. CDSCO APPROVED AMR PREDICATE IVDS
34 Annexure I INSTRUCTIONS � Perform risk assessments and identify if any risk [(e.g. operational (chemical/physical), biohazard (biological infectious material) etc.] for which methods need to reviewed or written (ISO 14971:2009) • To minimise the hazards to users of the assay/test •A risk assessment should be performed prior to using any reagent in a diagnostic test; •Infection risk of biological test materials or analyte that may pose a health threat (such as Mycobacterium tuberculosis). •Risk assessment of reporting a false-positive or false-negative result that would result in significant health problem/risk to the patient or general public. � Work place health and safety • The developer must provide adequate data for reasonable assurance of safety and effectiveness of the test/assay/device; • There must be documented policies and procedures relating to workplace health & safety that are consistent with relevant national & jurisdictional workplace health & safety requirements • Waste disposal – to ensure safe disposal of biological waste � Ethical approval for the use of specimens should be taken care of the laboratory undertaking diagnostic testing for validation. � Quality control procedures should be followed strictly by the laboratory. � During validation, quality of raw reagents for testing, storage condition for kits and reagents be defined or recommended. � The storage conditions of reference materials used in testing should be followed as per the instructions given by standard-provider to preserve the quality and stability of the reference material. These should be documented in the dossier.
35 Annexure II
CHECKLIST
• Product information including specifications and Instructions for use must be provided by the developer. The lot number/batch number of the product should also be mentioned. • Test requirements: Information regarding the reagents/chemicals required for the validation study should be provided. Data sheets for all the reagents required including safety data and details of the strains to be used must also be included. • Equipment required to conduct tests must be mentioned by the developer of the test and the same may be supplied if required • Details of the specimen to be used for the validation including the details of the cold chain and storage conditions must be provided. Specimen panel details should be mentioned • SOPs for conducting tests must also be provided • Interpretation criterion must also be provided by the test developer to the validating centre • Statistical advice/analyses
36 Annexure III.
DEFINITIONS
Analytical
sensitivity
Smallest amount of substance in a sample that can
accurately be measured by an assay
Analytical
specificity
Ability of an assay to measure on particular organism or
substance, rather than others, in a sample
Diagnostic
sensitivity
Percentage of persons who have a given disorder who
are identified by the assay as positive for the disorder
Diagnostic
specificity
Percentage of persons who do not have a given
condition who are identified by the assay as negative for
the condition
Accuracy
Closeness of agreement between the test results and an
accepted reference value
Precision
Closeness of agreement between results of replicate
measurements. It is also defined as level of concordance
of the individual test results within a single run (intra -
assay precision) and from one run to another (inter -
assay precision). It is usually characterised in terms of
the standard deviation of the measurements and relative
standard variation (coefficient of variation or %CV)
Reproducibility
Ability to produce essentially the same diagnostic
result/consistent results, under different conditions
(different operators, test batch, different apparatus -
laboratory or validated ancillary equipment, different
laboratories and/or after different intervals of time).
Repeatability is used to indicate within-run
reproducibility.
Linearity
Determination of the linear range of quantification for a
test or test system. It is achieved when measured results
are directly proportional to the concentration of the
analytes (microorganisms or nucleic acid) in the test
sample, within a given range.
Reportable range
Highest and lowest test values that can be analysed
while maintaining accuracy without dilution or
concentration.
Reference range
Range of test values expected for a designated
population of individuals.
37
Positive predictive
value (PPV)
PPV is the probability that when a test is positive,
the specimen does contain the
designated pathogen.
Negative predictive
value (NPV)
NPV is the probability that when a test is negative, the
specimen does not have the designated pathogen.
False Positive
False positive is a result that indicates a given condition
exists when it does not.
False Negative
False negative is a result which wrongly indicates that a
condition does not hold.
Cut off value
For diagnostic or screening tests, the value used to
divide continuous results into categories; typically
positive and negative
38 Annexure IV.
PROCESS FOR SUBMISSION OF HARD COPY OF APPLICATION
Figure 1: Approval process for Application received in Hard copy with respect to In Vitro Diagnostic Division. (Retrieved from CDSCO website https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/medical- device/Approval_process_flowchart_MD_hardcopy2.pdf)
39 Annexure V.
PROCESS FOR APPLICATION THROUGH ONLINE SUGAM PORTAL
Figure 2: Approval process for Application received through Online Sugam Portal for grant of permissions with respect to In Vitro Diagnostics. (A) Step-1:Registration of applicant with MD portal (B) Step-2:Submission and processing of application. (Retrieved from CDSCO website: https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/medical- device/Approval_process_flowchart_MD_Online1.pdf )
40 ANNEXURE VI.
VALIDATION REPORT SUMMARY FORMAT
Name of the product (Brand /generic):
Name and address of the legal manufacturer:
Name and address of the actual manufacturing site:
Type of test:
Lot No./Batch No.
Manufacturing date
Expiry date
Number of tests received:
Intended use
Regulatory Approval: Test License/Manufacturing License License No: Issue date: Valid upto:
Brief details of the test
Reference standard/Product
Samples used
Controls used
Reference method used
Brief details of the method validation plan
Relevant SOPs (Provide SOP Nos. and titles)
Calculation:
Clinical sensitivity: Sensitivity (%) = True Positives X 100 True Positives + False Negatives
41 Clinical specificity: Specificity (%) = True Negatives X 100 True Negatives + False Positives
Positive predictive value (PPV) :
PPV = (prevalence) (sensitivity)
(prevalence) (sensitivity) + ( prevalence) (1- sensitiv
Negative predictive value (NPV):
NPV = (1 - prevalence) (specificity)
(1 - prevalence) (specificity) + ( prevalence) (1-
sensitivity)
Accuracy:
Accuracy = (True Negatives + True Positives)
(True Negatives + True Positives + False
Negatives + False Positives)
Results: S.No Testing Parameter Criteria / specification Result obtained Remark
Conclusion:
Signature of the Analyst
Signature of the Lab. Head Name:
Name: Designation:
Designation: Date:
Date:
42 ANNEXURE VII. Examples of CDSCO approved antimicrobial resistant IVDs (August 2024)* (source: https://cdscomdonline.gov.in/NewMedDev/ListOfIvdMdApprovedDevices)
S. No. Manufacturer/ Importer Name Name of Device Device Class Intended Use Issuing Authority 1 Tulip Diagnostics Private Limited Microplate ELISA Reader Class A Antibiotic Susceptibility Testing machine designed for clinical ELISA test analysis using water-soluble samples and reagent. SLA - Goa 2 Copan India Private Limited WASPLab® System Class A Used for the incubation and the digital imaging of agar culture plates. CLA - CDSCO 3 BioMerieux India Pvt. Ltd.
ETEST
Azithromycin
Class B
For determining the MIC of
antimicrobial agents
CLA -
CDSCO
4
ETEST
Cefixime
Class B
CLA -
CDSCO
5
ETEST
Clarithromycin
Class B
CLA -
CDSCO
6
ETEST
Daptomycin
Class B
CLA -
CDSCO
7
ETEST
Erythromycin
Class B
CLA -
CDSCO
8
ETEST
Fluconazole
Class B
CLA -
CDSCO
9
ETEST
Flucytosine
Class B
CLA -
CDSCO
10
ETEST
Itraconazole
Class B
CLA -
CDSCO
11
Etest
Levofloxacin
Class B
CLA -
CDSCO
12
ETEST
Linezolid
Class B
CLA -
CDSCO
13
ETEST
Moxifloxacin
Class B
CLA -
CDSCO
14
ETEST
Spectinomycin
Class B
CLA -
CDSCO
15
ETEST
Teicoplanin
Class B
CLA -
CDSCO
16
ETEST
Tetracycline
Class B
CLA -
CDSCO
17
ETEST
Trimethoprim
Sulfamethoxazo
le
Class B
CLA -
CDSCO
18
ETEST
Voriconazole
Class B
CLA -
CDSCO
19
CEFTAZIDIME
/ AVIBACTAM
CZA
Class B
CLA -
CDSCO
20
ETEST
AZITHROMYC
IN
Class B
CLA -
CDSCO
43 21 ETEST BENZYLPENI CILLIN Class B CLA - CDSCO 22 ETEST CHLORAMPH ENICOL Class B CLA - CDSCO 23 ETEST CIPROFLOXA CIN Class B CLA - CDSCO 24 ETEST COLISTIN Class B CLA - CDSCO 25 ETEST DAPTOMYCIN Class B CLA - CDSCO 26 ETEST DORIPENEM Class B CLA - CDSCO 27 ETEST ERTAPENEM Class B CLA - CDSCO 28 ETEST IMIPENEM Class B CLA - CDSCO 29 ETEST LINEZOLID Class B CLA - CDSCO 30 ETEST MEROPENEM Class B CLA - CDSCO 31 ETEST MOXIFLOXAC IN Class B CLA - CDSCO 32 ETEST POLYMYXIN B Class B CLA - CDSCO 33 ETEST TEICOPLANIN Class B CLA - CDSCO 34 ETEST VANCOMYCI N Class B CLA - CDSCO 35 Ceftazidime/ceft azidime + clavulanic acid Class B CLA - CDSCO 36 Etest Cefotaxime/ Cefotaxime+ clavulanic acid, Class B CLA - CDSCO 37 Etest Imipenem/ Imipenem +EDTA, Class B CLA - CDSCO 38 Etest Meropenem/ Meropenem +EDTA Class B CLA - CDSCO 39 Etest Piperacillin/ Tazobactam Class B CLA - CDSCO 40 Becton Dickinson India Private Limited BD BACTEC Myco/F Lytic Culture Vials Class B Nonselective culture medium to be used as an adjunct to aerobic blood culture media for the recovery of CLA - CDSCO
44
mycobacteria, yeast and fungi from
blood.
41
BD BACTEC
Plus Aerobic/F
Culture Vials
Class B
Used in a qualitative procedure for
the aerobic culture and recovery of
microorganisms (bacteria and yeast)
from blood.
CLA -
CDSCO
42
BD BACTEC
Plus
Anaerobic/F
Culture Vials
Class B
BD BACTEC Plus Anaerobic/F
medium is used in a qualitative
procedure for the anaerobic culture
and recovery of microorganisms
(bacteria and yeast) from blood. The
principal use of these media is with
the BD BACTEC fluorescent series
instruments.
CLA -
CDSCO
43
BD BBL Sensi-
Disc
Ethionamide -
25 µg
Class B
These discs are used in qualitative
susceptibility testing procedures in
culture media. They serve as a
convenient method for addition of
antimicrobial agents to culture
media, especially for qualitative
studies of mycobacteria and related
organisms
CLA -
CDSCO
44
BD BBL Sensi-
Disc Rifampin,
RA25, 25ug
Class B
These discs are used in qualitative
susceptibility testing procedures in
culture media. They serve as a
convenient method for addition of
antimicrobial agents to culture
media, especially for qualitative
studies of mycobacteria and related
organisms
CLA -
CDSCO
45
HiMedia
Laboratories
Pvt. Ltd.
Positive Blood
Cultures
Pretreatment
Reagent
Class B
This is a pretreatment reagent which
used for the identification of positive
blood culture microorganisms using
the AUTOF MS. It is used in
conjunction with other clinical and
diagnosis procedures as an aid in the
early diagnosis of, for example,
bloodstream infection.
CLA -
CDSCO
46
Suyog
Diagnostic
Private Ltd
URO QUICK
SCREENING
KIT
Class B
Semi-quantitative single use in vitro
diagnostic kit, general
microbiological culture liquid media,
intended to be used by professional
users only for the detection of the
microbial growth in human urine.
CLA -
CDSCO
47
HiMedia
Laboratories Pvt
Ltd.
Dehydrated
Culture
Media(DCM),
HiMedia,
HiCrome
Class B
Devices intended t to grow meant to
grow/ isolate/identify and handle
microorganisms /infectious agent.
SLA -
Nashik
Division
48
Dehydrated
Culture Media
(DCM),
HiMedia,
HiVeg,
Granulated
Class B
Devices intended t to grow meant to
grow/ isolate/identify and handle
microorganisms /infectious agent.
SLA -
Nashik
Division
45 49 CML BIOTECH LIMITED Candida Agar Plate Class B A selective differential media used for rapid isolation and identification of Candida species from mixed cultures in clinical and non-clinical samples. SLA - Kerala 50 HiMedia Laboratories Pvt Ltd.
Anaerobic
Blood Agar
Plate w/
Neomycin
Class B
Anaerobic Blood Agar Plate w/
Neomycin is recommended for
isolation and cultivation of Group A
and Group B Streptococci from
throat cultures and other clinical
samples.
SLA -
Kokan
Division
51
Antimicrobial
Susceptibility
System(ASS)-
HiMic Plate Kit
(MPK071 -
Amphotericin
B)
Class B
To determine Minimum Inhibitory
Concentration (MIC) of antibiotic.
SLA -
Kokan
Division
52
Antimicrobial
Susceptibility
System(ASS)-
HiMic Plate Kit
(MPK012 -
Ceftazidime)
Class B
To determine Minimum Inhibitory
Concentration (MIC) of antibiotic.
SLA -
Kokan
Division
53
Antimicrobial
Susceptibility
System(ASS)-
HiMic Plate Kit
(MPK709 -
Levonadifloxaci
n)
Class B
To determine Minimum Inhibitory
Concentration (MIC) of antibiotic.
SLA -
Kokan
Division
54
Antimicrobial
Susceptibility
System(ASS)-
HiMic Plate Kit
(MPK001 -
Amikacin)
Class B
To determine Minimum Inhibitory
Concentration (MIC) of antibiotic.
SLA -
Kokan
Division
55
Bio-Rad
Laboratories
(India) Pvt Ltd
Ceftolozane +
Tazobactam
30/10 ug
Class B
Antibiotic disks are used to perform a
semi-quantitative antimicrobial
susceptibility testing using disk
diffusion method.
CLA -
CDSCO
56
M/s.Transasia
Bio-Medicals
Ltd.
Suspension
Medium MIC
Class B
Supplementary preparation for MIC
kits (MIC G-I, G-II, URINE,
NEFERM, STAPHY), which are
designed for antibiotic susceptibility
testing.
CLA -
CDSCO
57
BioMerieux
India Pvt. Ltd.
Mueller Hinton
2 agar + 5%
sheep blood
Class B
Susceptibility of pneumococci and
other streptococci to antibiotics.
CLA -
CDSCO
58
Rivaara Labs
Private Limited
AMR Direct
Flow Chip Kit
(Manual and
Auto)
Class C
Allows a quick detection of twenty
AMR gene families, which are
associated with multidrug-resistant
organisms (MRO)
CLA -
CDSCO
46
59
Abbott
Diagnostics
Medical Private
Limited
SD Bioline TB
Ag MPT64
Rapid
Class C
For the detection of antigen MPT64
MTB complex in samples from
liquid or solid culture media.
CLA -
CDSCO
60
HiMedia
Laboratories
Pvt. Ltd.
NG-Test
CARBA 5
Class C
For the detection of the KPC, OXA,
VIM, IMP, NDM carbapenemases in
a bacterial colony obtained from
culture.
CLA -
CDSCO
61
KILPEST
INDIA Ltd
UTI
Antimicrobial
Susceptibility
Testing PCR Kit
Class C
For the detection & differentiation of
AMR genes in uropathogens
CLA –
CDSCO
62
Q-Line Biotech
Private Ltd
MTB RT-PCR
Kit
Class C
Used for the detection of MTB DNA
from suspected samples
CLA -
CDSCO
63
EMPE
Diagnostic Pvt
Ltd
mfloDx MDR-
TB AMP Kit
Class C
Used for amplification and detection
of MTB and its resistance to
Rifampicin and Isoniazid genes.
CLA –
CDSCO
64
Roche
Diagnostics
India Pvt. Ltd.
cobas® 4800
MRSA/SA
Amplification/D
etection Kit
Class C
For the rapid in vitro qualitative
detection of MRSA and SA DNA
from nasal swabs
CLA -
CDSCO
65
Instrumentation
Laboratory India
Pvt Ltd
HemosIL Liquid
Anti-Xa
Class C
For the qualitative detection of MTB
complex DNA in smear positive or
smear negative specimens
CLA - CDSCO 66 Abbott Healthcare Pvt. Ltd.
Abbott
RealTime MTB
Amplification
Reagent Kit
Class C
For the qualitative detection of MTB
complex DNA in smear positive or
smear negative specimens
CLA -
CDSCO
67
Abbott
RealTime MTB
Control Kit
Class C
To establish run validity of the
Abbott RealTime MTB assay
CLA -
CDSCO
*Table 1 and Annexure VII provide examples of CDSCO approved IVDs purely for reference purposes only. None of the examples indicate any endorsement through this document.
47
References
48
Chapter 5 References
Álvarez, S. I., & Andreu, F. A. B. (2011). Procedures for Validation of Diagnostic Methods in Clinical Laboratory Accredited by ISO 15189 (A. B. Eldin, Ed.). IntechOpen. https://doi.org/10.5772/829 Baumfeld Andre, E., Carrington, N., Siami, F. S., Hiatt, J. C., McWilliams, C., Hiller, C., Surinach, A., Zamorano, A., Pashos, C. L., & Schulz, W. L. (2022). The Current Landscape and Emerging Applications for Real-World Data in Diagnostics and Clinical Decision Support and its Impact on Regulatory Decision Making. Clinical Pharmacology and Therapeutics, 112(6), 1172–1182. https://doi.org/10.1002/cpt.2565 Brunden, M. N., Zurenko, G. E., & Kapik, B. (1992). Modification of the error-rate bounded classification scheme for use with two MIC break points. Diagnostic Microbiology and Infectious Disease, 15(2), 135–140. https://doi.org/10.1016/0732-8893(92)90037-t Burd, E. M. (2010). Validation of laboratory-developed molecular assays for infectious diseases. Clinical Microbiology Reviews, 23(3), 550–576. https://doi.org/10.1128/CMR.00074-09 Caliendo, A. M., Gilbert, D. N., Ginocchio, C. C., Hanson, K. E., May, L., Quinn, T. C., Tenover, F. C., Alland, D., Blaschke, A. J., Bonomo, R. A., Carroll, K. C., Ferraro, M. J., Hirschhorn, L. R., Joseph, W. P., Karchmer, T., MacIntyre, A. T., Reller, L. B., Jackson, A. F., & Infectious Diseases Society of America (IDSA). (2013). Better tests, better care: improved diagnostics for infectious diseases. Clinical Infectious Diseases : An Official Publication of the Infectious Diseases Society of America, 57 Suppl 3(Suppl 3), S139-70. https://doi.org/10.1093/cid/cit578 CDSCO. (2023). CDSCO updated list of Laboratories for conducting Performance Evaluation of In
- Vitro Diagnostic Medical Device. Https://Cdsco.Gov.in/Opencms/Export/Sites/CDSCO_WEB/Pdf-Documents/Medical- Device/List-of-Laboratories-for-Conducting-Performance-Evaluation-OfIVDdated- 26.09.2023.Pdf. CDSCO/IVD/GD/Stability/01/2022. (2022). Guidance on Stability Studies of In-vitro Diagnostic Medical Device (IVDMD). Https://Medtech.Citeline.Com/-/Media/Supporting- Documents/Medtech-Insight/2022/08/D0822ind_1.Pdf. https://medtech.citeline.com/- /media/supporting-documents/medtech-insight/2022/08/d0822ind_1.pdf Cembrowski, G., & Sullivan, A. (n.d.). Quality control and statistics (Vol. 1992). Cleophas, T. J., & Zwinderman, A. H. (2009). Summary of Validation Procedures for Diagnostic Tests. In T. J. Cleophas, A. H. Zwinderman, T. F. Cleophas, & E. P. Cleophas (Eds.), Statistics Applied to Clinical Trials (pp. 555–568). Springer, Dordrecht. https://doi.org/https://doi.org/10.1007/978-1-4020-9523-8_37 Clodi-Seitz, T., Baumgartner, S., Turner, M., Mader, T., Hind, J., Wenisch, C., Zoufaly, A., & Presterl, E. (2024). Point-of-Care Method T2Bacteria®Panel Enables a More Sensitive and Rapid Diagnosis of Bacterial Blood Stream Infections and a Shorter Time until Targeted Therapy than Blood Culture. Microorganisms, 12(5). https://doi.org/10.3390/microorganisms12050967 CLSI. (2008). Development of in vitro susceptibility testing criteria and quality control parameters: approved guideline, 3rd ed. CLSI document M23-A3. . CLSI. (2018a). Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically; approved standard—10th ed. M07-A11. .
49 CLSI. (2018b). Performance standards for antimicrobial disk susceptibility tests; approved standard—12th ed. M02-A13. . CLSI. (2018c). Performance standards for antimicrobial susceptibility testing; 28th ed. M100-S28. CLSI and IFCC. (2008). C28-A3 document; Defining, establishing and verifying reference intervals in the clinical laboratory: approved guideline-third edition. https://doi.org/28:1-76 CLSI EP25. (2024). Evaluation of Stability of In Vitro Medical Laboratory Test Reagents, 2nd Edition. CLSI/NCCLS. (2003). Evaluation of the linearity of quantitative measurement procedures: a statistical approach. Approved guideline. CLSI document EP6-A. . Crowther, J. R., Unger, H., & Viljoen, G. J. (2006). Aspects of kit validation for tests used for the diagnosis and surveillance of livestock diseases: producer and end-user responsibilities. Revue Scientifique et Technique (International Office of Epizootics), 25(3), 913–935. FDA. (2009). Guidance for industry and FDA: class II special controls guidance document: antimicrobial susceptibility test (AST) systems. U.S Department of Health and Human Services. Food and Drug Administration Center for Devices and Radiological Health, Rockville, MD. GUIDELINE 2.1. (2012). LABORATORY METHODOLOGIES FOR BACTERIAL ANTIMICROBIAL SUSCEPTIBILITY TESTING . OIE Terrestrial Manual 2012. https://www.woah.org/fileadmin/Home/eng/Our_scientific_expertise/docs/pdf/GUIDE_2.1_A NTIMICROBIAL.pdf Hajian-Tilaki, K. (2014). Sample size estimation in diagnostic test studies of biomedical informatics. Journal of Biomedical Informatics, 48, 193–204. https://doi.org/10.1016/j.jbi.2014.02.013 Humphries, R. M., Ambler, J., Mitchell, S. L., Castanheira, M., Dingle, T., Hindler, J. A., Koeth, L., Sei, K., & CLSI Methods Development and Standardization Working Group of the Subcommittee on Antimicrobial Susceptibility Testing. (2018). CLSI Methods Development and Standardization Working Group Best Practices for Evaluation of Antimicrobial Susceptibility Tests. Journal of Clinical Microbiology, 56(4). https://doi.org/10.1128/JCM.01934-17 Humphries, R. M., Miller, L., Zimmer, B., Matuschek, E., & Hindler, J. A. (2023). Contemporary Considerations for Establishing Reference Methods for Antibacterial Susceptibility Testing. Journal of Clinical Microbiology, 61(6), e0188622. https://doi.org/10.1128/jcm.01886-22 ISO 15189. (2022). Medical laboratories — Requirements for quality and competence. ISO 20916. (2019). In vitro diagnostic medical devices — Clinical performance studies using specimens from human subjects — Good study practice. Jennings, L., Van Deerlin, V. M., Gulley, M. L., & College of American Pathologists Molecular Pathology Resource Committee. (2009). Recommended principles and practices for validating clinical molecular pathology tests. Archives of Pathology & Laboratory Medicine, 133(5), 743–755. https://doi.org/10.5858/133.5.743 Marimuthu, S. S., Radhakrishnan, A., M, V., & Kuppuswamy, G. (2019). Regulatory on stability studies for In-vitro diagnostic medical device. International Journal for Pharmaceutical Research Scholars. McAdam, A. J. (2000). Discrepant analysis: how can we test a test? Journal of Clinical Microbiology, 38(6), 2027–2029. https://doi.org/10.1128/JCM.38.6.2027-2029.2000 Medical Devices Rules, 2017. (n.d.). OIE. (2019). PRINCIPLES AND METHODS OF VALIDATION OF DIAGNOSTIC ASSAYS FOR INFECTIOUS DISEASES. In Manual of Diagnostic Tests for Aquatic Animals. Patel, J. B., Sharp, S., & Novak-Weekley, S. (2013). Verification of Antimicrobial Susceptibility Testing Methods: a Practical Approach. Clinical Microbiology Newsletter, 35(13), 103–109. https://doi.org/https://doi.org/10.1016/j.clinmicnews.2013.06.001
50 Rabenau, H. F., Kessler, H. H., Kortenbusch, M., Steinhorst, A., Raggam, R. B., & Berger, A. (2007). Verification and validation of diagnostic laboratory tests in clinical virology. Journal of Clinical Virology : The Official Publication of the Pan American Society for Clinical Virology, 40(2), 93–98. https://doi.org/10.1016/j.jcv.2007.07.009 Sharma, M., Gangakhedkar, R. R., Bhattacharya, S., & Walia, K. (2021). Understanding complexities in the uptake of indigenously developed rapid point-of-care diagnostics for containment of antimicrobial resistance in India. BMJ Global Health, 6(9). https://doi.org/10.1136/bmjgh-2021-006628 van Belkum, A., Bachmann, T. T., Lüdke, G., Lisby, J. G., Kahlmeter, G., Mohess, A., Becker, K., Hays, J. P., Woodford, N., Mitsakakis, K., Moran-Gilad, J., Vila, J., Peter, H., Rex, J. H., Dunne, W. M., & JPIAMR AMR-RDT Working Group on Antimicrobial Resistance and Rapid Diagnostic Testing. (2019). Developmental roadmap for antimicrobial susceptibility testing systems. Nature Reviews. Microbiology, 17(1), 51–62. https://doi.org/10.1038/s41579- 018-0098-9
51
ACKNOWLEDGEMENT
We acknowledge the contributions of the following experts for their guidance:
Dr. Camilla Rodrigues, P. D. Hinduja Hospital & Medical Research Centre, Mumbai
Dr. Pallab Ray, Postgraduate Institute of Medical Education and Research, Chandigarh
Dr. V. Balaji, Christian Medical College, Vellore
Dr. Sanjay Bhattacharya, Tata Medical Center, Kolkata
Dr. Ranjan Kumar Choudhury, National Health Systems Resource Centre, Delhi
Dr. Varsha Gupta, Govt. Medical College & Hospital, Chandigarh
Dr. Ravikrishnan Elangovan, Indian Institutes of Technology, Delhi
Dr. Jane Cunningham, Diagnostic Advisor, Medecins Sans Frontieres
Mr. Mohammad Ameel, WHO Regional Office for South-East Asia
Dr. Debjit Chakraborty, ICMR - National Institute for Research in Bacterial Infections, Kolkata
Dr. Rajlakshmi Vishwanathan, ICMR-National Institute of Virology, Pune
CDSCO Team Sh. Meshram Pramod Anand Rao, Deputy Drugs Controller Dr. Sella Senthil, Assistant Drugs Controller Smt P. Priyadharsini, Assistant Drugs Controller Dr. Md Omair Anwar, Drugs Inspector (Medical Device)
ICMR Team Dr. (Maj Gen) V C Ohri, Former Consultant Microbiologist, ICMR-AMR Surveillance Network Dr. R R Gangakhedkar, Former Head, Division of Epidemiology and Communicable Diseases Dr. Kamini Walia, Scientist G & Head, Division of Descriptive Research Dr. Monica Sharma, Former Scientist C (Project) Dr. Hina Singh, Consultant (Project) Dr. Shilpi Malhotra, Consultant (Project) Dr. Meenu Jain, Former Scientist C (Project)